← Product Code [PMC](/submissions/SU/subpart-e%E2%80%94surgical-devices/PMC) · K191166

# DigniCap Delta (K191166)

_Dignitana, Inc. · PMC · Jun 26, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PMC/K191166

## Device Facts

- **Applicant:** Dignitana, Inc.
- **Product Code:** [PMC](/submissions/SU/subpart-e%E2%80%94surgical-devices/PMC.md)
- **Decision Date:** Jun 26, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4360
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The DigniCap Delta® Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.

## Device Story

DigniCap Delta is a scalp cooling system designed to reduce chemotherapy-induced alopecia. The device cools fluid to a set temperature and circulates it through a cooling cap worn by the patient. It features a solid-state thermoelectric cooling unit with an integrated control system operated via a touchscreen monitor. The system monitors fluid temperature via sensors in the fluid lines to ensure consistent cooling. It is intended for use in a clinical setting. By maintaining scalp hypothermia during chemotherapy, the device aims to reduce hair loss. Healthcare providers use the touchscreen interface to manage treatment parameters. The device is a single-cap system, an evolution from the predicate's dual-cap design.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), accessory performance, stable fluid temperature verification, alarm system testing, electrical safety (IEC 60601-1:2005(R)2012, IEC 62133-2:2017), electromagnetic compatibility (IEC 60601-1-2:2014), and software verification and validation.

## Technological Characteristics

Solid-state thermoelectric cooling unit; touchscreen interface; single-cap configuration; cooling fluid (isopropyl alcohol/water mixture); temperature sensors in fluid lines; cooling range -7 to 1.5°C. Biocompatible materials per ISO 10993-1. Electrical safety per IEC 60601-1:2005(R)2012 and IEC 62133-2:2017. EMC per IEC 60601-1-2:2014.

## Regulatory Identification

A scalp cooling system to reduce the likelihood of chemotherapy-induced alopecia is a prescription device intended to reduce the frequency and severity of alopecia during chemotherapy in which alopecia-inducing chemotherapeutic agents are used.

## Special Controls

*Classification* —Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device meets all design specifications and performance requirements, and that the device performs as intended under anticipated conditions of use. This information must include testing to demonstrate accuracy of the temperature control mechanism.
(2) Performance testing must demonstrate the electromagnetic compatibility and electrical safety of the device.
(3) Software verification, validation, and hazard analysis must be performed.
(4) The patient contacting components of the device must be demonstrated to be biocompatible. Material names must be provided.
(5) Labeling must include the following:
(i) A statement describing the potential risk of developing scalp metastasis.
(ii) Information on the patient population and chemotherapeutic agents/regimen for which the device has been demonstrated to be effective.
(iii) A summary of the non-clinical and/or clinical testing pertinent to use of the device.
(iv) A summary of the device technical parameters, including temperature cooling range and duration of cooling.
(v) A summary of the device- and procedure-related adverse events pertinent to use of the device.
(vi) Information on how the device operates and the typical course of treatment.
(6) Patient labeling must be provided and must include:
(i) Relevant contraindications, warnings, precautions, and adverse effects/complications.
(ii) Information on how the device operates and the typical course of treatment.
(iii) Information on the patient population for which there is clinical evidence of effectiveness.
(iv) The potential risks and benefits associated with use of the device.
(v) Postoperative care instructions.
(vi) A statement describing the potential risk of developing scalp metastasis.

## Predicate Devices

- Dignitana DigniCap Scalp Cooling System ([K170871](/device/K170871.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Dignitana Inc. % Valerie Defiesta-Ng Vice President, Regulatory Affairs Experien Group 224 Airport Parkway, Suite 250 San Jose, California 95110

June 26, 2019

#### Re: K191166

Trade/Device Name: DigniCap Delta Regulation Number: 21 CFR 878.4360 Regulation Name: Scalp Cooling System to Reduce the Likelihood of Chemotherapy-Induced Alopecia Regulatory Class: Class II Product Code: PMC Dated: April 30, 2019 Received: May 1, 2019

Dear Valerie Defiesta-Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Neil R.P. Ogden, MS Assistant Director THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K191166

Device Name DigniCap Delta® Scalp Cooling System

Indications for Use (Describe)

The DigniCap Delta® Scalp Cooling System is indicated toreduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.

Type of Use (Select one or both, as applicable)

XX Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, induding suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Notification K191166

#### GENERAL INFORMATION

#### Applicant:

Dignitana, Inc. 10925 Estate Lane, Suite W-185 Dallas, Texas 75238 Phone: 469-917-5555 info@dignitana.com

#### Contact Person:

Valerie Defiesta-Ng Vice President, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA

Date Prepared: April 30, 2019

DEVICE INFORMATION

Trade/Proprietary Name: DigniCap Delta®

### Generic/Common Name:

Scalp Cooling System

# Classification:

21 CFR §878.4360

#### Product Code: PMC

Establishment Registration: 3012146457- Dignitana, Inc. Dallas, Texas 2019, Complaint File Establishment

# Manufacturing Facility: Dignitana, Inc.

10925 Estate Lane, Suite W-185 Dallas, Texas 75238

{4}------------------------------------------------

#### PREDICATE DEVICE(S)

DigniCap Delta is substantially equivalent to the predicate device, the Dignitana DigniCap Scalp Cooling System, cleared under K170871. DigniCap Delta is intended for use in the same population, with the same mechanism of action and Indications for Use. Any differences in the technological characteristics between the devices do not raise different questions of safety or effectiveness.

#### DEVICE DESCRIPTION

The DigniCap Delta Scalp Cooling System ("DigniCap Delta")is a device that is intended to function as a cooling system to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors. The proposed therapy of the new device is comparable to the therapy of the predicate device, DigniCap Scalp Cooling System (K170871).

DigniCap Delta cools fluid to a prescribed set temperature and circulates that cooled fluid through a cooling wrap and then back to the device. This mode of operation is equivalent to other scalp cooling devices. User operation of the DigniCap Delta system is carried out on the illuminated graphics display with the integrated touch control, located on the front of the unit. The navigation and selection buttons on the display are used to interface with the device.

#### INDICATIONS FOR USE

The DigniCap Delta® Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

DigniCap Delta is substantially equivalent to the predicate device, the Dignitana DigniCap® Scalp Cooling System, cleared under K170871. Any differences in the technological characteristics between the devices do not raise different questions of safety or effectiveness. DigniCap Delta has the same Indications for Use, population and mechanism of action as the predicate device. Thus, DigniCap Delta is substantially equivalent to the predicate device.

The tables below compare the regulatory information (Table 5.1) and technical characteristics (Table 5.2) for the proposed and predicate devices.

{5}------------------------------------------------

# 510(k) SUMMARY

|                                             | DigniCap Delta® Scalp Cooling System<br>(Subject Device)                                                                                                                                                                                           | DigniCap® Scalp Cooling System<br>(Predicate Device)                                                                                                                                                 | Analysis of<br>Differences                                                                                                                                                                                                                                                                                                              |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature                                     |                                                                                                                                                                                                                                                    |                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                         |
| 510(k)<br>Number                            | To Be Assigned                                                                                                                                                                                                                                     | K170871                                                                                                                                                                                              | N/A                                                                                                                                                                                                                                                                                                                                     |
| Classification                              | II                                                                                                                                                                                                                                                 | II                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                    |
| Product<br>Code                             | PMC                                                                                                                                                                                                                                                | PMC                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                    |
| Regulatory<br>Number                        | \$878.4360                                                                                                                                                                                                                                         | \$878.4360                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                    |
| Regulation<br>Name                          | Scalp Cooling System                                                                                                                                                                                                                               | Scalp Cooling System                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                    |
| Indications<br>for Use                      | The DigniCap Delta® Scalp Cooling System is indicated to<br>reduce the likelihood of chemotherapy-induced alopecia in<br>cancer patients with solid tumors.                                                                                        | The DigniCap® Scalp Cooling System is indicated to<br>reduce the likelihood of chemotherapy-induced alopecia in<br>cancer patients with solid tumors.                                                | Same                                                                                                                                                                                                                                                                                                                                    |
| Mechanism<br>of Action                      | The DigniCap Delta Scalp Cooling System transports<br>temperature controlled cooled fluid from the device to a<br>cooling cap to cool the patient's scalp thus reducing<br>chemotherapy-induced alopecia                                           | The DigniCap Scalp Cooling System transports<br>temperature controlled cooled fluid from the device to a<br>cooling cap to cool the patient's scalp thus reducing<br>chemotherapy-induced alopecia.  | Same                                                                                                                                                                                                                                                                                                                                    |
| Population                                  | Chemotherapy Patients with Solid Tumors                                                                                                                                                                                                            | Chemotherapy Patients with Solid Tumors                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                    |
|                                             | DigniCap Delta® Scalp Cooling System<br>(Subject Device)                                                                                                                                                                                           | DigniCap® Scalp Cooling System<br>(Predicate Device)                                                                                                                                                 | Analysis of Differences                                                                                                                                                                                                                                                                                                                 |
| Feature                                     |                                                                                                                                                                                                                                                    |                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                         |
| Technological<br>Features                   | The DigniCap Delta Scalp Cooling System<br>consists of a solid-state thermoelectric<br>cooling unit with integral control system<br>operated via a touchscreen monitor, capable<br>of precisely controlling the temperature of<br>one cooling cap. | The DigniCap Scalp Cooling System<br>consists of a refrigerator unit with integral<br>control system operated via a touch screen<br>monitor and capable of controlling two<br>separate cooling caps. | Any differences in the technological<br>characteristics between the devices do not<br>raise different questions of safety or<br>effectiveness.                                                                                                                                                                                          |
| User Interface                              | Touchscreen                                                                                                                                                                                                                                        | Touchscreen                                                                                                                                                                                          | Same<br>Updated for user ease.                                                                                                                                                                                                                                                                                                          |
|                                             | Cooling Cap Features                                                                                                                                                                                                                               |                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                         |
| Inner Cooling<br>Cap                        | Yes, includes adjustable tabs                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                  | The differences in the technological<br>characteristics between the devices are<br>minor and do not raise different questions<br>of safety or effectiveness.                                                                                                                                                                            |
| Detachable<br>Coolant Lines                 | Yes                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                    |
| Outer Cover/<br>Material                    | Yes/Neoprene                                                                                                                                                                                                                                       | Yes/Neoprene                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                    |
| Cooling Fluid<br>Required                   | Yes                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                  | Cooling fluid is similar; the DigniCap Delta<br>uses an isopropyl alcohol/water mixture,<br>while the predicate uses MPG<br>(monopropylene glycol). The cooling fluids<br>are used for the same purpose. This<br>difference does not raise different questions<br>of safety or effectiveness as demonstrated<br>by performance testing. |
| Fluid<br>Temperature                        | Fluid temperatures are regulated through<br>temperature sensors in the fluid lines.                                                                                                                                                                | Fluid temperatures are regulated through<br>temperature sensors in the cooling cap.                                                                                                                  | The sensors are located in a different region<br>but are used for the same purpose and have<br>been tested to prove that they work as<br>effectively as the predicate device. This<br>difference does not raise different questions<br>of safety or effectiveness as demonstrated<br>by performance testing.                            |
| Feature                                     | DigniCap Delta® Scalp Cooling System<br>(Subject Device)                                                                                                                                                                                           | DigniCap® Scalp Cooling System<br>(Predicate Device)                                                                                                                                                 | Analysis of Differences                                                                                                                                                                                                                                                                                                                 |
| Pre /Post<br>Cooling Phase                  | Yes                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                    |
| Coolant<br>Temperature<br>Range             | -7 to 1.5°C                                                                                                                                                                                                                                        | -15°C to 5°C                                                                                                                                                                                         | This range is within the range originally<br>cleared with the predicate device but is<br>tighter to provide more consistent cooling<br>to the patient's scalp. This difference does<br>not raise different questions of safety or<br>effectiveness.                                                                                     |
| Outer Cooling<br>Cap                        | Single Patient Use                                                                                                                                                                                                                                 | Multiple Patient Use                                                                                                                                                                                 | This difference does not raise different<br>questions of safety or effectiveness.                                                                                                                                                                                                                                                       |
| Inner Cooling<br>Cap - Single<br>Use        | Single Patient Use                                                                                                                                                                                                                                 | Multiple Patient Use                                                                                                                                                                                 | This difference does not raise different<br>questions of safety or effectiveness.                                                                                                                                                                                                                                                       |
| Device Testing                              |                                                                                                                                                                                                                                                    |                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                         |
| Biocompatibility<br>Testing                 | Yes. Cap materials are biocompatible in<br>accordance with ISO 10993-1.                                                                                                                                                                            | Yes. Cap materials are biocompatible in<br>accordance with ISO 10993-1.                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                    |
| Electrical Safety<br>Testing                | Yes.<br>IEC 60601-1:2005/(R)2012<br>IEC 62133-2:2017                                                                                                                                                                                               | Yes.<br>IEC 60601-1:2007                                                                                                                                                                             | Electrical Safety Testing was conducted to<br>the current standard.                                                                                                                                                                                                                                                                     |
| Electromagnetic<br>Compatibility<br>Testing | Yes.<br>IEC60601-1-2:2014                                                                                                                                                                                                                          | Yes.<br>IEC 60601-1-2:2007                                                                                                                                                                           | Electrical Safety Testing was conducted to<br>the current standard.                                                                                                                                                                                                                                                                     |

# Table 5.1: Substantial Equivalence Table – Regulatory Information

{6}------------------------------------------------

# 510(k) SUMMARY

# Table 5.2: Substantial Equivalence Table – Technological Information

{7}------------------------------------------------

# 510(k) SUMMARY

# Table 5.2: Substantial Equivalence Table – Technological Information (cont.)

{8}------------------------------------------------

#### SUBSTANTIAL EQUIVALENCE

DigniCap Delta has the same Indications for Use, patient population and mechanism of action as the predicate device (K170871). Any differences in the technological characteristics between the devices do not raise different questions of safety or effectiveness. Thus, DigniCap Delta is substantially equivalent to the predicate device.

#### PERFORMANCE DATA

All necessary non-clinical performance testing was conducted on DigniCap Delta to support a determination of substantial equivalence to the predicate device.

#### Nonclinical Testing Summary:

DigniCap Delta was tested to confirm the device met all performance requirements and is considered safe and effective for its intended use. Testing was completed per relevant FDA guidance documents. Specifically, the device was evaluated for the following through nonclinical, bench testing:

- . Biocompatibility Testing of Cooling Cap (Cap materials are biocompatible in accordance with ISO 10993-1)
- Accessory Performance Testing ●
- Stable Fluid Temperature Verification
- Notifications and Alarms System Testing
- Electrical Safety in accordance with IEC 60601-1:2005(R)2012 and IEC 62133-2:2017 ● and Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2014
- Software verification and validation testing

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of DigniCap Delta meet the established specifications necessary for consistent performance during its intended use. The bench testing has demonstrated that the DigniCap Delta can maintain and adjust cooling fluid temperature as needed to meet the predetermined treatment specifications. The collective bench testing demonstrates that DigniCap Delta does not raise new questions of safety or effectiveness for scalp cooling when compared to the predicate devices.

#### Clinical Testing Summary:

Not applicable. Clinical testing was not performed to support this 510(k) submission.

#### CONCLUSIONS

The collective results of the performance testing demonstrate that DigniCap Delta meets the established specifications necessary for consistent performance during its intended use. In addition, the collective performance testing demonstrate that DigniCap Delta does not raise different questions of safety or effectiveness when compared to the predicate device.

Based on the non-clinical performance testing results, the differences in the technological characteristics between the subject and predicate devices do not raise any new issues of safety or 

{9}------------------------------------------------

effectiveness. DigniCap Delta is intended for use in the same population and with the same Indications for Use as its predicate device. Thus, DigniCap Delta is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PMC/K191166](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PMC/K191166)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
