Sentimag System, Sentimark Magnetic Marker Systerm
K153044 · Endomagnetics Ltd., · PBY · Mar 2, 2016 · General, Plastic Surgery
Device Facts
| Record ID | K153044 |
| Device Name | Sentimag System, Sentimark Magnetic Marker Systerm |
| Applicant | Endomagnetics Ltd., |
| Product Code | PBY · General, Plastic Surgery |
| Decision Date | Mar 2, 2016 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4300 |
| Device Class | Class 2 |
Intended Use
The Endomag Sentimark® Magnetic Marker is intended to be placed percutaneously in the breast to mark temporarily (< 30 days) a lumpectomy site intended for surgical removal. Using imaging guidance (such as ultrasound or radiography) or aided by nonimaging guidance (Endomag Sentimag® System) the Endomag Sentimark® Magnetic Marker is located and surgically removed with the target tissue. The Endomag Sentimag® System is intended for the non-imaging detection and localization of the "Endomag Sentimark® Magnetic Marker" that has been implanted in a lumpectomy site intended for surgical removal.
Device Story
System comprises Sentimark Magnetic Marker and Sentimag System console/probe. Marker placed percutaneously in breast using imaging guidance (ultrasound/radiography) to mark lumpectomy site. During surgery, handheld Sentimag probe emits alternating magnetic field; detects magnetic response of implanted marker. Base unit converts signal into visual/audible feedback for surgeon. Used in operating room by trained physicians for non-palpable lesion localization. System aids surgeon in detecting marker material and locating target excision sites. Benefits include precise intraoperative localization of lumpectomy site.
Clinical Evidence
No clinical data. Evidence consists of bench testing and pre-clinical studies: dimensional verification, insertion/deployment/withdrawal force testing, MRI compatibility, simulated use, and a 28-day GLP implantation trial.
Technological Characteristics
System includes console with electronics/user interface and handheld probe. Probe connects to console via flexible cable. Emits alternating magnetic field for marker detection. Provides visual/audible proximity feedback. Marker is percutaneously implanted.
Indications for Use
Indicated for patients requiring temporary (< 30 days) marking of a breast lumpectomy site prior to surgical removal of target tissue.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
- Cianna Medical Tissue Marker and Delivery System (K120804)
Related Devices
- K141318 — CIANNA MEDICAL SGS SYSTEM · Cianna Medical, Inc. · Aug 27, 2014
- K163541 — Magseed Magnetic Marker System · Endomagnetics Ltd., · Jan 13, 2017
- K171767 — Cianna Medical SAVI Scout Reflector and SAVI Scout System · Cianna Medical, Inc. · Oct 31, 2017
- K161507 — Cianna Medical SAVI Scout Reflector and SAVI Scout System · Cianna Medical, Inc. · Jul 29, 2016
- K222832 — Sentimag System · Endomagnetics Ltd., · Jan 21, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 2, 2016
Endomagnetics Ltd Dr. Andrew Shawcross Chief Operations Officer The Jeffreys Building, Cowley Road Cambridge, UK CB4 0WS
Re: K153044
Trade/Device Name: Sentimag System, Sentimark Magnetic Marker System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: PBY Dated: January 29, 2016 Received: February 1, 2016
Dear Dr. Shawcross:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 4. INDICATIONS FOR USE STATEMENT
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Device Name: Endomag Sentimark® Magnetic Marker
Indications for Use:
The Endomag Sentimark® Magnetic Marker is intended to be placed percutaneously in the breast to mark temporarily (< 30 days) a lumpectomy site intended for surgical removal. Using imaging guidance (such as ultrasound or radiography) or aided by nonimaging guidance (Endomag Sentimag® System) the Endomag Sentimark® Magnetic Marker is located and surgically removed with the target tissue. The Endomag Sentimag® System is intended for the non-imaging detection and localization of the "Endomag Sentimark® Magnetic Marker" that has been implanted in a lumpectomy site intended for surgical removal.
Prescription Use __ V
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(21 CFR 801 Subpart C)
(Part 21 CFR 801 Subpart D)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### 5 510(K) SUMMARY
### 5.1 SUBMITTER INFORMATION
| Submitter's Name: | Endomagnetics Ltd. |
|------------------------|--------------------------------------------------------------------------------|
| Address: | The Jeffreys Building<br>Cowley Road<br>Cambridge<br>CB4 0WS<br>United Kingdom |
| Contact Person: | Andrew Shawcross<br>Chief Operations Officer |
| Tel: | +44 1223 652540 |
| Email: | ashawcross@endomag.com |
| Date summary prepared: | 2nd March 2016 |
#### 5.2 DEVICE INFORMATION
| Trade name: | Sentimark Magnetic Marker and Sentimag System |
|------------------------|---------------------------------------------------------------|
| Common name: | Tissue Marker, Marker Delivery System and Detection<br>System |
| Classification name: | Implantable Clip |
| Regulation: | 21 CFR 878.4300 |
| Device Classification: | Class II |
| Product Code: | PBY |
#### 5.3 PREDICATE DEVICE
Cianna Medical Tissue Marker and Delivery System (K120804).
{4}------------------------------------------------
#### DESCRIPTION OF DEVICE 5.4
The Sentimag System is intended for the non-imaging detection and localization of the Sentimark Magnetic Marker that has been implanted in a lumpectomy site intended for surqical removal.
The Sentimag System and Sentimark Magnetic Marker System are designed for use in an operating room environment by suitably trained physicians who are experienced in diagnosis and treatment of breast lesions.
The Sentimag System aids the surgeon to detect surgically invasively magnetic marker material that has been placed for the purpose of detecting a non-palpable lesion, and to locate target excision sites.
Prior to a lumpectomy procedure, the Sentimark Magnetic Marker is placed percutaneously into the breast, using imaging quidance such as ultrasound or radiography, to temporarily mark a site intended for surgical removal. During a surgical procedure, the hand-held Sentimaq probe emits an alternating magnetic field that detects the magnetic response of the Sentimark magnetic marker, this signal is converted by the base unit into a visual and audible response that is similar in use to the predicate device.
#### INTENDED USE 5.5
The Endomag Sentimark® Magnetic Marker is intended to be placed percutaneously in the breast to mark temporarily (< 30 days) a lumpectomy site intended for surgical removal. Using imaging guidance (such as ultrasound or radiography) or aided by nonimaging quidance (Endomag Sentimag® System) the Endomag Sentimark® Magnetic Marker is located and surgically removed with the target tissue.
The Endomag Sentimag® System is intended for the non-imaging detection and localization of the "Endomag Sentimark® Magnetic Marker" that has been implanted in a lumpectomy site intended for surgical removal.
### 5.6 SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The K120804 predicate uses imaging technology to locate the device. This can be a handheld Ultrasound transducer. The Sentimark Magnetic Marker System utilizes nonimaging technologies that are comprised of a console that incorporates electronics and a simple user interface, plus a probe handpiece.
In both cases, a location marker is placed percutaneously in situ at the clinical target site by a delivery system. Both devices then employ the handpiece for the intraoperative detection and localization of the implanted marker.
In both systems the handpiece is connected by a flexible cable to a console unit that provides the user with a visual indication of the presence and proximity of the marker.
{5}------------------------------------------------
The detailed technological characteristics of the two systems have been identified and compared. The minor differences in the technological characteristics do not raise any new questions of safety or effectiveness.
#### DISCUSSION OF NON-CLINICAL TESTS SUBMITTED 5.7
Performance testing was conducted to evaluate and characterize the performance of the Sentimag System and Sentimark Magnetic Marker System. Pre-clinical testing included:
- . Dimensional Verification
- . Insertion, Deployment and Withdrawal Force
- MRI Compatibility
- . Simulated Use
- 28-Day GLP Implantation Trial
#### 5.8 CONCLUSION
Endomagnetics believes that the Cianna Medical Tissue Marker and Delivery System (K120804) is the closest predicate device because it has the same intended use and very similar technological characteristics.
The Sentimag Magnetic Marker System has the same Intended Use as the predicate device. The different technological characteristics do not raise any new questions of safety or effectiveness. The test, verification and validation data presented in this submission demonstrate substantial equivalence of the Sentimag Magnetic Marker System.