CIANNA MEDICAL TISSUE MARKER AND DELIVERY SYSTEM
K120804 · Cianna Medical · PBY · Feb 4, 2013 · General, Plastic Surgery
Device Facts
| Record ID | K120804 |
| Device Name | CIANNA MEDICAL TISSUE MARKER AND DELIVERY SYSTEM |
| Applicant | Cianna Medical |
| Product Code | PBY · General, Plastic Surgery |
| Decision Date | Feb 4, 2013 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4300 |
| Device Class | Class 2 |
Intended Use
The Cianna Medical Tissue Marker is intended to be placed percutaneously in the breast to temporarily (< 30 days) mark a lumpectomy site intended for surgical removal. Using ultrasound or radiography imaging guidance, the Cianna Tissue Marker is located and surgically removed with the target tissue.
Device Story
Cianna Medical Tissue Marker and Delivery System consists of an implantable marker and a delivery system. Marker comprises 1 mm diameter titanium beads constrained on a heat-shaped nitinol wire. Preloaded in delivery system lumen; percutaneously deployed into breast tissue where it maintains a coiled shape (approx. 4.5mm diameter x 5mm length). Device is echogenic under ultrasound and radiopaque under radiography. Used by physicians in clinical settings to mark lumpectomy sites for surgical removal. Marker remains in situ for < 30 days and is removed along with target tissue. Healthcare providers use ultrasound or radiography to locate the marker, facilitating precise identification of the surgical site. Benefits include improved localization of target tissue for excision.
Clinical Evidence
No clinical data. Bench testing only. Testing included dimensional conformance, visual inspection, design verification (imaging equivalency, MRI compatibility), and biocompatibility per ISO 10993-1 (cytotoxicity, intracutaneous reactivity, sensitization, systemic toxicity, pyrogenicity, mutagenicity, and 30-day rabbit implantation study).
Technological Characteristics
Materials: Titanium beads, nitinol wire. Form factor: Coiled marker (4.5mm x 5mm). Imaging: Echogenic (ultrasound), radiopaque (radiograph). Delivery: Needle-based percutaneous system. Sterilization: Sterile, single-use. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for percutaneous placement in the breast to temporarily (< 30 days) mark a lumpectomy site intended for surgical removal. Used with ultrasound or radiography imaging guidance for subsequent surgical removal with target tissue.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
- Mixed Media Marker (K102506)
- TuMark Flex Tissue Marking System (K111692)
- Hydromark (K090501)
Related Devices
- K132463 — CIANNA MEDICAL PERMANENT TISSUE MARKER AND DELIVERY SYSTEM · Cianna Medical, Inc. · Dec 3, 2013
- K182082 — Tumark for Eviva, Tumark for Brevera · Somatex Medical Technologies GmbH · Oct 31, 2018
- K232615 — LumiMARK Biopsy Site Marker (LM0215T (Tulip Shape)); LumiMARK Biopsy Site Marker (LM0215L (Lotus Shape)); LumiMARK Biopsy Site Marker (LM0215R (Rose shape)) · Devicor Medical Products, Inc. · Jan 12, 2024
- K240429 — Trilogy Tissue Marker · Inrad, Inc. · Nov 7, 2024
- K183503 — EasyMark Tissue Marker · Inrad, Inc. · Oct 17, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
# Section 6
# 510(k) Summary
K120804 Page 1 of 3
## 510(k) Summary
This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Cianna Medical<br>6 Journey # 125<br>Aliso Viejo, CA 92656 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DATE PREPARED: | 2-3-2013 |
| CONTACT PERSON: | Gary Mocnik<br>Cianna Medical<br>6 Journey<br>Aliso Viejo, CA 92656<br>Phone: (949)360.0059 x134<br>Fax: (949) 297.4527<br>Email: gmocnik@ciannamedical.com |
| TRADE NAME: | Cianna Medical Tissue Marker and Delivery System |
| COMMON NAME: | Temporary Tissue Marker |
| CLASSIFICATION NAME: | Implantable Clip, 21 CFR 878.4300 |
| DEVICE CLASSIFICATION: | Class II |
| PRODUCT CODE | PBY |
| PREDICATE DEVICES: | Mixed Media Marker- K102506<br>TuMark Flex Tissue Marking System- K111692<br>Hydromark (K090501) |
#### Substantially Equivalent To:
The Cianna Medial Tissue Marker and Delivery System is substantially equivalent in intended use, principal of operation and technological characteristics to the Mixed Media Marker (K102506), TuMark Flex Tissue Marking System (K111692), and the Hydromark (K090501). The Cianna Medical Tissue Marker differs from the predicates only in its intended implant duration. The predicates are intended to be long term implants while the Cianna Medical Tissue Marker is intended for implant duration of less than 30 days and is removed with the target tissue.
#### Description of the Device Subject to Premarket Notification:
The Cianna Medical Tissue Marker and Delivery System is comprised of an implantable marker and a delivery system.
The marker is comprised of a series of 1 mm diameter titanium "beads" which are constrained onto a heat shaped nitinol wire. The marker is preloaded within the lumen of the delivery system. Once percutaneously deployed from the delivery system, the marker maintains a coiled shape (approximately 4.5mm diameter by 5mm length) in the tissue. The marker is echogenic under Ultrasound, radiopaque under radiograph imaging.. The Cianna Medical Tissue Marker and Delivery System is provided sterile, and is for single use only.
{1}------------------------------------------------
# Section 6
K120804 page 2/3
510(k) Summary
K120804 Page 2 of 3
#### Indication for Use:
The Cianna Medical Tissue Marker is intended to be placed percutaneously in the breast to temporarily (< 30 days) mark a lumpectomy site intended for surgical removal. Using ultrasound or radiography imaging guidance, the Cianna Tissue Marker is located and surgically removed with the target tissue.
# Discuss of Technological Characteristics:
The Cianna Medical Tissue Marker and Delivery System has the same technological characteristics and is similar in overall design, materials and configuration compared to the predicates. The Cianna Medical Tissue Marker and Delivery System and the identified predicates are all radiographically visible marker elements positioned into tissue for visualization of the tissue site. The technical designs of the Cianna Medical Tissue Marker and the predicate devices are very similar, being composed of an array of biocompatible, radiopaque metals which are delivered into tissue via a needle delivery system.
## Non-Clinical Performance Data
Performance testing was conducted to evaluate and characterize the performance of the Cianna Medical Tissue Marker System. Preclinical testing conducted included dimensional conformance evaluation, visual inspections, and design verification to confirm imaging equivalency, MRI compatibility testing, and biocompatibility testing based on the applicable elements of ISO 10993-1 shown below.
| Test Performed | Standard | Test Result/Conclusion |
|-------------------------------------------------------------|----------------------------|----------------------------------------------------------------|
| ISO MEM Elution Assay with L-<br>929 Mouse Fibroblast Cells | ISO 10993-5 | Passed.<br>Non-cytotoxic |
| ISO Intracutaneous Reactivity Test | ISO 10993-10 | Passed.<br>The test requirements were met |
| Sensitization: Guinea Pig<br>Maximization | ISO 10993-10 | Passed<br>Negative for evidence of<br>sensitization |
| ISO Acute Systemic Injection Test | ISO 10993-11 | Passed<br>The test requirements were met |
| Material Mediated Rabbit Pyrogen<br>Test | ISO 10993-10 | Passed<br>Non-Pyrogenic |
| Bacterial Mutagenicity Test-Ames<br>Assay | ISO 10993-3 | Passed<br>Non-Mutagenic |
| In Vitro Mouse Lymphoma Assay | ISO 10993-3 | Passed<br>Non-Mutagenic (non-genotoxic and<br>non-clastogenic) |
| In Vivo Mouse Micronucleus<br>Assay | ISO 10993-3 | Passed<br>Non-Mutagenic |
| 30 Day Rabbit Implantation with<br>Subchronic Toxicity | ISO 10993-6<br>ISO10993-11 | Passed<br>No local or systemic signs of toxicity |
{2}------------------------------------------------
# Section 6
K120504 Page 3/3
# 510(k) Summary
K120804 Page 3 of 3
# Clinical Data
This submission does not rely on clinical data to determine substantial equivalency to the predicate devices.
# Basis for Determination of Substantial Equivalence:
| Product | Intended Use | Principle of Operation | Overall Technological Characteristics |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|--------------------------------------------------|
| Cianna<br>Medical Tissue<br>Marker and<br>Delivery<br>System | The Cianna Medical Tissue Marker is intended<br>to be placed percutaneously in the breast to<br>temporarily (<30 days) mark a lumpectomy<br>site intended for surgical removal. Using<br>ultrasound or radiography imaging guidance,<br>the Cianna Tissue Marker is located and<br>surgically removed with the target tissue. | Marker is<br>implanted into<br>tissue site for<br>visualization of<br>tissue site | Radiographically<br>visible marker<br>element(s) |
| Mixed Media<br>Marker<br>K102506 | The Mixed Media Markers (MMM) are<br>intended to be implanted into the body in<br>situations where the location of specific<br>anatomy, normal or diseased, needs to be<br>marked for a future medical procedure. The<br>MMM can be visualized using medical imaging<br>devices; the MMM provides a reference from<br>which treatment can be guided. MMM's not<br>intended for use with ultrasonography. | SAME | SAME |
| TuMark Flex<br>Tissue<br>marking<br>System<br>K111692 | The TuMark Flex is intended for<br>radiographically and radiologically<br>percutaneous marking of soft tissue, especially<br>breast tissue, via a clip marker. The TuMark<br>Flex is not indicated to be used with magnetic<br>resonance imaging (MRI) techniques. | SAME | SAME |
| HydroMark<br>Biopsy Site<br>Marker<br>K090501 | The Biopsy Science LLC, HydroMark Biopsy<br>Site Marker is intended to mark tissue during a<br>percutaneous breast biopsy procedure, be<br>visible under ultrasound for at least 6 weeks<br>and be permanently visible by x-ray and MRI | SAME | SAME |
#### Conclusions Drawn:
As shown, the Cianna Medical Tissue Marker has the following similarities to the predicate devices:
- Similar intended use .
- Similar design characteristics .
- . Same operating principal
- . Same mechanism of action
- Similar technological characteristics .
Upon reviewing the safety and efficacy information provided in this submission and comparing intended use, principle of operation and overall technological characteristics, the Cianna Medical Tissue Marker System is determined by Cianna Medical, to be substantially equivalent to existing legally marketed devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure. The symbol is composed of three curved lines that create a sense of movement and flow.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Cianna Medical % Mr. Gary Mocnik Vice President, Regulatory Affairs and Quality Systems 6 Journey, Suite 125 Aliso Viejo, California 92656
Re: K120804
Trade/Device Name: Cianna Medical Tissue Marker System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: PBY Dated: January 18, 2013 Received: January 30, 2013
Dear Mr. Mocnik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
February 4, 2013
{4}------------------------------------------------
Page 2 - Mr. Gary Mocnik
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number: K120804
Device Name: Cianna Medical Tissue Marker System
Indications for Use:
The Cianna Medical Tissue Marker is intended to be placed percutaneously in the breast to temporarily (< 30 days) mark a lumpectorny site intended for surgical removal. Using ultrasound or radiography imaging guidance, the Cianna Tissue Marker is located and surgically removed with the target tissue.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use X (Per 21 CFR 801 Subpart D)
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
Page 1 of 1
# David Krause
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K120804