PleuraFlow System with FlowGlide

K203394 · Clearflow, Inc. · OTK · Mar 2, 2021 · General, Plastic Surgery

Device Facts

Record IDK203394
Device NamePleuraFlow System with FlowGlide
ApplicantClearflow, Inc.
Product CodeOTK · General, Plastic Surgery
Decision DateMar 2, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4780
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The PleuraFlow® System with FlowGlide® is indicated for use during cardical procedures and chest trauma. Its Active Clearance Technology® proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.

Device Story

PleuraFlow System with FlowGlide is a wound drain catheter system designed to maintain chest tube patency. It consists of a chest tube and a clearance apparatus (guide tube with magnetic shuttle and PTFE-coated clearance wire with distal loop). Used in clinical settings by healthcare providers to evacuate blood/fluid post-surgery. The clearance apparatus is inserted into the chest tube; the magnetic shuttle allows the user to advance/retract the clearance wire/loop to break up clots or obstructions. The PFFG3 model features a spring-activated button to increase magnetic force for shuttle operation. By maintaining tube patency, the device reduces retained blood, potentially improving patient outcomes post-cardiothoracic surgery or trauma.

Clinical Evidence

Bench testing only. No clinical data was deemed necessary based on risk management assessment and over 10 years of clinical history with the predicate technology.

Technological Characteristics

System comprises a chest tube and clearance apparatus (guide tube, magnetic shuttle, PTFE-coated wire with loop). Materials are biocompatible per ISO 10993-1. Sterilization via ISO 11135:2014. Features a spring-activated button for adjustable magnetic force. Single-use, sterile device.

Indications for Use

Indicated for adult and pediatric patients (including infants, preadolescents, adolescents) undergoing cardiothoracic surgical procedures or experiencing chest trauma to prevent/minimize chest tube occlusion and facilitate fluid evacuation.

Regulatory Classification

Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is written in blue letters next to it. March 2, 2021 ClearFlow, Inc, Serrah Namini VP RA/QA/CA 140 Technology Drive, Suite 100 Irvine, California 92618 Re: K203394 Trade/Device Name: PleuraFlow System with FlowGlide Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OTK, GBX Dated: February 3, 2021 Received: February 4, 2021 Dear Serrah Namini: We have reviewed your Section 5100k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Form Approved: OMB No. 0990-0120 Expiration Date: 06/30/2023 See PRA Statement below. #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K203394 Device Name PleuraFlow® System with FlowGlide® #### Indications for Use (Describe) The PleuraFlow® System with FlowGlide® is indicated for use during cardical procedures and chest trauma. Its Active Clearance Technology® proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings. Type of Use (Select one or both, as applicable) | <div> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |--------------------------------------------------------------------------------| | <div> <span></span> Over-The-Counter </div> | Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K203394 # 510(k) Summary (807.92(c)) The following information is provided as required by 21 CFR § 807.92(c) for PleuraFlow® System with FlowGlide™ 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based. | Date of Submission: | Nov 17, 2020 | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant: | ClearFlow, Inc.<br>140 Technology Drive, Suite 100<br>Irvine, CA 92618 | | Primary Contact Person: | Serrah Namini<br>ClearFlow, Inc.<br>140 Technology Drive, Suite 100<br>Irvine, CA 92618<br>Phone: 714-916-5014<br>Fax: 714-916-5019<br>Email: snamini@clearflow.com | | CEO/President: | Paul Molloy<br>ClearFlow, Inc.<br>Phone: 714-905-5271<br>Fax: 714-916-5019<br>Email: pmolloy@clearflow.com | | Device Proprietary Name: | PleuraFlow® System with FlowGlide® | | Device Common Name: | Wound drain catheter system | | Regulatory Class and Name: | Class II, 878.4780 Powered suction pump | | Product Codes: | OTK and GBX | {4}------------------------------------------------ Predicate Device: Predicate device is the PleuraFlow Catheter System (K163139 and number K182067) as well as K191733 by ClearFlow, Inc. The PleuraFlow® System with FlowGlide(R) PFFG3 iss a Device Description: modification to our current PleuraFlow® System with FlowGlide(R) (predicate). The primary components of the System are the Chest Tube and the Clearance Apparatus. The Chest Tube is connected to a Clearance Apparatus, which is connected to the tubing from the drainage canister. > The Clearance Apparatus that is part of the PleuraFlow System with FlowGlide consists of a Guide Tube with a magnetic shuttle and a PTFE-coated Clearance Wire with a Loop set on its distal end. The Clearance Apparatus is advanced into the PleuraFlow Chest Tube using a magnetic Shuttle. When indicated, the Clearance Wire and Loop is advanced and retracted within the PleuraFlow with FlowGlide Chest Tube to proactively prevent or break up and clear any tube obstructions or clogging to keep the tube patent. The PleuraFlow System with FlowGlide, PFFG3, has a new shuttle design that includes a spring activated button to allow the user to increase the magnetic force when needed. The range of magnetic force is the same as predicate. The PleuraFlow® System with FlowGlide™ is indicated for use Indication For Use: during cardiothoracic surgical procedures and chest trauma. Its Active Clearance Technology proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings. {5}------------------------------------------------ #### K203394 # Technological Characteristics and Performance Data: The performance of the new models of the PleuraFlow System with FlowGlide® was shown to be substantially equivalent to the cleared models (predicate) through bench testing. Performance of new configuration of the PleuraFlow System was verified using the following testing as summarized in the submission: - . Test for actuation and tracking of the Clearance Wire and Loop through the coated chest tube tortuous path - Functional Testing of the Shuttle o - o Coupling force testing of the Shuttle to Clearance Wire and Loop - Force to actuate, force to move the Shuttle along the guide tube o of the clearance apparatus while coupled to the Clearance Wire and Loop - Magnetic Flux testing, magnetic field in proximity to the device ● - Button cycle testing, test for button function after repeated uses ● - . Force to separate the shuttle enclosure - Transportation simulation. ● - Tensile strength of the drain tubing to drain barb o The biocompatibility of materials remains the same and as such, the new device meets ISO 10993-1 applicable requirements. The new device is provided sterile and single use and meets ISO 11135:2014 requirements. The safety and effectiveness of the predicate have been previously demonstrated through design validation and verification that were cleared under 510(k) premarket notification (K163139 and number K182067). No clinical performance was deemed necessary according to Risk Management {6}------------------------------------------------ assessment and evaluation. Use of the predicate devices for more than 10 years has demonstrated safety and effectiveness for patients recovering after surgery and who were placed with the PleuraFlow System with Active Clearance Technology® (ACT®). The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate device Results from performance testing of the new models of PleuraFlow System with FlowGlide® PFFG3, demonstrates that these are suitable for the intended use and did not raise new issues of safety and effectiveness when compared to the predicate models. The Indication for Use for the new configurations is same as Conclusion: the predicate. The fundamental design and technological characteristics are the same as the predicate. Risk assessment, verification and validation of the PleuraFlow System with FlowGlide®, PFFG3, do not raise any additional concerns regarding safety and effectiveness and they are substantially equivalent to the predicate system.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...