← Product Code [ONG](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG) · K221770

# Fiber Laser Treatment System (K221770)

_Shanghai Apolo Medical Technology Co., Ltd. · ONG · Oct 13, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG/K221770

## Device Facts

- **Applicant:** Shanghai Apolo Medical Technology Co., Ltd.
- **Product Code:** [ONG](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG.md)
- **Decision Date:** Oct 13, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Fiber Laser Treatment System is intended for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue.

## Device Story

Desktop Erbium fiber laser system (HS-230) for dermatological procedures; produces 1550nm pulsed near-infrared light. System comprises mainframe with color LCD touch-screen, power supply, cooling system, fiber treatment handpiece, and footswitch. Operator selects parameters via touch-screen; activates laser via footswitch. Laser beam delivered to treatment zone through optical fiber and handpiece. Device performs fractional skin resurfacing and soft tissue coagulation. Used in clinical settings by trained professionals. Benefits include precise, controlled thermal effects on skin tissue.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), and performance requirements for surgical/cosmetic laser equipment (IEC 60601-2-22). Biocompatibility complies with ISO 10993-1.

## Technological Characteristics

Erbium glass fiber laser; 1550nm wavelength; 15W output power; 1-70mJ pulse energy; 10ms pulse width; 110μm beam diameter. Desktop form factor with color LCD touch-screen interface. Power: 100-240VAC, 50/60Hz. Compliance: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22. Biocompatibility: ISO 10993-1.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Medical Non-ablative Fractional Laser Systems ([K192350](/device/K192350.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

October 13, 2022

Shanghai Apolo Medical Technology Co., Ltd. Felix Li RA Supervisor Contact Address

Re: K221770

Trade/Device Name: Fiber Laser Treatment System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONG Dated: July 6, 2022 Received: July 13, 2022

Dear Felix Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K221770

Device Name Fiber Laser Treatment System

Indications for Use (Describe)

The Fiber Laser Treatment System is in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue.

Type of Use (Select one or both, as applicable)

| <span style="font-size:10pt"> <span style="font-family:Wingdings">✓</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt"> <span style="font-family:Arial">▢</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span>      |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K221770 Section 3-510(k) summary

#### I Submitter

Shanghai Apolo Medical Technology Co., Ltd. 4F, Building A, No.388, Yindu Road, Xuhui District, Shanghai 200231,China

Establishment Registration Number: 3007120647

Contact person: Felix Li Position: Regulatory Affairs Phone: +86-138 4919 0618 Fax: +86-21-34622840 E-mail: liqiang@apolo.com.cn

### II Proposed Device

| Trade Name of Device: | Fiber Laser Treatment System                             |
|-----------------------|----------------------------------------------------------|
| Common name:          | Laser Surgical Instrument For Use In General And Plastic |
|                       | Surgery And In Dermatology                               |
| Regulation Number:    | 21 CFR 878.4810                                          |
| Regulatory Class:     | Class II                                                 |
| Product code:         | ONG                                                      |
| Review Panel:         | General & Plastic Surgery                                |

#### III Predicate Devices

| 510(k) Number:  | K192350                                                                                |
|-----------------|----------------------------------------------------------------------------------------|
| Trade name:     | Medical Non-ablative Fractional Laser Systems                                          |
| Common name:    | Laser Surgical Instrument For Use In General And Plastic<br>Surgery And In Dermatology |
| Classification: | Class II                                                                               |
| Product Code:   | ONG                                                                                    |
| Manufacturer    | Wingderm Electro-Optics Ltd.                                                           |

### IV Device description

The Fiber laser Treatment system (HS-230) is the desktop type. It consist of mainframe(touch-screen data input system, power supply control system and cooling system), fiber treatment handpiece (laser output system ) and footswitch.

{4}------------------------------------------------

The Fiber Laser Treatment System is a Erbium fiber laser, producing a pulsed beam of near-infrared light (1550nm) upon activation by a footswitch. The beam is then directed to treatment zone by mean of an optical fiber couple to a handpiece.

## V Indication for use

The Fiber Laser Treatment System is intended for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue.

| VI Comparison of technological characteristics with the predicate devices |                                                                                                                                                          |                                                                                                                                                                                          |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item                                                                      | Proposed device                                                                                                                                          | Predicate device<br>(K130193)                                                                                                                                                            |
| Product name                                                              | Fiber Laser Treatment System                                                                                                                             | Medical Non-ablative Fractional<br>Laser Systems                                                                                                                                         |
| Product Code                                                              | ONG                                                                                                                                                      | ONG                                                                                                                                                                                      |
| Regulation No.                                                            | 21 CFR 878.4810                                                                                                                                          | 21 CFR 878.4810                                                                                                                                                                          |
| Class                                                                     | Class II                                                                                                                                                 | Class II                                                                                                                                                                                 |
| Model                                                                     | HS-230                                                                                                                                                   | WFB-01                                                                                                                                                                                   |
| Indication for use                                                        | The Fiber Laser Treatment Systems is intended for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue. | The Medical Non-ablative Fractional Laser Systems (Model: WFB-01) is intended for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue. |
| Laser Type                                                                | Erbium glass                                                                                                                                             | Erbium glass                                                                                                                                                                             |
| Components                                                                | Laser system, Color touch screen, Fiber handpiece, Footswitch                                                                                            | Laser system, Color touch screen, Fiber, handpiece Footswitch.                                                                                                                           |
| Wavelength                                                                | 1550nm                                                                                                                                                   | 1550nm                                                                                                                                                                                   |
| Output power                                                              | 15W                                                                                                                                                      | 15W                                                                                                                                                                                      |
| Pulse energy                                                              | 1-70mJ                                                                                                                                                   | Up to 70mJ                                                                                                                                                                               |
| Pulse width                                                               | 10ms                                                                                                                                                     | 10ms                                                                                                                                                                                     |
| Beam delivery system                                                      | Fiber and handpiece                                                                                                                                      | Fiber and handpiece                                                                                                                                                                      |
| Laser beam diameter                                                       | 110μm                                                                                                                                                    | 110μm                                                                                                                                                                                    |
| Beam diameter variability                                                 | Fixed                                                                                                                                                    | Fixed                                                                                                                                                                                    |
| Dot Density                                                               | 25~484 (18 level)                                                                                                                                        | 500                                                                                                                                                                                      |
| (Dot/cm²)                                                                 |                                                                                                                                                          |                                                                                                                                                                                          |
| Laser<br>emission<br>control                                              | Footswitch                                                                                                                                               | Footswitch                                                                                                                                                                               |
| Scanning area                                                             | Adjustable,<br>Maximum treatment area :<br>20mm*20mm                                                                                                     | AccuTip: 10mm*10mm,<br>EffiTip: 20mm*20mm<br>GrowTip: 10mm*20mm                                                                                                                          |
| Laser class                                                               | Class 4                                                                                                                                                  | Class 4                                                                                                                                                                                  |
| Operation<br>interface                                                    | Color LCD touch screen                                                                                                                                   | Color LCD touch screen                                                                                                                                                                   |
| Power supply                                                              | 100-240VAC, 50/60Hz                                                                                                                                      | 100-240VAC, 50/60Hz,                                                                                                                                                                     |
| Software                                                                  | Yes                                                                                                                                                      | Yes                                                                                                                                                                                      |
| EMC                                                                       | Comply with IEC 60601-1-2                                                                                                                                | Comply with IEC 60601-1-2                                                                                                                                                                |
| Safety                                                                    | Comply with IEC 60601-1                                                                                                                                  | Comply with IEC 60601-1                                                                                                                                                                  |
| Performance                                                               | Comply with IEC 60601-2-22 and IEC<br>60825-1                                                                                                            | Comply with IEC 60601-2-22 and<br>IEC 60825-1                                                                                                                                            |
| Material<br>Biocompatibility                                              | Comply with ISO 10993-1                                                                                                                                  | Comply with ISO 10993-1                                                                                                                                                                  |

VI Comparison of technological characteristics with the predicate devices

{5}------------------------------------------------

## VII Non-Clinical Testing

A series of tests to verity that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:

### Electrical safety and electromagnetic compatibility

IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2014 Medical electrical equipment-Part1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests

IEC 60825-1:2014 Safety of Laser products-Part 1:Equipment classification and requirements

IEC 60601-2-22:2007 Medical electrical equipment-Part 2-22: Particular requirements for the safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment

### VIII Clinical Testing

lt is not applicable.

{6}------------------------------------------------

IX Conclusion
The proposed device has the same/similar indication for use, has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG/K221770](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG/K221770)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
