SR-1 Skin Treatment System

K221268 · AVAVA, Inc. · ONG · Nov 21, 2022 · General, Plastic Surgery

Device Facts

Record IDK221268
Device NameSR-1 Skin Treatment System
ApplicantAVAVA, Inc.
Product CodeONG · General, Plastic Surgery
Decision DateNov 21, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The MIRIA Skin Treatment System is indicated for use in dermatologic procedures requiring the coagulation of soft tissue, as well as for skin resurfacing procedures. The MIRIA Skin Treatment System is intended to be used by medical professionals and staff who are of lasers and who are familiar with the technology, operation of the system, and safety precautions. The MIRIA Skin Treatment System is a prescription device. Federal law restricts this device to sale by or on the order of a physician or any practitioner licensed by state law to use or order the use of this device.

Device Story

1550nm erbium glass diode-pumped fiber laser system; creates targeted spatially selective conical lesions (CTZs) in skin; distinguishes from cylindrical column injury lasers. Components: console (electronics, power, cooling), tablet (touchscreen GUI), patient interface (optics, scanner, aperture, sapphire contact cooling). Operator enters parameters on tablet; places interface on skin; initiates laser energy delivery. Used in professional settings by trained personnel. Contact cooling protects skin surface. Output: focused laser energy at depths 0-1500 micrometers. Benefits: enables treatment of all skin tones; allows extended energy range for resurfacing/coagulation.

Clinical Evidence

Bench testing only. Verification included: system focus accuracy (0-1500 micrometers), energy delivery (3-150 mJ), and a healing study demonstrating creation of conical microscopic treatment zones (CTZs) and post-treatment healing using worst-case parameter combinations. Software verification and validation performed per IEC 62304.

Technological Characteristics

1550nm erbium glass diode-pumped fiber laser; 20W output power; 150mJ max pulse energy; 12ms max pulse width. Features reusable sapphire cooled tip for continuous contact cooling. Software-controlled via touchscreen GUI. System architecture includes console, tablet, and patient interface. Complies with IEC 62304 for software life cycle.

Indications for Use

Indicated for dermatologic procedures requiring soft tissue coagulation and skin resurfacing. Intended for use by trained medical professionals and staff in a professional setting. Prescription only.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 21, 2022 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. AVAVA, Inc Jay Bhawalkar Chief Technology Officer 275 Second Avenue, Floor 3 Waltham, Massachusetts 02451 Re: K221268 Trade/Device Name: MIRIA Skin Treatment System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONG Dated: October 31, 2022 Received: October 31, 2022 Dear Jay Bhawalkar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Colin K. Chen -S for Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K221268 Device Name MIRIA Skin Treatment System #### Indications for Use (Describe) The MIRIA Skin Treatment System is indicated for use in dermatologic procedures requiring the coagulation of soft tissue, as well as for skin resurfacing procedures. The MIRIA Skin Treatment System is intended to be used by medical professionals and staff who are of lasers and who are familiar with the technology, operation of the system, and safety precautions. The MIRIA Skin Treatment System is a prescription device. Federal law restricts this device to sale by or on the order of a physician or any practitioner licensed by state law to use or order the use of this device. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary K221268 | General Provisions | | |--------------------|--| | 510(k) Owner's Name: | AVAVA, Inc. | |-----------------------------|-------------------------------------------------------------------------------------| | Address: | 275 Second Avenue, Floor 3<br>Waltham, MA 02451 | | Contact Person: | Jay Bhawalkar, PhD | | | Chief Technology Officer | | Phone Number: | Office: (617) 377-7945 | | Fax Number: | Not Applicable | | Classification Name: | Laser Surgical Instrument for Use in General and Plastic<br>Surgery and Dermatology | | Regulation: | 21 CFR § 878.4810 | | Regulatory Class: | II | | Product Code: | ONG | | Proprietary Name: | MIRIA Skin Treatment System | | Common Name: | Powered Laser Surgical Instrument<br>with Focal Point Technology | | Date Summary Prepared: | October 28, 2022 | | Primary Predicate Device(s) | | | SB-1 Skin Treatment System: | K202884 | #### Device Description The MIRIA Skin Treatment System falls under a class of intradermally focused lasers creating a targeted spatially selective conical lesion in the skin. This intradermal conical lesion profile distinguishes this class of lasers from those that generate a cylindrical column of injury. The ability of the beam to be focused to different depths is important for treatment of all skin tones and safety in the extended energy range in the MIRIA. The MIRIA is a 1550nm-based laser system that includes three (3) main components: Console, Tablet, and Patient Interface. The Console houses the system control electronics, power distribution, contact cooling, and laser. The Tablet is the primary user interface for controlling the system through a touch screen graphical user interface. The Patient Interface contains the focusing optics, scanner, laser aperture, and contact cooling interface. {4}------------------------------------------------ The MIRIA Skin Treatment System is a software-controlled device. The operator enters treatment parameters on the Tablet and places the Patient Interface on the treatment site. A treatment is initiated by the operator to cause laser energy to be projected into the skin of a patient. The device is intended to be used by suitably trained personnel in a professional setting. There are no sterile components. ### Indications for Use The MIRIA Skin Treatment System is indicated for use in dermatologic procedures requiring the coagulation of soft tissue, as well as for skin resurfacing procedures. The MIRIA Skin Treatment System is intended to be used by medical professionals and staff who are trained in the use of lasers and who are familiar with the technology, operation of the system, and safety precautions. The MIRIA Skin Treatment System is a prescription device. Federal law restricts this device to sale by or on the order of a physician or any practitioner licensed by state law to use or order the use of this device. ### Summary of Technological Similarities/Differences The intended use, technological characteristics, and operating principles of the subject device and the predicate device are the same with the exception of the pulse energy range (and the corresponding pulse width), as shown in the following table. This differences do not raise different questions of safety or efficacy for the MIRIA Skin Treatment System as shown in the performance testing results. | | Subject Device<br>(MIRIA Skin Treatment System) | Predicate Device<br>(SR-1 Skin Treatment System)<br>K202884 | |-------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------| | Operating principal | Scanned pulsed 1550nm laser<br>energy focused into the skin via an<br>operator-controlled delivery system. | Scanned pulsed 1550nm laser<br>energy focused into the skin via an<br>operator-controlled delivery system. | | Laser source | 1550nm erbium glass diode pumped<br>fiber laser | 1550nm erbium glass diode pumped<br>fiber laser | | Output power | 20W | 20W | | Maximum pulse<br>energy | 150 mJ | 70 mJ | | Maximum pulse<br>width | 12 ms | 10 ms | | Tissue contact | Reusable sapphire cooled tip | Reusable sapphire cooled tip | | Cooling mechanism | Continuous contact cooling | Continuous contact cooling | | Software control | Yes | Yes | | User interface | Touchscreen | Touchscreen | #### Table 8-1. Device Comparison {5}------------------------------------------------ #### Performance Testing The following verification methods were used to evaluate performance at the extended pulse energy range. - . Verification of the ability of the system to focus the energy to a depth of 0 to 1500 micrometers into the skin. - Verification of energy delivery from 3 to 150 mJ. - . A healing study to demonstrate device performance for the creation of conical microscopic treatment zones (CTZs) and to demonstrate healing post-treatment using worst-case parameter combinations. - . Software verification and validation for those elements impacted by the change in the values for the pulse energy in according to the FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (2005) and IEC 62304:2006 +A1:2015, Ed. 1.0 Medical device software - Software life cycle processes Note: The software permits only certain combinations of microbeam energy and focus depth to be selected by the user. #### Conclusion The MIRIA Skin Treatment System has the same intended use, technological characteristics, and principle of operation as the predicate device. The expanded pulse energy range raised no different issues of safety or efficacy as demonstrated the performance data. The data show that the MIRIA Skin Treatment System performs in accordance with its specifications and requirements for both safety and effectiveness and does not raise concerns about the safety and efficacy of the device in comparison to the predicate. Thus, the MIRIA Skin Treatment System and the predicate device are substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...