K130459 · Tria Beauty, Inc. · ONG · Oct 25, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130459
Device Name
TRIA FAN SYSTEM
Applicant
Tria Beauty, Inc.
Product Code
ONG · General, Plastic Surgery
Decision Date
Oct 25, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Tria FAN is indicated for the treatment of periorbital wrinkles, which may result in smoother appearing skin in the treated area.
Device Story
Tria FAN System is a semiconductor diode laser; delivers infrared light at 1450 nm ± 50 nm. Used for treatment of periorbital wrinkles. Device operates as a non-ablative fractional laser; induces tissue response similar to predicate non-ablative fractional devices. Intended for over-the-counter use. Provides cosmetic benefit by smoothing skin in treated area.
Clinical Evidence
Prospective, open-label study of 60 subjects evaluated safety and effectiveness for periorbital wrinkle treatment. Results demonstrated the device is as safe and effective as predicate non-ablative fractional devices. Non-clinical GLP-compliant histology study in hairless guinea pigs showed tissue response equivalent to predicate devices at immediate, 5-day, and 14-day post-treatment intervals.
Technological Characteristics
Semiconductor diode laser; infrared light output at 1450 nm ± 50 nm. Non-ablative fractional delivery mechanism. Class II medical device.
Indications for Use
Indicated for treatment of periorbital wrinkles to achieve smoother skin appearance.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
K222685 — NIRA Model 2 · Dermal Photonics Corp · Feb 8, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
K130459
## 510(k) SUMMARY
#### Tria Beauty's Tria FAN System
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Tria Beauty, Inc. 4160 Dublin Blvd, Ste 200 Dublin, CA 94568 Phone: 925-452-2500 Facsimile: 925-452-2598 Contact Person: Tobin Island, Ph.D. Date Prepared: September 12, 2013
#### Name of Device and Name/Address of Sponsor
Tria FAN System Tria Beauty, Inc. 4160 Dublin Blvd, Ste 200 Dublin, CA 94568
#### Common or Usual Name
OCT 2 5 2013
Diode Laser
## Classification Name
Laser Instrument, Surgical, Powered Regulation Number: 21 CFR§878.4810 Product Code: ONG
## Predicate Devices
Palomar LOI System (K090525) Cynosure Iluminage Diode Laser (K111454) Cynosure Affirm Laser Family (K080006; K101601) Solta Fraxel Re:Fine Laser (K091420; K060310; K063808; K062303; K070284; K050841; K042319; K040617; K031795)
#### Intended Use / Indications for Use
Tria FAN is indicated for the treatment of periorbital wrinkles, which may result in smoother appearing skin in the treated area.
### Technological Characteristics
The Tria FAN System is a semiconductor diode laser system that delivers infrared light at a wavelength of 1450 nm ± 50 nm.
{1}------------------------------------------------
## Performance Data
Performance data was submitted with this 510(k) notification to support the determination of substantial equivalence for the Tria FAN System relative to predicate devices.
### Performance Testing (Non-Clinical):
Performance testing was conducted to demonstrate that the Tria FAN System performs according to specifications and functions as intended.
#### Animal:
A GLP compliant histology study was conducted. Hairless guinea pigs were treated with the Tria FAN System and tissue sampling occurred immediately, 5 days, and 14 days post-treatment. The tissue response was found to be equivalent to the predicate non-ablative fractional devices.
### Clinical:
A 60-subject, prospective, open-label safety and effectiveness study was conducted to evaluate the Tria FAN System for the treatment of periorbital wrinkles. The Tria FAN System was found to be as safe and effective as the predicate non-ablative fractional devices.
#### Substantial Equivalence
The Tria FAN System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Tria FAN System and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Tria FAN System is as safe and effective as the predicate devices. Thus, the Tria FAN System is substantially equivalent.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Tria Beauty, Incorporated % Mr. Tobin Island, Ph.D. 4160 Dublin Boulevard. Suite 200 Dublin, California 94568
Re: K130459
Trade/Device Name: Tria Fan Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONG Dated: September 23, 2013 Received: September 23, 2013
Dear Mr. Island:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
Image /page/2/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.
October 25, 2013
{3}------------------------------------------------
Page 2 - Mr. Tobin Island, Ph.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mark Ni Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K130459
Device Name: Tria Fan
Indications For Use:
Tria FAN is indicated for the treatment of periorbital wrinkles, which may result in smoother appearing skin in the treated area.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
× Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Neil R Oggen
2013.10.25 113:17:42 -04'00' (Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number K130459
Page 1 of __ 1 __ 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.