← Product Code [ONF](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF) · K220385

# Intense Pulsed Light Equipment (K220385)

_Beijing Globalipl Development Co., Ltd. · ONF · Aug 12, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF/K220385

## Device Facts

- **Applicant:** Beijing Globalipl Development Co., Ltd.
- **Product Code:** [ONF](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF.md)
- **Decision Date:** Aug 12, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The US 500 Intense Pulsed Light Equipment is indicated for use in surgical and aesthetic applications in permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.

## Device Story

US 500 Intense Pulsed Light Equipment uses Xenon lamp to emit intense pulsed light for hair reduction; light filtered to specific wavelengths (640nm-1200nm). Device comprises console with touchscreen interface, water-cooled handpiece, and footswitch. Operated by clinicians in clinical settings. Clinician selects parameters via interface; light pulses delivered to skin via sapphire delivery system. Thermal effect targets hair follicles to achieve long-term, stable reduction in hair regrowth. Benefits patient through permanent hair reduction.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Xenon lamp light source; 640nm-1200nm wavelength; sapphire delivery system; 1-40J/cm² energy density; 1.0-9.9ms pulse width; water/air cooling. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, IEC 62471, ISO 10993-5, ISO 10993-10.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- IPL Therapy Machine ([K161286](/device/K161286.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

August 12, 2022

Beijing Globalipl Development Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China

Re: K220385

Trade/Device Name: Intense Pulsed Light Equipment Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: February 9, 2022 Received: February 10, 2022

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220385

Device Name US 500 Intense Pulsed Light Equipment

Indications for Use (Describe)

The US 500 Intense Pulsed Light Equipment is indicated for use in surgical and aesthetic applications in permanent hair removal. Permanent hair removal is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.

**Type of Use (Select one or both, as applicable):**

| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>             |

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

- 1. Date of Preparation:2022/08/12
- 2. Sponsor Identification

## Beijing Globalipl Development Co., Ltd.

F-8 Qunyinghui Building, Jinyuan Rd.32, Daxing Economic Development Zone, Beijing, China 102628

Contact Person: Liu Jun Position: QA Manager Tel: +86-15169727366 Fax: +86-10-60212336 Email:register@globalipl.com

- 3. Designated Submission Correspondent
Mr. Ray Wang

## Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China

Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com

{4}------------------------------------------------

#### 4. Identification of Proposed Device

Trade Name: Intense Pulsed Light Equipment

Common Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect

### Regulatory Information

Classification Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect Classification: II Product Code: ONF Regulation Number: 878.4810 Review Panel: General & Plastic Surgery

#### Indication For Use Statement:

The US 500 Intense Pulsed Light Equipment is indicated for use in surgical and aesthetic applications in permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.

#### Device Description:

The USS00 Intense Pulsed Light Equipment is a device that emits intense pulsed light in order to achieve hair reduction. The light source is a Xenon lamp that emits light which is filtered to include suitable wavelengths for hair reduction. The device is electrically powered, and main components include the console which includes electrical components and a touchscreen user interface, the IPL handpiece and cable, and a footswitch. The light pulses are emitted from the handpiece which is water cooled. The device is for prescription use only.

- న. Identification of Predicate Device(s)
510(k) Number: K161286 Product Name: IPL Therapy Machine Manufacturer: Beijing ADSS Development Co., Ltd

#### 6. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device conforms to the following standards:

- > IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance.
- A EC 60601-1-2:2014, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance Collateral Standard: Electromagnetic Compatibility-Requirements And Tests

{5}------------------------------------------------

- > IEC 60601-2-57:2011, Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- > IEC 62471:2006 Photobiological safety of lamps and lamp system
- > ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- > ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- 7. Clinical Test Conclusion

A clinical study was not considered to be needed for this 510(k).

- 8. Substantially Equivalent (SE) Comparison

| Item                         | Proposed Device                                                                                                                                                                                                                                                                                                                                            | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                         | Remark                                                                                 |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Device Name                  | Intense Pulsed Light<br>Equipment                                                                                                                                                                                                                                                                                                                          | IPL Therapy Machine                                                                                                                                                                                                                                                                                                                                                                                      | /                                                                                      |
| Classification<br>Regulation | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                            | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                          | SAME                                                                                   |
| Classification Panel         | General & Plastic<br>Surgery                                                                                                                                                                                                                                                                                                                               | General & Plastic<br>Surgery                                                                                                                                                                                                                                                                                                                                                                             | SAME                                                                                   |
| Class                        | II                                                                                                                                                                                                                                                                                                                                                         | II                                                                                                                                                                                                                                                                                                                                                                                                       | SAME                                                                                   |
| Product Code                 | ONF                                                                                                                                                                                                                                                                                                                                                        | ONF                                                                                                                                                                                                                                                                                                                                                                                                      | SAME                                                                                   |
| Common Name                  | Powered Light Based Non-Laser<br>Surgical Instrument With Thermal Effect                                                                                                                                                                                                                                                                                   | Powered Light Based Non-Laser Surgical Instrument With Thermal Effect                                                                                                                                                                                                                                                                                                                                    | SAME                                                                                   |
| Indication for use           | The US 500 Intense Pulsed<br>Light Equipment is indicated<br>for use in surgical and aesthetic<br>applications in permanent hair<br>reduction. Permanent hair<br>reduction is defined as the<br>long-term, stable reduction in<br>the number of hairs regrowing<br>when measured at 6, 9, and 12<br>months after the completion of<br>a treatment regimen. | The VE2000 device is indicated<br>for use in surgical, aesthetic<br>applications in permanent hair<br>reduction, reduction of benign<br>pigmented lesions and benign<br>vascular lesions.<br>Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6, 9,<br>and 12 months after the<br>completion of a treatment<br>regimen. | SIMILAR<br>The subset<br>indication<br>for use for<br>hair<br>reduction<br>is the same |

#### Table 6-1 General Comparison

{6}------------------------------------------------

| ITEM                     | Proposed Device                                   | Predicate Device                                                              | Remark  |
|--------------------------|---------------------------------------------------|-------------------------------------------------------------------------------|---------|
| Light source             | Intense pulsed light                              | Intense pulsed light                                                          | SAME    |
| Wavelength               | 640nm-1200nm,                                     | 480nm – 1200nm;<br>590nm – 1200nm;<br>640nm – 1200nm                          | SIMILAR |
| Deliver system           | Sapphire                                          | Sapphire                                                                      | SAME    |
| Energy density           | 1-40J/cm²                                         | 1-50J/cm²                                                                     | SIMILAR |
| Pulse Width              | 1.0-9.9ms, adjustable                             | 1-25ms                                                                        | SIMILAR |
| Pulse Interval           | 1 – 99 ms                                         | 1 – 50 ms                                                                     | SIMILAR |
| Light output<br>interval | 1 – 4 s                                           | 1 – 4 s                                                                       | SAME    |
| Pulse Number             | 1 - 6                                             | 1-6                                                                           | SAME    |
| Max. Power               | 2200VA                                            | 2200VA                                                                        | SAME    |
| Spot size                | 15 x 50mm                                         | 12 x 30 mm                                                                    | SIMILAR |
| Cooling mode             | close-cycle water cooling+ air<br>Contact cooling | Compressor closed circuit water<br>chiller with integrated heat<br>exchanger. | SIMILAR |
| Power supply             | AC 110/230V 50/60Hz                               | 110V, 50Hz                                                                    | SIMILAR |

Table 6-2 Performance Comparison

#### 9. Conclusion

The US500 Intense Pulse Light Equipment proposed device in this 510(k) uses the same IPL technology that is used in the predicate device K161286. Differences between the proposed device and predicate device do not raise new types of questions regarding safety and effectiveness, and performance testing demonstrates that the proposed device can be used safely and effectively for the proposed indications for use. The proposed US500 Intense Pulse Light Equipment device is considered to be substantially equivalent to the predicate K161286 device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF/K220385](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF/K220385)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
