← Product Code [ONF](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF) · K220214

# Painless IPL System - KDT750 (K220214)

_Shanghai Omni Laser Skinology Co., Ltd. · ONF · Feb 18, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF/K220214

## Device Facts

- **Applicant:** Shanghai Omni Laser Skinology Co., Ltd.
- **Product Code:** [ONF](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF.md)
- **Decision Date:** Feb 18, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The IPL System - KDT750 is intended for use in aesthetic applications requiring selective photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialties of general and plastic surgery, and dermatology.

## Device Story

The IPL System - KDT750 is a non-invasive, multi-wavelength pulsed light system used in dermatology and plastic surgery. It utilizes two interchangeable handpieces (SR and HR) to deliver light energy for selective photothermolysis of soft tissue. The system consists of a console, electronics, software, and a cooling system. It is operated by healthcare professionals in clinical settings. The device targets specific chromophores in the skin to treat vascular/pigmented lesions or to achieve permanent hair reduction. Clinicians select treatment parameters based on the patient's skin type and condition; the device delivers controlled pulses of light to the target area. The thermal effect of the light energy coagulates the target tissue, providing aesthetic benefits to the patient.

## Clinical Evidence

No clinical data was performed for this submission. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), non-laser light source safety (IEC 60601-2-57), and biocompatibility (ISO 10993-5, ISO 10993-10).

## Technological Characteristics

Multi-wavelength non-laser pulsed light system. Employs AFT (Advanced Fluorescence Technology) and EDF (Equal Distribution Fluence). Wavelengths: 530-950nm (SR) and 640-950nm (HR). Energy density: 1-12 J/cm² (SR) and 1-7 J/cm² (HR). Pulse repetition rate: 2-3 Hz. Cooling system integrated. Software-controlled. Biocompatibility per ISO 10993. Electrical safety per IEC 60601 series.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Alma Harmony XL Multi-Application Platform ([K072564](/device/K072564.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 18, 2023

Shanghai Omni Laser Skinology Co., Ltd. % Helen Nan General Manager New Risen Enterprise Management Consulting Co., Ltd. Room 302, Building 3, Hangqian Mansion, Hangqian Street, Lucheng District Wenzhou, Zhejiang 325000 China

Re: K220214

Trade/Device Name: IPL System - KDT750 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: January 19, 2023 Received: January 19, 2023

Dear Helen Nan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220214

Device Name IPL System - KDT750

#### Indications for Use (Describe)

The IPL System - KDT750 is intended for use in aesthetic applications requiring selective photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialties of general and plastic surgery, and dermatology.

Indications for use for IPL System with 515-980nm IPL Handpiece (SR Handpiece).

- · Benign pigmented epidermal and cutaneous lesions including warts, melasma, epithelides (freckles) and lentigines.
- · Benign cutaneous vascular lesions, including port wine stains, facial and truncal telangiectasias, rosacea, erythema of rosacea and poikiloderma of Civatte.
- · For use on skin types(I-V)

Indications for use for IPL System with 640-950nm IPL Handpiece (HR Handpiece).

- · The removal of unwanted hair to effect stable long-term or permanent hair reduction.
- · For use on Fitzpatrick skin types (I-V)

* Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

| Type of Use (Select one or both, as applicable)                                  |
|----------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)             |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a green shamrock with three leaves. A blurred, light purple arc surrounds the shamrock, giving the image a sense of motion or a halo effect. The background is white, which makes the green and purple colors stand out.

# K220214 510(k) Summary (As required by 21 CFR 807.92)

#### 1.0 Submitter Information

| Company: | Shanghai Omni Laser Skinology Co., Ltd.                                                     |
|----------|---------------------------------------------------------------------------------------------|
| Address  | Floor 3, Building 3, NO. 227, Mingqiang Road<br>Songjiang District, Shanghai, 201612, CHINA |
| Phone:   | +86-021-54847192                                                                            |
| E-mail:  | avril@omni-laser.com                                                                        |
| Contact  | Avril Ouyang                                                                                |
| Title:   | General Manager<br>• Date of Preparation: Feb. 17, 2023                                     |

#### 2.0 Device Information

| Trade/Device Name:      | IPL System                                                                              |
|-------------------------|-----------------------------------------------------------------------------------------|
| Model:                  | KDT750                                                                                  |
| Regulation Description: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology. |
| Device:                 | Powered Light Based Non-Laser Surgical Instrument<br>With Thermal Effect                |
| Review Panel:           | General & Plastic Surgery                                                               |
| Product Code:           | ONF                                                                                     |
| Submission Type:        | Traditional 510(k)                                                                      |
| Regulation Number:      | CFR 878.4810                                                                            |
| Device Class:           | Class II                                                                                |

#### 3.0 Predicate Device Information

| Trade/Device Name: | Alma Harmony XL Multi-Application Platform |
|--------------------|--------------------------------------------|
| 510k Number:       | K072564                                    |
| Submitter:         | Alma Lasers, Ltd.                          |

#### 4.0 Device Description

The IPL System - KDT750 is a multi-wavelength non-invasive system for IPL skin treatments, treatment of vascular and pigmented lesions, and hair removal using two wavelength ranges: 530-950nm and 640-950nm. The IPL System consists of a system console, electronics and software, cooling system, and two hand pieces with different wavelength.

#### 5.0 Indications for Use

The IPL System - KDT750 is intended for use in aesthetic applications requiring selective photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialties of general and plastic surgery, and dermatology.

#### Indication for use for the 530-950nm wavelength handpiece (SR Handpiece)

- · Benign pigmented epidermal and cutaneous lesions including warts, melasma, epithelides (freckles) and lentigines.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a green shamrock with three leaves. A blurred, light purple arc surrounds the shamrock, creating a circular effect. The shamrock is the focal point, with the arc adding a subtle, decorative element to the design.

- · Benign cutaneous vascular lesions, including port wine stains, facial and truncal telangiectasias, rosacea, erythema of rosacea and poikiloderma of Civatte.
- For use on skin types(I-V).

## Indication for use for the 640-950nm wavelength handpiece (HR Handpiece)

- · The removal of unwanted hair to effect stable long-term or permanent hair reduction.
- For use on Fitzpatrick skin types (I-V).
- * Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a green three-leaf clover with a light purple arc behind it. The clover is in the center of the image and has a slight shadow effect. The purple arc is positioned behind the clover, creating a sense of depth. The background is white.

### 6.0 Comparison of Technological Characteristics with the Predicate Device

Table 1 - Comparison table of the Subject device (SR Handpiece) and Predicate device (SSR Module AFT Handpiece)

| Device Feature           | IPL System-<br>SR Handpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HarmonyXL-SSR Module AFT<br>Handpieces                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Comparison |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Wavelength               | 530-950 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 540-950 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Note 1     |
| Indications for use      | The IPL System - KDT750 is intended for use in aesthetic applications requiring selective photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialties of general and plastic surgery, and dermatology.  Benign pigmented epidermal and cutaneous lesions including warts, melasma, epithelides (freckles) and lentigines. Benign cutaneous vascular lesions, including port wine stains, facial and truncal telangiectasias, rosacea, erythema of rosacea and poikiloderma of Civatte. For use on skin types (I-V). | Intended for use in aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery, and dermatology  The treatment of moderate inflammatory acne vulgaris. The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma and ephelides (freckles), lentigines, nevi, melasma, and café-au-lait macules. The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations. Use on all skin types (Fitzpatrick I-VI). | Note 2     |
| Light Source             | Pulsed light with AFT and EDF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Pulsed light with AFT and EDF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same       |
| Timers                   | 1, 3, 12 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 1, 3, 30 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Note 3     |
| Energy Density (Fluence) | 1-12 J/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 1-15 J/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same       |
| Spot Size                | 8*40 mm (3.2 cm2)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 3 cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |
| Pulse Repetition Rate    | 2 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 2 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same       |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a green shamrock inside of a white circle. A blurred purple arc surrounds the top and left side of the circle. The shamrock is a three-leaf clover, a symbol often associated with Ireland and Saint Patrick's Day. The overall design is simple and clean, with the green and purple colors providing a soft contrast.

Table 2 - Comparison table of the Subject device (HR Handpiece) and Predicate device (HR Module AFT Handpiece, SHR Module Handpieces)

| Device Feature              | IPL System-<br>HR Handpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | HarmonyXL-HR Module AFT<br>Handpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | HarmonyXL-SHR Module<br>Handpieces                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comparison |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Wavelength                  | 640-950 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 650-950 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 780-950 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Note 1     |
| Indication for use          | The IPL System - KDT750 is<br>intended for use in aesthetic<br>applications requiring<br>selective photothermolysis<br>(photocoagulation or<br>coagulation) of soft tissue in<br>the medical specialties of<br>general and plastic surgery,<br>and dermatology.<br>The removal of unwanted<br>hair to effect stable long-<br>term or permanent hair<br>reduction. For use on Fitzpatrick skin<br>types (I-V).              * Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6. | Intended for use in aesthetic and<br>cosmetic applications requiring<br>selective photothermolysis<br>(photocoagulation or<br>coagulation) of soft tissue in the<br>medical specialties of general<br>and plastic surgery, and<br>dermatology  The treatment of moderate<br>inflammatory acnevulgaris. The treatment of benign<br>pigmented epidermal<br>lesions, including<br>dyschromia,<br>hyperpigmentation,<br>melasma and<br>ephelides(freckles). The treatment of face and<br>body vascular and                                    | Intended for use in aesthetic and<br>cosmetic applications requiring<br>selective photothermolysis<br>(photocoagulation or<br>coagulation) and hemostasis of<br>soft tissue in the medical<br>specialties of general and plastic<br>surgery, and dermatology  The treatment of moderate<br>inflammatory acne<br>vulgaris. The treatment of benign<br>pigmented epidermal<br>lesions including<br>dyschromia,<br>hyperpigmentation,<br>melasma, and ephelides<br>(freckles). The treatment of cutaneous                                                                                               | Note 2     |
|                             | 9, and 12 months after the<br>completion of a treatment<br>regime.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | • pigmentedlesions.<br>• The treatment of cutaneous<br>lesions, including scars and<br>striae.<br>• The treatment of benign<br>cutaneous vascular lesions,<br>including port wine stains,<br>hemangiomas, facial,<br>truncal and leg<br>telangiectasias, rosacea,<br>melasma, angiomas and<br>spider angiomas,<br>poikiloderma of Civatte,<br>leg veins and venous<br>malformations.<br>• The removal of unwanted<br>hair to effect stable long-<br>term or permanent hair<br>reduction.<br>• Use on all skin types<br>(Fitzpatrick I-VI) | • lesions including warts,<br>scars and striae.<br>• The treatment of benign<br>cutaneous vascular lesions<br>including port wine stains,<br>hemangiomas, facial,<br>truncal and leg<br>telangiectasias, rosacea,<br>erythema. of rosacea,<br>angiomas and spider<br>angiomas, poikilodermna<br>of Civatte, leg veins and<br>venous malformnations.<br>• The treatment of<br>pseudofolliculitis barbae<br>(PFB).<br>• The removal of unwanted<br>hair and to effect stable<br>long-term or permnanent<br>hair reduction.<br>• Use on all skin types<br>(Fitzpatrick I-VI),<br>including tanned skin. |            |
| Light Source                | Pulsed light with AFT and EDF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Pulsed light with AFT and EDF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Near Infrared pulsed light                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same       |
| Timers/Pulse width          | 1,3 and 30 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 30, 40, 50 msec                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 1, 3 and 30 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same       |
| Energy Density<br>(Fluence) | 1-7 J/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 5-25 J/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 1-7 J/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Note 4     |
| Spot Size                   | 15*50 mm (7.5 cm²)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 6.4 cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 3 cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Note 5     |
| Pulse Repetition<br>Rate    | 3 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 2/3 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 3 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same       |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a green shamrock with three leaves. A light purple arc surrounds the shamrock, adding a sense of motion or emphasis to the design. The overall impression is a simple, clean logo or emblem, possibly related to Irish culture or St. Patrick's Day.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a green shamrock with three leaves, which is a symbol often associated with Ireland and Saint Patrick's Day. The shamrock is encircled by a curved, light purple shape, possibly suggesting movement or a stylized frame. The background is plain white, which makes the green and purple colors stand out.

#### Difference Analysis:

#### Note 1 Wavelength

The wavelength range of proposed devices is very closed to predicate difference is considered to have no negative effect on effectiveness and safety, and the bench tests conducted on the proposed device substantial equivalence to the predicate device.

#### Note 1 Indications for Use

Though the Harmony's has more indications than the KDT750's, the intended use of subject device is covered by the predicate device, which forms the foundation of their comparison, which will not result in negative safety and effectiveness, this forms the foundation of Substantial Equivalence.

#### Note 3: Timers

The proposed device (K220214, SR Handpiece) has a similar timer range as the predicate device (K072564, SSR Handpiece). The slight difference in settings do not result in negative safety and effectiveness. The bench tests conducted on the proposed device support substantial equivalence to the predicate device

#### Note 4: Energy Density (Fluence)

The proposed device has an energy density that is similar to the differences do not result in negative safety and efficacy effects. The bench tests conducted on the proposed device support substantial equivalence to the predicate device.

#### Note 5: Spot Size

The proposed device has a larger spot size with the predicate device. However the differences in spot sizes are not considered to produce a negative effect on safety and safety. The bench tests conducted on the proposed device substantial equivalence to the predicate device.

The differences noted between IPL System - KDT750 and the predicate device, Alma Harmony™ Multi-Application Platform (K072564), do not present any new or different questions related to safety and effectiveness.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows a green shamrock with three leaves. The shamrock is positioned in the center of the image and is surrounded by a light purple, curved shape that partially encircles it. The background is white, which makes the shamrock and the purple shape stand out.

#### 7.0 Discussion of Tests Performed

#### 7.1 Clinical Tests

Clinical testing was not performed for the subject device as part of the submission.

#### 7.2 Non-Clinical Tests

The following performance data are provided in support of the substantial equivalence determination:

- > Electrical Safety and Electromagnetic Compatibility (EMC)
#### IEC60601-1:2005+CORR.1:2006+CORR.2:2007+A1:2012

Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance;

IEC 60601-1-2:2014 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility

IEC 60825-1:2007, Safety of laser products - Part 1: Equipment classification and requirements.

IEC 60601-2-57:2011, Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

#### A Biocompatibility

ISO 10993-5:2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity.

ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

The conclusion from the testing is the device is safe and effective for its intended use, and performs as well or better than the predicate devices.

#### 8.0 Software

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

#### 9.0 Conclusion

The IPL System-KDT750 is as safe and effective as its predicate devices.

The IPL System-KDT750 has the same intended use and same technological characteristics and specifications as its predicate devices. Thus, the IPL System is substantially equivalent to its predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF/K220214](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF/K220214)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
