K241522 · Guard Medical, Inc. · OKO · Aug 15, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241522
Device Name
NPseal (20 and 25)
Applicant
Guard Medical, Inc.
Product Code
OKO · General, Plastic Surgery
Decision Date
Aug 15, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4683
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NPseal is indicated for patients who would benefit from wound management via application of negative pressure, particularly as the device may promote wound healing through the removal of small amounts of exudates from closed surgical incisions.
Device Story
NPseal is a single-use, non-powered negative pressure wound therapy (NPWT) system; integrates a mechanical pump with a multilayer dressing. Input: negative pressure generated by mechanical pump; output: sustained negative pressure (-75 to -125 mmHg) applied to closed surgical incisions. Dressing consists of hydrophilic foam pad for exudate absorption and breathable film for evaporation. Used in clinical or home settings; applied by healthcare providers. Device manages exudate via absorption and evaporation; promotes wound healing. No external power source or electronic components; purely mechanical operation.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by bench testing, including pressure and exudate handling, pull testing, and packaging validation conducted on the largest device size (NPseal 25).
Technological Characteristics
Non-powered, integrated mechanical pump; negative pressure range -75 to -125 mmHg (± 17.5 mmHg). Materials: Polyurethane film with acrylic adhesive, hydrophilic polyurethane foam pad, thermoplastic elastomer pump. Single-use; sterile (gamma irradiation). Biocompatibility per ISO 10993-1. Dimensions vary by model (20 cm or 25 cm incision capacity).
Indications for Use
Indicated for patients requiring wound management via negative pressure to promote healing of closed surgical incisions by removing small amounts of exudates.
Regulatory Classification
Identification
A non-powered suction apparatus device intended for negative pressure wound therapy is a device that is indicated for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. It is further indicated for management of wounds, burns, flaps, and grafts.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT).” See § 878.1(e) for the availability of this guidance document.
K200305 — NPseal · Guard Medical, Inc. · May 26, 2020
K212971 — NPseal · Guard Medical, Inc. · Feb 10, 2022
K211130 — NPseal · Guard Medical, Inc. · May 14, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 15, 2024
Guard Medical, Inc. % Eric Bannon Vice President of Regulatory and Clinical Affairs AlvaMed, Inc. 935 Great Plain Avenue Suite 166 Needham, Massachusetts 02492
Re: K241522
Trade/Device Name: NPseal (20 and 25) Regulation Number: 21 CFR 878.4683 Regulation Name: Non-Powered Suction Apparatus Device Intended For Negative Pressure Wound Therapy Regulatory Class: Class II Product Code: OKO Dated: May 29, 2024 Received: July 18, 2024
Dear Eric Bannon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Yu-chieh Chiu -S
Yu-Chieh Chiu, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241522
Device Name
NPseal (20 and 25)
Indications for Use (Describe)
The NPseal is indicated for patients who would benefit from wound management via application of negative pressure, particularly as the device may promote wound healing through the removal of small amounts of exudates from closed surgical incisions.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size:20px;"> </span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="font-size:20px;"> </span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### SPECIAL 510(K) SUMMARY
This Special 510(k) for the NPseal is submitted based on the FDA Guidance document "The Special 510(k) Program: Guidance for Industry and Food and Drug Administration Staff" (issued September 13, 2019).
- 1. Name and Address of Sponsor
Guard Medical, Inc. 1221 Brickell Avenue, Suite 900 Miami, FL 33131 USA Phone : +1 (888) 417-3644 Machiel van der Leest CEO m.vanderleest@guard-medical.com
- 2. Correspondent/Primary Contact Person
Eric Bannon Vice President of Regulatory and Clinical Affairs
AlvaMed, Inc. (consultant to Guard Medical) 935 Great Plain Avenue, Unit 166 Needham, MA 02492 USA ebannon@alvamed.com Phone: +1 (781) 710-8243 Fax: +1 (617) 249-0955
- 3. Submission Information
| Date Summary Prepared: | August 15, 2024 |
|------------------------|------------------------------------------------------------------|
| Name of Device: | NPseal |
| Common or Usual Name: | Negative Pressure Wound Therapy Non-Powered<br>Suction Apparatus |
| Classification: | Class II |
| Product Code: | OKO (21 CFR 878.4683) |
| Predicate Device: | NPseal (K240244) |
- 4. Device Description
The NPseal Negative Pressure Advanced System is a single-use device that includes an integrated, mechanical pump system. The NPseal maintains Negative Pressure Wound Therapy (NPWT) in the -75 mmHg to -125 mmHg nominal range.
{5}------------------------------------------------
The NPseal is intended for 7 days of use. Therapy duration of the system may be less than indicated if clinical practice or other factors such as time to wound closure and rate of exudate, result in earlier removal or need for system change.
The NPseal 20 is intended for surgically closed incisions up to 20 cm x 0.5 cm.
The NPseal 25 is intended for surgically closed incisions up to 25 cm x 0.5 cm.
- 5. Indications for Use
The NPseal is indicated for patients who would benefit from wound management via application of negative pressure, particularly as the device may promote wound healing through the removal of small amounts of exudates from closed surgical incisions.
- 6. Comparison of Manufacturer's Cleared Device and Modified Device
| | Subject<br>NPseal 20, NPseal 25 | Predicate Device<br>NPseal 5, NPseal 10, NPseal 15 |
|---------------------------|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K241522 | K240244 |
| 510k Submitter/Holder | Same as predicate | Guard Medical |
| Product Code | Same as predicate | OKO |
| Regulation No. | Same as predicate | 878.4683 |
| Regulation Description | Same as predicate | Non-Powered suction apparatus device<br>intended for negative pressure wound<br>therapy. |
| Common<br>Name | Same as predicate | Negative Pressure Wound Therapy non-<br>powered suction apparatus |
| | Subject<br>NPseal 20, NPseal 25 | Predicate Device<br>NPseal 5, NPseal 10, NPseal 15 |
| Indications for Use | Same as predicate | The NPseal is indicated for patients who<br>would benefit from wound management<br>via application of a negative pressure,<br>particularly as the device may promote<br>wound healing through the removal of<br>small amounts of exudates from closed<br>surgical incisions. |
| Wound types | Closed surgical incisions,<br>NPseal 20: Up to 20 cm x 0.5 cm<br>NPseal 25: Up to 25 cm x 0.5 cm | Closed surgical incisions,<br>NPseal 5: Up to 5 cm x 0.5 cm<br>NPseal 10: Up to 10 cm x 0.5 cm<br>NPseal 15: Up to 15 cm x 0.5 cm |
| Single Use | Same as predicate | Yes |
| Negative Pressure Range | Same as predicate | -75 to -125 mmHg (± 17.5 mmHg) |
| Device Technology | Same as predicate | Nonpowered, integrated mechanical<br>pump to generate negative pressure.<br>Multilayer pad composed of hydrophilic<br>foam with high fluid absorbency and top<br>breathable film designed to collect and<br>move exudate away from the wound bed. |
| Management of<br>Exudates | Same as predicate | Managed by the dressing itself - via<br>combination of absorption into the foam<br>pad and evaporation through the<br>breathable upper film. |
| Materials | Same as predicate | Film: Polyurethane coated with adhesive<br>acrylic,<br>Pad: Hydrophilic polyurethane,<br>Pump: Thermoplastic elastomer |
| Wear Time | Same as predicate | Up to 7 days |
| Sterility | Same as predicate | Sterile - Gamma irradiation |
| Biocompatibility | Same as predicate | Complies with ISO 10993-1 |
Table 1. Comparison of Modified Device to Cleared Device
{6}------------------------------------------------
K241522 Page 3 of 4
- 7. Summary of Modification
The dimensions of the current NPseal devices cleared initially in K212971 and most recently in K240244 are:
{7}------------------------------------------------
- NPseal 5 (Small) is 11 x 8 cm with a 7.5 x 4.5 cm foam pad for use on closed surgical incisions up to 5 cm in length and 0.5 cm in width.
- NPseal 10 (Medium) is 16 x 8 cm with a 12.7 x 4.7 cm foam pad for use on closed surgical incisions up to 10 cm in length and 0.5 cm in width.
- -NPseal 15 (Large) is 21 x 8 cm with a 17.7 x 4.7 cm foam pad for use on closed surgical incisions up to 15 cm in length and 0.5 cm in width.
The subject devices are two longer versions:
- -NPseal 20 is 26 x 8 cm with a 22.7 x 4.7 cm foam pad for use on closed surgical incisions up to 20cm in length and 0.5 cm in width.
- -NPseal 25 is 31 x 8 cm with a 27.7 x 4.7 cm foam pad for use on closed surgical incisions up to 25cm in length and 0.5 cm in width.
All other dimensions including the foam thickness, as well as the pump system and device materials remain identical to the NPseal Small predicate.
- 8. Summary of Functional and Performance Testing
The NPseal dressing length increase for the NPseal 25 was assessed based on risk and the Design Controls in 21 CFR 820.
The following verification and validation testing was performed exclusively on the NPseal 25 as it represents the most challenging device size:
- Pressure and Exudate Handling Over Time -
- -Pull testing
- -Benchtop Usability Testing
- -Packaging Validation
- 9. Summary of Clinical Testing
No clinical testing was applicable to this submission.
- 10. Conclusion
Like the three sizes (NPseal 5, NPseal 10, and NPseal 15) of the predicate device cleared by the FDA in K212971 and updated in K240244, the longer NPseal 20 and NPseal 25 have been shown to be safe through bench testing. Size increases made to the device do not raise any unanswered questions of safety or effectiveness. Verification and validation data support substantial equivalence of the NPseal 25 to the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.