IPL Hair Removal Device (I6 S1, I6 D1, I6 M1)

K252618 · Alovea Healthcare Co. , Ltd. · OHT · Nov 14, 2025 · General, Plastic Surgery

Device Facts

Record IDK252618
Device NameIPL Hair Removal Device (I6 S1, I6 D1, I6 M1)
ApplicantAlovea Healthcare Co. , Ltd.
Product CodeOHT · General, Plastic Surgery
Decision DateNov 14, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair

Device Story

IPL Hair Removal Device (models I6 S1, I6 D1, I6 M1) uses intense pulsed light (IPL) from a xenon arc lamp to remove unwanted body/facial hair. Device operates below skin surface without cutting/pulling. Features include skin contact sensor to prevent light emission unless fully contacted, and sapphire ice cooling function to manage treatment area temperature. User-operated via finger switch or auto-flash mode. Powered by external adapter. Output intensity levels (1-3) and pulse modes (single, dual, or switchable) vary by model. Healthcare provider/user observes hair removal results over time. Benefits include hair removal with minimal pain.

Clinical Evidence

Bench testing only. Includes biocompatibility (ISO 10993-5, -10, -23), electrical safety/EMC (IEC 60601-1, -1-2, -1-11, -2-83), eye safety (IEC 62471), and software verification/validation.

Technological Characteristics

Materials: ABS (host), AL2O3 (sapphire cold head). Energy: Xenon arc lamp (IPL). Wavelength: 600-1200nm. Spot size: 3.3cm². Power: 100-240V AC. Connectivity: Standalone. Sterilization: Not required. Software: Microprocessor-controlled, basic level of concern.

Indications for Use

Indicated for removal of unwanted body and/or facial hair in over-the-counter home-use settings.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION November 14, 2025 Alovea Healthcare Co. Ltd. % Moy Meng RA Specialist Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Qianhai Road Shenzhen, Guangdong 518052 China Re: K252618 Trade/Device Name: IPL Hair Removal Device (I6 S1, I6 D1, I6 M1) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: August 19, 2025 Received: August 19, 2025 Dear Moy Meng: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252618 - Moy Meng Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K252618 - Moy Meng Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Tanisha Hithe Digitally signed by Tanisha Hithe Date: 2025.11.14 14:54:34 -05'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K252618 | | | Device Name IPL Hair Removal Device (I6 S1, I6 D1, I6 M1) | | | Indications for Use (Describe) The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair | | | Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Alovea Healthcare Co. Ltd. 510(k)s -510(k) Summary # K252618 - 510(k) Summary "510(k) Summary" as required by 21 CFR Part 807.92. Date prepared: 2025-8-19 ## I. Submitter Alovea Healthcare Co. Ltd. Room 703, Building 1, No. 188 Fu Chun Jiang Rd. New District, Suzhou, 215163, P.R. China Bingying Hou Vice General Manager Tel: +86 13811391439 Email: houbingying@ka-care.com ## II. Device Information Name of Device: IPL Hair Removal Device Model(s): I6 S1, I6 D1, I6 M1 Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulation Number: 21 CRF 878.4810 Product Code: OHT Review Panel: General & Plastic Surgery ## III. Predicate Device and Reference Device ### Predicate device 1: | Manufacturer | Predicate Device | 510(k) Number | Cleared Date | | --- | --- | --- | --- | | Shenzhen Yang Wo Electronic Co., Ltd | IPL Hair Removal Device Model: BE965A, BE965B, BE965C | K250171 | March 19, 2025 | ### Predicate device 2: | Manufacturer | Predicate Device | 510(k) Number | Cleared Date | | --- | --- | --- | --- | | Shenzhen Ulike Smart Electronics Co.,Ltd. | Ice Cooling IPL Hair Removal Device Model: UI20 DB, UI20 RE, UI20 GP, UI20 PW, UI20S DB, UI20S RE, UI20S PW, UI20S GP,UI20 WH, UI20 GR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20WG, UI20S WH, UI20S GR, UI20S BS, UI20S MP | K241998 | December 13, 2024 | ### Reference device: | Manufacturer | Reference Device | 510(k) Number | Cleared Date | | --- | --- | --- | --- | Page 1 of 7 {5} Alovea Healthcare Co. Ltd. 510(k)s -510(k) Summary # IV. Device Description IPL Hair Removal Device, is an over-the-counter, home-use device for unwanted hair removing by using Intense Pulsed Light (IPL), and it has been designed three models with the same IPL technology for hair removal, which is model I6 S1, I6 D1 and I6 M1. The device works below the skin's surface and does not involve any cutting or pulling, removing hair growth with minimal pain. Model I6 S1 adopts single pulse technology, I6 D1 adopts dual pulse technology, and I6 M1 can switch between single pulse and dual pulse. The device is only powered by the external power adapter and its IPL emission activation is by finger switch or auto flash mode. The IPL Hair Removal Device has an irreplaceable light exit and it can cover an area of $3.3\mathrm{cm}^2$ that is suitable for multiple hair removal areas, such as limbs, armpits, upper lip, back, chest, bikini area. The device contains a skin sensor to detect appropriate skin contact, if the flash window is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the IPL Hair Removal Device has the sapphire ice cooling function, which will be activated throughout the whole hair removal process to cool down the treatment area's temperature and provide the user with a better using experience. # V. Indications for Use The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair. # VI. Materials | Component name | Material of Component | Body Contact Category | Contact Duration | | --- | --- | --- | --- | | Host of machine (including air outlet, treatment window, air inlet, buttons) | ABS | Surface-contacting device: Intact skin | Less than 24 hours | | Sapphire cold head (within treatment window) | AL2O3 | Surface-contacting device: Intact skin | Less than 24 hours | | Sapphire cold head (with in treatment window) | AL2O3 | Surface-contacting device: Intact skin | Less than 24 hours | {6} Alovea Healthcare Co. Ltd. 510(k)s -510(k) Summary # VII. Comparison of Technological Characteristics With the Predicate Device The IPL Hair Removal Device has the same intended use as the predicate. The technological characteristics such as wavelength, energy density, spot size and pulse duration, are similar to the predicate devices and reference device. Any minor differences between the subject device and the listed predicate devices and reference device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate devices and reference devices. | Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Reference Device | Remark | | --- | --- | --- | --- | --- | --- | | 510(k) Number | Pending | K250171 | K241998 | K250194 | / | | Trade name | IPL Hair Removal Device | IPL Hair Removal Device | Ice Cooling IPL Hair Removal Device | Ice Cooling IPL Hair Removal Device | / | | Model | I6 S1, I6 D1, I6 M1 | BE965A, BE965B, BE965C | UI20 DB, UI20 RE, UI20 GP, UI20 PW, UI20S DB, UI20S RE, UI20S PW, UI20S GP,UI20 WH, UI20 GR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20WG, UI20S WH, UI20S GR, UI20S BS, UI20S MP | UI06S PR, UI06S PN, UI06S WH | / | | Product code | OHT | OHT | OHT | OHT | Same | | Device classification | Class II | Class II | Class II | Class II | Same | | Indication for use/ Intended use | The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair. | The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair. | Ice Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at | Ice Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing | Same as predicate device 1 | Page 3 of 7 {7} Alovea Healthcare Co. Ltd. 510(k)s -510(k) Summary | Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Reference Device | Remark | | --- | --- | --- | --- | --- | --- | | 510(k) Number | Pending | K250171 | K241998 | K250194 | / | | | | | 6, 9 and 12 months after the completion of a treatment regime. | when measured at 6, 9 and 12 months after the completion of a treatment regime. | | | Prescription or OTC | OTC | OTC | OTC | OTC | Same | | Power supply | 100-240V~, 50Hz/60Hz | 100-240V, 50/60Hz | 100-240V~, 50/60Hz | 100-240V~, 50Hz/60Hz | Same | | Dimension | 182x60x32mm | 181*73*62mm | 206.73mm*68.68mm*54.29mm | 179.0*58.2*37.2mm | Different | | Sterilization | Not required | Not required | Not required | Not required | Same | | Light source | Intense Pulsed Light | Intense Pulsed Light (IPL) | Intense Pulsed Light (IPL) | Intense Pulsed Light | Same | | Energy medium | Xenon Arc lamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Same | | Wavelength range | 600-1200nm | 600nm-1200nm | 550-1200nm | 550nm~1200nm | Same as predicate device 1 | | Energy density | I6 S1: 2.25J/cm²~5.625J/cm² I6 D1: 2.75J/cm²~6.56J/cm² I6 M1: (single pulse) 2.25J/cm²~5.625J/cm², (dual pulse) 2.75J/cm²~6.56J/cm² | Max. 6.56J/cm² (range in 2.75~6.56J/cm²) | 1.67~6.67J/cm² | 2.42-7.27J/cm² | Similar | | Spot size | 3.3cm² | 3.2 cm² | 3.9 cm² | 3.3cm² | Same as reference | | Pulse duration | I6 S1 (single pulse): 0.7ms~5ms I6 D1 (dual pulse): 0.92ms~6.02ms I6 M1: (single pulse) 0.7ms~5ms/(dual pulse): 0.92ms~6.02ms | 0.82ms~5.32ms | 0.88~3.20ms Multipulse | 1.82-8.07ms multipulse | Similar | | Pulsing control | Finger switch | Finger switch | Finger switch | Finger switch | Same | Page 4 of 7 {8} Alovea Healthcare Co. Ltd. 510(k)s -510(k) Summary | Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Reference Device | Remark | | --- | --- | --- | --- | --- | --- | | 510(k) Number | Pending | K250171 | K241998 | K250194 | / | | Output intensity level | Level 1~3 | 3 levels | 1~10 levels | 4 levels | Same as predicate device 1 | | Software/Firmware/Microprocessor Control? | Yes | Yes | Yes | Yes | Same | | Electrical safety | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 60601-2-83 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 60601-2-83 | Same | | Eye safety | IEC 62471 | IEC 62471 | IEC 62471 | IEC 62471 | Same | | Biocompatibility | ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-5 ISO 10993-10 ISO 10993-23 | Same | Page 5 of 7 {9} Alovea Healthcare Co. Ltd. 510(k)s -510(k) Summary ## VIII. Performance Data The following performance data were provided in support of the substantial equivalence determination. ### 1) Biocompatibility Evaluation The biocompatibility evaluation for the body-contacting components of the IPL Hair Removal Device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process'", Document Issued on September 4, 2020, as recommended by FDA. - ISO 10993-5: 2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity - ISO 10993-10: 2021, Biological evaluation of medical devices – Part 10: Tests for skin sensitization - ISO 10993-23: 2021, Biological evaluation of medical devices – Part 23: Tests for irritation ### 2) Electrical Safety and EMC Electrical safety and EMC testing was performed to, and passed, as per the following standards: - IEC 60601-1-2: 2020, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: electromagnetic compatibility. - IEC 60601-1: 2020, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. - IEC 60601-1-11: 2020, Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. - IEC 60601-2-83: 2022, Medical electrical equipment – Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment. ### 3) Eye Safety - IEC 62471:2006, Photobiological safety of lamps and lamp systems ### 4) Software Verification and Validation Software documentation consistent with **Basic Documentation** of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. Page 6 of 7 {10} Alovea Healthcare Co. Ltd. 510(k)s –510(k) Summary ## IX. Conclusions Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate devices. Page 7 of 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...