K232932 · Shenzhen Greatro Electronic Technology Co., Ltd. · OHT · Dec 4, 2023 · General, Plastic Surgery
Device Facts
Record ID
K232932
Device Name
Intense Pulse Light Therapeutic Apparatus
Applicant
Shenzhen Greatro Electronic Technology Co., Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Dec 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intense Pulse Light Therapeutic Apparatus is an over-the-counter device intended for removal of unwanted body hair.
Device Story
Personal, light-based hair reduction device for OTC use. Uses Xenon arc flashlamp to emit intense pulsed light (IPL) to reduce hair growth; operates below skin surface without cutting or pulling. User-operated via finger switch; requires external power adapter. Includes built-in skin sensor to ensure proper contact before light pulse emission. Device provides direct illumination to tissue. Intended for home use to reduce hair growth with minimal pain. Benefits include convenient, non-invasive hair removal.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility testing (ISO 10993-5, -10, -23), electrical safety and EMC (IEC 60601-1, -1-2, -1-11, -2-83), photobiological safety (IEC 62471), and software verification/validation.
Technological Characteristics
Xenon arc flashlamp energy source; wavelength 510-1200nm; max energy density 4.92 J/cm²; spot size 3.3±0.25 cm². Powered by external AC/DC adapter (100-240V input, 12V 3A output). Features built-in skin contact sensor. Software level of concern: moderate. No sterilization required.
Indications for Use
Indicated for removal of unwanted body hair in individuals with Fitzpatrick Skin Types I-V.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
K233237 — xeo+ Family · Cutera, Inc. · Dec 15, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left, and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.
December 4, 2023
Shenzhen Greatro Electronic Technology Co., Ltd. % Bing Huang Registration Engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Center, No. 3101-90 Qianhai Road, Nanshan District Shenzhen, Guangdong 518000 China
Re: K232932
Trade/Device Name: Intense Pulse Light Therapeutic Apparatus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: September 19, 2023 Received: September 20, 2023
Dear Bing Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Hithe -Tanisha L. و 2023.12.04 Hithe -S 20:58:14 -05'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232932
Device Name
Intense Pulse Light Therapeutic Apparatus, Model(s): IPL-1 PRO, IPL-1, IPL-12, IPL-17
Indications for Use (Describe)
Intense Pulse Light Therapeutic Apparatus is an over-the-counter device intended for removal of unwanted body hair.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary K232932
# ''510(k) Summary'' as required by 21 CFR Part 807.92.
### Date: 2023-11-24
#### I. Submitter
Shenzhen Greatro Electronic Technology Co., Ltd East Side, 4/F Building A1, Xu'xing'da Industrial Park, Shiyan Sub-district, Bao'an District , Shenzhen, China Post code: 518108 Tel.: +86 755 86321189
Xinhua Yue Project Manager Tel.: + 86 139 2374 2027 Email: janis@jrdmic.com
### II. Device
Name of Device: Intense Pulse Light Therapeutic Apparatus Model(s): IPL-1 PRO, IPL-1, IPL-12, IPL-17 Common or Usual Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810
## III. Predicate Device
Primary Predicate device: 510(k) number: K230097 Manufacturer: Shenzhen BSX Technology Electronics Co., Ltd. Trade name: IPL Hair Removal Device, BSXT101 Product code: OHT Approval date: April 6, 2023
Reference device: 510(k) number: K220669 Manufacturer: Shenzhen Junbobeauty Technology Co., Ltd. Trade name: IPL HAIR REMOVAL HANDSET, MODEL: IPL-666 Product code: OHT Approval date: May 16, 2022
{5}------------------------------------------------
# IV. Device Description
The Intense Pulse Light Therapeutic Apparatus is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology, and it works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.
The Intense Pulse Light Therapeutic Apparatus is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and its built-in skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the treated skin, the device will not emit the light pulse.
There are IPL-1 PRO, IPL-1, IPL-12, and IPL-17 four models in this application. Their work principle, function, intended use, structure, and composition are the same, with differences being product appearance, size are slight differences, but these parameters are within the predicate device and do not affect or change the intended use of the device.
## V. Indications for Use
Intense Pulse Light Therapeutic Apparatus is an over-the-counter device intended for removal of unwanted body hair.
## VI. Comparison of Technological Characteristics With the Predicate Device
The subject device Intense Pulse Light Therapeutic Apparatus has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the subject device may be found substantially equivalent to its predicate device.
| Comparison<br>Elements | Subject Device | Predicate device<br>K230097 | Reference device<br>K220669 |
|------------------------|----------------------------------------------|-------------------------------------|-------------------------------------------------|
| K Number | Pending | K230097 | K220669 |
| Trade name | Intense Pulse Light<br>Therapeutic Apparatus | IPL Hair Removal<br>Device, BSXT101 | IPL HAIR REMOVAL<br>HANDSET , Model:<br>IPL-666 |
| Regulation number | 21CFR 878.4810 | 21CFR 878.4810 | 21CFR 878.4810 |
| Device class | II | II | II |
| Product code | OHT | OHT | OHT |
| Applicable skin | Fitzpatrick Skin Type I-<br>V | Fitzpatrick Skin Type I-<br>V | Fitzpatrick Skin Type I-<br>V |
| Wavelength range | 510~1200nm | 470~1200nm | 470~1000nm |
| Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp |
| Energy density | Max 4.92 J/cm² | Max 5.0J/cm² | 1.3~2.49 J/cm² |
The subject device is compared with the following Predicate Device in terms of intended use, design, specifications, and performance:
{6}------------------------------------------------
| Comparison<br>Elements | Subject Device | Predicate device<br>K230097 | Reference device<br>K220669 |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Spot size | $3.3+/-0.25 cm^2$ | $3.0cm^2+/-0.5 cm^2$ | $3 cm^2$ |
| Output energy | Level 1: 6.0J<br>Level 2: 7.5J<br>Level 3: 9.5J<br>Level 4: 11.0J<br>Level 5: 12.5J<br>Error is ±20% | Range in 5~12.5J (±20%) | Range in 3.9J~7.47 (±20%) |
| Output energy level | 1 to 5 level | 5 level | 5 level |
| Pulse duration | 5.84~9.24ms | 4~10ms | 11.5~15ms |
| Pulsing control | Finger switch | Finger switch | Finger switch |
| Delivery device | Direct illumination to tissue | Direct illumination to tissue | Direct illumination to tissue |
| Indication for use/Intended use | The Intense Pulse Light Therapeutic Apparatus is an over-the-counter device intended for removal of unwanted body hair. | The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair. | The IPL HAIR REMOVAL DEVICE is an over-the-counter device intended for removal of unwanted body and/ or facial hair. |
| Location for use | OTC | OTC | OTC |
| Treatment time | Once a week for 12 weeks | Once a week for 12 weeks | Use one time every two weeks for up to six times;<br>Use one time every four weeks when using6-10 times. |
| Sterilization | Not required | No required | No required |
| Power supply | Input: AC100~240V 50/60z<br>Output: DC12V 3A | 100-240v, 50/60Hz | 100~240V AC Input<br>12V 3A DC output |
### VII.Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### 1) Biocompatibility Safety
The materials of the patient-directly contacting components of the subject device was performed the biocompatibility evaluation in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document issued on Sep. 4, 2020", as recognized by FDA. The battery of testing was performed to, and passed, including:
- > ISO 10993-5 Biological Evaluation of Medical Devices –Part 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10 Biological Evaluation of Medical Devices –Part 10: Tests for Skin Sensitization
{7}------------------------------------------------
- > ISO 10993-23 Biological Evaluation of Medical Devices –Part 23: Tests for Irritation 2) Electrical Safety and EMC
Electrical safety and EMC testing was performed to, and passed, the following standards:
- > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility
- > IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- > IEC 60601-2-83 Medical Electrical Equipment Part 2-83: Particular Requirements For The Basic Safety And Essential Performance Of Home Light Therapy Equipment
# 3) Eve Safety
- > IEC 62471 Photobiological safety of lamps and lamp systems
## 4) Software Verification and Validation
Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. Summary
Based on the above performance as documented in this application, the subject device was found to have a safety and effectiveness profile that is similar to the predicate device.
#### VIII. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device Intense Pulse Light Therapeutic Apparatus is to be concluded substantial equivalent to its predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.