← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K230090

# Diode Laser Hair Removal (K230090)

_Wuhan Lotuxs Technology Co., Ltd. · OHT · Apr 19, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K230090

## Device Facts

- **Applicant:** Wuhan Lotuxs Technology Co., Ltd.
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** Apr 19, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Diode Laser Hair Removal is an over-the-counter device intended for adjunctive use with shaving for hair removal sustained with periodic treatments. Diode Laser Har Removal is also intended for permanent reduction in hair regrowth defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

## Device Story

Diode Laser Hair Removal device (models LHR-V8S-810, LHR-V8-810, LHR-V4S-810, LHR-V4-810) emits 810nm infrared laser pulses; GaAs laser source; energy absorbed by melanin in hair follicles; localized thermal effect disrupts hair growth. Device includes skin contact sensor to prevent accidental emission; integrated skin cooling feature; powered by external AC/DC adapter. Used in home environment by consumer; device provides hair removal/reduction benefits. User applies device to skin; contact sensor enables laser emission; thermal energy targets follicles. Clinical benefit is long-term, stable reduction in hair regrowth.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1, 60601-1-11, 60601-1-2, 60601-2-22), and laser safety (IEC 60825-1).

## Technological Characteristics

Diode laser (GaAs) emitting 810nm; energy density 2.0-9.0 J/cm² depending on model; treatment window 30mm x 10mm. Powered by external AC/DC adapter. Includes skin contact sensor and cooling system. Biocompatibility per ISO 10993-5/10. Safety standards: IEC 60601-1, 60601-1-11, 60601-1-2, 60601-2-22, IEC 60825-1.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- SILKPRO Laser Hair Removal System ([K142845](/device/K142845.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 19, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".

Wuhan Lotuxs Technology Co., Ltd. % Na Wu Management Representative Tacro Guangzhou Branch 501/E2. Future city. No.999 High-Tech Avenue (Free Trade Zone/Wuhan Area) Wuhan, 430206 China

Re: K230090

Trade/Device Name: Diode Laser Hair Removal Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: January 4, 2023 Received: January 12, 2023

Dear Na Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Colin K. Chen -S

for

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230090

Device Name Diode Laser Hair Removal

#### Indications for Use (Describe)

Diode Laser Hair Removal is an over-the-counter device intended for adjunctive use with shaving for hair removal sustained with periodic treatments. Diode Laser Har Removal is also intended for permanent reduction in hair regrowth defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

"510(k) Summary" as required by 21 CFR Part 807.92.

## Date: 2023-04-14

#### I. Submitter

Wuhan Lotuxs Technology Co.,Ltd. 501/E2, Future city, No.999 high-tech Avenue, Wuhan 430206, China(Free Trade Zone/Wuhan Area) Tel.: +86 -27-87619668

Na Wu (Management Representative) Tel: +86 13871252776 Email: na.wu@lotuxs.com

## II. Correspondent Consultant

Tacro Guangzhou Branch Rm. 501, No.55 West Tiyu Rd., Tianhe Dist., Guangzhou, Guangdong

Gamma Zhang(RA manager) Tel: +86 13433933949 Email: wbzhang@tacro.cn

# III. Device

Type of 510(k): Traditional Common Name: Powered laser surgical instrument Trade Name: Diode Laser Hair Removal Model: LHR-V8S-810, LHR-V8-810, LHR-V4S-810, LHR-V4-810 Classification Name: Light Based Over-The-Counter Hair Removal Regulation Number: 21 CFR 878.4810 Review Panel: General & Plastic Surgery Regulatory Class: II Product Code: OHT

# IV. Predicate Device

| Applicant                                   | Predicate Device                     | 510(k) Number |
|---------------------------------------------|--------------------------------------|---------------|
| Wuhan Lotuxs<br>Technology<br>Company, Ltd. | SILKPRO Laser Hair<br>Removal System | K142845       |

{4}------------------------------------------------

## 510(k)s-Section 6. 510(k) Summary

## V. Device Description

The Diode Laser Hair Removal device models LHR-V8S-810, LHR-V8-810, LHR-V4S-810, and LHR-V4-810 emit pulses of invisible infrared laser light of 810mm wavelength that penetrates into the skin and is selectively absorbed by melanin in the hair follicles. This creates a localized thermal effect that disrupts hair growth from the hair follicles. The device contains a sensor that detects contact with the skin so that the device will only emit the infrared laser pulses when the sensor is in contact with the skin. The device also includes a skin cooling feature and the device is powered by an external AC/DC power adaptor.

## VI. Indications for Use

Diode Laser Hair Removal is an over-the-counter device intended for adjunctive use with shaving for hair removal sustained with periodic treatments. SILKPRO is also intended for permanent reduction in hair regrowth defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

## VII.Comparison of Technological Characteristics With the Predicate Device

| Comparison<br>Elements                 | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                                                                                                                                                                             |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number                               | Applying                                                                                                                                                                                                                                                                                                                                                                                                          | K142845                                                                                                                                                                                                                                                                                                                                                                      |
| Trade name                             | Diode Laser Hair Removal                                                                                                                                                                                                                                                                                                                                                                                          | SILKPRO Laser Hair Removal System                                                                                                                                                                                                                                                                                                                                            |
| Model                                  | LHR-V8S-810, LHR-V8-810, LHR-V4S-810, LHR-V4-810                                                                                                                                                                                                                                                                                                                                                                  | /                                                                                                                                                                                                                                                                                                                                                                            |
| Classification name                    | Light Based Over-The-Counter Hair Removal                                                                                                                                                                                                                                                                                                                                                                         | Light Based Over-The-Counter Hair Removal                                                                                                                                                                                                                                                                                                                                    |
| Product code                           | OHT                                                                                                                                                                                                                                                                                                                                                                                                               | OHT                                                                                                                                                                                                                                                                                                                                                                          |
| Intended<br>use/Indications for<br>Use | Diode Laser Hair Removal is an over-the-counter device intended for<br>adjunctive use with shaving for hair removal sustained with periodic<br>treatments. Diode Laser Hair Removal is also intended for permanent reduction in<br>hair regrowth defined as a long-term, stable reduction in the number of hairs<br>regrowing when measured at 6, 9, and 12 months after the completion of a<br>treatment regime. | SILKPRO is an over-the-counter device intended for adjunctive use with shaving<br>for hair removal sustained with periodic treatments. SILKPRO is also intended for<br>permanent reduction in hair regrowth defined as a long-term, stable reduction in<br>the number of hairs regrowing when measured at 6, 9, and 12 months after the<br>completion of a treatment regime. |
| Laser Type                             | Diode laser                                                                                                                                                                                                                                                                                                                                                                                                       | Diode laser                                                                                                                                                                                                                                                                                                                                                                  |
| Laser Material                         | GaAs                                                                                                                                                                                                                                                                                                                                                                                                              | GaAs                                                                                                                                                                                                                                                                                                                                                                         |
| Wavelength                             | 810nm                                                                                                                                                                                                                                                                                                                                                                                                             | 810nm                                                                                                                                                                                                                                                                                                                                                                        |
| Max Energy Density                     | LHR-V8S-810, LHR-V8-810: 9.0J/cm²<br>LHR-V4S-810, LHR-V4-810: 7.0J/cm²                                                                                                                                                                                                                                                                                                                                            | 25J/cm²                                                                                                                                                                                                                                                                                                                                                                      |

A technical comparison to the predicate is provided below:

{5}------------------------------------------------

| Comparison<br>Elements   | Subject Device                                                                                    | Predicate Device                                                                                  |
|--------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Energy Density           | LHR-V8S-810, LHR-V8-810: 2.2J/cm2,<br>3.4J/cm2, 4.5J/cm2, 5.6J/cm2, 7.0J/cm2,<br>9.0J/cm2         | 5J/cm2, 10J/cm2, 15J/cm2,                                                                         |
|                          | LHR-V4S-810, LHR-V4-810: 2.0J/cm2, 2.8J/cm2, 3.5J/cm2, 4.3J/cm2, 5.3J/cm2, 7.0J/cm2               | 20J/cm2, 25J/cm2                                                                                  |
| Treatment Window<br>Size | 30mm×10mm                                                                                         | 9mm×9mm                                                                                           |
| Safety feature           | Complied with IEC 60601-1,<br>IEC 60601-1-11, IEC60601-1-<br>2, IEC 60601-2-22 and IEC<br>60825-1 | Complied with IEC 60601-1,<br>IEC 60601-1-11, IEC60601-1-<br>2, IEC 60601-2-22 and IEC<br>60825-1 |
| Biocompatibility         | All body-contacting materials<br>are complied with ISO10993-5<br>and ISO 10993-10                 | All body-contacting materials<br>are complied with ISO10993-5<br>and ISO 10993-10                 |

#### VIII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

## 1)Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of Diode Laser Hair Removal was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on June 16, 2016", as recognized by FDA. The testing was performed to, and passed, including:

- > ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical Devices –Part 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10:2010, Biological Evaluation of Medical Devices –Part 10: Tests for Irritation and Skin Sensitization

# 2)Electrical, EMC, and laser output, Safety and Performance Testing

Electrical, EMC, and laser output, Safety and Performance Testing was performed to, and passed, the following standards:

{6}------------------------------------------------

#### 510(k)s-Section 6. 510(k) Summary

- > IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-11 Medical electrical equipment -Part 1-11: General requirements for basic safety and essential performance -Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility -Requirements and tests
- > IEC 60601-2-22 Medical electrical equipment Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- > IEC 60825-1 Safety of laser products Part 1: Equipment classification, and requirements

#### Summary

Based on the above performance as documented in this application, Diode Laser Hair Removal was found to have a safety and effectiveness profile that is same as the predicate device.

#### VIII. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, Diode Laser Hair Removal is considered to be substantially equivalent to the predicate device K142845.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K230090](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K230090)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
