← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K221679

# IPL Hair Remover, Model: G993, G996, G998 and G885 (K221679)

_Shenzhen Mywin Technology Co., Ltd. · OHT · Jul 7, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K221679

## Device Facts

- **Applicant:** Shenzhen Mywin Technology Co., Ltd.
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** Jul 7, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

IPL Hair Remover is an over-the-counter device intended for removal of unwanted body and/or facial hair.

## Device Story

Handheld IPL hair removal device (models G993, G996, G998, G885) for over-the-counter use. Device uses a Xenon lamp to emit intense pulsed light (510-1100 nm) to heat hair roots. User operates device by applying it to skin; integrated skin proximity sensor prevents pulse emission unless full skin contact is detected. Features 5 intensity levels (1.8-4.5 J/cm²) controlled via finger switch. Powered by AC adapter. Intended for large areas including legs, arms, back, and abdomen. Benefits include home-based removal of unwanted hair.

## Clinical Evidence

No clinical data provided. Safety and efficacy established via bench testing, electrical safety (IEC 60601-1 series), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10) leveraged from the predicate device. Usability study performed to support OTC use.

## Technological Characteristics

Intense Pulsed Light (IPL) technology using a Xenon arc flashlamp. Wavelength: 510-1100 nm. Max fluence: 4.5 J/cm². Spot size: 4.3 cm². Pulse duration: 3 ms. Powered by external AC adapter (100-240V). Dimensions: 188 x 76 x 49 mm. Weight: 215.1g. Includes skin proximity sensor for safety. Complies with IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, and IEC 60601-1-2.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- IPL Hair Remover ([K211368](/device/K211368.md))

## Reference Devices

- IPL Home Use Hair Removal Device ([K211185](/device/K211185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 7, 2022

Shenzhen Mywin Technology Co., Ltd. % Jet Li Regulation manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong China

Re: K221679

Trade/Device Name: IPL Hair Remover, Model: G993, G996, G998 and G885 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT Dated: June 8, 2022 Received: June 9, 2022

Dear Jet Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221679

Device Name

IPL Hair Remover, model: G993, G996, G998 and G885

Indications for Use (Describe)

IPL Hair Remover is an over-the-counter device intended for removal of unwanted body and/or facial hair.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary - K221679

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

## 1 Submitter Information

Company Name: Shenzhen Mywin Technology Co., Ltd.

Address: 2nd Floor, Building B, New Age gongrong Industrial Zone, No. 2, Shihuan Road,

Shilong community, Shiyan Street, Baoan District, Shenzhen 518103, China

Phone: +86-18902842323

Contact Person (including title): Meiyun Liu (General Manager)

E-mail: Szmywin@163.com

Application Correspondent: Jet Li Company: Guangzhou KEDA Biological Technology Co., Ltd E-mail: med-jl@foxmail.com Phone: 86-18588874857 Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhou City, China

# 2 Subject Device Information

Type of 510(k) submission: Special 510(k) Device Modification Common Name: IPL Hair Remover Model: G993, G996, G998 and G885 Common Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Review Panel: General & Plastic Surgery Product Code: OHT Requlation Number: 878.4810

{4}------------------------------------------------

Requlation Class: 2

### 3 Predicate device Information

### Legally existing device K211368

Company Name:

Common Name: IPL Hair Remover

Model: G993, G996, G998 and G885

Common Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect

Classification Name: Powered Light Based Non-Laser Surgical Instrument With Thermal

Effect

Review Panel: General & Plastic Surgery

Product Code: ONF

Regulation Number: 878.4810

Regulation Class: 2

### Reference device K211185

Common Name: IPL Home Use Hair Removal Device Model: D-1150, D-1171, D-1153, D-1155, D-1156, D-1126, D-1178, D-1187 Common Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Review Panel: General & Plastic Surgery Product Code: OHT Regulation Number: 878.4810 Regulation Class: 2

### 4 Device Description

IPL Hair Remover Device (Model: G993, G996, G998 and G885), is an over-the-counter use device for removal of unwanted body and/or facial hair based on Intense Pulsed Light

{5}------------------------------------------------

(IPL) treatment. The device package includes IPL DEVICE, Power supply and User manual. It is supplied by AC power supply (Input 100-240 V AC). The weight of the device is 215.1g, and the size is 188 x 76 x 49mm. The device incorporates Intense Pulse Light (IPL) technology. The purpose of the light is to heat the root where the hair grows.

The device is equipped with a Xenon Lamp and a skin proximity sensor to detect appropriate skin contact. If the IPL Hair Remover Device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered for pulse emitting. The device can be used for large areas on legs, arms, back and abdomen. The product can flash for 100,000 times. .

#### Intended Use ട്

IPL Hair Remover is an over-the-counter device intended for removal of unwanted body and/or facial hair.

#### 6 Complied Standards

IPL Hair Remover G993, G996, G998 and G885 complies with the following FDA recognized consensus standards:

- 网 Electrical safety test according to IEC 60601-1, IEC60601-1-11 and IEC 60601-2-57 standards
- 风 Electromagnetic compatibility test according to IEC 60601-1-2 standard
- 风 ISO 10993-5:2009/(R) 2014, Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity.
- 区 ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.

# 7 Device modification description

The only modification of subject device is Intended Use. The device has changed to overthe-counter use. Therefore, the User Manual and Box labelling has been adding

{6}------------------------------------------------

corresponding use instructions for lay person user.

#### 8 Performance Testing

As the modification of subject device as above, results in no technological characteristics changes, the tests and data utilized to demonstrate safety and efficacy of the predicate device (legally existing device) are suitable for use in the assessment of the subject devices.

As there have been no changes to the performance of the subject device from the predicate device, this submission leverages performance and electrical testing provided in previous submission.

#### Biocompatibility 9

All the modified device materials which come in direct contact with the patient skin are biocompatible and identical to the materials used in the predicate device manufacturing. No biocompatibility test report is provided in this submission.

# 10 Clinical performance

Clinical performance is not deemed necessary.

# 11 Comparison with predicate device

Compare with predicate device (IPL Hair Remover K211368), the subject device is same in design principle, functions, material and the applicable standards. And the intended use of subject device is same to Reference device K211185. The differences between subject device and predicate devices do not raise any new questions of safety or effectiveness.

| Item         | Subject Device                          | Predicate Device                        | Reference Device                                                 |
|--------------|-----------------------------------------|-----------------------------------------|------------------------------------------------------------------|
| Manufacturer | Shenzhen Myw in<br>Technology Co., Ltd. | Shenzhen Myw in<br>Technology Co., Ltd. | Shenzhen Beauty Every<br>Moment intelligent electric<br>Co.,Ltd. |
| K number     | TBD                                     | K211368                                 | K211185                                                          |

{7}------------------------------------------------

|                                | IPL Hair Remover                                                                                                                                                                                                                              | IPL Hair Remover                                                                                                                                                                                                                                                                                                                                                                                                                           | IPL Home Use Hair Removal<br>Device                                                                                                                                                                                                           |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name                   | G993, G996, G998 and<br>G885                                                                                                                                                                                                                  | G993, G996, G998 and<br>G885                                                                                                                                                                                                                                                                                                                                                                                                               | D-1150, D-1171, D-1153, D-<br>1155, D-1156, D-1126, D-<br>1178, D-1187                                                                                                                                                                        |
| Regulation &<br>Classification | Classification Name: Laser<br>surgical instrument for use in<br>general and plastic surgery<br>and in dermatology<br>Review Panel: General &<br>Plastic Surgery<br>Product Code: OHT<br>Regulation Number:<br>878.4810<br>Regulation Class: 2 | Classification Name:<br>Powered Light Based Non-<br>Laser Surgical Instrument<br>With Thermal Effect<br>Review Panel: General &<br>Plastic Surgery<br>Product Code: ONF<br>Regulation Number:<br>878.4810<br>Regulation Class: 2                                                                                                                                                                                                           | Classification Name: Laser<br>surgical instrument for use in<br>general and plastic surgery<br>and in dermatology<br>Review Panel: General &<br>Plastic Surgery<br>Product Code: OHT<br>Regulation Number:<br>878.4810<br>Regulation Class: 2 |
| Indications for<br>Use         | IPL Hair Remover is an over-<br>the-counter device intended<br>for removal of unwanted<br>body and/or facial hair.                                                                                                                            | The IPL Hair Remover<br>Device, Model: G993, G996,<br>G998 and G885 is indicated<br>for the removal of unwanted<br>hair under the direction of a<br>physician, after training by a<br>healthcare professional. The<br>device is also indicated for<br>the permanent reduction in<br>hair regrowth, defined as the<br>long-term, stable reduction in<br>the number of hairs re-<br>growing when measured at<br>6, 9 and 12 months after the | IPL Home Use Hair Removal<br>Device is an over-the-<br>counter device intended for<br>removal of unwanted body<br>and/or facial hair.                                                                                                         |

{8}------------------------------------------------

|                                 |                                                   | completion of a treatment<br>regime. The device is used<br>for adults. |                                                   |
|---------------------------------|---------------------------------------------------|------------------------------------------------------------------------|---------------------------------------------------|
|                                 | OTC Use                                           | Prescription Use                                                       | OTC Use                                           |
| Source Energy                   | Supplied by external adapter                      | Supplied by external adapter                                           | Supplied by external adapter                      |
| Device Type                     | Intense Pulsed Light                              | Intense Pulsed Light                                                   | Intense Pulsed Light                              |
|                                 |                                                   |                                                                        | Regular window : 530-                             |
| Wavelength<br>(nm)              | 510nm ~ 1100nm                                    | 510nm ~ 1100nm                                                         | 1100nm, 590-1100nm                                |
|                                 |                                                   |                                                                        | Filter window: 600-1100nm                         |
| Max. Fluence                    | Max 4.5 J/cm²                                     | Max 4.5 J/cm²                                                          | Max 4.5 J/cm²                                     |
| Spot Size                       | 4.3 cm²                                           | 4.3 cm²                                                                | 4.3 cm²                                           |
|                                 | Level 1: 1.8 J/cm²                                | Level 1: 1.8 J/cm²                                                     |                                                   |
|                                 | Level 2: 2.3 J/cm²                                | Level 2: 2.3 J/cm²                                                     |                                                   |
| Light Intensity                 | Level 3: 3.2 J/cm²                                | Level 3: 3.2 J/cm²                                                     | 2.0 to 4.5 J/cm²                                  |
|                                 | Level 4: 4.4 J/cm²                                | Level 4: 4.4 J/cm²                                                     |                                                   |
|                                 | Level 5: 4.5 J/cm²                                | Level 5: 4.5 J/cm²                                                     |                                                   |
| Pulse duration                  | 3 ms                                              | 3 ms                                                                   | 7.5~12 ms                                         |
| Energy medium                   | Xenon Arc Flashlamp                               | Xenon Arc Flashlamp                                                    | Xenon Arc Flashlamp                               |
| Pulsing Control                 | Finger switch                                     | Finger switch                                                          | Finger switch                                     |
| Number of<br>Output<br>Channels | One channel                                       | One channel                                                            | One channel                                       |
| Output Intensity<br>Level       | 5 levels                                          | 5 levels                                                               | /                                                 |
| Software<br>Control?            | Yes                                               | Yes                                                                    | /                                                 |
| Weight                          | 215.1g                                            | 215.1g                                                                 | /                                                 |
| Dimensions                      | 188*76*49 mm (H*W*D)                              | 188*76*49 mm (H*W*D)                                                   | /                                                 |
| Electrical                      | IEC 60601-1                                       | IEC 60601-1                                                            | IEC 60601-1                                       |
| safety, EMC                     | IEC 60601-1-11<br>IEC 60601-1-2<br>IEC 60601-2-57 | IEC 60601-1-11<br>IEC 60601-1-2<br>IEC 60601-2-57                      | IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57 |
| Biocompatibility                | ISO10993-5<br>ISO10993-10                         | ISO10993-5<br>ISO10993-10                                              | ISO10993-5<br>ISO10993-10                         |

{9}------------------------------------------------

Sponsor: Shenzhen Mywin Technology Co., Ltd. Subject Device: IPL Hair Remover, Model: G993, G996, G998 and G885

### Note

Although the indication for use of subject device is different to predicate device K211368, but it is only to simplify the statement of the legally existing predicate device and remove the prescription use statement, which is not involved with technical specifications. The revised indication for use of subject device in this submission also had been covered by the indication for use of the legally existing predicate device; which is same to Reference device K211185. And the subject device had been verify by usability study for its OTC use. So the difference does not affect the safety and effectiveness.

### 12 Summary Prepared Date

28 Jun. 2022

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K221679](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K221679)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
