IPL Home Use Hair Removal Device

K221001 · Shenzhen Beauty Every Moment Intelligent Electric Co., Ltd. · OHT · May 12, 2022 · General, Plastic Surgery

Device Facts

Record IDK221001
Device NameIPL Home Use Hair Removal Device
ApplicantShenzhen Beauty Every Moment Intelligent Electric Co., Ltd.
Product CodeOHT · General, Plastic Surgery
Decision DateMay 12, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

IPL Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair.

Device Story

Personal, light-based hair reduction device; uses Intense Pulsed Light (IPL) technology to reduce hair growth below skin surface without cutting or pulling. Device consists of Xenon lamp and skin sensor; powered by external adapter; operated via finger switch. User places device in full contact with skin to trigger pulse emission; skin sensor prevents firing if contact is insufficient. Used by consumers in home environment. Output is light pulse; intended to provide hair reduction with minimal pain.

Clinical Evidence

Bench testing only. Evidence includes biocompatibility assessment (materials identical to predicate), electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57), photobiological eye safety (IEC 62471), and software verification/validation.

Technological Characteristics

Materials: PC plastic and Aluminum alloy. Energy source: Xenon Arc Flashlamp. Wavelength: 550-1100nm. Energy density: 2.0-4.3J/cm². Pulse duration: 5-12ms. Connectivity: Standalone, powered by external adapter. Safety features: Skin contact sensor. Software: Moderate level of concern.

Indications for Use

Indicated for removal of unwanted body and/or facial hair in an over-the-counter, home-use setting.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom. May 12, 2022 Shenzhen Beauty Every Moment intelligent electric Co., Ltd. % Rain Yip Registered Engineer Feiying Drug & Medical Consulting Technical Service Group Contact Address Re: K221001 Trade/Device Name: IPL Home Use Hair Removal Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: March 21, 2022 Received: April 4, 2022 Dear Rain Yip: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221001 Device Name IPL Home Use Hair Removal Device, models: D-1198, D-1185, D-1186, D-1189, D-1197 Indications for Use (Describe) IPL Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K221001 # ''510(k) Summary'' as required by 21 CFR Part 807.92. # Date: 2022-03-21 #### I. Submitter Shenzhen Beauty Every Moment intelligent electric Co.,Ltd. Rm.101,201,501, Bldg. 4, Haoer Jiashitai Technology Park, Fengtang Avenue No.164, Zhancheng Community, Fuhai Street, Bao'an District, Shenzhen, Guangdong, China Post code: 518100 Tel .: +86 755 3388 3239 LUO HUAJUN Certification engineer Tel: +86755 3388 3239 Email: 309024340@qq.com #### II. Device Name of Device: IPL Home Use Hair Removal Device Model(s): D-1198, D-1185, D-1186, D-1189, D-1197 Common or Usual Name: Light Based Over-The-Counter For Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810 ## III. Predicate Device | Manufacturer | Predicate Device | 510(k) Number | Approval Date | |----------------------------------|----------------------------------------------------|---------------|---------------| | Shenzhen<br>Technology Co., Ltd. | Bosidin<br>IPL Home Use Hair Removal Device/D-1171 | K211185 | Dec. 20, 2021 | ## IV. Device Description The IPL Home Use Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered a pulse emitting. {4}------------------------------------------------ The IPL Home Use Hair Removal Device includes D-1198, D-1186, D-1189, D-1197 five models. Their intended use, performance, structure design and operation are basically identical, with the differences mainly contains PCB layout, product appearance and internal construction, but these differences do not affect or change the intended use of the device. # V. Indications for Use IPL Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair. # VI. Comparison of Technological Characteristics With the Predicate Device The IPL Home Use Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the IPL Home Use Hair Removal Device may be found substantially equivalent to its predicate device. | Comparison Elements | Subject Device K221001 | Predicate Device K211185 | |---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------| | K Number | Pending | K211185 | | Trade name | IPL Home Use Hair Removal<br>Device/D-1198, D-1185, D-1186,<br>D-1189, D-1197 | IPL Home Use Hair Removal<br>Device/D-1171 | | Wavelength range | 550-1100nm | 530-1100nm | | Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | | Energy density | 2.0~4.3J/cm² | 2.0~4.3J/cm² | | | D-1198: 2.7cm²<br>D-1185: 2.7cm² | Non-removable regular window:<br>3.6cm² | | Spot size | D-1186: 3.0cm²<br>D-1189: 3.0cm²<br>D-1197: 3.4cm² | Removable regular window:<br>1.9cm² | | Pulse duration | 5~12ms | 7.5-12ms | | Pulsing control | Finger switch | Finger switch | | Delivery device | Direct illumination to tissue | Direct illumination tissue | | Indication<br>for<br>use/Intended use | IPL Home Use Hair Removal<br>Device is an over-the-counter<br>device intended for removal of<br>unwanted body and/or facial hair. | IPL Home Use Hair Removal<br>Device is an over-the-counter<br>device intended for removal of<br>unwanted body and/or facial hair. | | Location for use | OTC | OTC | IPL Home Use Hair Removal Device is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance: {5}------------------------------------------------ # VII.Performance Data The following performance data were provided in support of the substantial equivalence determination. # 1) Biocompatibility Safety The materials of the patient-directly contacting components of the subject device is containing PC plastic and Aluminum alloy, and both materials is identical to the material of the device IPL Home Use Hair Removal Device, which is cleared by FDA with K211185, produced by us. So these components' biocompatibility is safe. # 2) Electrical Safety and Eve Safetv Electrical safety and Eye safety testing was performed to, and passed, the following standards: - > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility - > IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance - > IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment - > IEC 60601-2-57 Medical electrical equipment -Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use # 3) Eye Safety - > IEC 62471 Photobiological safety of lamps and lamp systems # 4) Software Verification and Validation Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. Summary Based on the above performance as documented in this application, the subject device IPL Home Use Hair Removal Device was found to have a safety and effectiveness profile that is similar to the predicate device. #### VIII. Conclusions In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device IPL Home Use Hair Removal Device is to be concluded substantial equivalent to its predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...