← Product Code [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS) · K253712

# LED Light Therapy Mask (RB-051, RB-061, RB-071, RB-052, RB-062, RB-072, RB-053, RB-063, RB-073) (K253712)

_Shenzhen Rainbow Technology Co., Ltd. · OHS · Jan 23, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K253712

## Device Facts

- **Applicant:** Shenzhen Rainbow Technology Co., Ltd.
- **Product Code:** [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS.md)
- **Decision Date:** Jan 23, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Red+Infrared light: Treatment of full-face wrinkles. Red light: Treatment of full-face wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm;relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Blue light: Treatment of mild to moderate inflammatory acne. Red+Blue light: Treatment of mild to moderate inflammatory acne. The device is intended to use LED light for the treatment of wrinkles and mild to moderate acne for adults in home healthcare environment.

## Device Story

Wearable LED phototherapy mask; includes controller and Type-C charging cable. Emits red (630nm), blue (465nm), and infrared (850nm) light. Five treatment modes: red+infrared, red, infrared, blue, and red+blue. Used by patients in home settings. Device provides phototherapy for wrinkle reduction and acne treatment, plus topical heating for muscle/joint relief. User selects mode via controller; device delivers light to facial skin. Benefits include non-invasive wrinkle/acne management and temporary relief of muscle/joint stiffness via increased local blood circulation.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through biocompatibility testing (ISO 10993-5, -10, -23) and electrical safety/EMC testing (IEC 60601-1, -1-2, -1-11, -2-83, IEC 62471, IEC 62133-2).

## Technological Characteristics

Materials: ABS and Silicone. Light source: LEDs (Red 630nm, Blue 465nm, Infrared 850nm). Power: 5V 1A rechargeable lithium battery. Form factor: Wearable mask with remote controller. Connectivity: Standalone. Sterilization: Not applicable (non-sterile). Software: Embedded firmware for mode control.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- LED Light Therapy Mask ([K243423](/device/K243423.md))

## Reference Devices

- Q-Rejuvalight Pro Facewear ([K230042](/device/K230042.md))
- LED Light Therapy Device (KFB290, KFB291) ([K241857](/device/K241857.md))
- LED Facial Mask ([K230351](/device/K230351.md))
- LED light therapy mask (FM-01, FM-02, FM-03) ([K242796](/device/K242796.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

January 23, 2026

Shenzhen Rainbow Technology Co., Ltd.
% Youshan Gong
RA specialist
Feiying Drug &amp; Medical Consulting Technical Service Group
Rm.2401 Zhenye International Business Center
# 3101-90, Qianhai Rd.
Shenzhen, Guangdong 518052
China

Re: K253712
Trade/Device Name: LED Light Therapy Mask (RB-051, RB-061, RB-071, RB-052, RB-062, RB-072, RB-053, RB-063, RB-073)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP, ILY
Dated: November 22, 2025
Received: November 24, 2025

Dear Youshan Gong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K253712 - Youshan Gong
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}

K253712 - Youshan Gong
Page 3

See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA
L. HITHE -S

Digitally signed by
TANISHA L. HITHE -S
Date: 2026.01.23
14:05:53 -05'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253712 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  LED Light Therapy Mask (RB-051, RB-061, RB-071, RB-052, RB-062, RB-072, RB-053, RB-063, RB-073)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  *Red+Infrared light: Treatment of full-face wrinkles.
*Red light: Treatment of full-face wrinkles.
*Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle
spasm;relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood
circulation.
*Blue light: Treatment of mild to moderate inflammatory acne.
*Red+Blue light: Treatment of mild to moderate inflammatory acne.
The device is intended to use LED light for the treatment of wrinkles and mild to moderate acne for adults in
home healthcare environment.  |   |   |
|  Please select the types of uses (select one or both, as
applicable). | ☐ Prescription Use (21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

510 (k) Summary K253712

This “510(k) Summary” of 510(k) safety and effectiveness information is submitted in accordance with requirements of Title 21, CFR Section 807.92.

(1) Applicant information:

510(k) owner's name: Shenzhen Rainbow Technology Co.,Ltd
Address: 3th Floor, No. 2, Dongwangyang Industry, Huangtian Community, Hangcheng Street, Bao'an District, Shenzhen City, Guangdong Province, China 518000
Contact person: Layla Li
Position: General Manager
Phone number: +86 135 9032 9742
Email: Layla@rainbowdo.com
Date of summary prepared: 2026-1-13

(2) Reason for the submission

New device, there were no prior submissions for the device.

(3) Proprietary name of the device

Trade name/model: LED light therapy mask/Model: RB-051, RB-061, RB-071, RB-052, RB-062, RB-072, RB-053, RB-063, RB-073
Regulation name: Light Based Over The Counter Wrinkle Reduction(OHS), Over-The-Counter Powered Light Based Laser For Acne(OLP), Infrared, Therapeutic Heating(ILY)
Regulation number: 21 CFR 878.4810, 21 CFR 890.5500
Product code: OHS, OLP, ILY
Review panel: General &amp; Plastic Surgery
Regulation class: Class II

(4) Predicate devices

&gt; Predicate device

|  Sponsor | Shenzhen Rainbow Technology Co., Ltd.  |
| --- | --- |
|  Device Name and Model | LED Light Therapy Mask (Model(s): RB-008, RB-030, RB-081, RB-008G, RB-008GB, RB-008J, RB-008JB)  |
|  510(k) Number | K243423  |

Page 1 of 12

{5}

Page 2 of 12

|  Product Code | OHS, OLP, ILY  |
| --- | --- |
|  Regulation Number | 21 CFR 878.4810, 21 CFR 890.5500  |
|  Regulation Class | II  |

&gt; Reference device 1

|  Sponsor | Shenzhen Kaiyan Medical Equipment Co., Ltd  |
| --- | --- |
|  Device Name and Model | Q-Rejuvalight Pro Facewear  |
|  510(k) Number | K230042  |
|  Product Code | OHS, OLP  |
|  Regulation Number | 21 CFR 878.4810  |
|  Regulation Class | II  |

&gt; Reference device 2

|  Sponsor | Dongguan Boyuan Intelligent Technology Co.,Ltd  |
| --- | --- |
|  Device Name and Model | LED Light Therapy Device(Models: KFB290, KFB291)  |
|  510(k) Number | K241857  |
|  Product Code | OHS, OLP, ILY  |
|  Regulation Number | 21 CFR 878.4810, 21 CFR 890.5500  |
|  Regulation Class | II  |

&gt; Reference device 3

|  Sponsor | Shenzhen SUNGPO HI-TECH Electronic Co., Ltd  |
| --- | --- |
|  Device Name and Model | LED Facial Mask  |
|  510(k) Number | K230351  |
|  Product Code | OHS, OLP  |
|  Regulation Number | 21 CFR 878.4810  |
|  Regulation Class | II  |

&gt; Reference device 4

|  Sponsor | Guangdong Newdermo Biotech Co.,Ltd  |
| --- | --- |
|  Device Name and Model | LED light therapy mask, Models: FM-01, FM-02, FM-03  |
|  510(k) Number | K242796  |
|  Product Code | OHS, OLP, ILY  |
|  Regulation Number | 21 CFR 878.4810, 21 CFR 890.5500  |
|  Regulation Class | II  |

(5) Description/ Design of device:
LED Light Therapy Mask (Model: RB-051, RB-061, RB-071, RB-052, RB-062, RB-072,

{6}

RB-053, RB-063, RB-073) is a home use wearable LED phototherapy device which can help reduce wrinkles and acne and provide topical heating. LED Light Therapy Mask is consisting of main unit(mask), controller, Type-C charging cable and so on. There are 3 kinds of light, which include Red light (wavelength 630nm), Blue light (wavelength 465nm), Infrared light (wavelength 850nm).

All models can output 5 kinds of treatment modes: red+infrared, red, infrared, blue and red+blue light.

# (6) Indications for use:

*Red+Infrared light: Treatment of full-face wrinkles.
*Red light: Treatment of full-face wrinkles.
*Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
*Blue light: Treatment of mild to moderate inflammatory acne.
*Red+Blue light: Treatment of mild to moderate inflammatory acne.

The device is intended to use LED light for the treatment of wrinkles and mild to moderate acne for adults in home healthcare environment.

(7) Materials

|  Component name | Material of Component | Body Contact Category | Contact Duration  |
| --- | --- | --- | --- |
|  LED Light Therapy Mask main unit and controller (RB-051, RB-061, RB-071, RB-052, RB-062, RB-072, RB-053, RB-063, RB-073) | ABS, Silicone | Surface-contacting device: Intact skin | >24 hours, ≤ 30 days  |
|  Straps (RB-051, RB-061, RB-071, RB-052, RB-062, RB-072, RB-053, RB-063, RB-073) (The main unit and straps are integrated) | Silicone | Surface-contacting device: Intact skin | >24 hours, ≤ 30 days  |

We have tested the device for biocompatibility by a reliable third-party lab. For details, please refer to "Biocompatibility Evaluation Report".

{7}

(8) Technological characteristics and substantial equivalence:

The LED Light Therapy Mask have the same intended use as the predicate device and reference devices. The technological characteristics such as wavelength, irradiance, power supply, are similar to the predicate device and reference devices. Any minor differences between the subject device and the listed predicate device and reference devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device and reference devices for its intended use.

Therefore, the LED Light Therapy Mask may be found substantially equivalent to its predicate devices and reference devices.

|  Item | Subject device | Predicate device K243423 | Reference device 1 K230042 | Reference device 2 K241857 | Reference device 3 K230351 | Reference device 4 K242796 | Remar k  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  510 (k) number | Pending | K243423 | K230042 | K241857 | K230351 | K242796 | /  |
|  Trade name | LED light therapy mask | LED light therapy mask | Q-Rejuvalight Pro Facewear | LED Light Therapy Device(Models: KFB290, KFB291) | LED Facial Mask | LED light therapy mask, Models: FM-01, FM-02, FM-03 | /  |
|  Manufacturer | Shenzhen Rainbow Technology Co.,Ltd | Shenzhen Rainbow Technology Co.,Ltd | Shenzhen Kaiyan Medical Equipment Co., Ltd | Dongguan Boyuan Intelligent Technology Co.,Ltd | Shenzhen SUNGPO HI-TECH Electronic Co., Ltd | Guangdong Newdermo Biotech Co.,Ltd | /  |
|  Regulation number | 21 CFR 878.4810
21 CFR 890.5500 | 21 CFR 878.4810
21 CFR 890.5500 | 21 CFR 878.4810 | 21 CFR 878.4810
21 CFR 890.5500 | 21 CFR 878.4810 | 21 CFR 878.4810
21 CFR 890.5500 | Same  |
|  Regulation Name | Light Based Over The Counter Wrinkle Reduction | Light Based Over The Counter Wrinkle Reduction (OHS); | Light Based Over The Counter Wrinkle | Light Based Over The Counter Wrinkle | Light Based Over The Counter Wrinkle | Light Based Over The Counter Wrinkle | Same  |

Page 4 of 12

{8}

|  Item | Subject device | Predicate device K243423 | Reference device 1 K230042 | Reference device 2 K241857 | Reference device 3 K230351 | Reference device 4 K242796 | Remark  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   | (OHS); Over-The-Counter Powered Light Based Laser For Acne(OLP); Infrared, Therapeutic Heating(ILY) | Over-The-Counter Powered Light Based Laser For Acne(OLP); Infrared, Therapeutic Heating(ILY) | Reduction(OHS), Over-The-Counter Powered Light Based Laser For Acne(OLP) | Over-The-Counter Powered Light Based Laser For Acne(OLP), Infrared Therapeutic Heating(ILY) | Reduction (OHS); Over-The-Counter Powered Light Based Laser For Acne(OLP); | Reduction(OHS), Over-The-Counter Powered Light Based Laser For Acne(OLP), Infrared Therapeutic Heating(ILY) |   |
|  Product code | OHS, OLP, ILY | OHS, OLP, ILY | OHS, OLP | OHS, OLP, ILY | OHS, OLP | OHS, OLP, ILY | Same  |
|  Class | Class II | Class II | Class II | Class II | Class II | Class II | Same  |
|  Indications for use/ Intended use | *Red+Infrared light: Treatment of full-face wrinkles. *Red light: Treatment of full-face wrinkles. *Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and | • Red+Infrared light: Treatment of full-face wrinkles. • Red light: Treatment of full-face wrinkles. • Infrared light: Provide topical heating for the purpose of elevating tissue temperature; | The QRejuvalight Pro Facewear (Model: P19-0023) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne. | - Red light: Treatment of full-face wrinkles. - Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase | LED Facial Mask is an over the counter device that is intended to use LED light for the treatment of wrinkles and mild to moderate acne. | Red light: Treatment of full-face wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the | Same  |
|   | the use of the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and | the use of the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and | the use of the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and to the same method; and | the use of the same method; and to the same method; and to the same method; and | the use of the same method; and to the same method; and to the same method; and to the same method; and | the use of the same method; and |   |

{9}

|  Item | Subject device | Predicate device K243423 | Reference device 1 K230042 | Reference device 2 K241857 | Reference device 3 K230351 | Reference device 4 K242796 | Remark  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   | muscle spasm;relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
*Blue light: Treatment of mild to moderate inflammatory acne.
*Red+Blue light: Treatment of mild to moderate inflammatory acne.
The device is intended to use LED light for the treatment of wrinkles and mild | arthritis and muscle spasm;relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
• Blue light: Treatment of mild to moderate inflammatory acne.
• Red+Blue light: Treatment of mild to moderate inflammatory acne.
The device is intended to use LED light for the treatment of wrinkles and mild to |  | local blood circulation.
- Red+Infrared Light: Treatment of full-face wrinkles.
- Amber light: Treatment of full-face wrinkles.
- Blue light: Treatment of mild to moderate inflammatory acne.
- Mixed light(Red+Blue+Infrared): Treatment of mild to moderate inflammatory acne. |  | purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Mixed light:Treatment of mild to moderate inflammatory acne. |   |

Page 6 of 12

{10}

|  Item | Subject device | Predicate device K243423 | Reference device 1 K230042 | Reference device 2 K241857 | Reference device 3 K230351 | Reference device 4 K242796 | Remark  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   | to moderate acne for adults in home healthcare environment. | moderate acne for adults in home healthcare environment. |  |  |  |  |   |
|  Location for use | Face | Face | Face | Face | Face | Face | Same  |
|  OTC or prescription | OTC | OTC | OTC | OTC | OTC | OTC | Same  |
|  Power supply | Output: 5V 1A Rechargeable Lithium battery | Output: 5V 1A Rechargeable Lithium battery | Input: 5V, 50/60Hz, 2A Li-ion Polymer Battery: 3.7V, 600mAh, 2.22Wh | Input: 100 -240 V, 50 / 60 Hz Output: 5V, 1A Lithium ion battery: 1300mAh | An external adapter Input: AC 100-240V 50-60Hz 0.2A Output: DC 12V 0.5A | Input: 100-240V, 50/60Hz, 0.25 A Output: DC 5 V, 500 mA | Same  |
|  Light source |  | Light Emitting Diodes |  | Light Emitting Diodes | Light Emitting Diodes | Light Emitting Diodes | Same  |
|  Wavelength | Red: 620nm Blue: 465nm Infrared: 850nm | Red: 620nm Blue: 460nm Infrared: 850nm | 630nm, 880nm, 415nm 605nm, 660nm, | 635nm ± 5nm visible red light; 850nm ± 5nm Invisible red light; 465±5nm blue light; 605 ± 5nm amber | Blue: 465nm±5nm Red: 625nm±5nm Amber: 605nm±5nm | Red: 620nm Blue: 460nm Infrared: 850nm Mixed: 620nm and 850nm and 460nm | Similar r Note 1  |

Page 7 of 12

{11}

|  Item | Subject device | Predicate device K243423 | Reference device 1 K230042 | Reference device 2 K241857 | Reference device 3 K230351 | Reference device 4 K242796 | Remar k  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  |  |  | light |  |  |   |
|  Irradiance | 1) Red light: 20mw/cm²-10 %/+30%, that is 18~26mw/cm²
2) Infrared light: 4mW/cm²-10 %/+30%, that is 3.6~5.2mw/cm²
3) Blue light: 35mw/cm²-10 %/+30%, that is 31.5~45.5mw/cm²
4) Red + Infrared light: 35mw/cm²-10 %/+30%, that is | Red+IR: 45.5mW/cm²
Red light: 53.2mW/cm²
Infrared light: 4mW/cm²
Blue light: 58.7mW/cm²
Red+Blue light: 33.6mW/cm² | 605nm: 15±5mW/cm²
630nm: 20±5mW/cm²
660nm: 25±5mW/cm²
880nm: 10±5mW/cm²
415nm: 25±5mW/cm²
Total: 70mW/cm²(wrinkle)-(605nm+630nm+660nm+880nm)
45mW/cm²(acne)-(630nm+415nm) | Red: 25mW/cm²;
IR: 3mW/cm²;
Red+IR: 30mW/cm²;
Blue: 18mW/cm²;
Amber: 20mW/cm²;
Mixed light: 9mW/cm²; | Blue: 15~63mW/cm²
Red: 31~75mW/cm² | Red light: 2.0~3.0mW/cm²
Blue light: 2.0~4.0mW/cm²
Infrared light: 2.0~4.0 mW/cm²
Mixed light: 9.0~12.0 mW/cm² | Different
Note 2  |

Page 8 of 12

{12}

|  Item | Subject device | Predicate device K243423 | Reference device 1 K230042 | Reference device 2 K241857 | Reference device 3 K230351 | Reference device 4 K242796 | Remar k  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   | 31.5~45.5mw /cm²
5) Red + Blue light: 40mw/cm²-10 %/+30%, that is 36~52mw/cm² |  |  |  |  |  |   |
|  Treatment time | 10-minute at a time.
3-4 treatment a week, reduce to 1-2 treatment a week once the results shown. | 5-minute or 10-minute at a time.
3-4 treatment a week, reduce to 1-2 treatment a week once the results shown. | 3 minutes per treatment | For red, blue and red+infrared: 10, 20, 30 minutes
For infrared, amber light and mixed light: 10, 20 minutes. | 10 minutes/day, 3 times per week | Manual Mode: 15 minutes each time,
Automatic Mode: 10 minutes each time. | Same  |
|  Dimensions (mm) | LED Mask: About 63.4 X 22.8 X 15.4 (cm)
Remote control: About 4.5 X 13.3 X 3.2 (cm) | Model RB-030,RB-081,RB-008:
About 19.5 X 22.3 X 9.3 (cm)
Model RB-008G,RB-008GB:
About 27.5 X 20.5 X | No publicly available | KFB290: Approximately 302 mm x 197 mm x 22 mm
KFB291: Approximately 412 mm x 195 mm x 22 mm | Not publicly available | FM-01: 207 * 277 * 43 mm,
FM-02: 198 * 383 * 33.5 mm,
FM-03: 237.5 * 108 * 8.1 mm | Different
Note 3  |

Page 9 of 12

{13}

|  Item | Subject device | Predicate device K243423 | Reference device 1 K230042 | Reference device 2 K241857 | Reference device 3 K230351 | Reference device 4 K242796 | Remar k  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  | 0.3 (cm)
Model
RB-008J,RB-008JB:
About 27.8 X 20.8 X 0.38 (cm) |  |  |  |  |   |
|  Weight | LED Mask: About 338 g
Remote control: About 50 g | Model RB-030, RB-081, RB-008:
About 391 g
Model RB-008G, RB-008GB:
About 120 g
Model RB-008J,RB-008JB:
About 391 g | No publicly available | Not publicly available | Not publicly available | Not publicly available | Different
**Note 3**  |
|  Compliance with voluntary standards | IEC 60601-1;
IEC 60601-1-2;
IEC 60601-1-11;
IEC 60601-2-83;
IEC 62471 | IEC 60601-1;
IEC 60601-1-2;
IEC 60601-1-11;
IEC 60601-2-83;
IEC 62471 | IEC 60601-1;
IEC 60601-1-2;
IEC 60601-1-11;
IEC 60601-2-57;
IEC 62471;
IEC 62133-2 | IEC 60601-1;
IEC 60601-1-2;
IEC 60601-1-11;
IEC 60601-2-83;
IEC 62471 | IEC60601-1-2
IEC60601-1
IEC60601-1-11
IEC60601-2-57
IEC62471 | IEC 60601-1;
IEC 60601-1-2 | Same  |
|  Biocompatibility feature | All body-contacting materials are complied with | All body-contacting materials are complied with ISO10993-5, ISO 10993-10, ISO | ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | Comply with ISO10993-1, ISO10993-5 and ISO10993-10 | All body-contacting materials are complied with | Same  |

Page 10 of 12

{14}

Page 11 of 12

|  Item | Subject device | Predicate device K243423 | Reference device 1 K230042 | Reference device 2 K241857 | Reference device 3 K230351 | Reference device 4 K242796 | Remar k  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   | ISO10993-5, ISO 10993-10, ISO 10993-23 | 10993-23 |  |  |  | ISO10993-5 and ISO 10993-10 |   |

Comparison in details:

**Note 1:**
Though the wavelength of subject device is a little different from the predicate device, the wavelength of the subject device can be basically covered by the predicate devices and reference devices' range and they all comply with IEC 60601-2-83 and IEC 62471 requirements, so this difference will not raise any safety or effectiveness issue.

**Note 2:**
Though the irradiance of subject device is a little different from the predicate device and reference devices, the irradiance of subject device is within the range of the minimum and maximum value of the predicate device and reference devices, and the subject device complies with IEC 60601-2-83 and IEC 62471 requirements, so this difference will not raise any safety or effectiveness issue.

**Note 3:**
Though the dimension and weight are different from the predicate device and reference devices, this difference is insignificant and do not raise any safety or effectiveness problems.

**Conclusions**
LED Light Therapy Mask are substantially equivalent to the predicate device and reference devices.

{15}

(9) Non-clinical studies and tests performed:

Non-clinical testings have been conducted to verify that the LED Light Therapy Mask meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the subject device complies with the following standards:

- IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-83:2019/AMD1:2022, Medical Electrical Equipment - Part 2-83: Particular requirements for the basic safety and essential performance
- IEC 62471 First edition 2006-07, Photobiological safety of lamps and lamp systems
- IEC 62133-2 Edition1.0 2017-02, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems

The device has been tested for biocompatibility, it complies with the following standards.

- ISO 10993-5 Third edition 2009-06-01, Biological Evaluation of Medical Devices - Part 5: Tests for InVitro Cytotoxicity
- ISO 10993-10 Fourth edition 2021-11, Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization
- ISO 10993-23 First edition 2021-01 Biological Evaluation of Medical Devices - Part 23: Tests for Irritation

We have also conducted:

- Software verification and validation test according to the requirements of the FDA "Content of Premarket Submissions for Device Software Functions"

(10) Conclusion

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device LED Light Therapy Mask is as safe, as effective, and performs as well as the legally marketed predicate device.

Page 12 of 12

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K253712](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K253712)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
