NanoKnife Generator (H78720300351US0)

K253198 · Angiodynamics · OAB · Mar 24, 2026 · General, Plastic Surgery

Device Facts

Record IDK253198
Device NameNanoKnife Generator (H78720300351US0)
ApplicantAngiodynamics
Product CodeOAB · General, Plastic Surgery
Decision DateMar 24, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The NanoKnife System with six outputs is indicated for surgical ablation of soft tissue, including prostate tissue.

Device Story

The NanoKnife Generator is a software-controlled low-energy direct-current (LEDC) generator used for surgical ablation of soft tissue, including prostate tissue. The system utilizes up to six electrode probes placed in an array within the target tissue. The device delivers a series of voltage pulses between pairs of probes; the waveform is adjustable based on clinician-selected parameters including volts/cm, pulse length, pulse count, probe distance, and timing mode (90PPM or ECG synchronization). Activation is controlled via a foot-pedal. The system matches probes into pairs to deliver energy, resulting in tissue ablation. It is intended for use by clinicians in a surgical setting. The device is substantially equivalent to the predicate NanoKnife System, with modifications limited to updated internal components that do not alter the fundamental principle of operation, energy source, or mechanism of action.

Clinical Evidence

No clinical data submitted. Safety and efficacy were demonstrated through non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software validation/regression (IEC 62304), system functional testing, hardware design verification, reliability testing, and packaging/labeling verification (ASTM D4169-22; ISO 15223-1).

Technological Characteristics

Low-energy direct-current (LEDC) generator; software-controlled. Supports up to six electrode probes. Features adjustable pulse waveforms (volts/cm, pulse length, pulse count, timing). Connectivity includes ECG synchronization. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, ANSI C63.27, IEC TS 60601-4-2, ASTM D4169-22, ISO 15223-1. Cybersecurity features include threat assessment and attack surface minimization per June 2025 FDA guidance.

Indications for Use

Indicated for surgical ablation of soft tissue, including prostate tissue.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 24, 2026 Angiodynamics Brandon Brackett Senior Manager, Regulatory Affairs 26 Forest Street Suite 100 Marlborough, Massachusetts 01752 Re: K253198 Trade/Device Name: NanoKnife Generator (H78720300351US0) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and accessories Regulatory Class: II Product Code: OAB Dated: February 13, 2026 Received: February 13, 2026 Dear Brandon Brackett: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253198 - Brandon Brackett Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K253198 - Brandon Brackett Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark R. Kreitz -S for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253198 | ? | | Please provide the device trade name(s). | | ? | | NanoKnife Generator (H78720300351US0) | | | | Please provide your Indications for Use below. | | ? | | The NanoKnife System with six outputs is indicated for surgical ablation of soft tissue, including prostate tissue. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K253198, Page 1 of 2 | 510(k) #: K253198 | 510(k) Summary | Prepared on: 2026-03-11 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | Angiodynamics | | | Applicant Address | 26 Forest Street Suite 100 Marlborough MA 01752 United States | | | Applicant Contact Telephone | 978-493-7393 | | | Applicant Contact | Mr. Brandon Brackett | | | Applicant Contact Email | bbrackett@angiodynamics.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | NanoKnife Generator (H78720300351US0) | | | Common Name | NanoKnife System | | | Classification Name | Low Energy Direct Current Thermal Ablation System | | | Regulation Number | 878.4400 | | | Product Code(s) | OAB | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K242687 | NanoKnife System | OAB | | Device Description Summary | | 21 CFR 807.92(a)(4) | | The subject device is substantially equivalent to the predicate device. The NanoKnife Generator is part of the NanoKnife System, and is a software-controlled low-energy direct-current (LEDC) generator which surgically ablates soft tissue, including prostate tissue. Included for use within the NanoKnife System are the NanoKnife Single Electrode Probes and optional Probe Spacer. With the NanoKnife System, a voltage applied between pairs of probes in a series of pulses. The waveform of the voltage is adjustable as determined by clinician-chosen parameters. These parameters include volts/cm, pulse length, number of pulses to be delivered between electrode pairs, the distance between probes, and the timing mode (90PPM or ECG synchronization). Up to six probes may be placed in an array within the tissue. The probes of the array are matched as pairs by the system. When probes are activated via a foot-pedal, the scheduled voltage is delivered to tissue between subsequent pairs of probes. Soft tissue (including prostate tissue) between the probes is ablated. | | | | Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) | | The NanoKnife System with six outputs is indicated for surgical ablation of soft tissue, including prostate tissue. | | | | Indications for Use Comparison | | 21 CFR 807.92(a)(5) | | The Indications for Use for the predicate NanoKnife Generator and the proposed NanoKnife Generator are the same, and are not impacted by this proposed 510(k). | | | | Technological Comparison | | 21 CFR 807.92(a)(6) | | The proposed NanoKnife Generator differs slightly as compared to the predicate NanoKnife Generator, as the proposed NanoKnife | | | {5} K253198, Page 2 of 2 Generator has been updated with newly-selected components. While the components between the two may differ, the proposed and predicate devices remain substantially equivalence to one another per the definition within FDA Guidance. Specifically, the modifications do not impact the overall design of the product, principle of operation, energy source(s), mechanism of action, or any parameters associated with the use of the NanoKnife Generator. Functionality, safety, and performance have been demonstrated to not be impacted by means of a robust regiment of applicable testing, including but not limited to EMC testing, software testing, and design verification testing. # Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) In support of the proposed NanoKnife 3.5 Generator, Angiodynamics conducted a robust series of non-clinical testing demonstrating the safety and efficacy of the device. This testing included: 1) Basic Safety and Essential Performance (Electrical Safety) Testing (IEC 60601-1; IEC 60601-1-2) 2) EMC and Software Testing (IEC 60601-1; IEC 60601-1-2; IEC 62304; ANSI C63.27; IEC TS 60601-4-2, FDA Guidance for Industry and Food and Drug Administration Staff, "Electromagnetic Compatibility (EMC) of Medical Devices) 3) GUI Software Validation Testing (IEC 62304) 4) GUI Software Regression Testing (IEC 62304) 5) Generator System Functional Testing 6) Hardware Design Verification Testing 7) Generator Reliability Design Verification Testing 8) Generator Packaging and Labeling Design Verification Testing (ASTM D4169-22; ISO 15223-1) All testing demonstrated acceptable results and supports the conclusion that the proposed NanoKnife 3.5 Generator is safe and effective for its intended use. Not applicable - no Clinical Tests submitted. AngioDynamics, Inc. has also assessed the proposed NanoKnife 3.5 Generator from a cybersecurity perspective. Specifically, the proposed NanoKnife 3.5 Generator shares the same fundamental cybersecurity features as the predicate NanoKnife 3.0 Generator. The primary cybersecurity design objective was to minimize the system's attack surface. AngioDynamics, Inc. developed a Threat Assessment and implemented its findings in order to achieve this goal. Lastly, AngioDynamics, Inc. complies with "Guidance for Industry and Food and Drug Administration Staff - Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" (June 2025) Per the Substantial Equivalence discussion pertaining to the proposed NanoKnife 3.5 Generator, Angiodynamics has demonstrated via the testing it has conducted on the subject device that it is equivalently safe, effective, and performs as well as the predicate NanoKnife 3.0 Generator. The testing regiment of the proposed NanoKnife 3.5 Generator was design to ensure a side-by-side comparison of tests and results was possible, in order to arrive at the conclusions described above.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...