TheraVision Ultrasound Ablation System and ACOUSTx Applicators
K150019 · Acoustic Medsystems, Inc. · NTB · Jan 13, 2016 · General, Plastic Surgery
Device Facts
Record ID
K150019
Device Name
TheraVision Ultrasound Ablation System and ACOUSTx Applicators
Applicant
Acoustic Medsystems, Inc.
Product Code
NTB · General, Plastic Surgery
Decision Date
Jan 13, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TheraVision Ultrasound Ablation System, ACOUSTx Applicators, and accessories are intended for the laparoscopic, intraoperative, and percutaneous coagulation and ablation of soft tissue. It is not indicated for ablation of prostate tissue.
Device Story
System uses PC-based RF generator to drive piezoelectric transducers in sterile, single-use ACOUSTx applicators; applicators deliver ultrasound energy to induce thermal ablation in soft tissue. System includes cooling pump for degassed sterile water circulation through applicators for transducer cooling and acoustic coupling. Operates via GUI allowing clinicians to plan treatment, control power, and monitor temperature via integrated thermocouples. Compatible with external DICOM imaging (MRI, CT, US, Fluoroscopic X-Ray, PET-CT) for guidance. Used in OR/clinical settings by physicians. Provides thermal therapy for soft tissue lesions; benefits include directional control (180°/360° patterns) and adjustable ablation lengths via multi-transducer activation, allowing tailored treatment to targeted tissue volumes.
Clinical Evidence
No human clinical data. Evidence includes ex vivo testing (chicken breast, porcine liver/muscle/kidney) and in vivo porcine model testing. Evaluated ablation patterns, penetration depths (1.4–3.4 cm), and lengths (1.1–4.7 cm) across 5–11.5 MHz frequency range and 3–9 minute treatment times. Confirmed comparable lesion patterns to predicate devices.
Indicated for laparoscopic, intraoperative, and percutaneous coagulation and ablation of soft tissue. Contraindicated for prostate tissue ablation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Sonatherm 600i Ultrasonic Lesion Generating System (K070779)
Valleylab Microwave Ablation Generator and Microwave Antennas (K072687)
Valleylab Microwave Ablation Generator and Microwave Antennas (K011676)
Valleylab Microwave Ablation Generator and Microwave Antennas (K032702)
Related Devices
K200061 — HEATfx Microwave Ablation System, HEATfx Microwave Ablation System Cart, HEATfx 2.1 Needle, HEATfx 2.1 S Needle, Multi-Point Thermal Sensor (MTS) · Galil Medical, Inc. · Jul 9, 2020
K143203 — AveCure Microwave Coagulation-Ablation Generator, AveCure Microwave Large Antenna-Probe, AveCure Microwave Medium Antenna-Probe, AveCure Microwave Small Antenna-Probe, AveCure Microwave Mini Antenna-Probe · Medwaves Incorporated · Aug 26, 2015
K122217 — CERTUS 140 2.45GHZ ABLATION SYSTEM · Neuwave Medical, Inc. · Dec 19, 2012
K113237 — CERTUS 140 2.45GHZ ABLATION SYSTEM · Neuwave Medical, Inc. · Jan 4, 2012
K190052 — Radio Frequency Ablation System · Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. · Jun 25, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines above them that could be interpreted as hair or wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2016
Acoustic Medsystems Incorporated Ms. Yvonne Schleife Regulatory and Quality Systems Manager 208 Burwash Avenue Savoy, Illinois 61874
Re: K150019
Trade/Device Name: TheraVision Ultrasound Ablation System and ACOUSTx Applicators Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NTB Dated: November 23, 2015 Received: November 30, 2015
Dear Ms. Schleife:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150019
#### Device Name
Thera Vision Ultrasound Ablation System and ACOUSTx Applicators
Indications for Use (Describe)
The Thera Vision Ultrasound Ablation System, ACOUSTx Applicators, and accessories are intended for the laparoscopic, intraoperative, and perculation and ablation of soft tissue. It is not indicated for ablation of prostate tissue.
Type of Use (Select one or both, as applicable)
| > Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### 1.0 Submitter Information
Acoustic MedSystems, Inc. 208 Burwash Avenue Savoy, IL 61874 Contact: Yvonne Schleife Regulatory and Quality Systems Manager Phone: +1 (217) 239-0900 Fax: +1 (217) 239-0905 E-mail: yschleife@acousticmed.com Date summary prepared: 16 November 2015
#### 2.0 Name of Device
Trade Name: TheraVision™ Ultrasound Ablation System and ACOUSTx™ Applicators
Common/Classification Name: Electrosurgical cutting and coagulation device and accessories
Regulation Number: 878.4400
Device Class: II
Product Code: NTB
Reason for submission: New Device
K Number: K150019
#### 3.0 Predicate Devices
The TheraVision System and ACOUSTx Applicators are substantially equivalent to the following legally marketed devices:
- . Sonatherm 600i Ultrasonic Lesion Generating System [K070779] - Primary Predicate
- Cool-tip™ RF Ablation System [K053290] .
- Valleylab Microwave Ablation Generator and Microwave Antennas [K072687 and ● K011676/K032702, respectively]
{4}------------------------------------------------
### 4.0 Device Description
The TheraVision System consists of the following subsystems:
- 4.1 Computer System: The computer system consists of a PC Compatible type computer running Microsoft Windows® Operating system.
- 4.2 Visualization Options: Visualization options include utilization with compatible, legally marketed imaging systems. Patient images may be imported in DICOM format from MRI, CT, US, Fluoroscopic X-Ray, and PET-CT scanners.
- 4.3 Software: A set of user interface and display modules allow user interaction and high-level control over the system, including power generation and safety monitoring. Top-level modules provide top level menu options and implement next-generation Microsoft Windows® display functionality.
- 4.4 Thermometry System: Temperature monitoring is accomplished using an optional multichannel type T patient isolated thermocouple data acquisition system.
- 4.5 RF Generation System: A Multichannel RF power generator system provides the RF enerqy needed drive the piezoelectric transducers for generation of sufficient ultrasound acoustic energy from the applicators to raise the temperature to therapeutic levels in the targeted tissue
- 4.6 Cooling System: Degassed sterile water is pumped through the ultrasound transducers in the applicators and provides ultrasound energy coupling to tissue and also provides a means for cooling the applicator.
## ACOUSTx Applicators:
The ultrasound energy generated from ACOUSTx Applicators is absorbed during treatment producing thermal therapy within the targeted tissue region. The ACOUSTx Applicators are sterile and single-use and can be utilized with the TheraVision System and may contain up to four transducers each with angular insonation patterns of 360° or 180° for application of thermal therapy.
#### 5.0 Indications for Use
The TheraVision Ultrasound Ablation System, ACOUSTx Applicators, and accessories are intended for the laparoscopic, intraoperative, and percutaneous coagulation and ablation of soft tissue. It is not indicated for ablation of prostate tissue.
{5}------------------------------------------------
#### 6.0 Summary of Device Characteristics
TheraVision System: The TheraVision System includes a multichannel RF power generator system, cooling pump, monitor, power supply, mouse and keyboard. The TheraVision System includes custom software with a graphic user interface (GUI) that allows users to create or adjust treatment plans, control and direct the application of ultrasound energy to targeted areas, and review the resulting treatment data. A single use sterile cooling circuit is external to the TheraVision System and connect to the circulating cooling pump, water bag and applicators. Pumping sterile room temperature water through and around the ACOUSTx Applicators is needed for ultrasonic energy coupling and also provides an effective means for cooling the ultrasound transducers during power-on operation. Optional use of an imaging device can provide additional anatomical information for device guidance and monitoring.
ACOUSTx Applicators: During treatment, single or multiple applicators may be utilized simultaneously, depending on the volume of the targeted region. ACOUSTx applicators do not directly contact the patient due to insertion within a sterile, biocompatible, legally marketed closed-tip interstitial plastic catheter. Degassed sterile water flows through an embedded channel in the applicator and the return water exits the catheter by flowing back out of the catheter on the exterior surface of the applicator within the interstitial catheter. An embedded thermocouple in each applicator is provided as a water flow sensor, and applicator operating temperature indication. One transducer of an ACOUSTx Applicator can deliver an acoustic output power of up to 20 Watts.
#### 7.0 Safety and Performance Data
The TheraVision System, ACOUSTx Applicators and accessories were utilized to apply thermal therapy in both ex vivo and in vivo testing, confirming the device's safety and effectiveness. In vivo testing was performed using a porcine model and three tissue types were treated: liver, muscle, and kidney. Ex vivo testing was performed using 360° and 180° ACOUSTx Applicator models to apply thermal therapy to both chicken breast tissue and porcine tissue: liver, muscle, and kidney.
360° and 180° ACOUSTx Applicator models were also utilized for in vivo testing in order to apply thermal therapy between the frequency ranges of 5 to 11.5 MHz for the porcine tissue types (liver, muscle, and kidney) for different treatment times (3, 6, and 9 minutes). The following table provides a summary of the lesion pattern ranges (details provided in
{6}------------------------------------------------
submission). These results confirmed comparable ablation patterns and times with respect to the three listed predicate devices.
| Tissue & Treatment<br>Pattern | Pentratation Depth | Pattern Length<br>[For 1,2,3, or 4 Active<br>Transducers] |
|-------------------------------|--------------------|-----------------------------------------------------------|
| Liver / 360° | 1.4 - 3.4 cm | 1.1 - 4.7 cm |
| Muscle / 360° | 1.63 - 3.4 cm | 1.2 - 4.7 cm |
| Kidney / 360° | 1.5 – 2.7 cm | 1.1 - 2.5 cm* |
| l | Penetration Depth | Length |
| Liver / 180° | 1.9 – 2.8 cm | 1.1 - 4.6 cm |
| Muscle / 180° | 1.8 - 3.0 cm | 1.1 - 4.6 cm |
| Kidney / 180° | 1.6 - 2.7 cm | 1.1 - 2.4 cm* |
| Table 1 - Summary: Ablation Dimension Range over Frequency Range 5-11.5 MHz | |
|-----------------------------------------------------------------------------|--|
| at 6W acoustic power (For 3, 6, and 9 Minute Insonation Times) | |
*1 or 2 active transducers longitudinally
One to four transducers can be activated for any given ACOUSTx Applicator model and this allows the user to select different ablation pattern lengths as is most appropriate to the targeted area. Different active angular sectors provide directional control. This and other treatment tailoring allows for improved patient safety. The above ablation patterns were compared to the ablation patterns of the predicate devices (Sonatherm, Cool-tip, and Valleylab) which demonstrates the TheraVision System and ACOUSTx Applicators create comparable lesion patterns at comparable treatment times. Sonatherm penetration depth is up to 3 cm. Both ACOUSTx angular sector (180°) applicator and Sonatherm produce wedge-shaped ablation patterns. Cool-tip's ablation diameter is 2.3 – 2.6 cm at lengths of 2.6 – 3.7 cm. Valleylab's (Evident) ablation diameter is 1.8 - 3 cm at lengths of 3.9 - 5 cm. ACOUSTx 360° applicators, Cool-tip and Valleylab (Evident) all produce an ellipsoidal pattern. This data support the substantial equivalence of the TheraVision System and ACOUSTx Applicators to its predicates.
{7}------------------------------------------------
Other safety, performance, and bench testing was also conducted. Verification and validation testing included safety and performance testing of software and dosimetry testing in addition to risk evaluation and ship testing. This device was also tested to and complies with IEC 60601-1, IEC 60601-1-2, IEC 61000-3-2, and IEC 61000-3-3. The results from bench, animal, performance, and safety testing demonstrate the safety and efficacy of this device's intended use and technology, as well as its substantial equivalence to the predicates.
#### 8.0 Conclusion
The TheraVision Ultrasound Ablation System, ACOUSTx Applicators, and accessories do not raise new questions of safety or efficacy and is substantially equivalent to the predicate devices based upon the intended use, operating characteristics, and outcomes.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.