K240784 · Varian Medical Systems, Inc. · NEY · May 29, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240784
Device Name
Temperature Sensor Probe (ABL-18TP20)
Applicant
Varian Medical Systems, Inc.
Product Code
NEY · General, Plastic Surgery
Decision Date
May 29, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Indications for Use
The Temperature Sensor Probe, used with Varian ablation systems, is intended for monitoring tissue temperature near the ablation site.
Device Story
Temperature Sensor Probe is an optional, single-use, disposable accessory for Varian microwave ablation systems. It consists of a needle with a T-type thermocouple at the tip, a handle containing a PCBA, and a lead cable. The device is inserted into tissue near the ablation site by a clinician; the thermocouple measures tissue temperature in real-time. The handle's PCBA converts analog temperature signals to digital, which are then transmitted to the ablation system console for continuous display. This real-time monitoring allows clinicians to observe temperature changes during ablation, potentially aiding in procedure control. The device is rigid, 18G, and 20cm in length, featuring a trocar tip that eliminates the need for a separate cannula and stylet used in predicate designs.
Clinical Evidence
No clinical data or testing included. Performance validated via non-clinical bench testing (accuracy, reliability, functional support) and porcine soft tissue animal models. Biocompatibility evaluated per ISO 10993-1; electrical safety/EMC per IEC 60601 series; sterilization per ISO 11135.
Technological Characteristics
Materials: Stainless steel (needle), ABS (handle), PVC (cable jacket). Sensing: T-type thermocouple. Form factor: 18G, 20cm rigid needle with trocar tip. Connectivity: Wired to ablation system console. Sterilization: 100% ethylene oxide. Standards: ISO 10993-1, ISO 11135, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-6.
Indications for Use
Indicated for monitoring tissue temperature near the ablation site during ablation procedures using Varian ablation systems. Not intended for cardiac procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Ablation Remote Temperature Probe (RTP20) (K200796)
Related Devices
K031556 — VIVATHERM THEMPERATURE MEASUREMENT SYSTEM · Vivant Medical, Inc. · Jun 23, 2003
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
May 29, 2024
Varian Medical Systems, Inc. Lynn Allman Sr. Director Regulatory Affairs 3100 Hansen Way Palo Alto, California 94304
Re: K240784
Trade/Device Name: Temperature Sensor Probe (ABL-18TP20) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY Dated: March 20, 2024 Received: March 21, 2024
Dear Lynn Allman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm____identifies ____combination_ product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long H. Chen -- S Digitally signed by Long H. Chen -S
Date: 2024.05.29 Date: 2024.05.29 13:06:28 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240784
Device Name Temperature Sensor Probe
Indications for Use (Describe)
The Temperature Sensor Probe, used with Varian ablation systems, is intended for monitoring tissue temperature near the ablation site.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains three blue circles arranged horizontally. The circles are evenly spaced and appear to be the same size. The background is plain white, which makes the blue circles stand out.
Image /page/4/Picture/2 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "Varian" is in large, bold, black font. Below the word "Varian" is the text "A Siemens Healthineers Company" in a smaller font size.
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
## PREMARKET NOTIFICATION
510(k) Summary
### Temperature Sensor Probe
The following information is provided as required by 21 CFR 807.92
### Submitter's Information: l.
| Name and Address: | Varian Medical Systems Inc.<br>9825 Spectrum Drive, Building 2 Austin, TX 78717 |
|-------------------|---------------------------------------------------------------------------------|
| Contact Name: | Lynn Allman, Senior Director Regulatory Affairs |
| E-mail: | submissions.support@varian.com |
| Date Prepared: | May 29, 2024 |
### Device Information: ll.
| Proprietary Name: | Temperature Sensor Probe |
|-----------------------|----------------------------------------------------------------|
| Common/ Usual Name: | System, Ablation, Microwave and Accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Classification Panel: | General & Plastic Surgery |
| Regulation Number: | 21 CFR 878.4400 |
| Product Code | NEY |
#### III. Predicate Device 510(k):
Emprint™ Ablation System with Thermosphere Technology: K200796
Predicate Device within K200796 identified to claim Substantial Equivalence: Ablation Remote Temperature Probe (RTP20).
#### IV. Subject Device Description:
Temperature Sensor Probe is an optional accessory designed for use with compatible Varian ablation systems. Principal components of the temperature sensor probe include:
- Needle ●
- . Handle
- . Lead cable and connector
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a sequence of alphanumeric characters, specifically 'K240784'. The characters are presented in a clear, sans-serif font, making them easily readable. The arrangement suggests a code, serial number, or identifier.
Image /page/5/Picture/1 description: The image shows three blue circles in a row. The circles are all the same size and color. The circles are evenly spaced apart from each other. The background is white.
Image /page/5/Picture/2 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "Varian" is written in a bold, sans-serif font. Below the word "Varian" is the text "A Siemens Healthineers Company" in a smaller, sans-serif font. The logo is black and white.
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
### V. Intended Use/ Indications for Use Statement:
The Temperature Sensor Probe, used with Varian ablation systems, is intended for monitoring tissue temperature near the ablation site.
#### Substantial Equivalence Discussion: VI.
The following table compares the subject device to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows three blue circles in a row. The circles are evenly spaced and of the same size. The background is white.
Image /page/6/Picture/1 description: The image shows a sequence of alphanumeric characters, specifically 'K240784'. The characters are rendered in a simple, sans-serif font, and they are all aligned horizontally. The image appears to be a close-up view of the characters, with a plain white background.
Image /page/6/Picture/2 description: The image shows the word "varian" in a bold, sans-serif font. Below the word "varian" is the text "A Siemens Healthineers Company" in a smaller font. The text is all black and the background is white.
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
Table 1: Summary of the comparison of technological characteristics between the predicate and subject devices
| Features/<br>Characteristics | Subject Device<br>Temperature Sensor Probe | Predicate Device (K200796)<br>Ablation Remote Temperature Probe | Differences |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>representative image | Image: Temperature Sensor Probe | Image: Ablation Remote Temperature Probe | |
| Intended Use/<br>Indications for Use | The Temperature Sensor Probe, used with Varian ablation systems, is intended for monitoring tissue temperature near the ablation site. | The RTP20 remote temperature probe is for use with the Emprint Ablation System with Thermosphere Technology to monitor tissue temperature at or near the ablation site.<br><br>The remote temperature probe can be used with the Emprint Ablation Generator with Thermosphere Technology to automatically disable microwave | Both the subject and the predicate devices <b>have the same intended use of monitoring tissue temperature in ablation procedures</b> . The predicate's additional design feature mentioned in the indications of use statement has no impact on the overall intended use/ indications of use of the device. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the alphanumeric string "K240784" in a simple, sans-serif font. The characters are black against a white background. The string appears to be a code or identifier.
Image /page/7/Picture/1 description: The image shows three blue circles in a horizontal row. The circles are evenly spaced and appear to be the same size. The background is white, which makes the blue circles stand out. There are no other objects or elements in the image.
# varian
A Siemens Healthineers Company
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
| Features/<br>Characteristics | Subject Device | Predicate Device (K200796) | Differences |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| | Temperature Sensor Probe | Ablation Remote Temperature Probe<br>energy when it measures a<br>temperature of 45° C (default<br>setting). Use of the RTP20 in an<br>ablation procedure is optional. | |
| Compatible Systems<br>and their intended<br>use/ indications of<br>use | IntelliBlate Microwave Ablation<br>System under FDA review (K240480)<br>The IntelliBlate Microwave Ablation<br>System is intended for coagulation<br>(ablation) of soft tissue in<br>laparoscopic, intraoperative, and<br>percutaneous ablation procedures,<br>including partial or complete<br>ablation of non-resectable liver<br>tumors. | Emprint™ Ablation System with<br>Thermosphere Technology<br>The Emprint™ Ablation System is<br>intended for use in percutaneous,<br>laparoscopic, endoscopic, and<br>intraoperative coagulation (ablation)<br>of soft tissue, including partial or<br>complete ablation of nonresectable<br>liver tumors. | Same |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows three blue circles in a row. The circles are evenly spaced and of the same size. The background is white.
# varian
A Siemens Healthineers Company
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
| Features/<br>Characteristics | Subject Device<br>Temperature Sensor Probe | Predicate Device (K200796)<br>Ablation Remote Temperature Probe | Differences |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| | The IntelliBlate Microwave Ablation System is not intended for use in cardiac procedures. | The Emprint™ Ablation System is not intended for use in cardiac procedures. | |
| Operating Principle | Measures the tissue temperature in real time via a thermocouple within the tip of the probe. | Measures the tissue temperature in real time via a thermocouple within the tip of the probe. | Same |
| Product code, FDA<br>regulation and device<br>class | NEY, 21 CFR 878.4400 and Class II | NEY, 21 CFR 878.4400 and Class II | Same |
| Manufacturer | Varian Medical Systems | Covidien | Different manufacturer |
| Prescription (Rx only) | Yes | Yes | Same |
| Optional Accessory | Yes | Yes | Same |
| Measuring Site | Tissue near ablation site | At or near ablation site | Similar |
| Temperature Range<br>(Temperature<br>Accuracy) | $10°C-60°C (±2°C)$ | $37°C-60°C (±1.5°C)$ | Similar performance<br>specifications.<br>Representation of the<br>reported temperature on<br>the User Interface of the<br>compatible ablation |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the alphanumeric string "K240784" in a simple, sans-serif font. The characters are uniformly sized and spaced, creating a clear and legible sequence. The string appears to be a code or identifier, possibly a serial number or product key.
Image /page/9/Picture/1 description: The image shows three blue circles in a horizontal row. The circles are evenly spaced and of the same size. The background is white.
varian
A Siemens Healthineers Company
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
| Features/<br>Characteristics | Subject Device<br>Temperature Sensor Probe | Predicate Device (K200796)<br>Ablation Remote Temperature Probe | Differences |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| | | | system used with the<br>subject device are rounded<br>to nearest integer to<br>enhance usability. |
| Thermocouple-based<br>temperature sensor<br>at the tip of the<br>needle | Yes | Yes | Same |
| Signal Processing and<br>Display on the<br>compatible ablation<br>system | Real time, continuous display of<br>tissue temperature | Real time, continuous display of<br>tissue temperature | Same |
| Tissue Contact per<br>ISO 10993-1 | External communicating device, in<br>contact with tissue/bone/dentin and<br>limited contact duration (A) (<24<br>hours) | External communicating device, in<br>contact with tissue/bone/dentin and<br>limited contact duration (A) (<24<br>hours) | Same |
| Design and Construction | | | |
| Principle<br>Components | Needle, handle and lead cable | Needle, handle, lead cable and<br>cannula with stylet | Thicker needle shaft of the<br>subject device eliminates<br>the use of a cannula |
| Rigid Construction | Yes | Yes | Same |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows three blue circles in a horizontal row. The circles are evenly spaced and appear to be the same size. The background is white, which makes the blue circles stand out.
# varian
A Siemens Healthineers Company
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
| Features/<br>Characteristics | Subject Device<br>Temperature Sensor Probe | Predicate Device (K200796)<br>Ablation Remote Temperature Probe | Differences |
|----------------------------------------|--------------------------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Distal Tip<br>configuration | Trocar Tip | Rounded non-sharp tip on<br>Temperature Probe used together<br>with a sharp cannula | Sharp Trocar Tip replaces<br>the use of a cannula |
| Right Angle Probe | Yes | Yes | Same |
| Needle Shaft length | 20cm | 20 cm | Same |
| Needle Gauge<br>inserted in the tissue | 18G | 18G<br>Cannula (18G) with stylet inserted<br>with 20.5G needle. | Similar.<br>Overall gauge size is the<br>same, however the<br>predicate uses a cannula<br>with a stylet and thinner<br>probe needle.<br>Thicker wall of the subject<br>device's needle shaft<br>provides mechanical<br>stiffness that eliminates<br>the use of a cannula and<br>stylet. |
| Cannula and Stylet | No | Yes | Thicker wall of the needle<br>shaft of the subject device<br>provides mechanical<br>stiffness that eliminates<br>the use of a cannula and<br>stylet. |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the alphanumeric string "K240784" in a clear, sans-serif font. The characters are uniformly sized and spaced, making the entire string easily readable. The string appears to be a code or identifier, possibly a serial number or part number.
Image /page/11/Picture/1 description: The image contains three blue circles in a horizontal row. The circles are evenly spaced and appear to be the same size. The background is plain white, which makes the blue circles stand out. The image is simple and minimalist in design.
# varian
A Siemens Healthineers Company
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
| Features/<br>Characteristics | Subject Device<br>Temperature Sensor Probe | Predicate Device (K200796)<br>Ablation Remote Temperature Probe | Differences |
|---------------------------------------------------------------|--------------------------------------------|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Needle<br>Thermocouple | Soldered at the tip | Soldered at the tip | Same |
| Patient Contacting<br>Components | Needle Shaft and needle tip | Needle: Needle Shaft and needle tip<br>Cannula: Cannula tip<br>Stylet: Stylet tip | Same.<br>Thicker wall of the needle<br>shaft of the subject device<br>provides mechanical<br>stiffness that eliminates<br>the use of a cannula and<br>stylet |
| Probe handle with<br>Printed Circuit Board<br>Assembly (PCBA) | Yes | No | Different design.<br>The predicate device<br>converts the probe analog<br>temperature to digital<br>signal inside the console<br>and sends the information<br>to the display.<br>The subject device<br>converts the probe analog<br>temperature to digital<br>signal in the handle (PCBA)<br>and sends the information<br>to the console display. |
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image contains three blue circles arranged horizontally. The circles are evenly spaced and appear to be the same size. The background is white, which makes the blue circles stand out. There are no other objects or elements in the image.
# varian
A Siemens Healthineers Company
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
| Features/<br>Characteristics | Subject Device<br>Temperature Sensor Probe | Predicate Device (K200796)<br>Ablation Remote Temperature Probe | Differences |
|---------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Patient Contacting<br>Materials | Stainless Steel | Stainless Steel | Same |
| Lead Cable Outer<br>Jacket | PVC | TPV | Different polymer;<br>thermoplastic elastomer<br>similar to the predicate<br>device. |
| Handle Plastic | ABS | ABS | Same |
| Thermocouple | T-Type | T-Type | Same |
| Safety Standard<br>Compliance | IEC 60601-1, 60601-1-6, 60601-2-6,<br>60601-1-2 | IEC 60601-1, 60601-1-6, 60601-2-6,<br>60601-1-2 | Same |
| Biocompatibility | All the patient-contacting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO 10993-1. | All the patient-contacting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO 10993-1. | Same |
| Software | No | No | Same |
| Device Usage | Single-use, disposable, optional<br>accessory | Single-use, disposable, optional<br>accessory | Same |
| Sterilization | ETO | ETO | Same |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows three blue circles in a horizontal line. The circles are evenly spaced and appear to be the same size. The background is white.
# varian
A Siemens Healthineers Company
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
| Features/ | Subject Device | Predicate Device (K200796) | Differences |
|---------------------------------------------|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Characteristics | Temperature Sensor Probe | Ablation Remote Temperature Probe | |
| Packaging<br>configuration and<br>materials | Tyvek sterile barrier pouch within a<br>cardboard shelf carton box sealed<br>with a tamper tape. | Thermoform tray and single Tyvek<br>sterile barrier lid within a SBS (solid<br>bleached sulfate) shelf carton box<br>sealed with a tamper paper tape. | Different packaging (less<br>footprint) |
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image shows three blue circles in a row. The circles are evenly spaced and of the same size. The background is white.
Image /page/14/Picture/2 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "varian" is written in a bold, sans-serif font. Below the word "varian" is the text "A Siemens Healthineers Company" in a smaller font. The logo is black and white.
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
### Performance Data: VII.
## 1. Non-clinical Testing:
Temperature Sensor Probe underwent non-clinical testing to demonstrate the design and performance of the device meets the established design criteria. The subject device successfully completed functional intended use support and reliability testing. To support substantial equivalence in performance with the predicate device, the subject and predicate devices underwent side-by-side testing for performance i.e accuracy of temperature monitoring. The test results established that the subject device is substantially equivalent to the predicate device in performance and is as safe and effective as the predicate device.
### 2. Non-clinical Animal Testing:
Animal testing completed with Temperature Sensor Probe to validate usability and performance of the device on porcine soft tissue models complies with the United States Food and Drug Administration Good Laboratory Practice (GLP) Regulations, 21 CFR Part 58.
### 3. Clinical Data/ Testing:
No clinical data/ testing is included within the submission.
### 4. Sterilization:
Temperature Sensor Probe is sterilized utilizing 100% ethylene oxide gas to ensure a minimum sterility assurance level of 106 in accordance with ISO 11135, ISO 10993-7, ISO 11138-1, ISO 11737-1 and ISO 11737-2.
### 5. Software, Cybersecurity, and Interoperability
Temperature Sensor Probe does not contain software. Software verification related to functionality of Temperature Sensor Probe with IntelliBlate Microwave Ablation System complies with IEC 62304 – "Medical device software – software life cycle processes "and FDA Guidance "Content of Premarket Submissions for Device Software Functions."
### 6. Temperature Monitoring:
Temperature monitoring testing completed in accordance with FDA Guidance "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" demonstrates that the temperature sensing under simulated conditions meets design specification and performance requirements.
### 7. Biocompatibility:
Biocompatibility testing of the Temperature Sensor Probe in accordance with ISO 10993-1 demonstrates that the patient-contacting components are biocompatible.
## 8. Electrical Safety and Electromagnetic Compatibility:
Temperature Sensor Probe complies with applicable IEC 60601 series of standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-2-6, and IEC 60601-1-2.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows three blue circles in a row. The circles are all the same size and color. The circles are evenly spaced apart from each other. The background is white.
Image /page/15/Picture/2 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "Varian" is written in a bold, sans-serif font. Below the word "Varian" is the text "A Siemens Healthineers Company" in a smaller, sans-serif font. The logo is in black and white.
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
## 9. Packaging Integrity and Shelf Life:
Packaging integrity and shelf-life testing complies with ISO 11607-1 and ISO 11607-2.
### VIII. Conclusion:
The performance data for the Temperature Sensor Probe support the safety of the device and the verification and validation demonstrate that the subject device should perform as intended in the specified use conditions. Varian considers Temperature Sensor Probe to be as safe and effective as the predicate device and therefore, substantially equivalent to the predicate device, Ablation Remote Temperature Probe cleared along with Emprint Ablation System in K200796.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.