K190245 · Pdo Max, Inc. · NEW · Jun 4, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190245
Device Name
PDS Barbed Suture, PDO MAXX Threads
Applicant
Pdo Max, Inc.
Product Code
NEW · General, Plastic Surgery
Decision Date
Jun 4, 2019
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4840
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PDS Barbed Surgical Suture is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate. The anatomical location(s) of use are on the skin for dermatological applications only. The suture is not intended for interior body cavity applications and the suture is not intended for lifting and supporting tissues.
Device Story
PDO MAXX Threads are absorbable polydioxanone (PDO) monofilament sutures with cut barbs; preloaded in 19G/100 mm blunt needle cannula. Used by clinicians for soft tissue approximation in dermatological applications. Barbs allow tissue embedding without surgical knots. Device is sterile, single-use, and non-powered. Clinical benefit derived from secure tissue approximation without knot tying.
Clinical Evidence
No clinical data. Evidence provided via bench testing (biocompatibility, tensile strength, dimension, absorption, needle performance) and in vivo biodegradation study in Sprague-Dawley rats.
Technological Characteristics
Material: Polydioxanone (PDO) with D&C Violet No. 2 colorant. Form: Monofilament with cut barbs, USP 0 diameter, USP 2-0 tensile strength, 150 mm length. Delivery: Pre-loaded in 19G/100 mm blunt needle cannula. Energy source: None. Sterilization: Not specified.
Indications for Use
Indicated for soft tissue approximation in dermatological applications requiring absorbable sutures. Not for interior body cavity use or tissue lifting/support.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
K240389 — Secret Line up (FCL18100C160 FCL19100E160 FCL1990E150 FCL2190G150 ); i-Thread (FCL18100C160 FCL19100E160 FCL1990E150 FCL2190G150) · Hyundae Meditech Co., Ltd. · Mar 27, 2024
K210871 — PDO Max Suture with Dual Needle · Pdo Max, Inc. · Dec 17, 2021
K201139 — WEGO-PDO Barbed Suture · Foosin Medical Supplies Inc., Ltd. · Oct 20, 2020
K143413 — J+PRAS YOUNG'S Thread · Y. Jacobs Medical, Inc. · Sep 11, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 4, 2019
PDO MAX, Inc. % Ms. Mary Vater Regulatory Consultant Medical Device Academy, Inc. 345 Lincoln Hill Rd. Shrewsbury, Vermont 05738
Re: K190245
Trade/Device Name: PDS Barbed Suture, PDO MAXX Threads Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable Polydioxanone Surgical Suture Regulatory Class: Class II Product Code: NEW Dated: March 6, 2019 Received: March 6, 2019
Dear Ms. Vater:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For David Krause, Ph.D. Acting Division Director Division of Infection Control and Plastic Surgery Devices Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190245
Device Name
PDO MAXX Threads Barbed Surgical Suture
Indications for Use (Describe)
The PDS Barbed Surgical Suture is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate. The anatomical location(s) of use are on the skin for dermatological applications only. The suture is not intended for interior body cavity applications and the suture is not intended for lifting and supporting tissues.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
K190245 pg. 1 of 3
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
SUBMITTER 1. PDO Max, Inc. 4971 Bear Road Liverpool. New York 13088 Tel: +1.315.708.9913
| Contact Person: | Giovanna McCarthy |
|-----------------|-------------------|
| Date Prepared: | 4 February 2019 |
- II. DEVICE Name of Device: PDO MAXX Threads Barbed Suture Classification Name: Suture, Surgical, Absorbable, Polydioxanone Regulation: 21 CFR §878.4840 Regulatory Class: Class II Product Classification Code: NEW
- III. PREDICATE DEVICE
| Predicate Manufacturer: | Feeltech Co., Ltd. |
|-------------------------|--------------------|
| Predicate Trade Name: | Miracu |
| Predicate 510(k): | K172602 |
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
PDO MAXX THREADS™ barbed sutures are made of polydioxanone (PDO) and have the colorant D&C Violet No. 2 (21 CFR 74.3602). The sutures are USP 0 suture in diameter and USP 2-0 suture in tensile strength, and 150 mm in length, swaged to a 19G/100 mm "L" Blunt needle and preloaded in a needle cannula. The sutures are inserted into the cannula/needle. The barbed sutures have cut barbs along the axis of the monofilament that allows the suture to embed in the tissue after the surgeon has placed it; therefore, there is no need to tie a surgical knot. The sutures are absorbable in the body.
#### V. INDICATIONS FOR USE
The PDS Barbed Surgical Suture is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate. The anatomical location(s) of use are on the skin for dermatological applications only.
The suture is not intended for interior body cavity applications and the suture is not intended for lifting and supporting tissues.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVCE
The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence:
- . Indications for Use – The predicate and subject device have almost indications for use; both are indicated for use in soft tissue approximation where use of absorbable sutures is appropriate,
{4}------------------------------------------------
K190245 pg. 2 of 3
specifically designed for use in the skin.
- . Materials - The predicate and subject device materials are identical.
- . Design - The predicate and subject device are equivalent in design. They are both made of the same basic monofilament with similar barbs cut into the monofilament. Both the predicate and subject devices are supplied to the user pre-loaded in a hollow needle.
- Energy Source Neither the predicate nor the subject device requires an energy source. ●
- Performance Testing - Both the predicate and subject devices were subjected to the same biocompatibility and performance tests listed below.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
Suture Biocompatibility Testing
- · Cytotoxicity
- · Irritation
- · Acute Systemic
- · Material Mediated Pyrogenicity
- · Genotoxicity, Carcinogenicity, and Reproductive Test
- Endotoxin
- · Sensitization
### Needle Biocompatibility Testing
- · Cytotoxicity
- · Acute Systemic
- · Material Mediated Pyrogenicity
- · Intracutaneous Reactivity Test
- · Skin Sensitization Test
- · Hemolysis Test
### Suture Performance Testing
- · Suture Absorption
- · Dimension Test
- · Tensile Strength
### Needle Performance Testing
- Inner/Outside Structure
- · Dimension Test
- · Elasticity Test
- · Flexural Rigidity Test
- Draw Test
- · Extraction Test
### Electrical Safety and EMC Testing
Electrical safety and EMC testing were not applicable.
### Animal Testing
- In Vivo Biodegradation in Sprague-Dawley Rat Absorption and Tensile Strength over . time
### Clinical Testing
Clinical testing was not required to demonstrate the safety and effectiveness.
{5}------------------------------------------------
K190245 pg. 3 of 3
# VIII. CONCLUSIONS
Based on a comparison of technological characteristics, indications for use, and performance data, it can be concluded that the proposed PDS Barbed Suture is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.