K955863 · Molnlycke/Scott Health Care · NAD · Mar 13, 1996 · General, Plastic Surgery
Device Facts
Record ID
K955863
Device Name
ALLDRESS MULTI LAYERED WOUND DRESSING
Applicant
Molnlycke/Scott Health Care
Product Code
NAD · General, Plastic Surgery
Decision Date
Mar 13, 1996
Decision
SN
Submission Type
Traditional
Regulation
21 CFR 878.4020
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
For the dressing and management of chronic/open wounds.
Device Story
Sterile multi-layered wound dressing; absorbent layer with non-adherent contact surface; adhesive backing layer secures dressing to skin. Used in clinical settings for wound management. Modified version of predicate; improved backing layer composition enhances wound barrier properties. Applied by clinicians to cover wounds; provides barrier against bacteria and viruses. Benefits patient by protecting wound site and managing exudate.
Clinical Evidence
Bench testing only. Biocompatibility: intracutaneous toxicity, systemic toxicity, muscle implantation, primary skin irritation, hemolysis, cytotoxicity, and delayed hypersensitivity. Performance: bacterial strike-through (S. aureus, P. aeruginosa) and viral penetration (phiX174 bacteriophage) testing conducted.
Indicated for patients with chronic or open wounds requiring dressing and management.
Regulatory Classification
Identification
An occlusive wound dressing is a nonresorbable, sterile or non-sterile device intended to cover a wound, to provide or support a moist wound environment, and to allow the exchange of gases such as oxygen and water vapor through the device. It consists of a piece of synthetic polymeric material, such as polyurethane, with or without an adhesive backing. This classification does not include an occlusive wound dressing that contains added drugs such as antimicrobial agents, added biologics such as growth factors, or is composed of materials derived from animal sources.
{0}
MAR 13 1996
510 (k) SUMMARY
K955863
# 1. ADMINISTRATIVE
Submitter: Mölnlycke/Scott Health Care
500 Baldwin Tower
Eddystone, PA 19022
Contact Person: Mike Drummond
(610) 499-3375
Date of Preparation: December 26, 1995
# II. DEVICE NAME
Proprietary Name: AllDress®
Common Name: Sterile Multi-Layered Wound Dressing
Classification Name: Wound Dressing
Regulatory Class: Unclassified
Product Code: 79FRO
# III. PREDICATE DEVICE
AllDress® Multi Layered Wound Dressing (Scott Health Care; K945206)
# IV. DEVICE DESCRIPTION
The AllDress® Multi Layered Wound Dressing is a sterile wound dressing composed of an absorbent layer with a non-adherent wound contact surface. The absorbent layer is bonded to an adhesive backing layer which secures the dressing to the skin around the wound. Dressings are supplied in sizes of 2¼" x 2¼" (1" x 1" wound contact surface); 6" x 6" (4" x 4" wound contact surface); 4" x 4" (2" x 2" wound contact surface); and 6" x 8" (4" x 6" wound contact surface). The wound contact side of the dressing is covered with scored siliconized release paper to permit aseptic application.
The AllDress® Multi Layered Wound Dressing represents a modified version of the currently marketed AllDress® Multi Layered Wound Dressing (K945206) and is identical to the predicate device with respect to function and intended use. to the Comfeel® Plus Ulcer Dressing (K941263). The composition of the backing layer of the modified product has been changed to improve the wound barrier properties of the dressing.
Biocompatibility of the modified device has been established by intracutaneous toxicity test (rabbit), systemic toxicity test (mice), muscle implantation test (rabbit), primary skin irritation test (rabbit), hemolysis test (In Vitro), cytotoxicity test (In Vitro), and delayed hypersensitivity test (guinea pig). In addition, bacterial penetration (strike-through) studies with S. aureus and P. aeruginosa and ASTM
Premarket Notification: AllDress Multi-Layered Wound Dressing (modified)
11
{1}
IV. DEVICE DESCRIPTION (continued)
viral penetration test with $\phi X174$ bacteriophage were conducted to demonstrate the effective performance of the dressing as a bacterial and viral barrier.
V. INTENDED USE
For the dressing and management of chronic/open wounds.
VI. CONCLUSION
The technological characteristics of the modified AllDress™ Multi Layered Wound Dressing are the same as to those of the predicate device, except for compositional changes intended to improve its wound barrier properties of the product. Biocompatibility and performance testing of the modified device have demonstrated safety as well as effective wound barrier performance. Accordingly, Mölnlycke/Scott Health Care has concluded that the modified device is safe and effective for its intended use and performs at least as well as the legally marketed predicate device.
Premarket Notification: AllDress Multi Layered Wound Dressing (modified)
12
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.