← Product Code [MPN](/submissions/SU/subpart-e%E2%80%94surgical-devices/MPN) · K212246

# Exofin Precision Pen (K212246)

_Chemence Medical, Inc. · MPN · Sep 9, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/MPN/K212246

## Device Facts

- **Applicant:** Chemence Medical, Inc.
- **Product Code:** [MPN](/submissions/SU/subpart-e%E2%80%94surgical-devices/MPN.md)
- **Decision Date:** Sep 9, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4010
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Exofin® Precision Pen is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. Exofin® Precision Pen may be used in conjunction with, but not in place of, deep dermal sutures.

## Device Story

Exofin Precision Pen is a sterile, single-use topical skin adhesive for wound closure. It consists of a 2-octyl cyanoacrylate monomer formulation in an aluminum tube, integrated into a silicone bulb and polypropylene pen body for improved ergonomics. The device is operated by a clinician in a clinical setting. The user applies the liquid adhesive to skin edges via a polyethylene-based applicator tip (connector, porous disc, soft elastomeric brush). Upon application, the adhesive polymerizes within minutes to form a film that holds skin edges together and acts as a microbial barrier. The device is intended to assist in precise application and prevent unintended migration of the adhesive. It does not replace deep dermal sutures.

## Clinical Evidence

No clinical data was conducted. Substantial equivalence is supported by bench testing, including mechanical applicator testing and established biocompatibility of the unchanged adhesive formulation. In-vitro studies previously demonstrated the microbial barrier properties of the adhesive film.

## Technological Characteristics

2-octyl cyanoacrylate monomer; D&C Violet #2 colorant. Applicator: polyethylene connector, porous disc, soft elastomeric brush. Housing: aluminum tube, silicone bulb, polypropylene pen body. Packaging: PETG blister pack, Tyvek® backer. Sterilization: two-stage dry heat and ethylene oxide (SAL 10^-6). Shelf life: 12 months.

## Regulatory Identification

A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.

## Predicate Devices

- Exofin® High Viscosity Topical Skin Adhesive ([K200264](/device/K200264.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The Department of Health & Human Services logo is a stylized depiction of an eagle and a human figure. The FDA acronym is in a blue square, and the words "U.S. FOOD & DRUG ADMINISTRATION" are in blue text to the right of the square.

September 9, 2021

Chemence Medical, Inc. Charnelle Thomas Director of Regulatory Affairs 200 Technology Drive Alpharetta, Georgia 30115

Re: K212246

Trade/Device Name: Exofin Precision Pen Regulation Number: 21 CFR 878.4010 Regulation Name: Tissue Adhesive Regulatory Class: Class II Product Code: MPN Dated: August 9, 2021 Received: August 10, 2021

Dear Charnelle Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lixin Liu, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212246

Device Name Exofin® Precision Pen

#### Indications for Use (Describe)

Exofin® Precision Pen is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, traumainduced lacerations. Exofin® Precision Pen may be used in conjunction with, but not in place of, deep dermal sutures.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </span>                 |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the word "CHEMENCE" in a bold, sans-serif font. To the left of the word is a blue hexagon-like shape. The shape is made up of three overlapping, rounded rectangles that create a three-dimensional effect. The overall design is clean and modern.

# Special 510(k) for Exofin® Precision Pen (K212246)

## 510(k) Summary

This 510(k) summary is prepared in accordance with 21 CFR 807.92.

# 1. Submitter

| Submitted by:    | Chemence Medical, Inc.<br>200 Technology Drive<br>Alpharetta, GA 30005-3926<br>Phone: 844-633-4583<br>Fax: 678-820-3320            |
|------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:  | Charnelle Thomas<br>Director of Regulatory Affairs<br>Chemence Medical, Inc.<br>Phone: 678-690-0760<br>Email: cthomas@chemence.com |
| Date of Summary: | September 8, 2021                                                                                                                  |

- 2. Device

| Device Proprietary Name: | Exofin® Precision Pen             |
|--------------------------|-----------------------------------|
| Common or Usual Name:    | Topical Skin Adhesive             |
| Classification Name:     | Tissue Adhesive (21 CFR 878.4010) |
| Regulatory Class:        | Class II                          |
| Product Code:            | MPN                               |

# 3. Predicate Device

Legally marketed devices to which equivalence is claimed:

| Device Name:      | Exofin® High Viscosity Topical Skin Adhesive |
|-------------------|----------------------------------------------|
| 510(k) Clearance: | K200264                                      |

{4}------------------------------------------------

# Special 510(k) for Exofin® Precision Pen (K212246)

## 4. Device Description

Exofin® Precision Pen is a sterile liquid topical skin adhesive containing a monomeric (2-octyl cyanoacrylate) formulation for rapid polymerization, and the colorant D&C Violet #2 which aids in visualization during application. The adhesive is provided in a 1.0g size, single-use, aluminum, collapsible tube that is fitted with a polyethylene-based applicator tip. The applicator tip consists of three components, a connector fitted with a self-puncturing cap, porous disc and soft elastomeric brush. The aluminum tube is housed within a silicone bulb that is connected to a polypropylene pen body and held by the end user during application. The adhesive, applicator tip, silicone bulb and pen body are packaged together in a (PETG) plastic blister pack and sealed with a labeled Tyvek® blister backer. A total of 12 units are packaged in a tray which is covered by a sleeve. When applied to the skin, the adhesive is distributed through the applicator tip in a syrup-like viscosity and polymerizes within minutes. The increased viscosity assists in the unintended placement of the adhesive during application due to migration of the liguid adhesive from the wound site. The silicone bulb and pen body of the Exofin® Precision Pen were designed to improve ergonomics during application. In-vitro studies have shown that Exofin® Precision Pen acts as a barrier to microbial penetration when the adhesive film remains intact. Clinical studies were not conducted to demonstrate microbial barrier properties.

## 5. Intended Use

Exofin® Precision Pen is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. Exofin® Precision Pen may be used in conjunction with, but not in place of, deep dermal sutures.

## 6. Comparison of Technological Characteristics with the Predicate Device

Most technological characteristics of Exofin® Precision Pen and the predicate device are the same. Both devices:

- are 2-octyl cyanoacrylate-based, rapid polymerizing, liquid adhesive formulations
- contain D&C violet #2 colorant to aid in visualization during application ●
- provide an applicator tip that comprises of a connector, porous disc and soft elastomeric . brush
- polymerizes within minutes of application ●
- maintain skin edge approximation and provide a bacterial barrier ●
- are sterilized by a two-stage process with a sterility assurance level of 10° ●

The difference between Exofin® Precision Pen and the predicate is the addition of a pen (silicone bulb and pen body) for improved ergonomics during application and a change to increase the blister size to accommodate the pen body. Additionally, there is an increase in the number of device units in each tray from 10 to 12. The adhesive formulation, aluminum tube and applicator tip remain unchanged from the predicate device, Exofin High Viscosity Topical Skin Adhesive (K200264). These differences do not raise different questions of safety and effectiveness.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the word "CHEMENCE" in a bold, sans-serif font. To the left of the word is a blue hexagon-like shape. The shape is made up of three curved lines that overlap each other. The overall image appears to be a logo.

## Special 510(k) for Exofin® Precision Pen (K212246)

#### 7. Performance Data

Testing was performed in accordance with the FDA Class II Special Controls Guidance Document:Tissue Adhesive for the Topical Approximation of Skin.

#### Performance Testing

The following tests were performed on Exofin® Precision Pen to demonstrate substantial equivalence:

- · Mechanical Applicator Testing
In these studies, Exofin® Precision Pen met all performance criteria. Because the adhesive, aluminum tube and applicator tip remain unchanged, no additional performance test were done.

#### Biocompatibility Testing

The pen of subject device does not come into direct contact with the patient or adhesive, therefore,biocompatibility is not required. Because the adhesive is unchanged, no additional biocompatibility tests were performed.

## Sterilization and Shelf Life

Exofin® Precision Pen is sterilized in a two-step process by dry heat and ethylene oxide gas at a sterility assurance level (SAL) of 10°. The shelf life of the device has been determined through both real time and accelerated aging studies. The data from these studies support a 12-month shelf-life.

#### 8. Conclusion

Exofin® Precision Pen was evaluated in accordance with the Class II Special Controls Guidance Document: Tissue Adhesive for the Topical Approximation of Skin. Exofin® Pen is substantially equivalent to Exofin® High Viscosity Topical Skin Adhesive with regard to indications for use, mechanism of action and performance characteristics. Both devices contain the same principle chemical ingredient, 2-octyl cyanoacrylate. Both devices were shown to be equivalent in performance testing. Therefore, the change of adding a silicone bulb and pen body, for improved ergonomics during application and a change to packaging dimensions to accommodate the pen size, do not raise different questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/MPN/K212246](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/MPN/K212246)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
