← Product Code [LEJ](/submissions/SU/subpart-e%E2%80%94surgical-devices/LEJ) · K220500

# Ergoline Vitality Total Light 3, Planet Fitness 50/4 TLT by Ergoline, Planet Fitness 42/3 TLT by Ergoline, Ergoline Vitality Hybrid Light 3, Ergoline Vitality Hybrid Light, Planet Fitness 50/4 HLT by Ergoline, Planet Fitness 46/3 HLT by Ergoline, Ergoline Vitality Pure Light 3, Ergoline Vitality Pure Light, Planet Fitness 50/4 PLT III by Ergoline, Planet Fitness 50/4 PLT by Ergoline, Planet Fitness 46/3 PLT by Ergoline, Planet Fitness 42/3 PLT by Ergoline (K220500)

_Jk-Holding GmbH · LEJ · Jul 1, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/LEJ/K220500

## Device Facts

- **Applicant:** Jk-Holding GmbH
- **Product Code:** [LEJ](/submissions/SU/subpart-e%E2%80%94surgical-devices/LEJ.md)
- **Decision Date:** Jul 1, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4635
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

This sunlamp product is intended exclusively for aesthetic tanning of the human skin, for one person at a time, at the age of 18 or above.

## Device Story

Sunlamp products (booths) for aesthetic tanning; utilize artificial UV radiation sources (UV lamps) and mechanical structures. Devices emit UV-A and UV-B radiation; some models incorporate red/blue light LEDs or red light lamps for synergistic light spectra. Operated by users in clinics or salons; user selects exposure time. Output is UV radiation for skin tanning. Benefits include rapid superficial tanning (UV-A) and long-term tanning (UV-B). Safety features include compliance with 21 CFR 1040.20 and IEC 60601-1/1-2 standards.

## Clinical Evidence

Bench testing only. Compliance with ISO 10993 (biocompatibility), IEC 60601-1 (electrical/mechanical safety), IEC 60601-1-2 (EMC), and 21 CFR 1040.20 (performance standard for sunlamp products).

## Technological Characteristics

Sunlamp booths; UV-A/UV-B lamps; some models include red/blue LEDs or red light lamps. Electrical: 230V 3-phase or 2-phase. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993, 21 CFR 1040.20. Connectivity: Standalone. Sterilization: N/A.

## Regulatory Identification

A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.

## Special Controls

*Classification.* Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
*Labeling* —(i)*Sunlamp products.* (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement:      Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
*Sunlamp products and UV lamps intended for use in sunlamp products.* Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
*Performance standard.* Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.

## Predicate Devices

- Planet Fitness 42/4 Hybrid Light Technology ([K190173](/device/K190173.md))
- Planet Fitness 42/4 ([K151400](/device/K151400.md))
- Ergoline Prestige 1600 Hybrid Technology ([K180555](/device/K180555.md))
- Affinity 900 AP TT AM/A USA ([K151400](/device/K151400.md))
- Affinity 700 PF AP AC USA ([K151400](/device/K151400.md))

## Reference Devices

- Sunrise 7200 Hybrid Light Technology ([K183539](/device/K183539.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

July 1, 2022

JK-Holding GmbH Juergen Gerstenmeier Official Applicant Koehlershohner Strasse 60 Windhagen, Rheinland-Pfalz 53578 Germany

Re: K220500

Trade/Device Name: Ergoline Vitality Total Light 3, Planet Fitness 50/4 TLT by Ergoline, Planet Fitness 42/3 TLT by Ergoline Vitality Hybrid Light 3, Ergoline Vitality Hybrid Light, Planet Fitness 50/4 HLT by Ergoline, Planet Fitness 46/3 HLT by Ergoline, Ergoline Vitality Pure Light 3, Ergoline Vitality Pure Light, Planet Fitness 50/4 PLT III by Ergoline, Planet Fitness 50/4 PLT by Ergoline, Planet Fitness 46/3 PLT by Ergoline, Planet Fitness 42/3 PLT by Ergoline Regulation Number: 21 CFR 878.4635 Regulation Name: Sunlamp products and ultraviolet lamps intended for use in sunlamp products Regulatory Class: Class II Product Code: LEJ Dated: May 20, 2022 Received: May 31, 2022

Dear Juergen Gerstenmeier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of 

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220500

#### Device Name

Ergoline Vitality Total Light, Ergoline Vitality Hybrid Light, Ergoline Vitality Pure Light, Planet Fitness TLT by Ergoline, Planet Fitness HLT by Ergoline, Planet Fitness PLT by Ergoline

#### Indications for Use (Describe)

This sunlamp product is intended exclusively for aesthetic tanning of the human skin, for one person at a time, at the age of 18 or above.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Chapter 5 510(k) Summary of Safety and Effectiveness

{4}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness

In accordance with the requirements of the Safe Medical Device Act, JK-HOLDING GmbH herewith submits a Summary of Safety and Effectiveness.

This 510(k) summary includes the bundles for device trade names for Ergoline Vitality Total Light, Ergoline Vitality Hybrid Light, Ergoline Vitality Pure Light, Planet Fitness TLT by Ergoline, Planet Fitness HLT by Ergoline, Planet Fitness PLT by Ergoline & variant units which meet the requirements of 21 CFR 807.92

| Manufacturer Information: | JK-Produkts GmbH<br>Rottbitzer Strasse 69<br>53604 Bad Honnef<br>Germany<br>Phone: +49 (0) 2224 / 818-140<br>Fax: +49 (0) 2224 / 818-166                                                                                                                          |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                           | Establishment Registration Number: 9615023<br>Owner/Operator Number: 9034303                                                                                                                                                                                      |
| Submitter Information:    | JK-HOLDING GmbH<br>Koehlershohner Strasse 60<br>53578 Windhagen<br>Phone: +49 (0) 2224 / 818-0<br>Fax: +49 (0) 2224 / 818-116                                                                                                                                     |
| Official Correspondent:   | Juergen Gerstenmeier<br>Koehlershohner Strasse 60<br>53578 Windhagen<br>Germany<br>Phone: +49 (0) 2224 / 818-330<br>Email: juergen.gerstenmeier@jk-group.net                                                                                                      |
| US Agent (Contact):       | Marie Merrill<br>JK PRODUCTS & SERVICES, INC.<br>411 West Washington Ave Suite B<br>Jonesboro , AR 72401<br>Phone: 870 761 8906<br>Fax: 870 9359046<br>Email: marie.merrill@jkamerica.com                                                                         |
| Device trade name:        | Ergoline Vitality Total Light 3<br>Planet Fitness 50/4 TLT by Ergoline<br>Planet Fitness 42/3 TLT by Ergoline<br>Ergoline Vitality Hybrid Light 3<br>Ergoline Vitality Hybrid Light<br>Planet Fitness 50/4 HLT by Ergoline<br>Planet Fitness 46/3 HLT by Ergoline |

{5}------------------------------------------------

Ergoline Vitality Pure Light 3 Ergoline Vitality Pure Light Planet Fitness 50/4 PLT III by Ergoline Planet Fitness 50/4 PLT by Ergoline Planet Fitness 46/3 PLT by Ergoline Planet Fitness 42/3 PLT by Ergoline

Date Prepared: February 16th, 2022

{6}------------------------------------------------

#### Table of contents

| 1   | Table of contents              |                                               | 5  |
|-----|--------------------------------|-----------------------------------------------|----|
| 1.1 | Device(s) Identification:      |                                               | 5  |
|     | 1.1.1                          | Indications for use:                          | 5  |
|     | 1.1.2                          | Classification of all proposed device:        | 6  |
| 1.2 | Device trade name of bundle 1: |                                               | 6  |
|     | 1.2.1                          | Device Description:                           | 6  |
|     | 1.2.2                          | Legally marketed predicate device:            | 7  |
|     |                                | Legally marketed technical referenced device: | 7  |
|     | 1.2.3                          | Comparison bundle 1 to predicate device:      | 8  |
|     | 1.2.4                          | Summary of performance testing:               | 12 |
|     | 1.2.5                          | Conclusion:                                   | 13 |
| 1.3 | Device trade name of bundle 2: |                                               | 13 |
|     | 1.3.1                          | Device description:                           | 13 |
|     | 1.3.2                          | Legally marketed predicate device 2:          | 14 |
|     | 1.3.3                          | Comparison bundle 2 to predicate device:      | 15 |
|     | 1.3.4                          | Summary of performance testing:               | 17 |
|     | 1.3.5                          | Conclusion:                                   | 17 |
| 1.4 | Device trade name of bundle 3: |                                               | 17 |
|     | 1.4.1                          | Device description:                           | 17 |
|     | 1.4.2                          | Legally marketed predicate device:            | 18 |
|     |                                | Legally marketed technical referenced device: | 18 |
|     | 1.4.3                          | Comparison bundle 3 to predicate device:      | 19 |
|     | 1.4.4                          | Summary of performance testing:               | 27 |
|     | 1.4.5                          | Conclusion:                                   | 27 |
| 1.5 | Device trade name of bundle 4: |                                               | 27 |
|     | 1.5.1                          | Device description:                           | 28 |
|     | 1.5.2                          | Legally marketed predicate device 4:          | 28 |
|     | 1.5.3                          | Comparison bundle 4 to predicate device:      | 29 |
|     | 1.5.4                          | Summary of performance testing:               | 32 |
|     | 1.5.5                          | Conclusion:                                   | 32 |
| 1.6 | Device trade name of bundle 5: |                                               | 32 |
|     | 1.6.1                          | Device Description:                           | 33 |
|     | 1.6.2                          | Legally marketed predicate device 5:          | 33 |
|     | 1.6.3                          | Comparison bundle 5 to predicate device:      | 34 |
|     | 1.6.4                          | Summary of performance testing:               | 39 |
|     | 1.6.5                          | Conclusion:                                   | 39 |

{7}------------------------------------------------

# 1.1 Device(s) Identification:

In the following, 5 different bundles of devices will be presented and evaluated.

| Bundle No | Model No                                                                                                                                                            | Device names                                                                                                                                      |
|-----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| 1         | JK 164 / 42-3 TLT-3 AC<br>JK 164 / 42-3 TLT-3<br>JK 164 / 42-3 TLT AC<br>JK 164 / 42-3 TLT                                                                          | Planet Fitness 42/3 TLT by Ergoline<br>Ergoline Vitality Total Light 3                                                                            |
| 2         | JK 164 / 42-3 PLT AC<br>JK 164 / 42-3 PLT                                                                                                                           | Planet Fitness 42/3 PLT by Ergoline<br>Ergoline Vitality Pure Light                                                                               |
| 3         | JK 164 / 50-4 TLT-3 TT<br>JK 164 / 50-4 TLT-3<br>JK 164 / 50-4 HLT-3 TT<br>JK 164 / 50-4 HLT-3<br>JK 164 / 50-4 TLT TT<br>JK 164 / 50-4 TLT AC<br>JK 164 / 50-4 TLT | Planet Fitness 50/4 TLT by Ergoline<br>Ergoline Vitality Total Light 3<br>Planet Fitness 50/4 HLT by Ergoline<br>Ergoline Vitality Hybrid Light 3 |
| 4         | JK 164 / 50-4 PLT-3 AC<br>JK 164 / 50-4 PLT-3<br>JK 164 / 50-4 PLT AC<br>JK 164 / 50-4 PLT                                                                          | Planet Fitness 50/4 PLT III by Ergoline<br>Ergoline Vitality Pure Light 3<br>Planet Fitness 50/4 PLT by Ergoline<br>Ergoline Vitality Pure Light  |
| 5         | JK 164 / 46-3 PLT AC<br>JK 164 / 46-3 PLT<br>JK 164 / 46-3 HLT AC<br>JK 164 / 46-3 HLT                                                                              | Planet Fitness 46/3 PLT by Ergoline<br>Ergoline Vitality Pure Light<br>Planet Fitness 46/3 HLT by Ergoline<br>Ergoline Vitality Hybrid Light      |

All products have an identical intended use and comparable technology.

The different devices in the bundles have different trade names, number of lamps, a different combination of light spectra and different maximum exposure times.

The basic products can be upgraded with conversion kits, consisting of stickers with the private label naming, to Soltron or further private label sunlamp products. The conversion kit only serves to label the basic product as a Soltron or further private label sunlamp product. It does not alter any technical specification or aspect of the devices.

For further information concerning conversion kits please refer to chapter 11.

# 1.1.1 Indications for use:

This sunlamp product is intended exclusively for aesthetic tanning of the human skin, for one person at a time, at the age of 18 or above.

# 1.1.2 Classification of all proposed device:

| Device Classification Name: | Booth, Sun Tan            |
|-----------------------------|---------------------------|
| Product Code:               | LEJ                       |
| C.F.R. Section.:            | 878.4635                  |
| Classification Panel:       | General & Plastic Surgery |
| Device Class:               | Class II                  |

{8}------------------------------------------------

# 1.2 Device trade names of bundle 1:

Planet Fitness 42/3 TLT by Ergoline Ergoline Vitality Total Light 3

Common name: Sunlamp product

The following table presents the different variants of the JK 164 / 42-3 Total Light Technology sunlamp product.

All variants use the same basic device technology and differ only in private labelling, exterior design, decor lighting and comfort features.

| Model-No.                                     | Variant Name                        |
|-----------------------------------------------|-------------------------------------|
| JK 164 / 42-3 TLT-3 AC<br>JK 164 / 42-3 TLT-3 | Planet Fitness 42/3 TLT by Ergoline |
|                                               | Ergoline Vitality Total Light 3     |
| JK 164 / 42-3 TLT AC<br>JK 164 / 42-3 TLT     | Planet Fitness 42/3 TLT by Ergoline |
|                                               | Ergoline Vitality Total Light 3     |

# 1.2.1 Device Description:

# Total Light Technology

The primary technical components of sunlamp products are artificial sources of UV radiation as well as a mechanical structure with a defined active surface.

UV lamps emit different UV-A and UV-B proportions of the UV radiation. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. The UV-B radiation is primarily responsible for more long-term tanning results. Using traditional UV light with the combination of red and blue light the user indulges in a new luxurious tanning experience.

Optimal tanning results are achieved due to the synergistic combination of light spectra.

# 1.2.2 Legally marketed predicate device:

| Device Name:                | Planet Fitness 42/4 Hybrid Light Technology |
|-----------------------------|---------------------------------------------|
| 510k number:                | K190173                                     |
| Device Classification Name: | Booth, Sun Tan                              |
| Product Code:               | LEJ                                         |
| C.F.R. Section.:            | 878.4635                                    |
| Classification Panel:       | General & Plastic Surgery                   |
| Device Class:               | Class II                                    |

{9}------------------------------------------------

# Legally marketed technical referenced device:

Sunrise 7200 Hybrid Light Technology Device Name: 510k number: K183539 Device Classification Name: Booth, Sun Tan Product Code: LEJ C.F.R. Section.: 878.4635 General & Plastic Surgery Classification Panel: Device Class: Class II

{10}------------------------------------------------

### 1.2.3 Comparison bundle 1 to predicate device:

The following table presents the comparison of technological characteristics, functions and parameters of the identified predicate, the proposed and the technically referenced device.

| Parameter                                                         | Proposed device                                                                                                                                                | Proposed device                                                                                                                                                | Predicate device 1                                                                                                                                             | Technically<br>referenced<br>device       | Evaluation to<br>predicate device                                                                                                                                                                                                                                                                                 | Evaluation to<br>Technically<br>referenced<br>device                                                                                                                                                                                                                                                                             |                                                      |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Model-No.                                                         | JK 164 / 42-3 TLT-<br>3 AC<br>JK 164 / 42-3 TLT-<br>3                                                                                                          | JK 164 / 42-3 TLT<br>AC<br>JK 164 / 42-3 TLT                                                                                                                   | JK 126 / 42-4 HT AC                                                                                                                                            | JK 158 / 48-0<br>HT                       | N/A                                                                                                                                                                                                                                                                                                               | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Intended Use                                                      | This sunlamp<br>product is intended<br>exclusively for<br>aesthetic tanning of<br>the human skin, for<br>one person at a<br>time, at the age of<br>18 or above | This sunlamp<br>product is intended<br>exclusively for<br>aesthetic tanning of<br>the human skin, for<br>one person at a<br>time, at the age of<br>18 or above | This sunlamp product<br>is intended<br>exclusively for<br>aesthetic tanning of<br>the human skin, for<br>one person at a time,<br>at the age of 18 or<br>above | N/A                                       | Identical                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Number of body<br>lamps                                           | 42 (28 UV lamps<br>and 14 UV + Red<br>Light lamps)                                                                                                             | 42 (28 UV lamps<br>and 14 UV + Red<br>Light lamps)                                                                                                             | 42 (28 UV lamps and<br>14 UV + Red Light<br>lamps)                                                                                                             | N/A                                       | Identical                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Watts of body<br>lamps                                            | 200W                                                                                                                                                           | 160W                                                                                                                                                           | canopy 160W<br>bench 180W                                                                                                                                      | N/A                                       | Similar, the irradiance<br>characteristics are still<br>comparable as<br>provided under section<br>18. The safety has<br>been confirmed<br>additionally by<br>corresponding tests in<br>accordance with IEC<br>62471.<br>In terms of functionality<br>and safety there is no<br>impact due to this<br>difference. | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Parameter                                                         | Proposed device                                                                                                                                                | Proposed device                                                                                                                                                | Predicate device 1                                                                                                                                             | Technically<br>referenced<br>device       | Evaluation to<br>predicate device                                                                                                                                                                                                                                                                                 | Evaluation to<br>Technically<br>referenced<br>device                                                                                                                                                                                                                                                                             |                                                      |
| Lamp item<br>description                                          | GENESIS PLT<br>Pure Light<br>Technology 200W<br>GENESIS HLT<br>Hybrid Light<br>Technology 200W<br>(UV&Red light)                                               | GENESIS PLT<br>Pure Light<br>Technology 160W<br>GENESIS HLT<br>Hybrid Light<br>Technology 160W<br>(UV&Red light)                                               | GENESIS Type U<br>Hybrid Performance<br>160 W / 180W<br>GENESIS Type R<br>Hybrid Performance<br>160W / 180W                                                    | N/A                                       | Different description,<br>no impact, see above.                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Number of Total<br>Light-LED                                      | 72 LED (36 Red +<br>36 Blue)                                                                                                                                   | 36 LED (18 RED +<br>18 Blue)                                                                                                                                   | N/A                                                                                                                                                            | 60 Red LED                                | N/A                                                                                                                                                                                                                                                                                                               | Similar, the<br>irradiance<br>characteristics<br>are still<br>comparable as<br>provided under<br>section 18. The<br>safety has been<br>confirmed<br>additionally by<br>corresponding<br>tests in<br>accordance with<br>IEC 62471.<br>In terms of<br>functionality and<br>safety there is<br>no impact due to<br>this difference. |                                                      |
| Watts of Total<br>Light-LED                                       | 100W                                                                                                                                                           | 50W                                                                                                                                                            | N/A                                                                                                                                                            | 45W                                       | N/A                                                                                                                                                                                                                                                                                                               | No impact, see<br>above.                                                                                                                                                                                                                                                                                                         |                                                      |
|                                                                   | Parameter                                                                                                                                                      | Proposed device                                                                                                                                                | Proposed device                                                                                                                                                | Predicate device 1                        | Technically<br>referenced<br>device                                                                                                                                                                                                                                                                               | Evaluation to<br>predicate device                                                                                                                                                                                                                                                                                                | Evaluation to<br>Technically<br>referenced<br>device |
| UV-Facials                                                        | Number of<br>UV-facials                                                                                                                                        | 3 x 42 UVA-LED                                                                                                                                                 | 3 x 42 UVA-LED                                                                                                                                                 | 4 high pressure<br>lamps                  | N/A                                                                                                                                                                                                                                                                                                               | Similar, the irradiance<br>characteristics are still<br>comparable as<br>provided under section<br>18.<br>In terms of functionality<br>and safety there is no<br>impact due to this<br>difference.                                                                                                                               | N/A                                                  |
|                                                                   | Watts<br>facial lamp                                                                                                                                           | 3 x 80W                                                                                                                                                        | 3 x 80W                                                                                                                                                        | 4x400W                                    | N/A                                                                                                                                                                                                                                                                                                               | No impact, see below.                                                                                                                                                                                                                                                                                                            | N/A                                                  |
|                                                                   | Facial item<br>description                                                                                                                                     | Spectra UV-LED                                                                                                                                                 | Spectra UV-LED                                                                                                                                                 | High pressure lamp<br>Ergoline Ultra 520W | N/A                                                                                                                                                                                                                                                                                                               | Similar, the irradiance<br>characteristics are still<br>comparable as<br>provided under section<br>18.<br>In terms of functionality<br>and safety there is no<br>impact due to this<br>difference.                                                                                                                               | N/A                                                  |
|                                                                   | Filter facial<br>lamps                                                                                                                                         | N/A                                                                                                                                                            | N/A                                                                                                                                                            | Ultra Performance<br>862H                 | N/A                                                                                                                                                                                                                                                                                                               | No filters necessary<br>due to no use of high-<br>pressure lamps.                                                                                                                                                                                                                                                                | N/A                                                  |
|                                                                   | Number of<br>low<br>pressure<br>facial<br>lamps                                                                                                                | 2 x HLT                                                                                                                                                        | 2 x HLT                                                                                                                                                        | 3 x Type R                                | N/A                                                                                                                                                                                                                                                                                                               | Similar, the irradiance<br>characteristics are still<br>comparable as<br>provided under section<br>18. The safety has<br>been confirmed<br>additionally by<br>corresponding tests in<br>accordance with IEC<br>62471.<br>In terms of functionality<br>and safety there is no<br>impact due to this<br>difference.                | N/A                                                  |
| Parameter                                                         | Proposed device                                                                                                                                                | Proposed device                                                                                                                                                | Predicate device 1                                                                                                                                             | Technically<br>referenced<br>device       | Evaluation to<br>predicate device                                                                                                                                                                                                                                                                                 | Evaluation to<br>Technically<br>referenced<br>device                                                                                                                                                                                                                                                                             |                                                      |
| Watts low<br>pressure<br>facial lamp                              | 8W                                                                                                                                                             | 8W                                                                                                                                                             | 8W                                                                                                                                                             | N/A                                       | Identical                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Low<br>pressure<br>facial<br>lamps item<br>description            | GENESIS HLT<br>Hybrid Light<br>Technology 8W                                                                                                                   | GENESIS HLT<br>Hybrid Light<br>Technology 8W                                                                                                                   | Genesis Type R<br>Hybrid 8W (Mix of<br>UV+Red Light)                                                                                                           | N/A                                       | Different description,<br>no impact, see above.                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Max exposure<br>time [min]                                        | 12 minutes                                                                                                                                                     | 12 minutes                                                                                                                                                     | 12 minutes                                                                                                                                                     | N/A                                       | Identical                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Electrical<br>Requirements                                        | 230V 3Ø or 230V<br>2Ø                                                                                                                                          | 230V 3Ø or 230V<br>2Ø                                                                                                                                          | 230V 3Ø or 230V 2Ø                                                                                                                                             | N/A                                       | Identical                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Total power<br>consumption<br>[watts]                             | 11,600 W (with AC)<br>10,400 (w/o AC)                                                                                                                          | 8,800 W (with AC)<br>7,200 W (w/o AC)                                                                                                                          | 11,200 W                                                                                                                                                       | N/A                                       | Similar:<br>The power is<br>different due to<br>different lamps.<br>Performance is not<br>impaired.                                                                                                                                                                                                               | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Rated<br>overcurrent<br>protection<br>device (circuit<br>breaker) | 45A / 3-pole or 80<br>A /2-pole (with AC)<br>or<br>45A / 3-pole or 70A<br>/2-pole (w/o AC)                                                                     | 35 A / 3-pole or 60<br>A / 2-pole (with AC)<br>30A / 3-pole or<br>50A/ 2-pole (w/o<br>AC)                                                                      | 70A / 3-pole 3Ø or<br>100A / 2-pole 2Ø                                                                                                                         | N/A                                       | Similar, the<br>overcurrent protection<br>is less with the<br>proposed device due<br>to the use of LED<br>technology for facials<br>and different lamps.<br>The irradiance<br>Characteristics are still<br>comparable as<br>provided under section<br>18.                                                         | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Number of<br>wires                                                | 4 3Ø or 3 2Ø                                                                                                                                                   | 4 3Ø or 3 2Ø                                                                                                                                                   | 4 3Ø or 3 2Ø                                                                                                                                                   | N/A                                       | Identical                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Irradiance ratio<br>in accordance<br>with 1040.20                 | fulfilled                                                                                                                                                      | fulfilled                                                                                                                                                      | fulfilled                                                                                                                                                      | N/A                                       | Identical (refer to<br>section 18 for bench<br>test reports)                                                                                                                                                                                                                                                      | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Parameter                                                         | Proposed device                                                                                                                                                | Proposed device                                                                                                                                                | Predicate device 1                                                                                                                                             | Technically<br>referenced<br>device       | Evaluation to<br>predicate device                                                                                                                                                                                                                                                                                 | Evaluation to<br>Technically<br>referenced                                                                                                                                                                                                                                                                                       |                                                      |
|                                                                   |                                                                                                                                                                |                                                                                                                                                                |                                                                                                                                                                |                                           |                                                                                                                                                                                                                                                                                                                   | device                                                                                                                                                                                                                                                                                                                           |                                                      |
| Electrical safety                                                 | IEC 60601-1                                                                                                                                                    | IEC 60601-1                                                                                                                                                    | IEC 60601-1                                                                                                                                                    | N/A                                       | Identical                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |
| Electromagnetic compatibility                                     | IEC 60601-1-2                                                                                                                                                  | IEC 60601-1-2                                                                                                                                                  | IEC 60601-1-2                                                                                                                                                  | N/A                                       | Identical                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                              |                                                      |

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

# 1.2.4 Summary of performance testing:

The proposed devices have been tested in respect to biocompatibility in accordance with the ISO 10993-series, electrical and mechanical safety in accordance with IEC 60601-1 and EMC in accordance with IEC 60601-1-2.

The proposed devices are in compliance with U.S. performance standard 21CFR 1040.20.

# 1.2.5 Conclusion:

JK-Holding GmbH believes that the JK 164 / 42-3 Total Light Technology are substantially equivalent to currently legally marketed devices. Also, all special controls are fulfilled. These devices do not introduce new indications for use, has the same technological characteristics and does not introduce new potential hazards or safety risks.

# 1.3 Device trade names of bundle 2:

Planet Fitness 42/3 PLT by Ergoline Ergoline Vitality Pure Light

Common name: Sunlamp product

The following table presents the different variants of the JK 164 / 42-3 Pure Light Technology sunlamp product.

All variants use the same basic device technology and differ only in private labelling, exterior design, decor lighting and comfort features.

| Model-No.            | Variant Name                        |
|----------------------|-------------------------------------|
| JK 164 / 42-3 PLT AC | Planet Fitness 42/3 PLT by Ergoline |
| JK 164 / 42-3 PLT    | Ergoline Vitality Pure Light        |

#### 1.3.1 Device description:

### Pure Light Technology

The primary technical components of sunlamp products are artificial sources of UV radiation (UV lamps) and a mechanical structure.

UV lamps emit different UV-A and UV-B proportions of the UV radiation. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly, the UV-B radiation is primarily responsible for more long-term tanning results.

{16}------------------------------------------------

# 1.3.2 Legally marketed predicate device 2:

| Device Name:                | Planet Fitness 42/4            |
|-----------------------------|--------------------------------|
| 510k number:                | K151400 CV344_PF_100003421_V01 |
| Device Classification Name: | Booth, Sun Tan                 |
| Product Code:               | LEJ                            |
| C.F.R. Section.:            | 878.4635                       |

Device Class:

Class II

{17}------------------------------------------------

# 1.3.3 Comparison bundle 2 to predicate device:

The following table presents the comparison of technological characteristics, functions and parameters of the identified predicate and the proposed device.

|                                                          | Parameter                                     | Proposed device                                                                                                                                    | Predicate device                                                                                                                                   | Evaluation to predicate device                                                                                                                                                                                                              |
|----------------------------------------------------------|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                          | Model-No.                                     | JK 164 / 42-3 PLT AC<br>JK 164 / 42-3 PLT                                                                                                          | JK 126 / 42-4                                                                                                                                      | N/A                                                                                                                                                                                                                                         |
|                                                          | Intended Use                                  | This sunlamp product is intended<br>exclusively for aesthetic tanning of<br>the human skin, for one person at a<br>time, at the age of 18 or above | This sunlamp product is intended<br>exclusively for aesthetic tanning of<br>the human skin, for one person at a<br>time, at the age of 18 or above | Identical                                                                                                                                                                                                                                   |
|                                                          | Number of body lamps                          | 42 UV lamps                                                                                                                                        | 42 UV lamps                                                                                                                                        | Identical                                                                                                                                                                                                                                   |
|                                                          | Watts of body lamps                           | 160W                                                                                                                                               | 200 W                                                                                                                                              | Similar, the irradiance characteristics<br>are still comparable as provided<br>under section 18.<br>In terms of functionality and safety<br>there is no impact due to this<br>difference.                                                   |
|                                                          | Lamp item description                         | GENESIS PLT Pure Light<br>Technology 160W                                                                                                          | GENESIS VHP12<br>Dynamic Power 200W                                                                                                                | Different description, no impact, see<br>above.                                                                                                                                                                                             |
| UV-<br>Facials                                           | Number of UV-facials                          | 3 x 42 UVA-LED                                                                                                                                     | 4 high pressure lamps                                                                                                                              | Similar, the irradiance characteristics<br>are still comparable as provided<br>under section 18.<br>In terms of functionality and safety<br>there is no impact due to this<br>difference.                                                   |
|                                                          | Watts facial lamp                             | 3 x 80W                                                                                                                                            | 4 x 400W                                                                                                                                           | No impact, see below.                                                                                                                                                                                                                       |
|                                                          | Facial item description                       | Spectra UV-LED                                                                                                                                     | High pressure lamp<br>Ergoline Ultra 520W                                                                                                          | Similar, the irradiance characteristics<br>are still comparable as provided<br>under section 18.<br>In terms of functionality and safety<br>there is no impact due to this<br>difference.                                                   |
|                                                          | Parameter                                     | Proposed device                                                                                                                                    | Predicate device                                                                                                                                   | Evaluation to predicate device                                                                                                                                                                                                              |
| UV-Facials                                               | Filter facial lamps                           | N/A                                                                                                                                                | Ultra Performance 862                                                                                                                              | No filters necessary due to no use of<br>high-pressure lamps.                                                                                                                                                                               |
|                                                          | Number of low pressure<br>facial lamps        | 2 x PLT                                                                                                                                            | 3 x Type U                                                                                                                                         | Similar, the irradiance characteristics<br>are still comparable as provided<br>under section 18.<br>In terms of functionality and safety<br>there is no impact due to this<br>difference.                                                   |
|                                                          | Watts low pressure facial<br>lamp             | 8W                                                                                                                                                 | 8W                                                                                                                                                 | Identical                                                                                                                                                                                                                                   |
|                                                          | Low pressure facial lamps<br>item description | GENESIS PLT Pure Light<br>Technology 8W                                                                                                            | Genesis VHP B-Power E9 8W                                                                                                                          | Different description, no impact, see<br>above.                                                                                                                                                                                             |
|                                                          | Max exposure time [min]                       | 12 minutes                                                                                                                                         | 12 minutes                                                                                                                                         | Identical                                                                                                                                                                                                                                   |
|                                                          | Electrical Requirements                       | 230V 3Ø or 230V 2Ø                                                                                                                                 | 230V 3Ø or 230V 2Ø                                                                                                                                 | Identical                                                                                                                                                                                                                                   |
| Total power consumption [watts]                          |                                               | 9,000 W (with AC)<br>7,800 W (w/o AC)                                                                                                              | 13,200W                                                                                                                                            | Similar, the total power consumption<br>is less with the proposed device due<br>to the use of LED technology for<br>facials and different lamps. The<br>irradiance characteristics are still<br>comparable as provided under<br>section 18. |
| Rated overcurrent protection<br>device (circuit breaker) |                                               | 35 A / 3-pole or 60 A / 2-pole (with<br>AC)<br>30A / 3-pole or 50A/ 2-pole (w/o<br>AC)                                                             | 70A / 3-pole 3Ø or 100A / 2-pole 2Ø                                                                                                                | Similar, die overcurrent protection is<br>less with the proposed device due to<br>the use of LED technology for facials<br>and different lamps. The irradiance<br>characteristics are still comparable<br>as provided under section 18.     |
| Number of wires                                          |                                               | 4 3Ø or 3 2Ø                                                                                                                                       | 4 3Ø or 3 2Ø                                                                                                                                       | Identical                                                                                                                                                                                                                                   |
| Irradiance ratio in accordance<br>with 1040.20           |                                               | fulfilled                                                                                                                                          | fulfilled                                                                                                                                          | Identical (refer to section 18 for<br>bench test reports)                                                                                                                                                                                   |
| Electrical safety                                        |                                               | IEC 60601-1                                                                                                                                        | IEC 60601-1                                                                                                                                        | Identical                                                                                                                                                                                                                                   |
|                                                          | Electromagnetic compatibility                 | IEC 60601-1-2                                                                                                                                      | IEC 60601-1-2                                                                                                                                      | Identical                                                                                                                                                                                                                                   |

{18}------------------------------------------------

{19}------------------------------------------------

# 1.3.4 Summary of performance testing:

The proposed devices have been tested in respect to biocompatibility in accordance with the ISO 10993-series, electrical and mechanical safety in accordance with IEC 60601-1 and EMC in accordance with IEC 60601-1-2.

The proposed devices are in compliance with U.S. performance standard 21CFR 1040.20.

# 1.3.5 Conclusion:

JK-Holding GmbH believes that the JK 164 / 42-3 Pure Light Technology are substantially equivalent to currently legally marketed devices. Also, all special controls are fulfilled. These devices do not introduce new indications for use, has the same technological characteristics and does not introduce new potential hazards or safety risks.

# 1.4 Device trade names of bundle 3:

Planet Fitness 50/4 TLT by Ergoline Ergoline Vitality Total Light 3 Planet Fitness 50/4 HLT by Ergoline Ergoline Vitality Hybrid Light 3

Common name: Sunlamp product

The following table presents the different variants of the JK 164 / 50-4 Total Light Technology and JK 164 / 50-4 Hybrid Light Technology sunlamp product.

All variants use the same basic device technology and differ only in private labelling, exterior design, decor lighting, combination of light spectra and comfort features.

| Model-No.              | Variant Name                        |
|------------------------|-------------------------------------|
| JK 164 / 50-4 TLT-3 TT | Planet Fitness 50/4 TLT by Ergoline |
| JK 164 / 50-4 TLT-3    | Ergoline Vitality Total Light 3     |
| JK 164 / 50-4 HLT-3 TT | Planet Fitness 50/4 HLT by Ergoline |
| JK 164 / 50-4 HLT-3    | Ergoline Vitality Hybrid Light 3    |
| JK 164 / 50-4 TLT TT   | Ergoline Vitality Total Light 3     |
| JK 164 / 50-4 TLT AC   | Planet Fitness 50/4 TLT by Ergoline |
| JK 164 / 50-4 TLT      | Ergoline Vitality Total Light 3     |

#### 1.4.1 Device description:

# Total Light Technology

The primary technical components of sunlamp products are artificial sources of UV radiation as well as a mechanical structure with a defined active surface.

{20}------------------------------------------------

UV lamps emit different UV-A and UV-B proportions of the UV radiation. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. The UV-B radiation is primarily responsible for more long-term tanning results. Using traditional UV light with the combination of red and blue light the user indulges in a new luxurious tanning experience.

Optimal tanning results are achieved due to the synergistic combination of light spectra.

# Hybrid Light Technology

The primary technical components of sunlamp products are artificial sources of UV radiation as well as a mechanical structure with a defined active surface.

UV lamps emit different UV-A and UV-B proportions of the UV radiation. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. The UV-B radiation is primarily responsible for more long-term tanning results.

Traditional UV light with the addition of (beauty) red light results in a natural looking tan. Optimal tanning results are achieved due to the synergistic combination of light spectra.

### 1.4.2 Leqally marketed predicate device:

| Device Name:                | Ergoline Prestige 1600 Hybrid Technology |
|-----------------------------|------------------------------------------|
| 510k number:                | K180555                                  |
| Device Classification Name: | Booth, Sun Tan                           |
| Product Code:               | LEJ                                      |
| C.F.R. Section.:            | 878.4635                                 |
| Classification Panel:       | General & Plastic Surgery                |
| Device Class:               | Class II                                 |

### Legally marketed technical referenced device:

| Device Name:                | Sunrise 7200 Hybrid Light Technology |
|-----------------------------|--------------------------------------|
| 510k number:                | K183539                              |
| Device Classification Name: | Booth, Sun Tan                       |
| Product Code:               | LEJ                                  |
| C.F.R. Section:             | 878.4635                             |
| Classification Panel:       | General & Plastic Surgery            |
| Device Class:               | Class II                             |

{21}------------------------------------------------

### 1.4.3 Comparison bundle 3 to predicate device:

The following table presents the comparison of technological characteristics, functions and…

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/LEJ/K220500](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/LEJ/K220500)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
