MICRO ENDOSCOPE-STAINLESS STEEL, MODEL MMD-1300, MICRO ENDOSCOPE-POLYIMIDE,MODEL MMD-1400
K983882 · Micro-Medical Devices, Inc. · HXR · Mar 15, 1999 · General, Plastic Surgery
Device Facts
Record ID
K983882
Device Name
MICRO ENDOSCOPE-STAINLESS STEEL, MODEL MMD-1300, MICRO ENDOSCOPE-POLYIMIDE,MODEL MMD-1400
Applicant
Micro-Medical Devices, Inc.
Product Code
HXR · General, Plastic Surgery
Decision Date
Mar 15, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4800
Device Class
Class 1
Indications for Use
Micro Arthroscope is intended for insertion into a small incision or puncture (through a cannula) to view the surgical site of small and large joints in conjunction with cameras. Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
Device Story
Micro Endoscope is a rigid arthroscope used for joint visualization. Device is inserted through small incisions or cannulas to view surgical sites in wrists, ankles, elbows, knees, or shoulders. Operates in conjunction with external cameras and xenon light sources (max 300W). Used by physicians in clinical/surgical settings. Provides optical visualization to assist surgeons in joint procedures. Contraindicated for bending or prying, which damages optics. Requires specific compatibility with Micro Medical Devices, Inc. cameras.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use comparisons.
Technological Characteristics
Materials: Stainless steel and polyimide. Form factor: Rigid arthroscope for insertion through cannula. Energy source: External xenon light source (max 300W). Sterilization: Single-use, Ethylene Oxide (EO) to SAL 10^-6.
Indications for Use
Indicated for visualization of surgical sites in small and large joints (wrists, ankles, elbows, knees, or shoulders) via small incision or cannula. Prescription use only.
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
K030096 — VIDEO ARTHROSCOPE, MODELS AR-3050-30, AR-3050T-30, AR-3050-70, AR-3051-30, AR-3052-30, AR3030AN, AR-3030AS, AR3030AW · Arthrex, Inc. · Mar 18, 2003
K960805 — ARTHOSCOPE · United States Surgical, A Division of Tyco Healthc · Apr 10, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS MICRO MEDICAL DEVICES, INC. MICRO ENDOSCOPE
The Summary of Safety and Effectiveness on Micro Endoscope reflects data available and presented at the time the submission was prepared, but caution should be exercised in interpreting the data. The results of future studies may require alterations of the conclusions or recommendations set forth.
## Intended Use
Micro Arthroscope is intended for insertion into a small incision or puncture (through a cannula) to view the surgical site of small and large joints in conjunction with cameras.
### Caution
Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
#### Contraindication
Bending or prying with scope will damage optics.
#### Precautions
- Surgical laser and mechanical cutting instruments can damage the scope. ●
- Do not allow the arthroscope tip to come into contact with hard or sharp objects. .
- Use this device only with a xenon light source (maximum 300 watts) and Micro Medical . Devices, Inc camera.
#### Substantial Equivalency Information
The Micro Arthroscope is similar to the Galileo Arthroscope currently offered by Galileo Electro-Optics Corporation.
| Materials | Galileo<br>Stainless Steel | Micro Medical<br>Stainless Steel<br>Polyimide |
|---------------|----------------------------|-----------------------------------------------|
| Sterilization | Reusable: EO and<br>Steris | Single Use: EO<br>SAL 10-6 |
The intended use and technological characteristics of these devices do not vary significantly. The safety and effectiveness of the Micro Medical Endoscope is comparable to that of the Galileo Electro-Optics Corporation Arthroscope
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 5 1999
Mr. Jim Ohneck General Manger Micro Medical Devices, Inc. 11000 Cedar Avenue Suite 445 Cleveland, Ohio 44106
Re: K983882 Trade Name: Micro Endoscope Regulatory Class: II Product Code: HXR Dated: February 3, 1999 Received: February 5, 1999
Dear Mr. Ohneck:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Mr. Jim Ohneck
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):_
Micro Endoscope
983882
Devicc Namc:_______
Indications For Use:
Micro Endoscope is intended for insertion into a small incision or puncture (through a cannula) to view the surgical site of small and large joints of the wrists, ankles, elbows, knees, or shoulders in conjunction with cameras.
CAUTION: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
CONTRAINDICATIONS: Bending or prying with scope will damage optics.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) -
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of General Restorative Devices | |
| 510(k) Number | K983882 |
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.