STRYKER UNIVERSAL DRILL ATTACHMENTS & ACCESSORIES
Device Facts
| Record ID | K962800 |
|---|---|
| Device Name | STRYKER UNIVERSAL DRILL ATTACHMENTS & ACCESSORIES |
| Applicant | Stryker Instruments |
| Product Code | HWE · General, Plastic Surgery |
| Decision Date | Sep 11, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4820 |
| Device Class | Class 1 |
| Attributes | Therapeutic |
Indications for Use
The Stryker Universal Drill Attachments and Accessories are intended for use in the cutting, drilling, reaming, decorticating, and smoothing of bone and other bone related tissue in a variety of surgical procedures including: ENT, orthopedic, maxillofacial, neuro, and plastics.
Device Story
System consists of surgical drill attachments and accessories used with Stryker 75K Drill motor; powered by DC electric current via Total Performance System Console. Attachments include extra-long variants (up to 400 mm) and Hudson Chuck (reduces speed from 75,000 rpm to 700-1200 rpm) for cranial perforators. Used by surgeons in OR settings for bone tissue manipulation. Provides mechanical cutting/drilling action; output is physical bone modification. Benefits include surgical versatility and compatibility with standard Hudson shank fittings.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Mechanical surgical attachments; includes Hudson Chuck for speed reduction. Materials and cutting configurations equivalent to predicate surgical accessories. Powered by DC electric motor (Stryker 75K). Sterile or non-sterile delivery.
Indications for Use
Indicated for patients requiring bone cutting, drilling, reaming, decorticating, or smoothing during ENT, orthopedic, maxillofacial, neuro, or plastic surgical procedures.
Regulatory Classification
Identification
Surgical instrument motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. Accessories or attachments may include a bur, chisel (osteotome), dermabrasion brush, dermatome, drill bit, hammerhead, pin driver, and saw blade.
Reference Devices
- Midas Rex powered instrument attachments
- Codman powered instrument attachments
- Stryker 75K Drill
- Zimmer/Hall cutting accessories
Related Devices
- K961970 — COMMAND 2 MICOELECTRIC SURGICAL SYSTEM · Stryker Corp. · Jul 26, 1996
- K210377 — Stryker iBur hubs and cutting accessories · Stryker Corporation · Jun 30, 2021
- K954690 — STRYKER ORAL MAX SYSTEM · Stryker Corp. · Aug 20, 1996
- K233153 — Traus SSG10 Surgical System · Saeshin Precision Co., Ltd. · Sep 20, 2024
- K133604 — ANSPACH EG1 HIGH SPEED SYSTEM, G1 ATTACHMENTS · The Anspach Effort, Inc. · Jan 10, 2014