← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K252564

# Medical Diode Laser Systems (VeinCure) (K252564)

_Baoding Te'Anzhou Electronic Technology Co., Ltd. · GEX · Nov 7, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K252564

## Device Facts

- **Applicant:** Baoding Te'Anzhou Electronic Technology Co., Ltd.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Nov 7, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The “VeinCure” is indicated for use in the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

## Device Story

VeinCure is a GaAlAs diode laser system (1470nm) for endovenous laser treatment. Device consists of laser unit, color touch screen, and footswitch. Physician operates device to deliver laser energy via a disposable sterile fiber (600μm core, SMA905 connector) into the saphenous vein. Laser energy induces thermal ablation to treat venous reflux. Aiming beam (650nm) assists in positioning. System features microprocessor control for operation modes (CW, single pulse, repeat pulse). Used in clinical settings; output affects clinical decision-making by enabling minimally invasive treatment of varicose veins, potentially reducing patient recovery time compared to traditional surgery.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing and adherence to international safety standards for medical electrical equipment and laser safety.

## Technological Characteristics

GaAlAs diode laser; 1470nm wavelength; 0.1W-15W output power; 650nm aiming beam (<3mW). Microprocessor-controlled. Operation modes: CW, single pulse, repeat pulse. Pulse width 10ms-2.50s; repetition rate 0.2Hz-50Hz. Power: 100-240VAC. Dimensions: 320x280x320mm; Weight: 7.1kg. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1. IPX1 (unit), IPX8 (footswitch).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Medical Diode Laser Systems ([K211977](/device/K211977.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

November 7, 2025

Baoding Te'anzhou Electronic Technology Co., Ltd.
% Long Yang
General Manager
Shenzhen Hlongmed Biotech Co., Ltd.
16th floor, Tianming Technology Building
No.8 Wushitou Road, North High-tech District, Nanshan
Shenzhen, 518000
China

Re: K252564
Trade/Device Name: Medical Diode Laser Systems (VeinCure)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: August 14, 2025
Received: August 14, 2025

Dear Long Yang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}

K252564 - Long Yang
Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-

{2}

K252564 - Long Yang
Page 3

devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE-S
Digitally signed by
TANISHA L. HITHE-S
Date: 2025.11.07
22:16:50 -05'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)

K252564

Device Name

Medical Diode Laser Systems (VeinCure)

Indications for Use (Describe)

The “VeinCure” is indicated for use in the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”

{4}

510(k) Summary

(as required by 807.92(c))

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR &amp; 807.92

The assigned 510(K) number is: K252564

Date of Summary: Oct 10, 2025

1. Submitter information

Manufacturer Name: Baoding Te'anzhou Electronic Technology Co., Ltd.

Address: Room 1702, 17th Floor, Zhibo Technology Comprehensive R&amp;d Building, No. 89, Hengbin Road, Baoding, Hebei, China

Contact Person and Title: Ji chuan Cao (General manager)

Tel: 0086-15631235764

Email: lvan@tazlaser.com

2. Contact person

2.1 Primary Contact Person

Long Yang (General manager)

Shenzhen Hlongmed Biotech Co., Ltd.

16th floor, Tianming Technology Building, No.8 Wushitou Road, North High-tech District, Nanshan, Shenzhen, PRC

Tel: 0086-755-86664986

Fax: 0086-755-86664933

E-mail: yanglong@hlongmed.com

2.2 Secondary Contact Person

Ji chuan Cao (General manager)

1 / 7

{5}

Baoding Te'anzhou Electronic Technology Co., Ltd.

Room 1702, 17th Floor, Zhibo Technology Comprehensive R&amp;d Building, No. 89,

Hengbin Road, Baoding, Hebei, China

Tel: 0086-15631235764

E-mail: lvan@tazlaser.com

# 3. Device information

|  Type of 510(k) submission | Traditional  |
| --- | --- |
|  Device name | Medical Diode Laser Systems  |
|  Model | VeinCure  |
|  Common name | Medical Diode Laser Systems  |
|  Review Panel | General & Plastic Surgery  |
|  Product Code | GEX  |
|  Regulation Number | 21 CFR 878.4810  |
|  Device Class | II  |

# 4. Predicate device information

|  510(k) number | K211977  |
| --- | --- |
|  Device name | Medical Diode Laser Systems  |
|  Sponsor | Wuhan Dimed Laser Technology Co., Ltd.  |
|  Product code | GEX  |

# 5. Device Description

Diode laser is a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source. The diode laser for this unit is GaAlAs (Gallium-aluminum-arsenic) diode bar, and the wavelength is  $1470\mathrm{nm}$ . It features impact structure, high efficiency and long lifetime. Generally the beam shall be shamed as the big beam divergence of the laser from the diode. When the

{6}

coaxiality of laser and fiber meet the requirements, the laser beam can be coupled efficiently into the fiber.

# 6. Intended Use/Indications for Use

The "VeinCure" is indicated for use in the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

# 7. Technological characteristics and Substantial Equivalence

The following is a comparison between the Medical Diode Laser Systems (model:VeinCure) (Subject Device) and legally marketed device. The result shows the conformance of Subject Device to the Predicate Devices.

|  No. | Item of Comparison | Predicate device | Subject device | Discussion  |
| --- | --- | --- | --- | --- |
|  1 | 510K Number | K211977 | / | NA  |
|  2 | Proprietary Name | Medical Diode Laser Systems | Medical Diode Laser Systems | NA  |
|  3 | Model | CHERYLAS-15N; CHERYLAS-20N | VeinCure | NA  |
|  4 | Manufacturer | Wuhan Dimed Laser Technology Co., Ltd. | Baoding Te'anzhou Electronic Technology Co., Ltd. | NA  |
|  5 | Product picture |  |  | NA  |

{7}

|  6 | Classification name | Powered Laser Surgical Instrument | Powered Laser Surgical Instrument | similar  |
| --- | --- | --- | --- | --- |
|  7 | Product Code | GEX | GEX | similar  |
|  8 | Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | similar  |
|  9 | Intended use/Indication for use | The “CHERYLAS-15N and CHERYLAS-20N” are indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities. | The “VeinCure” is indicated for use in the treatment of reflux of the saphenous veins associated with varicose veins and varicosities. | similar  |
|  10 | Laser Type | diode laser | diode laser | similar  |
|  11 | Components | Laser system, Color touch screen, Foot switch | Laser system, Color touch screen, Foot switch | similar  |
|  12 | Wavelength | 1470nm | 1470nm | similar  |
|  13 | Output power | CHERYLAS-15N: 0.1W-15W; CHERYLAS-20N: 0.1W-20W; | VeinCure: 0.1W-15W | similar  |
|  14 | Operation mode | CW, single pulse, repeat pulse | CW, single pulse, repeat pulse | similar  |
|  15 | Pulse width | 10ms- 3s | 10ms-2.50s | similar  |
|  16 | Pulse repetition rate | 0.2Hz-50Hz | 0.2Hz-50Hz | similar  |
|  17 | Application / Light delivery | It is recommended to use the disposable sterile fiber (K124003, | It is recommended to use the disposable sterile fiber (K124003, | similar  |

4 / 7

{8}

|   | system | MED-Fibers, Inc.) registered in U.S.A. The parameters must meet the following requirements:
•bare fiber, long as 3m
•Fiber core diameter: 600μm
•NA ≥0.22
•With SMA905 connector
•Single used | MED-Fibers, Inc.) registered in U.S.A. The parameters must meet the following requirements:
•bare fiber, long as 3m
•Fiber core diameter: 600μm
•NA ≥0.22
•With SMA905 connector
•Single used |   |
| --- | --- | --- | --- | --- |
|  18 | Aiming Beam | Diode laser of 650nm, power <3mW, adjustable brightness. | Diode laser of 650nm, power <3mW, adjustable brightness. | similar  |
|  19 | Laser Class | 4 | 4 | similar  |
|  20 | Operation interface | Color LCD touch screen | Color LCD touch screen | similar  |
|  21 | Power Supply | 100-240VAC, 50/60Hz, 1.4A | 100-240VAC,50/60Hz,2. 6-1.0A | Different 1  |
|  22 | Safety classification | ClassI Type B | ClassI Type B | similar  |
|  23 | Dimensions | 380(W)*430(L)*220(H) mm | 320(W)*280(L)*320(H) mm | Different 2  |
|  24 | Weight | 11kg | 7.1 Kg | Different 3  |
|  25 | Waterproof level | IPX1 | IPX1 | similar  |
|  26 | Footswitch Waterproof level | IPX8 | IPX8 | similar  |
|  27 | Standard | IEC60601-1
IEC60601-1-2
IEC60601-2-22
IEC60825-1 | IEC60601-1
IEC60601-1-2
IEC60601-2-22
IEC60825-1 | similar  |

5 / 7

{9}

|  28 | Non-sterile | Yes | Yes | similar  |
| --- | --- | --- | --- | --- |
|  29 | MicroprocessorControl | Yes | Yes | similar  |

The subject device has same intended use, similar product design, same performance effectiveness, and performance safety as the predicate device.

The differences between the subject device and predicate device are minor and both products meet the standard IEC60601-1, IEC60601-1-2, IEC60601-2-22 and IEC60825-1. The differences between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.

# Argument for Substantial Equivalence to Predicate Devices

As seen in the comparison tables, the subject and predicate device have the same intended use and similar technological characteristics. The main technological differences between the subject and predicate device are minor differences, and do not raise different questions of safety or effectiveness. Information contained in this submission demonstrates that any differences in their characteristics do not raise any new questions of safety or effectiveness.

Thus, the subject device is substantially equivalent to the predicate device.

# 8. Performance Data

# Clinical data

Not applicable.

# Non-clinical data:

The Medical Diode Laser Systems (Model: VeinCure) conforms to applicable standards included IEC 60601-1:2005+A1:2012+A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance;

{10}

IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests; IEC60601-2-22:2019 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment and IEC60825-1:2014 Safety of laser products - Part 1: Equipment classification, and requirements.

## 9. Conclusion

The non-clinical data support the safety of the device and the software verification and validation demonstrate that the Medical Diode Laser Systems (model:VeinCure) should perform as intended in the specified use conditions, and all the data demonstrate that the subject device performs comparably to the predicate device that is currently marketed for the same intended use. In other words, the subject device Medical Diode Laser Systems (model:VeinCure) is substantially equivalent to the predicate device K211977.

7 / 7

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K252564](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K252564)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
