← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K012033

# MULTIFUNCTION ENDOSCOPIC INSTRUMENT (K012033)

_Seedling Enterprises., LLC · GEI · Sep 20, 2001 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K012033

## Device Facts

- **Applicant:** Seedling Enterprises., LLC
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Sep 20, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The MEI is a laparoscopic and endoscopic, bipolar, electrosurgical device intended for grasping, dissecting, coagulating and transecting tissue in laparoscopic and open procedures. The types of surgery indicated are: General surgery Laparoscopic procedures Endoscopic procedures Laryngeal coagulation Orthopedic coagulation Thorascopic coagulation Neurosurgical coagulation Gynecological coagulation, (except for use in female sterilization) Ear, Nose and Throat coagulation

## Device Story

Multifunction Endoscopic Instrument (MEI) is a single-patient-use, bipolar electrosurgical device combining scissors and a grasper. Designed for use with 5-10mm trocar cannulae in laparoscopic and open procedures. Operates via manual handle linkages to pivot a central stationary member and a separate scissors blade/forceps jaw for mechanical cutting and grasping. Bipolar electrical energy, delivered via external electrosurgical generators, enables tissue coagulation. Used by surgeons in OR settings to facilitate tissue mobilization and transection. Output is mechanical tissue manipulation and thermal coagulation. Benefits include consolidated functionality for dissection and hemostasis in minimally invasive surgery.

## Clinical Evidence

Bench testing only. Compliance with ANSI/AAMI HF18-1993, IEC 60601-1-2, IEC 601-2-18, and IEC 60601-2-2 standards demonstrated.

## Technological Characteristics

Bipolar electrosurgical device; manual handle-actuated linkages; grasper jaws and scissors blades; compatible with standard electrosurgical generators. Complies with ANSI/AAMI HF18-1993, IEC 60601-1-2, IEC 601-2-18, and IEC 60601-2-2. Single-patient use.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Everest Evershears Bipolar Scissors ([K000496](/device/K000496.md))
- Circon Tripolar Forceps ([K884306](/device/K884306.md))
- Clearglide Bipolar Device ([K003587](/device/K003587.md))
- Bipolar Trigger Flex Electrode ([K003126](/device/K003126.md))
- Select-Sutter Micro-Bipolar Forceps ([K992760](/device/K992760.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

2.

## SEP 2 0 2001

## 510K Summary of Safety and Effectiveness

# K012033

1. Sponsor Name Seedling Enterprises, LLC 150 California Street Newton, MA 02458

- Device Name Multifunction Endoscopic Instrument Proprietary Name: Electrosurgical Cutting and Coagulation Common/Usual Name: Device and Accessories Electrosurgical Cutting and Coagulation Classification Name: Device and Accessories Class II by the General and Plastic Surgery Panel (21 CFR 878.4400) Product Code GEI.
- Identification of Predicate or Legally Marketed Device 3. The Multifunction Endoscopic Instrument is substantially equivalent in intended use and/or function to the following predicate devices:
	- Everest Evershears Bipolar Scissors K 000496, manufactured by Everest . Medical
	- Circon Tripolar Forceps 510(k) number unknown, believed to be legally . marketed under K884306, manufactured by Circon Corporation
	- Clearglide Bipolar Device K 003587, manufactured by Ethicon, Inc. .
	- Bipolar Trigger Flex Electrode K003126, manufactured by ellman . international inc
	- Select-Sutter Micro-Bipolar Forceps K992760, manufactured by Select . Sutter GmbH

#### Device Description 4.

The Multifunction Endoscopic Instrument is a single patient use device, which is a combination scissors and bipolar coagulator-grasper. It is intended for use in canjunction with presently employed 5-10mm trocar cannulae. The instrument grasping and cutting functions are operated easily using one hand. The device is indicated for use in a variety of open and minimally invasive procedures. It is designed to facilitate grasping, mobilization, dissection, coagulation and transection of tissue.

{1}------------------------------------------------

The grasping and cutting functions of the distal end are accomplished by employing three elements.

- A central stationary member serves as both a scissors blade and a O forceps jaw.
- A separate scissors blade and a forceps jaw are attached to the o A separate some that they can be independently pivoted to open and close thereby performing the cutting and grasping functions.
- The pivoting scissors blade and forceps jaw are connected to the ০ handle and actuated via linkages.
- 5. Intended Use

The MEI is a laparoscopic and endoscopic, bipolar, electrosurgical device intended for grasping, dissecting, coagulating and transecting tissue in laparoscopic and open procedures.

The types of surgery indicated are:

General surgery Laparoscopic procedures Endoscopic procedures Laryngeal coagulation Orthopedic coagulation Thorascopic coagulation Neurosurgical coagulation Gynecological coagulation, (except for use in female sterilization) Ear, Nose and Throat coagulation

### Comparison of Technological Characteristics 6.

The Multifunction Endoscopic Instrument is substantially equivalent to the predicate devices listed, which provide the same or similar functions. All of the devices are bipolar scissors or graspers that permit the coagulation of soft tissue through the use of bipolar energy. All of the devices use sharp objects (scissors or knives) to permit the surgeon to cut the desiccated tissues. All are connected to the same or similar electrosurgical generators and use similar power ranges for operation.

{2}------------------------------------------------

The technological characteristics are the same as the predicates in that they are all electrosurgical devices that use bipolar electrical energy to trovide coagulation powered by electrosurgical generators, and provide cutting with either a knife or a scissors. They are also similar in construction in that they have a handle, shaft and end effectors (working oonstraonomin that that that the predicate devices are comprised of grasping or scissors jaws with the electrode surfaces, and embedded knives or scissors. The working end of the MEI is comprised of grasper jaws with the electrode surfaces, and scissor blades.

The intended use, statement of indications and technological and performance characteristics of the Multifunction Endoscopic Instrument support the concept of substantial equivalence.

- 7. Performance Testing
The Multifunction Endoscopic Instrument complies with the following voluntary standards:

- ANSI/AAMI HF18-1993 Electrosurgical Devices .
- IEC 60601-1-2 Medical Electrical Equipment General Requirements for . Safety
- IEC 601-2-18 Medical Electrical Equipment Part 2: Particular . Requirements for safety of Endoscopic Equipment
- IEC 6060 1-2-2 Medical Electrical Equipment Part 2-2: Particular . Requirements for the Safety of High Frequency Surgical Equipment

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines beneath them.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 0 2001

Seedling Enterprises, LLC c/o Ms. Debbie Iampietro QRC Consulting Associates 7 Tiffany Trail Hopkinton, Massachusetts 01748

Re: K012033

Trade/Device Name: Multifunction Endoscopic Instrument Regulation Number: 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: June 27, 2001 Received: June 29, 2001

Dear Ms. Iampietro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 as stated in the encreated) is the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may , are erovisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Debbie Iampietro

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Lisa Walker, us

Celia M. Witten, Ph.D., M.D.

Image /page/4/Picture/5 description: The image shows a close-up, hand-drawn sketch of a person's face. The drawing is simple, with thick lines defining the contours of the face, including the nose, mouth, and chin. The style is minimalist, focusing on essential features rather than detailed rendering.

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): K O12033

Seedling Enterprises Multifunction Endoscopic Instrument Device Name:

Indications For Use:

The MEI is a laparoscopic and endoscopic, bipolar, electrosurgical device intended for grasping, dissecting, coagulating and transecting tissue in laparoscopic and open procedures.

The types of surgery indicated are:

General surgery Laparoscopic procedures Endoscopic procedures Laryngeal coagulation Orthopedic coagulation Thorascopic coagulation Neurosurgical coagulation Gynecological coagulation, (except for use in female sterilization) Ear, Nose and Throat coagulation

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number_Kosz 2033

000010

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K012033](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K012033)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
