K013417 · Misonix, Inc. · GBX · Jan 10, 2002 · General, Plastic Surgery
Device Facts
Record ID
K013417
Device Name
MISONIX IRRIGATION SYSTEM MOEDL BC20P
Applicant
Misonix, Inc.
Product Code
GBX · General, Plastic Surgery
Decision Date
Jan 10, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4200
Device Class
Class 1
Indications for Use
The Misonix Irrigation System Model BC20P is indicated for use in the delivery of fluids in the following surgical specialties: General Surgical Procedures. The Misonix Irrigation System (BC20P) is not intended for the administration of parenteral fluids, infusion of drugs or for any life sustaining purposes.
Device Story
Liquid transfer system; utilizes standard/custom tubing sets connected to IV bottles, bags, or reservoirs. Pump head draws and pressurizes liquid for transfer to a cannula or orifice for site flushing and cleaning. Used to refill small containers from larger reservoirs. Operated in surgical settings. Output is pressurized fluid flow; assists healthcare providers in site irrigation. Benefits patient by facilitating surgical site cleaning.
Clinical Evidence
No clinical data. Substantial equivalence determined via bench testing including output power, irrigation flowrate, vacuum flowrate, pressure measurements, and electrical safety testing.
Technological Characteristics
Liquid transfer pump system. Components include pump head and tubing sets. Electrical components subject to EMI and dielectric testing. No specific materials or software algorithms described.
Indications for Use
Indicated for fluid delivery during general surgical procedures. Not for parenteral fluid administration, drug infusion, or life-sustaining purposes.
Regulatory Classification
Identification
An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions. Examples include irrigation and drainage catheters, pediatric catheters, peritoneal catheters (including dialysis), and other general surgical catheters. An introduction/drainage catheter accessory is intended to aid in the manipulation of or insertion of the device into the body. Examples of accessories include adaptors, connectors, and catheter needles.
K974233 — LYSONIX IRRIGATION SYSTEM · Lysonix, Inc. · Feb 3, 1998
K962090 — KARL STORZ MODEL 203210 O1 UNIMAT PLUS · KARL STORZ Endoscopy-America, Inc. · Sep 30, 1996
K071588 — DISPENSER DP 20 · NOUVAG AG · Nov 26, 2007
K063326 — CONMED LINVATEC 24K IRRIGATION SYSTEM · Conmed Linvatec · Jan 31, 2007
K990724 — CONWAY STUART MODEL 100 IRRIGATION PUMP & ACCESSORIES · Conway Stuart Medical, Inc. · Aug 13, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
## 510K SUMMARY - MISONIX IRRIGATION SYSTEM MODEL BC20P
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of SMSA 1990 and 21CFR 807.92.
The assigned 510(k) number is: K013417
#### Submitter's Identification 1.
| Name: | MISONIX, INC. |
|-------------------|-----------------------------------------|
| Address: | 1938 New Highway, Farmingdale, NY 11735 |
| Telephone Number: | (516) 694-9555 |
| Contact Person: | Ronald R. Manna |
| Date Prepared: | 5 October 2001 |
| Date Revised: | 8 February 2002 |
#### 2. Name of Device
Misonix Irrigation System (Model BC20P) Proprietary Name:
BC20P Common / Usual Name:
Classification Name: Pump, Irrigation
GBX Product Code:
#### 3. Predicate Device Information
LySonix Irrigation System K974233 Predicate Device
#### 4. Device Description:
The Misonix Irrigation System (BC20P) is a liquid transfer system. It is designed to utilize standard or custom tubing sets, which mate to IV bottles or bags, open reservoirs or other liquid canisters. The liquid is drawn through the pump head and then pressurized, allowing the liquid to be transferred from the source. The outlet of the tube set may be connected to a cannula or other orifice to allow flushing and cleaning of the site. It can be used to refill small containers from a larger reservoir. It is not intended for drug delivery or the administration of parenteral fluids.
#### પં Intended Use:
The Misonix Irrigation System Model BC20P is indicated for use in the delivery of fluids in the following surgical specialties:
General Surgical Procedures
#### 6. Comparison to Predicate Device:
The Misonix Irrigation System is identical in all aspects to the LySonix Irrigation System since both units have been designed and manufactured to the same specifications by Misonix.
{1}------------------------------------------------
### Discussions of Non-Clinical Tests Performed for Determination of Substantial 7. Equivalence are as follows:
Output Power Measurements (No Load to Maximum Load) Irrigation Flowrate Measurements Life Tests Vacuum Flowrate and Pressure Measurements Input Power Measurements EMI Tests Dielectric Tests on Mains Circuits Patient Current Leakage and Patient Sink Current Measurements Power Line Ground Leakage Measurements Dielectric Tests on Patient Circuits.
#### Discussions of Clinical Tests Performed 8.
N/A
#### 9. Conclusions
Based upon an analysis of the operating characteristic specifications, Output of Engineering Tests, Hazard Analysis, Voluntary Consensus Standard Investigations and the designing activities, Misonix, Inc. has concluded that the Misonix Irrigation System is substantially equivalent to the LySonix Irrigation System.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three curved lines representing its wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JAN 1 0 2002
Misonix, Inc. Albert F. Clancy, Jr. Manager, QA/RA 1938 New Highway Farmingdale, New York 11735
Re: K013417
Trade Name: Misonix Irrigation System (Model BC20P) Regulation Number: 878.4200; 876.5220 Regulation Name: Introduction/Drainage Catheter and Accessories; Colonic Irrigation System Regulatory Class: II Product Code: GBX; KPL Dated: October 5, 2001 Received: October 15, 2001
Dear Mr. Clancy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Albert Clancy
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, «Misbranding by reference to premarket notification» (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Celia M. Witten, Ph.D., M.D.
Witten, Ph.D Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# PREMARKET NOTIFICATION
### INDICATIONS FOR USE
510k Number: K013417
Device Name: Misonix Irrigation System (Model BC20P)
Indications for Use/ Intended Use:
The Misonix Irrigation System Model BC20P is indicated for use in the delivery of fluids in the following surgical specialties: .
## General Surgical Procedures
The Misonix Irrigation System (BC20P) is not intended for the administration of parenteral fluids, infusion of drugs or for any life sustaining purposes.
# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restora
and Neurological Devices
| 510(k) Number | K012417 |
|---------------|---------|
|---------------|---------|
| Prescription Use (Per 21CRF 801.109) | <div>✓</div> |
|--------------------------------------|--------------|
| OR | |
| Over the Counter Use | |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.