← Product Code [GAW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAW) · K193530

# Deklene MAXX (K193530)

_Teleflex Medical · GAW · Mar 19, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAW/K193530

## Device Facts

- **Applicant:** Teleflex Medical
- **Product Code:** [GAW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAW.md)
- **Decision Date:** Mar 19, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.5010
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Deknatel® Deklene® Maxx™ Polypropylene Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular and neurological procedures, but not for use in ophthalmic procedures.

## Device Story

Nonabsorbable, sterile, monofilament polypropylene surgical suture; used for soft tissue approximation and ligation in cardiovascular and neurological procedures. Available in USP sizes 8-0 through 0; supplied with or without 300/400 series stainless steel needles and PTFE felt pledgets. Used by surgeons in clinical settings. Provides mechanical tissue closure; maintains tensile strength over time. MR safe (needles removed).

## Clinical Evidence

No clinical trials were conducted. Substantial equivalence is supported by bench testing (USP non-absorbable suture requirements) and biocompatibility evaluation per ISO 10993-1:2018.

## Technological Characteristics

Monofilament polypropylene (synthetic linear polyolefin). USP sizes 8-0 to 0. 300/400 series stainless steel needles. PTFE felt pledgets. Sterilized via Ethylene Oxide (SAL 10^-6). Compliant with USP <861>, <871>, <881> and ISO 10993-1. MR safe (needles removed).

## Regulatory Identification

Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Deknatel® Deklene® Maxx™ Polypropylene Suture ([K930738](/device/K930738.md))
- Deknatel® Deklene® Maxx™ Polypropylene Suture ([K153076](/device/K153076.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains the words "Convert to PDF" in a large, bold font. The text is centered and appears to be the main focus of the image. The background is a light gray color, which provides a neutral backdrop for the text.

Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 19, 2020

Teleflex Medical Robin Haden Regulatory Affairs Specialist 3015 Carrington Mill Blvd Morrisville, North Carolina 27560

Re: K193530

Trade/Device Name: Deklene MAXX Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable Polypropylene Surgical Suture Regulatory Class: Class II Product Code: GAW Dated: December 19, 2019 Received: December 20, 2019

Dear Robin Haden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193530

#### Device Name

Deknatel® Deklene® MAXX™ Polypropylene Surgical Suture

Indications for Use (Describe)

Deknatel® Deklene® MAXX™ Polypropylene Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, and neurological procedures, but not for use in ophthalmic procedures.

Type of Use (Select one or both, as applicable)

| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)             |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

| Submission Date:<br>K193530 | 12/20/2019                                                     |
|-----------------------------|----------------------------------------------------------------|
| SUBMITTER INFORMATION:      |                                                                |
| Company Name:               | Teleflex Medical, Inc.                                         |
| Company Address:            | 3015 Carrington Mill Blvd, Morrisville, NC 27560               |
| Contact Person:             | Robin Haden<br>Phone: 919-544-8000<br>robin.haden@teleflex.com |
| Device Trade Name:          | Deknatel® Deklene® Maxx™ Polypropylene Surgical Suture         |
| Device Common Name:         | Polypropylene nonabsorbable surgical suture                    |
| Class:                      | Class II                                                       |
| Classification:             | 21 CFR 878.5010                                                |
| Product Code:               | GAW                                                            |

# Predicate Devices:

The predicate is Deknatel® Deklene® Maxx™ Polypropylene Suture devices cleared under K930738 (July 26, 1994) and K153076 (June 16, 2016).

# Device Description:

Deknatel® Deklene® Maxx™ Polypropylene Suture is a nonabsorbable, sterile surgical suture composed of a strand of polypropylene, a synthetic linear polyolefin. Deknatel® Deklene® MAXX™ Polypropylene Surgical Suture is available dyed (blue) or undyed (colorless).

Deknatel® Deklene® Maxx™ Polypropylene Surgical Suture meets all nonabsorbable surgical suture requirements established by the United States Pharmacopeia (USP) except for suture sizes 7-0 and 8-0 which differ from USP in diameter only.

Deknatel® Deklene® Maxx™ Polypropylene Surgical Suture is available in USP sizes 8-0 through 0. The suture is provided in a variety of lengths, with and without 300 or 400 series stainless steel needles, with and without PTFE felt pledgets, and may be supplied in a variety of cut lengths. Finished suture may be packaged in cartons as single packs, multipacks or procedure packs.

{4}------------------------------------------------

# Intended Use:

Deknatel® Deklene® Maxx™ Polypropylene Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular and neurological procedures, but not for use in ophthalmic procedures.

{5}------------------------------------------------

# Technological Characteristics

The proposed Deknatel® Deklene® Maxx™ Polypropylene Suture is substantially equivalent to its predicate Deknatel® Deklene® Maxx™ Polypropylene Suture devices because there are no differences in technological or performance characteristics between the proposed and predicate devices. The difference between the proposed device and the predicate device is the supplier from which the resin used for extrusion is acquired. The resin from the new supplier is equivalent to that from the current supplier and does not introduce any new issues of safety and effectiveness. The results of the comparison between the technological characteristics of Deknatel® Deklene® Maxx™ Polypropylene Suture manufactured with the proposed and predicate material demonstrate that the proposed Deknatel® Deklene® Maxx™ Polypropylene Sutures are equivalent to the marketed predicate devices in performance characteristics.

# Materials

All patient contacting materials are in compliance with the requirements of ISO 10993-1. The proposed modification is replacing the source of the polypropylene resin used to manufacture the Deknatel® Deklene® Maxx™ Polypropylene Suture.

# Summary of Testing

Deknatel® Deklene® Maxx™ Polypropylene Suture is tested in accordance with the requirements of USP – non-absorbable surgical sutures for suture diameter, , tensile strength and needle attachment, and meet the requirements of Class II Special Controls Guidance: Surgical Sutures; Guidance for Industry and FDA; June 3, 2003.

All materials used in the fabrication of the Deknatel® Maxx™ Polypropylene Suture were evaluated through biological qualification safety tests as outlined in ISO 10993-1:2018 – Biological Evaluation of Medical Devices - Part I: Evaluation and Testing.

Deknatel® Deklene® Maxx™ Polypropylene Suture is tested to demonstrate it is "MR Safe" and poses no hazards in an MR environment with any needles removed.

{6}------------------------------------------------

# Substantial Equivalence:

The proposed Deknatel® Deklene® Maxx™ Polypropylene Surgical Suture is substantially equivalent in intended use and fundamental scientific technology to the Deknatel® Deklene® Maxx™ Polypropylene Suture devices cleared under K930738 (July 26, 1994) and K153076 (June 16, 2016). The substantial equivalence table for the comparison of the previously cleared devices and the proposed devices is included in the Table below.

| Comparative<br>Characteristics                      | Predicate Device<br>(K153076, K930738)                                                                                                                                                                                      | Proposed Device                   | Explanation<br>of Differences        |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|--------------------------------------|
| Manufacturer                                        | Teleflex Medical                                                                                                                                                                                                            |                                   | Identical                            |
| Product Code                                        | GAW                                                                                                                                                                                                                         |                                   | Identical                            |
| Regulation                                          | 878.5010                                                                                                                                                                                                                    |                                   | Identical                            |
| Device Class                                        | 2                                                                                                                                                                                                                           |                                   | Identical                            |
| Indications for Use                                 | Polypropylene Surgical Sutures are indicated for use in<br>general soft tissue approximation and/or ligation, including<br>use in cardiovascular, and neurological procedures, but not for<br>use in ophthalmic procedures. |                                   | Identical                            |
| Contraindications                                   | None known.                                                                                                                                                                                                                 |                                   | Identical                            |
| Trade Names                                         | Deklene® Maxx™                                                                                                                                                                                                              |                                   | Identical                            |
| Deklene®<br>Maxx™ USP<br>Size                       | 8-0 to 2                                                                                                                                                                                                                    | 8-0 to 0                          | Equivalent<br>within stated<br>range |
| Suture Diameter                                     | All sizes except 7-0 and 8-0 meet USP <861> requirements<br>for nonabsorbable surgical suture                                                                                                                               |                                   | Identical                            |
| Suture Tensile<br>Strength                          | All sizes meet USP <881> requirements for nonabsorbable<br>surgical suture                                                                                                                                                  |                                   | Identical                            |
| Needle Attachment<br>Strength                       | All sizes meet USP <871> requirements for nonabsorbable<br>surgical suture                                                                                                                                                  |                                   | Identical                            |
| Material                                            | Polypropylene (synthetic linear polyolefin)                                                                                                                                                                                 |                                   | Identical                            |
| Suture Construction                                 | Monofilament                                                                                                                                                                                                                |                                   | Identical                            |
| Coated                                              | No                                                                                                                                                                                                                          |                                   | Identical                            |
| Uncoated                                            | Yes                                                                                                                                                                                                                         |                                   | Identical                            |
| Undyed White                                        | Yes                                                                                                                                                                                                                         |                                   | Identical                            |
| Dyed Blue with<br>[phthalocyaninato(2<br>-)] copper | Yes                                                                                                                                                                                                                         |                                   | Identical                            |
| Impregnated PTFE<br>Pledget                         | Yes                                                                                                                                                                                                                         |                                   | Identical                            |
| Pledget Material                                    | polytetrafluoroethylene (PTFE)                                                                                                                                                                                              |                                   | Identical                            |
| Needle Material                                     |                                                                                                                                                                                                                             | 300 or 400 Series Stainless Steel | Identical                            |
| Package Materials                                   | Solid Sulfite Board (SBS)/ Uncoated Tyvek Pouch                                                                                                                                                                             |                                   | Identical                            |
| Sterility Method                                    | Ethylene Oxide (EO);<br>SAL 10-6                                                                                                                                                                                            |                                   | Identical                            |
| Biocompatibility                                    | Per ISO 10993-1                                                                                                                                                                                                             |                                   | Identical                            |
| Shelf Life                                          |                                                                                                                                                                                                                             | 5 Years                           | Identical                            |

{7}------------------------------------------------

# Clinical Tests Performed:

# No clinical trials were conducted

# Conclusion:

The proposed Deknatel® Deklene® Maxx™ Polypropylene Suture is substantially equivalent to its predicate Deknatel® Deklene® Maxx™ Polypropylene Suture devices because there are no differences in technological or performance characteristics between the proposed and predicate devices.

The proposed modification is replacing the source of the polypropylene resin used to manufacture the Deknatel® Deklene® Maxx™ Polypropylene Suture. It also has the same design being a sterile, flexible, monofilament nonabsorbable thread meeting the requirements of the United States Pharmacopeia.

The biocompatibility data and the results of performance testing presented, demonstrate the substantial equivalence of Deklene® Maxx™ Polypropylene Suture to that of the predicate devices. It further demonstrates conformance with the USP, ISO 10993 and FDA Guidance for Surgical Suture 510(k).

The new device Deknatel® Deklene® MAXX™ Polypropylene Surgical Suture is substantially equivalent in intended use and fundamental scientific technology to the predicate devices Deknatel® Deklene® Maxx™ Polypropylene Suture cleared under K930738 and K153076.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAW/K193530](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAW/K193530)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
