← Product Code [GAP](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAP) · K162396

# SILK Surgical Suture (K162396)

_Teleflexmedical, Inc. · GAP · Nov 23, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAP/K162396

## Device Facts

- **Applicant:** Teleflexmedical, Inc.
- **Product Code:** [GAP](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAP.md)
- **Decision Date:** Nov 23, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5030
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Deknatel® Silk Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.

## Device Story

Deknatel Silk Surgical Suture is a sterile, nonabsorbable, braided or twisted suture composed of fibroin protein derived from Bombyx mori. It is waxed with beeswax or a beeswax/paraffin mixture and available in dyed (black) or undyed (natural) forms. Used by surgeons in clinical settings for soft tissue approximation and ligation. The device is provided in various lengths, with or without needles/pledgets, or on ligating reels. It functions as a mechanical support for tissue healing. The current modification involves replacing the mordant Sodium Dichromate with Ferrous Sulfate in the dyeing process. The suture is intended for permanent implantation (>30 days).

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing performed in accordance with USP standards for nonabsorbable sutures, including diameter, needle attachment, and tensile strength. Biocompatibility testing was conducted per ISO 10993, including cytotoxicity, sensitization, irritation, systemic toxicity, subchronic toxicity, genotoxicity, implantation, haemocompatibility, and pyrogen testing.

## Technological Characteristics

Nonabsorbable surgical suture composed of fibroin protein (Bombyx mori). Waxed with beeswax or beeswax/paraffin. Available in USP sizes 10-0 through 5. Mordant: Ferrous Sulfate. Meets USP <861> (Diameter), <871> (Needle Attachment), and <881> (Tensile Strength) standards. Sterile, permanent implant.

## Regulatory Identification

Natural nonabsorbable silk surgical suture is a nonabsorbable, sterile, flexible multifilament thread composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori ) of the family Bombycidae. Natural nonabsorbable silk surgical suture is indicated for use in soft tissue approximation. Natural nonabsorbable silk surgical suture meets the United States Pharmacopeia (U.S.P.) monograph requirements for Nonabsorbable Surgical Suture (class I). Natural nonabsorbable silk surgical suture may be braided or twisted; it may be provided uncoated or coated; and it may be undyed or dyed with an FDA listed color additive.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Deknatel® Nonabsorbable Surgical Sutures ([K153076](/device/K153076.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 23, 2016

Teleflex Medical, Inc Ms. Natalie Hichak Sr. Regulatory Affairs Specialist 3015 Carrington Mill Blvd Morrisville, North Carolina 27560

Re: K162396

Trade/Device Name: Silk Surgical Suture Regulation Number: 21 CFR 878.5030 Regulation Name: Natural Nonabsorbable Silk Surgical Suture Regulatory Class: Class II Product Code: GAP Dated: August 24, 2016 Received: August 26, 2016

Dear Ms. Hichak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162396

Device Name Deknatel® Silk Surgical Sutures

Indications for Use (Describe)

Deknatel® Silk Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.

| Type of Use (Select one or both, as applicable)                            |
|----------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </div> |
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) SUMMARY

## Deknatel® Silk Surgical Sutures

#### I. SUBMITTER

Teleflex Medical, Inc. 3015 Carrington Mill Blvd Morrisville, NC 27560 Phone: 919-433-8049 Fax: 919-433-4996

Contact Person: Natalie Hichak, Senior Regulatory Affairs Specialist Date Prepared: August 24, 2016

### II. DEVICE

| Name of Device      | Deknatel Silk Surgical Suture              |
|---------------------|--------------------------------------------|
| Common Name         | Suture, Nonabsorbable, Silk                |
| Classification Name | Natural Nonabsorbable Silk Surgical Suture |
| Regulatory Class    | II                                         |
| Product Code        | GAP                                        |
| Regulation          | 878.5030                                   |

### III. PREDICATE DEVICE

Deknatel® Nonabsorbable Surgical Sutures, K153076

### IV. DEVICE DESCRIPTION

Deknatel Silk Suture is a nonabsorbable, sterile, surgical suture composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. Silk Suture is waxed, braided or twisted, and provided dyed (black) or undyed (natural white).

Silk Suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP sizes 10-0 through 5 (metric sizes 0.2 through 7). The suture is supplied sterile, waxed, and braided or twisted. The suture is provided in a variety of lengths, with and without needles, with and without pledgets, and may be supplied in a variety of cut lengths or on ligating reels. Finished suture may be packaged in cartons as single packs, multipacks or procedure packs.

# V. INDICATIONS FOR USE

Deknatel Silk Surgical Sutures is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.

{4}------------------------------------------------

The Indications for Use statement for Deknatel Silk Surgical Sutures is identical to the predicate device.

## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The proposed Deknatel Silk Surgical Sutures have the same technology, indications for use and functional characteristics as the predicate system. The proposed modification is a dying process change to replace the current mordant, Sodium Dichromate with Ferrous Sulfate.

A comparison of the technological characteristics of the proposed Deknatel Silk Surgical Sutures and the predicate has been performed. The results of this comparison demonstrate that the Silk Surgical Sutures are equivalent to the marketed predicate devices in performance characteristics.

### VII. PERFORMANCE DATA

Non-clinical testing has been performed in accordance USP (United States Pharmacopeia) Nonabsorbable Sutures, 36-NF 31 <861> Sutures -Diameter, <871> Sutures- Needle Attachment, and <881>Tensile Strength in order to verify mordant material change of the proposed Deknatel Silk Surgical Sutures are substantially equivalent to the predicate devices.

### Biocompatibility Testing

The biocompatibility evaluation for the Deknatel Silk Surgical Suture was conducted in accordance with FDA Blue Book Memorandum #G95-1 "Use of International Standards ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standards ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by the FDA. The battery of testing included the following tests:

- . Cytotoxicity
- Sensitization ●
- Irritation ●
- Systemic toxicity ●
- Subchronic toxicity
- Genotoxicity ●
- Implantation
- Haemocompatibility
- Pyrogen testing

The Deknatel Silk Surgical Sutures is considered a permanent implant (> 30 days) and the nature of body contact is tissue/bone and blood. Deknatel Silk

{5}------------------------------------------------

Suture is composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. Silk Suture is waxed with beeswax or a beeswax and paraffin mixture. Black silk suture is dyed with Logwood dye.

### VII. CONCLUSIONS

Based upon the comparative test results, the proposed Deknatel Silk Surgical Sutures are substantially equivalent in performance to the predicate devices cleared to market via 510(k)s K153076.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAP/K162396](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAP/K162396)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
