← Product Code [GAM](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM) · K181652

# Coated Vicryl Plus Antibacterial (Polyglactin 910) Absorable Suture (K181652)

_ETHICON, Inc. · GAM · Sep 28, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM/K181652

## Device Facts

- **Applicant:** ETHICON, Inc.
- **Product Code:** [GAM](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM.md)
- **Decision Date:** Sep 28, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4493
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Coated VicryI™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues.

## Device Story

Coated Vicryl™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is a sterile, synthetic, absorbable surgical suture; composed of 90% glycolide and 10% L-lactide copolymer; coated with a mixture of Polyglactin 370 and calcium stearate; contains IRGACARE® MP (triclosan) as an antibacterial agent. Used by surgeons in clinical settings for soft tissue approximation and ligation. The device functions as a mechanical support for tissue healing; the antibacterial agent is intended to inhibit bacterial colonization on the suture. The device is absorbed by the body over time, eliciting mild tissue reaction. No changes to materials, construction, or manufacturing compared to the predicate; labeling updated to include recent meta-analysis and guidelines regarding triclosan-coated sutures for surgical site infection reduction.

## Clinical Evidence

No clinical data provided; bench testing not required as technological characteristics are identical to the predicate device.

## Technological Characteristics

Synthetic absorbable suture composed of 90% glycolide/10% L-lactide copolymer. Coated with Polyglactin 370 and calcium stearate. Contains antibacterial agent IRGACARE® MP (triclosan) at ≤ 472 µg/m. Available in dyed (D&C Violet No. 2) or undyed (natural) forms. Nonantigenic and nonpyrogenic.

## Regulatory Identification

An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Coated Vicryl™ Plus Antibacterial (Polyglactin 910) Absorable Suture ([K132580](/device/K132580.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 28, 2018

Ethicon, Inc. % Ms. Donna Marshall Regulatory Affairs Manager Route 22 West, P.O. Box 151 Somerville, New Jersey 08876-0151

Re: K181652

Trade/Device Name: Coated VicryI™ Plus Antibacterial (Polyglactin 910) Absorbable Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: August 30, 2018 Received: August 31, 2018

Dear Ms. Marshall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products, and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K181652

Device Name

Coated Vicryl™ Plus Antibacterial (Polyglactin 910) Absorable Suture

Indications for Use (Describe)

Coated VicryI™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues.

Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K181652 p. 1 of 2

Image /page/3/Picture/1 description: The image shows the logo for Ethicon, Inc., a Johnson & Johnson company. The word "ETHICON" is written in large, red, sans-serif letters, with the ",INC." in smaller, red letters to the right. Below the company name is the phrase "a Johnson+Johnson company" in a cursive font.

510(k) Summary

| Submitter:      | Ethicon Inc. a Johnson & Johnson company<br>P.O. Box 151<br>Route 22 West<br>Somerville, NJ 08876-0151<br>USA                                                           |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Donna Marshall<br>Manager, Regulatory Affairs<br>Ethicon, Inc. a Johnson & Johnson company<br>Ph: (908) 541-3990<br>Fax: (908) 218-2595<br>e-mail: dmarsha2@its.jnj.com |

| Date Prepared: | September 26, 2018 |
|----------------|--------------------|
|----------------|--------------------|

| Device Trade Name:   | Coated VICRYL™ Plus Antibacterial (Polyglactin 910)<br>Synthetic Absorbable Suture |
|----------------------|------------------------------------------------------------------------------------|
| Device Common Name:  | Suture, Surgical, Absorbable                                                       |
| Class:               | II                                                                                 |
| Classification Name: | Absorbable Poly(glycolide/L-lactide) Surgical Suture<br>(21 CFR 878.4493)          |
| Product Code:        | GAM                                                                                |
| Panel:               | General and Plastic Surgery Devices                                                |

### Predicate Device:

| Device Trade Name                                                       | 510(k) Number |
|-------------------------------------------------------------------------|---------------|
| Coated Vicryl™ Plus Antibacterial<br>(Polyglactin 910) Absorable Suture | K132580       |

## Device Description:

Coated VicryI™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is a sterile, synthetic absorbable surgical suture (dyed and undyed) and is composed of a copolymer made from 90% glycolide and 10% L-lactide. Coated Vicryl™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is coated with a mixture composed of equal parts of a copolymer of glycolide and lactide (Polyglactin 370) and calcium stearate. The suture contains IRGACARE®* MP (triclosan), a broad spectrum antibacterial agent, at no more than 472 µg/m. The copolymers in

{4}------------------------------------------------

the product have been found to be nonantigenic, nonpyrogenic and elicit only a mild tissue reaction during absorption. The suture is available undyed(natural) or dyed (D&C Violet No. 2).

## Indications for Use:

Coated Vicry|™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues.

## Performance Data:

The technological characteristics of the proposed device is identical to the predicate device, therefore, performance data are not necessary to establish substantial equivalence.

### Substantial Equivalence:

The subject device is identical to the predicate device with respect to functionality, technological characteristics and intended use. There is no material, device construction, performance specification, packaging, sterilization or manufacturing process changes to the currently marketed devices. The device differs only in the labeling (Instructions for Use) that have been revised to include the references of the latest publication of meta-analysis and guidelines for consideration of the use of triclosan-coated sutures to lower surgical site infection rates in the Actions Section. The proposed device does not raise new questions of safety or effectiveness as the predicate device and therefore substantially equivalent.

### Conclusion:

Based on the intended use, fundamental scientific technology, technological characteristics and the intended use, the following subject device Coated Vicry]™ Plus Antibacterial (Polyglactin 910) Absorable Suture is considered to be substantially equivalent to the predicated device.

* Trademark IRGACARE®* MP (triclosan) "Registered Trademark of BASF Group"

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM/K181652](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM/K181652)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
