Disposable Polymer Ligation Clips

K250643 · Hangzhou Kangji Medical Instrument Co., Ltd. · FZP · Nov 20, 2025 · General, Plastic Surgery

Device Facts

Record IDK250643
Device NameDisposable Polymer Ligation Clips
ApplicantHangzhou Kangji Medical Instrument Co., Ltd.
Product CodeFZP · General, Plastic Surgery
Decision DateNov 20, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4300
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Disposable polymer ligation clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.

Device Story

Disposable, non-absorbable, non-active implantable polymer clips; used for vessel or tissue ligation in general surgery. Clips housed in cartridges; manually loaded into reusable, non-sterile clip appliers (Kangji-manufactured). Surgeon places clip around tissue; closes via applier grip. Output is a secured, ligated vessel/tissue structure. Benefits include reliable hemostasis or tissue closure. Used in clinical/surgical settings by surgeons. Device supplied sterile (EO); single-use.

Clinical Evidence

No clinical data. Bench testing only: appearance, dimension, assembly firmness, matching, locking, toughness, clamping, tensile, and pressure endurance performance. Biocompatibility per ISO 10993 (cytotoxicity, sensitization, irritation, pyrogenicity, implantation, chemical characterization, toxicological risk assessment). Sterilization and shelf-life validated per ISO/EN standards.

Technological Characteristics

Material: medical grade acetyl homopolymer. Principle: mechanical ligation via clip applier. Form factor: clips in cartridges (various sizes M, ML, L, XL). Sterilization: EO (10^-6 SAL). Standards: ISO 10993 (biocompatibility), ISO 11607 (packaging), EN ISO 11135 (sterilization), ISTA 2A/3A (transportation).

Indications for Use

Indicated for ligation of vessels or tissue structures during general surgical procedures. No specific age or gender restrictions; contraindicated if the clip size is inappropriate for the vessel or tissue structure.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION November 20, 2025 Hangzhou Kangji Medical Instrument Co., Ltd. % Esther Zhang Official Correspondent Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District Shanghai, China Re: K250643 Trade/Device Name: Disposable Polymer Ligation Clips Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: October 21, 2025 Received: October 21, 2025 Dear Esther Zhang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K250643 - Esther Zhang Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250643 - Esther Zhang Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Alexander Nguyen -S for Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250643 | | | Device Name Disposable polymer ligation clips | | | Indications for Use (Describe) The Disposable Polymer Ligation Clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K250643 Page 1 of 7 510(k) summary I Submitter Device submitter: Hangzhou Kangji Medical Instrument Co., Ltd. No. 1668 Chunjiang East Road, Economic Development Zone, Tonglu, Hangzhou, 311501, China. Contact person: Martin Sun Manager of Regulatory Affairs Phone: +86-0571-69901712 Fax: +86-0571-69901712 Email: ra01@hzkangji.com Date prepared: Nov 20, 2025 II Device Trade Name of Device: Disposable polymer ligation clips Common Name: ligation clip Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: II Product code: FZP Review Panel: General and Plastic Surgery III Correspondent Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District, Shanghai, P.R.China Contact: Esther ZHANG Email: Esther.zhang@llins-tech.com IV Device description The Disposable polymer ligation clips includes Polymer Clips and Multiple polymer clips. The devices are sterilized by EO, single-use, non-absorbable, non-active implantable devices designed for use in general surgical procedures that require vessel or tissue ligation. The Disposable polymer ligation clips are compatible with the reusable, non-sterile Clip appliers which are made by Kangji. The clips are placed around the tissue and closed with the grip of a clip applier. These clips are made of medical grade acetyl homopolymer and supplied in different sizes and packaged in different clip quantity. {5} K250643 Page 2 of 7 Polymer Clips are housed in a colored cartridge which is consist of base and platen. The clips shall be placed and fired in clip applicer one by one. Polymer Clips are available in four sizes: M, ML, L, XL. Multiple polymer clips (model: KJ-JZJDML, KJ-JZJDL, KJ-JZJDXL) are available in 3 sizes (ML, L, XL), housed in a colored cartridge and has 3 clips which can be placed in clip applicer in one time and fired constantly totally 3 times. Multiple polymer clips (except model: KJ-JZJDML, KJ-JZJDL, KJ-JZJDXL) are available in ML and L size, housed in a striped cartridge. The cartridge which contains 5-15 clips will be placed in clip applicer. The clips will be delivered individually advance after each firing. ## V Indications for use The Disposable polymer ligation clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure. ## VI Substantial Equivalence The proposed Disposable polymer ligation clips are substantially equivalent to the predicate devices: | Model of proposed device | Predicate device | Manufacturer | 510(k) No. | | --- | --- | --- | --- | | Polymer Clips | Hem-O-Lok® ligation clips | Weck Closure Systems | K993157 | | Multiple Polymer Clips | Weck Auto Endo5® 5mm Automatic Endoscopic 35cm Applier (Hem-o-lok® ligating clips) | Teleflex Medical Incorporated | K230480 | Table 1 compares the model-Polymer clips to the Hem-O-Lok® ligation clips(Weck Closure Systems) with respect to intended use, and technological characteristics, providing detailed information regarding the basis for the determination of substantial equivalence. {6} K250643 Page 3 of 7 Table 1 Comparison of Polymer clips to the predicate device | Device feature | Predicate Device K993157 (Hem-O-Lok®) | Subject Device (Polymer Clips, Multiple Polymer Clips (model: KJ-JZJDML, KJ-JZJDL, KJ-JZJDXL)) | Comment | | --- | --- | --- | --- | | Indications for use | Hem-O-Lok® ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure. | It is intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure. | Equivalent | | Product code | FZP | FZP | Equivalent | | Reuse durability | Single use | Single use | Equivalent | | Sterility | EtO at 10-6SAL | EtO at 10-6SAL | Equivalent | | Size | M, ML, L, XL | M, ML, L, XL | Equivalent | | Composition | Polyacetal | Acetyl homopolymer (which is the same as Polyacetal) | Equivalent | | Processing | Injection molding | Injection molding | Equivalent | # Conclusion: The subject and predicate device have the same intended use. The technological characteristics of this product are believed to be substantially equivalent as those for the predicate device. This device and its predicate are substantially equivalent in material and manufacturing methods. The subject and predicate device designs are nearly identical. There are no new questions raised regarding to safety and effectiveness. {7} K250643 Page 4 of 7 Table 2 compares the model-Multiple polymer clips to the Weck Auto Endo5® 5mm Automatic Endoscopic 35cm Applier (Hem-o-lok® ligating clips)(Teleflex Medical Incorporated) with respect to intended use, and technological characteristics, providing detailed information regarding the basis for the determination of substantial equivalence. Table 2 Comparison of Multiple polymer clips to the predicate device | Device feature | Predicate Device K230480 (Hem-o-lok® ligating clips) | Subject Device (Multiple Polymer Clips) | Comment | | --- | --- | --- | --- | | Indications for use | Hem-O-Lok® ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure. | It is intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure. | Equivalent | | Product code | FZP | FZP | Equivalent | | Reuse durability | Single use | Single use | Equivalent | | Sterility | EtO at 10-6SAL | EtO at 10-6SAL | Equivalent | | Size | ML | ML, L | Equivalent | | Amount of clip in a cartridge | 15 | 5 to 15 | Difference comment 1 | | Composition | Polymer | Acetyl homopolymer (which is the same as Polyacetal) | Equivalent | | Processing | Injection molding | Injection molding | Equivalent | | Packaging | The clips are housed in a cartridge and preloaded | The clips are housed in a cartridge and packed | Difference comment 2 | {8} K250643 Page 5 of 7 | Device feature | Predicate Device K230480 (Hem-o-lok® ligating clips) | Subject Device (Multiple Polymer Clips) | Comment | | --- | --- | --- | --- | | | inside the clip applier | in a dialysis paper with Tyvek coted | | Comment 1 Subject device has a wider range of amount of clip in a cartridge than predicate device, which can provide more choice for surgical clips quantity demand. Surgeons will make decision on how many clips are needed. This difference does not affect the basic design principle, usage, effectiveness and safety of the subject device. There are no new questions raised regarding to safety and effectiveness. Comment 2 The packing way of two devices is different. they are both housed in a cartridge, while the predicate device is housed in a cartridge and preloaded in the clip applier and the subject device comes separately from the clip applier in its own sterile packaging. The subject will be placed in a clip applier after tearing off the primary package. And then the use of clips relies on the clip applier which is same as the predicate device. The package was addressed by ISO 11607-1, ISO 11607-2 and ASTM F1980 to maintain the sterility of product. This difference does not affect the basic design principle, usage, effectiveness and safety of the subject device. There are no new questions raised regarding to safety and effectiveness. Conclusion: The Multiple Polymer Clips and Hem-o-lok® ligating clips in predicate device have the same intended use. The technological characteristics of this product are believed to be substantially equivalent as those for the predicate device. This device and its predicate are substantially equivalent in material and manufacturing methods. The subject and predicate device designs are nearly identical. The differences in technological characteristics between the subject and predicate devices (i.e., different size, amount, packing way) do not raise new questions of safety and effectiveness. {9} K250643 Page 6 of 7 ## Summary of comparison with the predicate devices: The subject and predicate devices have the same intended use and application. The subject and predicate device designs are nearly identical. They are single-use devices. The differences in technological characteristics between the subject and predicate devices (i.e., different size, packing way) do not raise new questions of safety and effectiveness. ## VII Summary of non-clinical testing ### Performance testing The Disposable polymer ligation clips has been evaluated through performance studies and bench testing, as attached in Appendix C. Testing encompassed appearance, dimension, assembly firmness, matching performance, locking performance, toughness, clamping performance, tensile performance, pressure endurance performance. All the test results were assessed and deemed substantially equivalent to the predicate device. ### Biocompatibility testing Biocompatibility of the Disposable polymer ligation clips was evaluated in accordance with ISO 10993-1:2018 for the body contact category. Table 3 List of biocompatibility test | Item | Biological endpoints | Standard | Test result | | --- | --- | --- | --- | | 1 | Cytotoxicity | ISO 10993-5:2009 | Pass | | 2 | Sensitization | ISO 10993-10:2021 | Pass | | 3 | Irritation | ISO 10993-23:2021 | Pass | | 4 | Material mediated pyrogenicity | ISO 10993-11:2017 | Pass | | 5 | Implantation effects | ISO 10993-6:2016 | Pass | | 6 | Chemical characterization | 10993-18:2020/Amd 1:2022 | Pass | | 7 | Toxicological risk assessment | ISO 10993-17:2023 | Pass | ### Sterilization and shelf life testing - EO sterilization validation per EN ISO 11135-1, 11737-1, 11737-2 - Transportation test per ISTA 2A:2011 and ISTA 3A:2018 - Packaging validation per ISO 11607-1/-2 - The 3 years shelf life of the device is determined based on stability study which includes ageing test. - Bacterial Endotoxin Testing per USP-NF:2023 <85> {10} K250643 Page 7 of 7 ## VIII Summary of clinical testing None. ## IX Conclusion The Disposable polymer ligation clips are substantially equivalent to its predicate devices. The differences between the predicate and subject device do not raise any new or different questions of safety or effectiveness. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...