AtriClip PRO, PRO2, and PRO V LAA Exclusion System devices with preloaded Gillinov-Cosgrove Clip
K163261 · AtriCure, Inc. · FZP · May 19, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163261
Device Name
AtriClip PRO, PRO2, and PRO V LAA Exclusion System devices with preloaded Gillinov-Cosgrove Clip
Applicant
AtriCure, Inc.
Product Code
FZP · General, Plastic Surgery
Decision Date
May 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The AtriClip LAA Exclusion System is indicated for the occlusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
Device Story
AtriClip LAA Exclusion System is a single-use, sterile, self-closing implantable clip preloaded on a disposable applier. Device used by surgeons during cardiac procedures to occlude the left atrial appendage (LAA). Surgeon operates applier to place clip at LAA base; clip applies uniform pressure to ensure secure, permanent occlusion. Available in 35mm, 40mm, 45mm, and 50mm lengths to accommodate varying LAA sizes. Provides mechanical exclusion of LAA to reduce risk of thrombus formation. Used in OR settings under direct visualization (with or without endoscopic assistance).
Clinical Evidence
Evidence includes multiple IDE clinical studies (Stroke G110145, Hybrid DEEP G090279, Staged DEEP G120056) and published literature. Total of 129 patients implanted via video-assisted thoracoscopic procedures. Primary endpoints: LAA occlusion post-deployment (99% success, 128/129) and LAA exclusion at 3-12 months post-op (96% success, 56/58). Zero device-related serious adverse events reported.
Technological Characteristics
Implantable clip and disposable applier (handle, shaft, suture anchors, end-effector). Clip lengths: 35-50 mm. Mechanical self-closing mechanism providing uniform pressure. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for occlusion of the left atrial appendage in patients undergoing cardiac surgical procedures requiring direct visualization of the heart.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip (K093679)
Reference Devices
AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip (K122276)
AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip (K131107)
AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip (K142120)
AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip (K150996)
AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove PRO-V Clip (K153500)
AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip (K160454)
Related Devices
K153500 — AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove PRO V Clip · AtriCure, Inc. · Jan 27, 2016
K142120 — ATRICLIP LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV-COSGROVE CLIP · AtriCure, Inc. · Aug 28, 2014
K173031 — AtriClip LAA Exclusion System with Preloaded PRO·V Clip · AtriCure, Inc. · Oct 25, 2017
K150996 — AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip · AtriCure, Inc. · Jun 17, 2015
K093679 — ATRICLIP LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV-COSGROVE CLIP, GILLINOV-COSGROVE SECTION GUIDE MODEL VARIOUS LAA0 · AtriCure, Inc. · Jun 10, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2017
AtriCure Inc. Jonathan McElwee Regulatory Affairs Manger 7555 Innovation Way Mason, Ohio 45040
Re: K163261
Trade/Device Name: AtriClip Pro, Pro2, and Pro V LAA Exclusion Systems With Preloaded Gillinov-Cosgrove Clip Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: FZP Dated: April 19, 2017 Received: April 20, 2017
Dear Mr. McElwee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
M.A. Hillemann
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163261
### Device Name
AtriClip LAA Exclusion System with Gillinov-Cosgrove Clip
Indications for Use (Describe)
The AtriClip LAA Exclusion System is indicated for the occlusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|--|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |
| <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AtriCure. The logo is in two colors, blue and orange. The first part of the name, "Atri," is in blue, while the second part, "Cure," is in orange. There is a blue dot in the middle of the "Cure" part of the name. There is a registered trademark symbol next to the word "Cure".
# 510(k) Summary
#### l. Submitter
| Manufacturer: | AtriCure, Inc.<br>7555 Innovation Way<br>Mason, OH 45040<br>P: 513-755-4100<br>F: 513-755-4108 |
|----------------------|------------------------------------------------------------------------------------------------|
| Contact Person: | Jonathan McElwee, RAC<br>Regulatory Affairs Manager |
| Date Submitted: | 11/18/2016 |
| II. Device | |
| Name of Devices: | AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove® Clip |
| Common Name: | Implantable Clip and Clip Applier |
| Classification Name: | Implantable Clip and Clip Applier (21 CFR 878.4300) |
| Regulatory Class: | Class II |
| Product Code: | FZP |
#### III. Predicate Device
Predicate Device:
K093679 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip
The following reference devices were also used in this submission:
- K122276 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip .
- K131107 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ●
- K142120 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ●
- K150996 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip
- AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove PRO-V Clip ● K153500
- K160454 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ●
{4}------------------------------------------------
### IV. Device Description
The AtriClip LAA Exclusion System consists of a single use, sterile, self-closing, implantable Clip preloaded on a Single Use Clip Applier along with a selection guide. When closed, the Clip applies uniform pressure over the length of the Clip to ensure consistent, reproducible, and secure occlusion of the left atrial appendage (LAA). The clip is then deployed and is left as a permanent implant. The Clip is available in the following lengths to accommodate different sizes of LAA: 35 mm, 40 mm, 45 mm, and 50 mm.
The Clip Appliers are disposable devices with a handle, shaft, suture anchors, and deployment end-effector containing the Clip.
#### V. Indications For Use
The AtriClip LAA Exclusion System is indicated for the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
### VI. Comparison Of Technological Characteristics With The Predicate Device
The AtriClip LAA Exclusion System devices have the following similarities to the technological characteristics of the previously cleared devices.
- The devices have the same intended use "occlusion of the heart's left atrial appendage" ●
- . Same operating principle
- Same fundamental scientific technology
- Same device designs .
- Same materials used
- The results of the verification and validation testing: .
- Demonstrated equivalency in performance o
- O Did not raise any new issues of safety
#### VII. Performance Data
Non-clinical Bench Testing Reliability Testing Usability Testing on a Cadaver Model
## Clinical Studies
AtriCure has collected scientific based evidence from multiple sponsored IDE clinical studies on the use of the AtriClip LAA Exclusion System during video assisted thoracoscopic procedures of the left atrial appendage (LAA). In the IDE Stroke clinical study (G110145) patients were implanted with the AtriClip device via video assisted thoracoscopic procedures without any adverse events. At three months followup, these patients had 100% LAA exclusion.
In addition, AtriCure also conducted two IDE studies, Hybrid DEEP (G090279) and Staged DEEP (G120056), where the AtriClip devices were implanted via using video without any adverse events.
{5}------------------------------------------------
Furthermore, AtriCure performed a published literature review evaluating the use of AtriClip LAA Exclusion System devices, which provided additional evidence the device can be placed safely at the base of the appendage via a video assisted thoracoscopic procedure of the LAA.
In totality of the clinical evidence based on the multiple IDE clinical studies and published clinical literature, 129 total patients were safely implanted during video assisted thoracoscopic procedures of the left atrial appendage.
A review of the IDE clinical studies and published literature evaluated the following endpoints when the Clip was deployed via a video assisted thoracoscopic procedures of the LAA: 1.) LAA occlusion directly following Clip deployment and 2.) LAA exclusion post-operatively. The results demonstrated: 1.) 99% (128/129) of the patients had LAA occlusion directly following Clip deployment and 2.) 96% (56/58) of patients had LAA exclusion evaluated at 3-12months post-operatively.
The IDE EXCLUDE clinical study, included in the original AtriClip 510(k) premarket notification - K093679, demonstrated effectiveness of 95.7% LAA occlusion directly following Clip deployment and 98.4% LAA exclusion 3 months post-operatively.
There were Zero (0) AtriClip device related serious adverse events reported in the clinical studies and published literature.
## Summary
Based on the clinical performance as documented in the clinical studies and published literature, the AtriClip LAA Exclusion System was found to have a safety and effectiveness profile similar to the predicate device. Substantial equivalence was based in part on this information.
#### VIII. Conclusions
The AtriCure sponsored clinical studies, published literature, and bench testing data included in this submission provide evidence demonstrating the AtriClip LAA Exclusion System, as labeled per the Indication for Use, provide A.) clinically significant results and at the same time B.) do not present an unreasonable risk of illness or injury associated with the use of the device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.