ATRICLIP LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV-COSGROVE CLIP, GILLINOV-COSGROVE SECTION GUIDE MODEL VARIOUS LAA0

K093679 · AtriCure, Inc. · FZP · Jun 10, 2010 · General, Plastic Surgery

Device Facts

Record IDK093679
Device NameATRICLIP LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV-COSGROVE CLIP, GILLINOV-COSGROVE SECTION GUIDE MODEL VARIOUS LAA0
ApplicantAtriCure, Inc.
Product CodeFZP · General, Plastic Surgery
Decision DateJun 10, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4300
Device ClassClass 2
AttributesTherapeutic

Intended Use

The AtriClip LAA Exclusion System is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.

Device Story

AtriClip LAA Exclusion System is a sterile, single-use, implantable device designed for permanent occlusion of the left atrial appendage (LAA). The system comprises a self-closing clip preloaded on a disposable applier and a selection guide. The clip features a frame assembly of two springs connecting opposing tubes covered with pressure pads and a knit fabric. During open cardiac surgery, a physician uses the applier to position the clip over the LAA; pulling a thumb lever opens the clip, while releasing it allows the clip to close, applying uniform pressure to ensure secure occlusion. The device is used in an OR setting by cardiac surgeons. By providing consistent, reproducible LAA closure, the device aims to reduce the risk of thromboembolic events associated with the LAA. Clinical performance is verified intra-operatively and via follow-up imaging.

Clinical Evidence

Clinical study of 70 subjects evaluated safety and efficacy. Primary safety endpoint event rate was 0%. Efficacy demonstrated via LAA exclusion rates: 95.1% composite (intra-operative and 3-month) and 98% at 3-month follow-up. No events linked to device or procedure per independent physician adjudication. Non-clinical testing included acute and chronic animal models (7 to 90 days) confirming complete appendage exclusion with no adverse events.

Technological Characteristics

Implantable clip with frame assembly (two springs, opposing tubes, pressure pads, knit fabric). Available in 35mm, 40mm, 45mm, and 50mm lengths. Biocompatibility tested per ISO 10993-1. Mechanical actuation via thumb lever on disposable applier. Sterile, single-use.

Indications for Use

Indicated for patients undergoing open cardiac surgical procedures requiring occlusion of the left atrial appendage (LAA) under direct visualization.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the word "AtriCure" in a stylized font. The letters "A," "t," "r," and "i" are in a bold, sans-serif font, while the "Cure" portion of the word is outlined. There is a solid black circle inside the "C" of "Cure." The "e" in "Cure" has a small trademark symbol next to it. KC093679 ﯿﻨﮧ ﮨ ## 510(k) Summary # JUN 1 0 2010 #### General Information | Classification | Class 2 | |---------------------|--------------------------------------------------------------------------------------------------------| | Trade name | AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip | | Common name | Implantable Clip | | Classification Name | Clip, Implantable (21 CFR 878.4300, Product Code FZP) | | Manufacturer | AtriCure, Inc.<br>6217 Centre Park Dr.<br>West Chester, OH 45069<br>P: 513-755-4100<br>F: 513-755-4108 | | Contact | James Lucky, RAC<br>Vice President Quality Assurance and Regulatory Affairs | | Date of Submission | November 27, 2009 | #### Intended Use The AtriClip LAA Exclusion System is indicated for the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures. #### Predicate Devices The predicate devices for the AtriClip LAA Exclusion System are the Weck Hem-O-Lok® Ligating Clip and Clip Applier (K030311), Medtronic VNUS U-Clip and Applier (K031623), Tyco AutoSuture TA and GIA Staplers (K032696), Power Medical SurgASSIST® Straight Linear 4 Row No Knife DLUs with Reloads (K040398), Demetech Braided Nonabsorbable Polyester Suture and needle driver (K023030) and Gore SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material (K043056). #### Device Description The AtriClip LAA Exclusion System consists of a single use, sterile, self-closing, implantable Clip preloaded on a Single Use Clip Applier, and a Selection Guide to aid in appropriate Clip size selection. The frame assembly of the implantable Clip consists of two springs connecting two opposing tubes which are covered with pressure pads. This assembly is covered with a knit fabric. When closed, the Clip applies uniform pressure over the length of the Clip to ensure consistent, reproducible, and secure occlusion of the left atrial appendage (LAA). {1}------------------------------------------------ The Clip is available in the following lengths to accommodate different sizes of LAA: 35 mm, 40 mm, 45 mm, and 50 mm. The Clip Applier is a disposable device with a handle, shaft, and an end effector which contains the Clip. Both sides of the Clip are free to move within the end effector when activated. Pulling the thumb lever proximally opens the Clip. The bias of the Clip allows both the clip and the lever to return to the closed position when the lever is released. #### Materials All materials in the AtriClip LAA Exclusion System are suitable for their intended use and have been used in previously cleared products. Testing was conducted in accordance with ISO 10993-1 to ensure appropriate biocompatibility of all materials. #### Predicate Comparison The technological characteristics of the device are the same as the predicate devices. The same materials are used in the AtriClip and predicate devices. The sizes of the devices are equivalent and are provided in similar ranges. The method of LAA closure, permanent approximation of the sides of the LAA, is the same for the AtriClip and the predicates. All of the devices may be repositioned on the LAA prior to deployment and once deployed exert similar pressures on the tissue. #### Non-Clinical Testing The AtriClip was tested on in both acute and chronic models ranging from 7 days to 90 days Testing presented in the 510(k) document demonstrated complete appendage duration. exclusion both acutely and long-term in all animals with no adverse events attributed to the Clip device. Post-implant verification of efficacy confirmed there was no communication between the LAA and left atrium in all animals. #### Clinical Testing The AtriClip has been evaluated for safety and efficacy in a controlled clinical study of seventy subjects. The results of the study demonstrate the excellent performance of the product in its ability to exclude the LAA in a safe and efficient manner. Efficacy has been demonstrated at a very high rate intra-procedurally and this rate has been shown to improve as further healing was achieved. The observed exclusion rate of 95.1% composite (both intra-operative and 3 month) and 98% at 3 month follow-up support that this procedure can achieve excellent efficacy results. The safety of the device and procedure has been well established through a careful scrutiny of all events reported in the study. Independent physician adjudication has demonstrated that there were no events that were in any way linked to the device or the procedure. The primary safety endpoint event rate in this trial is 0%. These results confirm the safety of the product and the procedure. The results further support the technical comparability of the product to other currently marketed options for LAA exclusion and therefore provides the clinical support to demonstrate a substantial equivalence of the AtriClip to the identified predicate devices. {2}------------------------------------------------ # AtriCure® #### Summary of Substantial Equivalence As demonstrated by the predicate comparison of technological characteristics, data provided in the non-clinical studies and clinical testing, the AtriClip LAA Exclusion System is equivalent to the predicate products. The data presented in K093679 demonstrate that the device is as safe, as effective and performs as well as or better than the predicate device. . {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like bird with its wings spread, positioned to the right of a circular inscription. The inscription reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement around the bird. The logo is presented in black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 JUN 1 0 2010 AtriCure, Inc. c/o Mr. James Lucky Vice President of Quality Assurance and Regulatory Affairs 6217 Centre Park Drive West Chester, OH 45069 Re: K093679 AtriClip™ LAA Exclusion System w/Pre-loaded Gillnov-Cosgrove™ Clip Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: II Product Code: FZP Dated: June 1, 2010 Received: June 2, 2010 Dear Mr. Lucky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. James Lucky Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # STATEMENT: INDICATION FOR USE K093679 510 (k) Number: Device Name: AtriClip LAA Exclusion System ### Indications for Use: The AtriClip LAA Exclusion System is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures. Prescription Use X Over the Counter Use (Part 21 CFR 801 Subpart D) and/or (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) W.W.A. ardiovascular Devices 510(k) Number K093679
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...