CARDICA C-PORT FLEX-A DISTAL ANASTOMOSIS SYSTEM, MODEL FG-000150
K073123 · Cardica, Inc. · FZP · Dec 6, 2007 · General, Plastic Surgery
Device Facts
Record ID
K073123
Device Name
CARDICA C-PORT FLEX-A DISTAL ANASTOMOSIS SYSTEM, MODEL FG-000150
Applicant
Cardica, Inc.
Product Code
FZP · General, Plastic Surgery
Decision Date
Dec 6, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Cardica® C-Port® Flex-A™ Distal Anastomosis System is intended for the creation of anastomoses in blood vessels and grafts, including use in coronary artery bypass grafting procedures.
Device Story
Sterile, single-use surgical instrument; creates end-to-side anastomosis between conduit and small vessel. System includes conduit loading accessories and anastomosis device. Surgeon loads conduit onto device; positions device against target vessel; pushes actuation button to deploy stainless steel clips. Used in clinical settings (e.g., OR) by surgeons. Facilitates reliable patent anastomosis; benefits patient by enabling precise vessel connection during bypass procedures.
Clinical Evidence
Bench testing only; no clinical data presented. In vitro and in vivo testing performed to ensure safety and effectiveness.
Technological Characteristics
Sterile, single-use mechanical device. Materials include stainless steel clips. Manual actuation mechanism. No software or electronic components.
Indications for Use
Indicated for patients requiring creation of anastomoses in blood vessels and grafts, including coronary artery bypass grafting procedures.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Cardica® C-Port® Flex-ATM Distal Anastomosis System (K070548)
Related Devices
K090872 — MODIFICATION TO CARDICA C-PORT XA DISTAL ANASTOMOSIS SYSTEM · Cardica, Inc. · Apr 21, 2009
K040832 — C-PORT ANASTOMOSIS SYSTEM (DEPLOYMENT SYSTEM, CLIP, ACCESSORIES), MODEL FG-000050 · Cardica, Inc. · Nov 10, 2005
K063644 — CARDICA C-PORT XA DISTAL ANASTOMOSIS SYSTEM · Cardica, Inc. · Jan 11, 2007
K073304 — CARDICA C-PORT X-CHANGE ANASTOMOSIS SYSTEM AND HANDLE AND XA X-CHANGE SUBASSEMBLIES · Cardica, Inc. · Dec 20, 2007
K053524 — C-PORT DISTAL ANASTOMOSIS SYSTEM · Cardica, Inc. · Nov 16, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
## DEC 0 6 2007
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| 510(k) Number | | |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | November 2, 2007 | |
| Applicant<br>Information | Cardica, Inc.<br>900 Saginaw<br>Redwood City, California 94063<br>Main: 650-364-9975<br>Fax: 650-364-3134 | |
| Contact Person | Iskra Mrakovic<br>Office: 650-331-7153<br>Fax: 650-364-3134<br>e-mail: mrakovic@cardica.com | |
| Establishment<br>Registration Number | 3004114958 | |
| Device Information | Classification Name:<br>Regulation Number:<br>Trade Name:<br>Common Name: | Clip, Implantable<br>21 CFR §878.4300<br>Cardica® C-Port® Flex-ATM Distal<br>Anastomosis System<br>Cardiovascular Surgical Instruments |
| Predicate Device(s) | Cardica® C-Port® Flex-ATM Distal Anastomosis System<br>(#K070548) | |
| Device Description | The Cardica® C-Port® Flex-ATM Distal Anastomosis System is a<br>sterile, single use device for creation of a reliably patent end-<br>to-side anastomosis between a conduit and a small vessel.<br>The product consists of accessories to assist in the conduit<br>loading and a device that completes the anastomosis with<br>stainless steel clips. Once the conduit has been loaded onto<br>the device and the device positioned against the target vessel,<br>the anastomosis is created by pushing the actuation button. | |
| Intended Use | The Cardica® C-Port® Flex-AT™ Distal Anastomosis System is<br>intended for the creation of anastomoses in blood vessels and<br>grafts, including use in coronary artery bypass grafting<br>procedures. | |
| Comparison to<br>Predicate Device | The Cardica® C-Port® Flex-AT™ Distal Anastomosis System is<br>substantially equivalent to the Cardica® C-Port® Flex-AT™ Distal<br>Anastomosis System (#K070548, 21 CFR §878.4300). The<br>subject device is substantially equivalent to the predicate<br>device with regard to the intended use, device characteristics,<br>method of use, materials, labeling, sterilization method and<br>biocompatibility. | |
| Device Testing<br>Results and<br>Conclusion | All necessary <i>in vitro</i> and <i>in vivo</i> testing has been performed<br>on the C-Port® Flex-AT™ Distal Anastomosis System and its<br>packaging to ensure substantial equivalence to the predicate<br>device, and to ensure the safety and effectiveness of the<br>device. | |
| Substantial<br>Equivalence<br>Summary | Cardica® C-Port® Flex-AT™ Distal Anastomosis System has the<br>same indications for use and the same technological<br>characteristics as the predicate device (#K070548). This<br>premarket notification has described the characteristics of the<br>modified device in sufficient detail to assure substantial<br>equivalence. For the few characteristics that may not be<br>precise enough to ensure equivalence, performance data was<br>collected, and this data demonstrates substantial equivalence.<br>In keeping with the format of a Special 510(k) for Device<br>Modification, performance data were not included in the<br>submission, but the declarations provide certification that the<br>data demonstrate equivalence. | |
| Conclusions | This Special 510(k) for Device Modification submission has<br>demonstrated Substantial Equivalence as defined and<br>understood in the Federal Food Drug and Cosmetic Act and<br>various guidance documents issued by the Center for Devices<br>and Radiological Health. | |
{1}------------------------------------------------
:
Cardica® and C-Port® are registered trademarks of Cardica, Inc.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, symbolizing health and human services. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the logo.
DEC 0 6 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Cardica, Inc. c/o Ms. Iskra Mrakovic Manager, Regulatory Affairs 900 Saginaw Dr. Redwood City, CA 94063
Re: K073123
Cardica® C-Port® Flex-A™ Distal Anastomosis System Regulation Number: 21 CFR 878.4300 Regulation Name: Clip, Implantable Regulatory Class: Class II (two) Product Code: FZP Dated: November 02, 2007 Received: November 06, 2007
Dear Ms. Mrakovic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Iskra Mrakovic
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
una R. V. Auner
Image /page/3/Picture/5 description: The image contains a handwritten symbol or character on the left side, which appears to be a combination of curved and looped lines. To the right of this symbol are the capital letters 'B' and 'D', stacked vertically. The letters are in a simple, sans-serif font and are positioned closely together.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
| 510(k) Number:<br>(if known) | K073123 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Cardica® C-Port® Flex-A™ Distal Anastomosis System |
| Indications for Use: | The Cardica® C-Port® Flex-A™ Distal Anastomosis System is intended for<br>the creation of anastomoses in blood vessels and grafts, including use in |
coronary artery bypass grafting procedures.
Prescription Use × (Part 21 CFR§801.109)
> : and the comments of the country
Over-The-Counter Use OR (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Suane R. V. Aunes
(Division Sign-Off) Division of Caralovascular Devices
Indications for Use Statement
510(k) Number_k0 7317 3
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.