K041751 · Niti Medical Technologies , Ltd. · FZP · Sep 16, 2004 · General, Plastic Surgery
Device Facts
Record ID
K041751
Device Name
COMPRESSION ANASTOMOSIS CLIP (CAC)
Applicant
Niti Medical Technologies , Ltd.
Product Code
FZP · General, Plastic Surgery
Decision Date
Sep 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The NiTi CAC (Compression Anastomosis Clip) is intended to be used to facilitate side-to-side anastomosis of the alimentary tract yielding an inverted serosa-to-serosa anastomosis. Once wound strength is sufficient to maintain the anastomosis, the NiTi CAC is passed from the body. The NiTi CAC is not applicable through Trocars in laparoscopic procedures.
Device Story
Sterile, single-use device for side-to-side compression anastomosis of alimentary tract; consists of double-ring clip and applier. Surgeon inserts clip into two cut tissue segments; clip applies compression to tissue. Over 7-10 days, compression induces necrosis of tissue under rings; device and necrosed tissue detach and are naturally expelled via stool. Facilitates inverted serosa-to-serosa anastomosis. Used in clinical settings; not for laparoscopic use via trocars.
Clinical Evidence
Animal studies performed. No human clinical data provided.
Technological Characteristics
Sterile, single-use implantable clip. Comprised of double-ring clip and applier. Mechanical compression principle. No performance standards established under Section 514.
Indications for Use
Indicated for patients requiring side-to-side anastomosis of the alimentary tract. Not for use in laparoscopic procedures via trocars.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Valtrac Biofregmentable Anastomosis Ring (K881484, K931056)
Related Devices
K033324 — COMPRESSION ANASTOMOSIS CLIP (CAC) · Niti Medical Technologies , Ltd. · May 10, 2004
K043115 — LAPAROSCOPIC COMPRESSION ANASTOMOSIS CLIP (LAPCAC) · Niti Medical Technologies , Ltd. · Dec 1, 2004
K062008 — ENDO CAR COMPRESSION ANASTOMOSIS RING (CAR) · Niti Medical Technologies , Ltd. · Aug 9, 2006
K093661 — COLONRING · Niti Surgical Solutions, Ltd. · Feb 10, 2011
K050356 — COMPRESSION ANASTOMOSIS RING (CAR) · Niti Medical Technologies , Ltd. · May 5, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(K) SUMMARY
K041757
彩神
1/3
# Compression Anastomosis Clip (CAC)
## 510(k) Number K_______________________________________________________________________________________________________________________________________________________________
### Applicant's Name:
NiTi Medical Technologies Ltd. 1 Hatzoran st., P.O.Box 8634 Netanya 42506, Israel Tel.: 972-9-865-0610 Fax: 972-9-835-0127
#### Contact Person:
Orly Maor 1 Hatzoran st., P.O.Box 8634 Netanya 42506, Israel Tel.: 972-9-865-0610 Fax: 972-9-835-0127 orly@nitimed.com
And / or
Jonathan S. Kahan, Esq. Hogan & Hartson, L.L.P. Columbia Square 555 Thirteenth Street, NW Washington, DC 20004-1109 Tel: (202) 637-5794 Fax: (202) 637-5910
#### Date Prepared:
June 2004
#### Trade Name:
Compression Anastomosis Clip (CAC)
## Classification Name:
IMPLANTABLE CLIP
{1}------------------------------------------------
K0417K1
2/3
#### Classification:
The FDA has classified implantable clip as class II device (product code FZP, Regulation No. 21 C.F.R. § 878.4300) and they are reviewed by the Division of General and Restorative Devices.
#### Predicate Device:
- 트 Compression Anastomosis Clip (CAC) (NiTi Medical Technologies Ltd.) cleared under K033324.
- Proximate™ Linear Cutter and Stapler (Ethicon Endo-Surgery, Inc. USA) cleared under K843603 and K020779.
- 피 GIA stapler (and Endo GIA) (United States Surgical Cooperation, UAS) cleared under K843603.
- 트 Valtrac Biofregmentable Anastomosis Ring (Devis + Geck) cleared under K881484 and K931056
#### Performance Standards:
No performance standards have been established for such device under Section 514 of the Federal Food, Drug, and Cosmetic Act.
#### Intended Use:
The NiTi CAC (Compression Anastomosis Clip) is intended to be used to facilitate side-to-side anastomosis of the alimentary tract yielding an inverted serosa-to-serosa anastomosis. Once wound strength is sufficient to maintain the anastomosis, the NiTi CAC is passed from the body. The NiTi CAC is not applicable through Trocars in laparoscopic procedures.
#### Device Description:
The Compression Anastomosis Clip (CAC) device is a sterile single use device. The CAC provides a simple method for the creation of side-to-side compression anastomosis of the alimentary tract. The CAC device is comprised of two components:
{2}------------------------------------------------
KO41751
3/3
- ▶ Clip double ring clip that is inserted into the two cut segments of the tissue to be anastomosed and performs the required compression on the tissue.
- ▶ Applier with which the Clip is introduced into the treated area.
After a period of 7-10 days, a compression-induced necrosis of the tissue sides underneath the rings occurs and the whole device, together with the necrosed tissue that was compressed by the rings, detaches and is naturally expelled with the stool.
#### Substantial Equivalence:
Based on validations and performance testing results, including animal studies, NiTi Medical Technologies Ltd. believes that the Compression Anastomosis Clip (CAC) is substantially equivalent to its predicate device cited above without raising new safety and/or effectiveness issues.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 6 2004
Ms. Orly Maor Regulatory Manger NiTi Medical Technologies Ltd. 1 Hatzoran St., P.O. Box 8634 Netanya 42506, Israel
Re: K041751
Trade/Device Name: Compression Anastomosis Clip (CAC) Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: II Product Code: FZP Dated: June 23, 2004 Received: June 29, 2004
Dear Ms. Maor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Orly Maor
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K041757.
#### INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: Compression Anastomosis Clip (CAC)
Indications for Use:
The NiTi CAC is intended to be used to facilitate side-toside anastomosis of the alimentary tract yielding an inverted serosa-to-serosa anastomosis. Once wound strength is sufficient to maintain the anastomosis, the NiTi CAC is passed from the body. The NiTi CAC is not applicable through Trocars in laparoscopic procedures.
Prescription Use イ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number_ K04/75/
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.