K230135 · Copioumed International, Inc. · FYC · Oct 4, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230135
Device Name
Copioumed Chemotherapy Isolation Gown
Applicant
Copioumed International, Inc.
Product Code
FYC · General, Plastic Surgery
Decision Date
Oct 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Copioumed Chemotherapy Isolation Gown is intended to be worn by healthcare personnel to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate. Copioumed Chemotherapy Isolation Gown is non-sterile and for single use only. Copioumed Chemotherapy Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). Copioumed Chemotherapy Isolation Gown is a disposable device intended to protect health care personal from exposure to chemotherapy drugs during preparation, handling, and administration.
Device Story
Disposable isolation gown; provides AAMI Level 3 barrier protection. Constructed from polypropylene laminated with polyethylene; features knit cuffs and three neck closure styles (tie, Velcro, sticker). Used by healthcare personnel in clinical settings during chemotherapy drug preparation, handling, and administration. Protects against transfer of microorganisms, body fluids, and particulates; provides resistance to chemotherapy drug permeation (tested per ASTM F739-20). Device is non-sterile and single-use. Clinical benefit: reduces risk of healthcare worker exposure to hazardous chemotherapy agents and cross-contamination.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing, including AAMI PB70:2012 liquid barrier performance, ASTM F739-20 chemotherapy drug permeation resistance, tensile/tear strength (ASTM D5034, D5587), flammability (16 CFR 1610), and biocompatibility (ISO 10993-5, -10, -23).
Technological Characteristics
Polypropylene laminated with polyethylene; knit cuffs. AAMI Level 3 barrier protection. Tested per ASTM F739-20 for chemotherapy permeation. Tensile strength (ASTM D5034), tear strength (ASTM D5587), hydrostatic pressure (AATCC 127), impact penetration (AATCC 42). Flammability Class 1 (16 CFR 1610). Biocompatible per ISO 10993. Non-sterile, single-use.
Indications for Use
Indicated for healthcare personnel to protect patients and themselves from microorganisms, body fluids, and particulates. Indicated for protection against chemotherapy drug exposure during preparation, handling, and administration. For single-use, non-sterile, OTC use in healthcare facilities.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
CoverU Disposable Gown with Tape - Chemo Gown (K210414)
Related Devices
K210414 — CoverU Disposable Gown with Tape AAMI Level 4 Isolation Gown CoverU Disposable Gown with Tape - Chemo Gown · Medtecs (Taiwan) Corp. · Oct 21, 2022
K214108 — Medline Orbis Chemo Surgical Gown · Medline Industries, LP · May 6, 2022
K193327 — Cardinal Health ChemoPlus Full Coverage Gown, Closed Back, Cardinal Health ChemoPlus Full Coverage Gown, Open Back, Cardinal Health ChemoPlus Full Coverage Sleeve · Cardinal Health200, LLC · Mar 20, 2020
K223304 — Disposable Isolation Gowns (S,M,L,XL,XXL,XXXL) · Xiamen Probtain Medical Technology Co., Ltd. · Feb 8, 2023
K212504 — Isolation Gown AAMI Level 3 · Kenpax International Limited · Nov 7, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
October 4, 2023
Copioumed International Inc. % Ming-Yie Jan Principle Consultant RusCert Technology Co., Ltd. 8F., No. 187, Lequn 2nd Rd., Zhongshan Dist. Taipei City, 10462 Taiwan
Re: K230135
Trade/Device Name: Copioumed Chemotherapy Isolation Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYC, QSO Dated: August 28, 2023 Received: September 5, 2023
Dear Ming-Yie Jan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Bifeng Qian -S
BiFeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K230135
Device Name Copioumed Chemotherapy Isolation Gown
#### Indications for Use (Describe)
Copioumed Chemotherapy Isolation Gown is intended to be worn by healthcare personnel to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate. Copioumed Chemotherapy Isolation Gown is non-sterile and for single use only.
Copioumed Chemotherapy Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70).
Copioumed Chemotherapy Isolation Gown is a disposable device intended to protect health care personal from exposure to chemotherapy drugs during preparation, handling, and administration.
The Copioumed Chemotherapy Isolation Gown has been evaluated for resistance to permeation of various chemotherapy drugs per ASTM F739-20 "Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact." Chemotherapy drug permeation resistance, average standardized breakthrough time in minutes. *:
Carmustine [3.3 mg/ml] >480 Cisplatin [1.0 mg/ml] >480 Cyclophosphamide [20 mg/ml] >480 Cytarabine Hydrochloride [100 mg/ml] >480 Dacarbazine [10 mg/ml] >480 Daunorubicin Hydrochloride [5 mg/ml] >480 Doxorubicin Hydrochloride [2 mg/ml] > 480 Etoposide [20 mg/ml] >480 Fluorouracil [50 mg/ml] >480 Ifosfamide [50 mg/ml] > 480 Methotrexate [25 mg/ml] >480 Mitomycin C [0.5 mg/ml] >480 Mitoxantrone [2.0 mg/ml] >480 Paclitaxel [6 mg/ml] >480 Thiotepa [10 mg/ml] >480 Vincristine Sulfate [1 mg/ml] >480
*No permeation was detected at either the minimum detectable permeation or 0.1 ug/cm2/min
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 新 萬 盟 有 限 公 司
### 510(k) SUMMARY
| A. 510(K) NUMBER | K230135 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B. DATE PREPARED | September 27, 2023 |
| C. SUBMITTER | COPIOUMED INTERNATIONAL INC. |
| D. CONTACT PERSON | Primary Contact<br>Title: General Manager<br>Name: James Liu<br>Tel: + 886 928 644 966<br>E-Mail: james.liu@gcv.com.tw<br>Official Correspondent<br>Title: Principal Consultant<br>Name: Ming-Yie Jan<br>Tel: + 8886 2 8785 8989<br>E-Mail: mingyie.jan@ruscert.net |
| E. SUBJECT DEVICE | Proprietary Name: Copioumed Chemotherapy Isolation Gown<br>Product Code: FYC, QSO<br>Classification Names: Gown, Isolation, Surgical; Medical Gowns With<br>Chemotherapy Labeling Claims - Tested For Use With<br>Chemotherapy Drugs<br>Regulation Number 21 CFR 878.4040<br>Regulation Name: Surgical Apparel<br>Device Class: Class II<br>Review Panel: General Hospital |
| F. INDICATIONS FOR<br>USE | Copioumed Chemotherapy Isolation Gown is intended to be worn by healthcare<br>personnel to protect health care patients and health care personnel from the transfer<br>of microorganisms, body fluids and particulate. Copioumed Chemotherapy<br>Isolation Gown is non-sterile and for single use only.<br>Copioumed Chemotherapy Isolation Gown meets the requirements of an AAMI<br>Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012<br>Liquid Barrier Performance Classification of Protective Apparel Drapes Intended<br>for Use in Health Care Facilities (ANSI/AAMI PB70).<br>Copioumed Chemotherapy Isolation Gown is a disposable device intended to<br>protect health care personal from exposure to chemotherapy drugs during<br>preparation, handling, and administration.<br>The Copioumed Chemotherapy Isolation Gown has been evaluated for resistance<br>to permeation of various chemotherapy drugs per ASTM F739-20 "Standard Test<br>Method for Permeation of Liquids and Gases Through Protective Clothing<br>Materials Under Conditions of Continuous Contact." |
{6}------------------------------------------------
# 新 萬 盟 有 限 公 司
| | | Chemotherapy drug permeation resistance, average standardized breakthrough time<br>in minutes. *:<br>Carmustine [3.3 mg/ml] >480<br>Cisplatin [1.0 mg/ml] >480<br>Cyclophosphamide [20 mg/ml] >480<br>Cytarabine Hydrochloride [100 mg/ml] >480<br>Dacarbazine [10 mg/ml] >480<br>Daunorubicin Hydrochloride [5 mg/ml] >480<br>Doxorubicin Hydrochloride [2 mg/ml] > 480<br>Etoposide [20 mg/ml] >480<br>Fluorouracil [50 mg/ml] >480<br>Ifosfamide [50 mg/ml] > 480<br>Methotrexate [25 mg/ml] >480<br>Mitomycin C [0.5 mg/ml] >480<br>Mitoxantrone [2.0 mg/ml] >480<br>Paclitaxel [6 mg/ml] >480<br>Thiotepa [10 mg/ml] >480<br>Vincristine Sulfate [1 mg/ml] >480 |
|--|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | *No permeation was detected at either the minimum detectable permeation or 0.1<br>µg/cm2/min |
| | G. PREDICATE<br>DEVICE | Proprietary Name: AMD Ritmed AssureWear VersaGown<br>Product Code: FYC<br>Regulation Number: 21 CFR 878.4040<br>Regulation Name: Surgical Apparel<br>510(k) Number: K190306<br>510(k) Submitter: AMD Medicom Inc.<br>Device Class: Class II<br>Review Panel: General Hospital |
| | H. REFERENCE<br>DEVICE | Proprietary Name: CoverU Disposable Gown with Tape - Chemo Gown<br>Product Code: FYC, QSO<br>Regulation Number: 21 CFR 878.4040<br>Regulation Name: Surgical Apparel<br>510(k) Number: K210414<br>510(k) Submitter: Medtecs (Taiwan) Corp.<br>Device Class: Class II<br>Review Panel: General Hospital |
| | I. DEVICE<br>DESCRIPTION | Copioumed Chemotherapy Isolation Gown is a disposable device intended to<br>protect health care personal from exposure to chemotherapy drugs during<br>preparation, handling, and administration. This device is non-sterile and for single<br>use only. Copioumed Chemotherapy Isolation Gown is a Polypropylene Gown<br>laminated with Polyethylene and Knit cuff in three styles: Tie neck closure, Velcro<br>neck closure and Sticker neck closure. |
K230135
{7}------------------------------------------------
# 新 萬 盟 有 限 公 司
| Elements of<br>Comparison | Proposed Device:<br>Copioumed Chemotherapy<br>Isolation Gown<br>K230135 | Predicate Device:<br>AMD Ritmed AssureWear<br>VersaGown<br>K190306 | Comparisons |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | FYC, QSO | FYC | Same for FYC |
| Intended<br>Use/Indications<br>for Use | Copioumed Chemotherapy<br>Isolation Gown is intended to be<br>worn by healthcare personnel to<br>protect health care patients and<br>health care personnel from the<br>transfer of microorganisms, body<br>fluids and particulate. Copioumed<br>Chemotherapy Isolation Gown is<br>non- sterile and for single use<br>only.<br><br>Copioumed Chemotherapy<br>Isolation Gown meets the<br>requirements of an AAMI Level 3<br>barrier protection for an isolation<br>gown per ANSI/AAMI<br>PB70:2012 Liquid Barrier<br>Performance Classification of<br>Protective Apparel Drapes<br>Intended for Use in Health Care<br>Facilities (ANSI/AAMI PB70).<br><br>Copioumed Chemotherapy<br>Isolation Gown is a disposable<br>device intended to protect health<br>care personal from<br>exposure to chemotherapy drugs<br>during preparation, handling, and<br>administration.<br><br>The Copioumed Chemotherapy<br>Isolation Gown has been evaluated<br>for resistance to permeation of<br>various chemotherapy drugs per<br>ASTM F739-20 "Standard Test<br>Method for Permeation of Liquids<br>and Gases Through Protective<br>Clothing Materials Under<br>Conditions of Continuous<br>Contact " | AMD Ritmed Assure Wear™<br>VersaGown is intended to be<br>worn by healthcare personnel<br>to protect health care patients<br>and health care personnel from<br>the transfer of<br>microorganisms, body fluids<br>and particulate material.<br>AssureWear™ VersaGown<br>meets the requirements of an<br>AAMI Level 3 barrier<br>protection for an isolation<br>gown per ANSI/AAMI<br>PB70:2012 Liquid Barrier<br>Performance Classification of<br>Protective Apparel Drapes<br>Intended for Use in Health<br>Care<br>Facilities (ANSI/AAMI<br>PB70). AMD Ritmed<br>AssureWear™ VersaGown is<br>a single use, non-sterile<br>disposable medical device and<br>not intended for use in<br>operating rooms. The medical<br>device will be available in 18<br>models in large and X-large<br>sizes.<br><br>Flexneck Models Flexneck<br>Plus Models Tie Models<br>A69964 A69933 A8300<br>A69965 A69934 A8301<br>A69966 A69935 A8302<br>A69967 A69936 A8303<br>A69968 A69937 A8304<br>A69969 A69938 A8305 | Same:<br>1. AAMI<br>Level 3<br>Isolation<br>Gowns<br>2. Meet the<br>requirements<br>of level 3<br>Liquid Barrier<br>Performance<br>as per AAMI<br>PB70:2012<br>3. Non-<br>sterile and<br>single use<br>only.<br>Similar¹:<br>1. Protection<br>description<br>Different¹:<br>1. Evaluated<br>for resistance<br>to permeation<br>of various<br>chemotherapy |
# J. COMPARISON OF CHARACTERISTICS WITH THE PREDICATE DEVICE
{8}------------------------------------------------
# 新 萬 盟 有 限 公 司
| | Chemotherapy<br>drug permeation resistance,<br>average standardized breakthrough<br>time in minutes.*:<br>Carmustine [3.3 mg/ml] >480<br>Cisplatin [1.0 mg/ml] >480<br>Cyclophosphamide [20 mg/ml]<br>>480<br>Cytarabine Hydrochloride [100<br>mg/ml]<br>>480<br>Dacarbazine [10 mg/ml] >480<br>Daunorubicin Hydrochloride [5<br>mg/ml]<br>>480<br>Doxorubicin Hydrochloride [2<br>mg/ml] ><br>480<br>Etoposide [20 mg/ml] >480<br>Fluorouracil [50 mg/ml] >480<br>Ifosfamide [50 mg/ml] > 480<br>Methotrexate [25 mg/ml] >480<br>Mitomycin C [0.5 mg/ml] >480<br>Mitoxantrone [2.0 mg/ml] >480<br>Paclitaxel [6 mg/ml] >480<br>Thiotepa [10 mg/ml] >480<br>Vincristine Sulfate [1 mg/ml]<br>>480<br>*No permeation was detected at<br>either the minimum detectable<br>permeation or 0.1 µg/cm2/min | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material<br>Composition | Polypropylene Gown laminated<br>with polyethylene | PP SMS non-woven + PE | Similar2 |
| Design Features | Knit cuff in three style(model) Tie<br>neck closure, Velcro neck closure<br>and Sticker neck closure | Thumb loop Elastic cuffs<br>Extended cuff (Thumb loop)<br>Flexneck™<br>Tie (neck)<br>Straight sleeve Inclined sleeve<br>Blue belt tie | Similar3 |
| Sterility | Non-sterile | Non-sterile | Same |
| Color | Blue | Blue | Same |
| Use | OTC | OTC | Same |
| Contact Durations | Surface, < 24 hours | Surface, < 24 hours | Same |
| Disposable | Disposable | Disposable | Same |
| Single<br>Use/Reusable | Single Use | Single Use | Same |
| Performance per<br>AAMI PB70:2012 | AAMI Level 3 | AAMI Level 3 | Same |
| Basis weight<br>ASTM D3776 | 51.2 ± 1.2 g/m²<br>(1.5 oz/yd² ± 0.04) | 39.97 ± 1.61 g/m²<br>(1.17 oz/yd² ± 0.05) | Similar5 |
| Liquid barrier<br>performance:<br>Hydrostatic<br>Pressure AATCC<br>127 | Pass.<br>Met the standard's requirement (≥<br>50 cm H2 O)<br>Testing targets: main body fabric,<br>sleeve seam, shoulder seam, belt,<br>armpit | Chest: 109.34 ± 0.34 cmH2O<br>Sleeve seams: 110.67± 3.84<br>cmH2O<br>Belt attachments: 104 ± 5.19<br>cmH2O<br>Body/sleeve/belt mean: 108 ±<br>3.1 cm H2O | Similar6 |
| Liquid barrier<br>performance:<br>Impact Penetration<br>AATCC 42 | Pass,<br>Met the standard's requirement. (≤<br>1.0 g) | Chest: <0.1 g<br>Sleeve seams: <0.1 g<br>Belt attachments: < 0.1 g<br>Body/sleeve/belt mean: <0.1 g | Same |
| Breaking strength<br>(MD)<br>ASTM D5034 | 15.8 ± 3.7 lbf…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.