Disposable Isolation Gown -Level 3, Large, Blue, Disposable Isolation Gown - Level 3, X-Large, Blue

K212633 · String King Lacrosse, LLC · FYC · Nov 24, 2021 · General, Plastic Surgery

Device Facts

Record IDK212633
Device NameDisposable Isolation Gown -Level 3, Large, Blue, Disposable Isolation Gown - Level 3, X-Large, Blue
ApplicantString King Lacrosse, LLC
Product CodeFYC · General, Plastic Surgery
Decision DateNov 24, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The StringKing Disposable Isolation Gown is intended to protect health care personnel and health care patients from the transfer of microorganisms, body fluids and particulate material. These gowns are not intended for use in the operating room. The StringKing Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). The String King Disposable Isolation Gown is a single use, disposable medical device and is provided non-sterile.

Device Story

Disposable, single-use isolation gown; constructed from blue polyethylene (LDPE). Features hook and loop neck closure, waist tie, and thumb-loops on sleeves to prevent sliding during donning. Serves as physical barrier against microorganisms, body fluids, and particulates. Used in healthcare facilities by personnel and patients; not for operating room use. Provides AAMI Level 3 barrier protection. Non-sterile, disposable device.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing including AAMI PB70 Level 3 barrier performance, hydrostatic pressure (AATCC 127), impact penetration (AATCC 42), flammability (16 CFR 1610), tear resistance (ASTM D5587), grab tensile strength (ASTM D5034), and seam strength (ASTM D1683). Biocompatibility testing (ISO 10993-5, ISO 10993-10) confirmed non-cytotoxic, non-irritating, and non-sensitizing properties.

Technological Characteristics

Material: Polyethylene (LDPE, CAS # 9002-88-4). Barrier performance: AAMI Level 3 per ANSI/AAMI PB70:2012. Design: Hook and loop neck closure, waist tie, thumb-loops. Sterilization: None (non-sterile). Connectivity: None. Energy source: None (passive).

Indications for Use

Indicated for health care personnel and patients to protect against transfer of microorganisms, body fluids, and particulate material. Not for use in operating rooms. Single-use, non-sterile.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 24, 2021 String King Lacrosse, LLC % Sharon Morrow Regulatory Affairs Consultant Medical Device Academy Inc. 345 Lincoln Hill Road Shrewsbury, Vermont 05738 Re: K212633 Trade/Device Name: Disposable Isolation Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYC Dated: August 3, 2021 Received: August 19, 2021 Dear Sharon Morrow: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212633 Device Name Disposable Isolation Gown ### Indications for Use (Describe) The StringKing Disposable Isolation Gown is intended to protect health care personnel and health care patients from the transfer of microorganisms, body fluids and particulate material. These gowns are not intended for use in the operating room. The StringKing Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). The String King Disposable Isolation Gown is a single use, disposable medical device and is provided non-sterile. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K212633 # 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92: #### l. SUBMITTER String Kina Lacrosse, LLC 19100 S Vermont Ave. Gardena, CA, 90248 USA Tel: +1.508.654.1988 Fax: N/A Contact Person: Thomas Frasca Date Prepared: November 23, 2021 #### DEVICE 11. | Name of Device: | Disposable Isolation Gown | |------------------------------|---------------------------| | Classification Name: | Surgical Isolation Gown | | Regulation: | 21 CFR § 878.4040 | | Regulatory Class: | Class II | | Product Classification Code: | FYC | ### = PREDICATE DEVICE | Predicate Manufacturer: | Wildcat PPE, LLC | |-------------------------|----------------------------------------------| | Predicate Trade Name: | Wildcat PE Surgical Isolation Gown Full Back | | Predicate 510(k): | K202310 | No reference devices were used in this submission. #### DEVICE DESCRIPTION IV. Disposable, single-use isolation gown. The StringKing Disposable Isolation Gown is made of blue polyethylene and serves as a barrier to the transfer of microorganisms, body fluids and particulate material. The gowns feature hook and loop neck closure with a strap for fastening the back of the gown. The gown sleeves have thumb-loops to keep sleeves from sliding when donning the gown. The gowns are available in Large sizes. The gowns are non-sterile, single-use devices. ### V. INDICATIONS FOR USE The StringKing Disposable Isolation Gown is intended to protect health care personnel and health care patients from the transfer of microorganisms, body fluids and particulate material. These gownsare not intended for use in the operating room. The StringKing Disposable Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). The String King Disposable Isolation Gown is a single use, disposable medical device and is provided nonsterile. {4}------------------------------------------------ ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVCE The following characteristics were compared between the subject device and the predicate device to demonstrate substantial equivalence in Table 1 below. | Device | Proposed Device -<br>Disposable Isolation<br>Gown | Predicate Device -<br>Wildcat PE Surgical<br>Isolation Gown Full<br>Back | Result | |----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------| | Manufacturer | StringKing | Wildcat PPE, LLC | N/A | | 510K Number | K212633 | K202310 | N/A | | Product Common<br>Name | Disposable Isolation Gown | Surgical Isolation Gown | N/A | | Product Code | FYC | FYC | Same | | Classification | Class II (21 CFR 878.4040) | Class II (21 CFR 878.4040) | Same | | Intended<br>Use/Indications for<br>Use | The StringKing Disposable<br>Isolation Gown is intended to<br>protect health care personnel and<br>health care patients from the<br>transfer of microorganisms, body<br>fluids and particulate material.<br>These gownsare not intended for<br>use in the operating room. The<br>StringKing Disposable Isolation<br>Gown meets the requirements of<br>an AAMI Level 3 barrier protection<br>for an isolation gown per<br>ANSI/AAMI PB70:2012 Liquid<br>Barrier Performance Classification<br>of Protective Apparel Drapes<br>Intended for Use in Health Care<br>Facilities (ANSI/AAMI PB70). The<br>String King Disposable Isolation<br>Gown is a single use, disposable<br>medical device and is provided<br>non-sterile. | Wildcat PE Surgical Isolation Gown<br>Full Back is intended to protect<br>health care patients and health care<br>personnel from the transfer of<br>microorganisms, body fluids and<br>particulate material. Not intended for<br>use in the operating room. In<br>addition, The Wildcat PE Surgical<br>Isolation Gown Full Back meets the<br>requirements of an AAMI Level 3<br>barrier protection for an isolation<br>gown per ANSI/AAMI PB70:2012<br>Liquid Barrier and Performance<br>Classification of Protective Apparel<br>and Surgical Drapes Intended for<br>Use in Health Care Facilities<br>(ANSI/AAMI PB70). The Wildcat PE<br>Surgical Isolation Gown Full Back is<br>a single use, disposable medical<br>device provided non-sterile. | Same | | Material<br>Composition | Polyethylene LDPE<br>CAS # 9002-88-4 | Linear Low-Density Polyethylene<br>(LLDPE) | Same | | Design Features | Hook and Loop Neck Closure<br>Closed back - tie at waist<br>Thumb Loops | Tabs Tie in the Middle of Back<br>Thumbhole at Cuff | Similar; predicate<br>device has ties at<br>neck while<br>subject device<br>has hook and<br>loop | | Color | Blue<br>CAS # 147-14-8 | Blue (CAS unknown) | Same | | OTC Use | Yes | Yes | Same | | Sterility | Non-Sterile | Non-Sterile | Same | | Use | Single Use, Disposable | Single Use, Disposable | Same | Table 1 – Comparison of Technological Characteristics {5}------------------------------------------------ ## VII. NON-CLINICAL PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. | Device | Proposed Device -<br>Disposable<br>Isolation Gown | Predicate Device -<br>Wildcat PE Surgical<br>Isolation Gown (K202310) | Result | |-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------| | ANSI/AAMI<br>PB70:2012 | Level 3 | Level 3 | Same | | Water Resistance<br>HydrostaticPressure<br>Test - AATCC<br>127:2017 (cm) | Fabric (body)<br>Mean 170.33<br>Min 128.87<br>Max 196.87<br>Sleeve<br>Mean 126.60<br>Min 54.83<br>Max 189.40<br>Armhole<br>Each lot tested had less<br>than 4 failures in the<br>armhole region; thus<br>meeting the<br>requirement of AQL 4.0. | Chest:<br>Mean = $>151$<br>Min = 155<br>Max = $>155$<br>Sleeve Seam: Mean = $>165$<br>Min = 171<br>Max = $>173$ | Both the subject and predicate<br>device passed at $\ge$ 50cm | | Water Resistance<br>Impact Penetration<br>Test AATCC 42 | Fabric (body)<br>Mean 0.01<br>Min 0.00<br>Max 0.11<br>Sleeve<br>Mean 0.03<br>Min 0.01<br>Max 0.38<br>Armhole<br>Mean 0.02<br>Min 0.00<br>Max 0.16 | Chest:<br>Mean = $<0.1$<br>Min = $<0.1$<br>Max = .5<br>Sleeve Seam: Mean = $<0.1$<br>Min = $<0.1$<br>Max = $<0.1$ | Both the subject and<br>predicate device passed at<br>$\le$ 1.0g | | Flammability of<br>Clothing Textiles - 16<br>CFR Part 1610 (a) | Mean 5.41<br>Min 1.63<br>Max 7.87 | Class I | Class 1: average > 3.5<br>seconds. Both the subject and<br>predicate device passed with<br>an average of $\ge$ 3.5 seconds. | | Tear Resistance -<br>ASTM D1004* | N/A | Max Load (lbf): Mean = 1.04<br>Min = .95<br>Max = 1.17<br><br>Max Extension (in):<br>Mean = 1.01<br>Min = .91<br>Max = 1.10 | Subject device not tested as<br>this method is not required<br>by ANSI/AAMI PB70:2012<br>or ASTM F2407-20 | | General Tensile<br>Testing - ASTMD882* | N/A | Breaking Factor (lbf/in):<br>Mean = 6.66<br>Min = 4.87<br>Max = 9.20<br><br>Tensile (Max) (MPa):<br>Mean = 24.8<br>Min = 19.7<br>Max = 33.4<br><br>Tensile (Break) (MPa):<br>Mean = 21.5<br>Min = 15.9<br>Max = 33.4<br><br>Elongation (%): Mean = 1061<br>Min = 945<br>Max = 1327<br><br>Modulus (MPa): Mean = 91.1<br>Min = 2.58<br>Max = 117 | Subject device not tested as<br>this method is not required<br>by ANSI/AAMI PB70:2012<br>or ASTM F2407-20 | | Trap Tear ASTM<br>D5587-15 | Length<br>Mean 3.80<br>Min 3.18<br>Max 4.46<br><br>Width<br>Mean 6.88<br>Min 6.26<br>Max 7.58 | Mean = 5.10<br>Ind Min = 3.2<br>Ind Max = 7.0 | Both the subject and<br>predicate device passed at<br>≥2.3lbf | | Grab Tensile CD<br>ASTM D5034 | Length<br>Mean 23.75<br>Min 20.54<br>Max 25.91<br><br>Width<br>Mean 19.63<br>Min 17.10<br>Max 21.59 | Mean = 21.95<br>Ind Min = 20.60<br>Ind Max = 23.30 | Both the subject and<br>predicate device passed at<br>≥7lbf | | | | | | | ASTM D1683<br>Seam Strength | Sleeve<br>Mean 11.95<br>Min 6.48<br>Max 18.18<br>Armhole<br>Mean 12.30<br>Min 6.95<br>Max 17.41 | Not included in 510k summary | The subject device passed at<br>≥7lbf; test data not provided<br>in 510k summary for<br>predicate device | | Liquid Barrier<br>Performance<br>Classification<br>Properties | Device was tested in<br>accordance with<br>AAMI PB70:2012<br>and meets Level 3<br>requirements for a<br>surgical gown. | Device was tested in<br>accordance with AAMI<br>PB70:2012 and meets Level 3<br>requirements for a surgical<br>gown. | Same | | Biocompatibility | | | | | Cytotoxicity: ISO<br>10993-5:2009 [FR<br>Recognition #2-245]<br>Study # JN2010831 | Pass<br>Test Articles:<br>Grade 0 (non-cytotoxic) | Details not included in 510k<br>summary; Summary states "Pass:<br>ISO 10993-1" | Same | | Skin Irritation: ISO<br>10993-<br>10:2010(R)2014 [FR<br>Recognition 2-174]<br>Study #: JN2010832 | Negligible (pass)<br>Specimen:0.0 non-irritating | Details not included in 510k<br>summary; Summary states "Pass:<br>ISO 10993-1" | Same | | Skin Sensitization:<br>ISO 10993-<br>10:2010(R)2014 [FR<br>Recognition 2-174]<br>Study #: JN2010833 | No sensitization noted<br>(pass)<br>Specimen:0.0<br>non- sensitizing | Details not included in 510k<br>summary; Summary states "Pass:<br>ISO 10993-1" | Same | | Sterilization Modality | None (Non-Sterile) | None (Non-Sterile) | Same | {6}------------------------------------------------ {7}------------------------------------------------ *These tests are specifically for tear and tensile properties of thin plastic sheeting. The following standards were utilized in performance and biocompatibility testing of the StringKing Disposable Isolation Gown. Table 2 – Use of Standards | | Standards<br>Organization | Standards Title | |----------|---------------------------|---------------------------------------------------------------------------------------------| | PB70 | AAMI | AAMI PB70 Liquid Barrier Performance Classification of Protective<br>Apparel | | TM 127 | AATCO | Water Resistance: Hydrostatic Pressure Test | | TM 42 | AATCO | Water Resistance: Impact Penetration Test | | F2407-20 | ASTM | Standard Specification for Surgical Gowns Intended for Use<br>in Healthcare Facilities | | D5034 | ASTM | Standard Test Method for Breaking Strength and Elongation<br>of Textile Fabrics (Grab Test) | {8}------------------------------------------------ | D5587 | ASTM | Standard Test Method for Tearing Strength of Fabrics by<br>Trapezoid Procedure | |-----------|------|--------------------------------------------------------------------------------------------------------------| | D1683 | ASTM | Standard Test Method for Failure in Sewn Seams of Woven Fabrics | | Part 1610 | CPSC | Standard for the Flammability of Clothing Textiles | | 10993-1 | ISO | ISO 10993-1 Biological Evaluation of Medical Devices- Part 1:<br>Evaluation and | | 10993-5 | ISO | Biological Evaluation of Medical Devices- Part 5: Tests for<br>In-Vitro Cytotoxicity | | 10993-10 | ISO | Biological Evaluation of Medical Devices- Part 10: Tests for<br>irritation and delayed-type hypersensitivity | ## Sterilization & Shelf-life Testing Not Applicable (This is a non-sterile device and shelf-life is not applicable to this device because of low likelihood of time-dependent product degradation.) ## Biocompatibility Testing Biocompatibility testing was performed in accordance with ISO 10993-1:2018. Specifically, the following testing endpoints were evaluated. ## Table 3 - Biocompatibility Testing | Biocompatibility Testing Endpoints | Acceptance Criteria | Result | |------------------------------------|---------------------|--------| | Cytotoxicity - ISO 10993-5 | Non-Cytotoxic | Pass | | Skin Sensitization - ISO 10993-10 | Non- Sensitizing | Pass | | Skin Irritation - ISO 10993-10 | Non-Irritating | Pass | ## Software Verification and Validation Testing Not Applicable (Passive Device) ## Electrical safety and electromagnetic compatibility (EMC) Not Applicable (Passive Device) ## Animal Study Animal performance testing was not required to demonstrate safety and effectiveness of the device. ## Human Clinical Performance Testing Clinical testing was not required to demonstrate the safety and effectiveness of the device. ### VIII. CONCLUSIONS The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device (K202310), manufactured by Wildcat PPE LLC.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...