K241214 · Standard Textile Co., Inc. · FYA · Sep 10, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241214
Device Name
ProPel SG3 Surgical Gown
Applicant
Standard Textile Co., Inc.
Product Code
FYA · General, Plastic Surgery
Decision Date
Sep 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
ProPel SG3TM Surgical Gown is intended to protect both surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material. They are classified as Level 3 surgical gowns in accordance with ANSI/AAMI PB70. The reusable Propel Surgical Gown is provided non-sterile and must be sterilized before use. Sterilization parameters are as follows: Prevacuum Steam Sterilization: 132°C/270°F (temperature)/ 4 minutes (exposure time)/15 minute (dry time)
Device Story
Reusable, woven surgical gown; 100% polyester with silicone coating in critical zones; includes electrostatic dissipative yarn (graphene) in inner plies and back panels; features polyester mesh and vented cape on back for ventilation. Provided non-sterile; requires steam sterilization by healthcare facility before use. Used in operating rooms by surgical personnel. Protects patients and staff from fluid/microorganism transfer. Reusable up to 60 cycles. Includes RFID chip and QCM label for use-life tracking.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including liquid barrier performance (ANSI/AAMI PB70), tensile/tear/seam strength (ASTM standards), flammability (16 CFR 1610), and biocompatibility (ISO 10993 series).
Technological Characteristics
Reusable woven gown; 100% polyester with silicone coating; graphene-based electrostatic dissipative yarn. ANSI/AAMI PB70 Level 3 barrier. Dimensions vary by size (SM-3XL). Sterilization: Prevacuum steam (132°C/4 min). Includes RFID chip for tracking. Standards: ASTM D5034, D5587, D1683, F1868, D4846, D3776; ISO 10993-5, 10, 11, 23; AATCC 42, 127.
Indications for Use
Indicated for operating room personnel and surgical patients to provide a liquid barrier (Level 3 per ANSI/AAMI PB70) against microorganisms, body fluids, and particulate material during surgical procedures.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
K201351 — Maxima Surgical Gown · Elevate Textiles · Mar 11, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 10, 2024
Standard Textile Co., Inc. % Catherine Kang Senior Principal Consultant Ram+ 2790 Mosside Blvd Suite 800 Monroeville, Pennsylvania 15146
Re: K241214
Trade/Device Name: ProPel SG3TM Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: April 30, 2024 Received: August 9, 2024
Dear Catherine Kang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Image /page/2/Picture/3 description: The image shows the name "Bifeng Qian -S" in a large, sans-serif font. The text is black against a white background. The letters are evenly spaced and the text is horizontally aligned.
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241214
Device Name ProPel SG3TM Surgical Gown
#### Indications for Use (Describe)
ProPel SG37M Surgical Gown is intended to protect both surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material. They are classified as Level 3 surgical gowns in accordance with ANSI/AAMI PB70.
The reusable Propel Surgical Gown is provided non-sterile and must be sterilized before use. Stertlization parameters are as follows:
Prevacuum Steam Sterilization: 132°C/270°F (temperature)/ 4 minutes (exposure time)/15 minute (dry time)
Item Number Gown Size Tag Color Gown Size 7565CC17 Yellow SM 7565CC16 Gold MD 7565CC15 Green LG 7565CC14 Red XL 7565CC13 Orange 2XL 7565CC12 Purple 3XL
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K241214
### DATE PREPARED
September 6, 2024
| Manufacturer Name | Standard Textile Co., Inc.<br>One Knollcrest Drive<br>Cincinnati, OH 45237 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Richard M. Holbert, Jr., PhD<br>Vice President, Research and Development<br>Telephone: (513) 384-4189<br>Email: rholbert@standardtextile.com |
| Representative/Consultant | Catherine Kang, PhD<br>RQM+<br>2790 Mosside Blvd Suite 800,<br>Monroeville, PA 15146<br>Telephone: (312) 318-4730<br>Email: ckang@rqmplus.com |
#### DEVICE INFORMATION
Device trade name, or proprietary name: ProPel SG3TM Surgical Gown Device common name: Surgical Gown Classification name: Gown, Surgical Product classification: II Recommended classification regulation: 21 CFR 878.4040 Product Code(s): FYA Panel: General Hospital
#### PREDICATE DEVICE IDENTIFICATION
The ProPel SG3™ Surgical Gown is substantially equivalent to the following predicate device:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate |
|---------------|---------------------------------------|----------------------|
| K922753 | ComPel XTR/Standard Textile Co., Inc. | ✓ |
The predicate device has not been subject to a design related recall.
{5}------------------------------------------------
### DEVICE DESCRIPTION
The ProPel SG3TM Surgical Gown is a reusable, woven surgical gown that provides an ANSI /AAMI PB70:2022 Level 3 Liquid Barrier Performance Barrier using 100% polyester with silicone coating in the critical zone. The critical zone inner ply and back of the surgical gown is made of polyester with electrostatic dissipative yarn (graphene) and silicone coating. The gown has polyester mesh and vented cape on the back panels of the gown for ventilation. The ProPel SG3TM Surgical Gown may undergo up to 60 reprocessing cycles.
The ProPel SG3™ Surgical Gown is available in small, medium, large, extra large (XL), 2XL and 3XL. The device will be sold as a non-sterile surgical gown that is to be laundered, sterilized and processed by the healthcare facility prior to first use and after each subsequent use.
### INDICATIONS FOR USE
ProPel SG3™ Surgical Gown is intended to protect both surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material. They are classified as Level 3 surgical gowns in accordance with ANSI/AAMI PB70.
The reusable Propel Surgical Gown is provided non-sterile and must be sterilized before use. Sterilization parameters are as follows:
Prevacuum Steam Sterilization: 132°C/270°F (temperature)/ 4 minutes (exposure time)/15 minute (dry time)
| Item<br>Number | Gown Size Tag<br>Color | Gown<br>Size |
|----------------|------------------------|--------------|
| 7565CC17 | Yellow | SM |
| 7565CC16 | Gold | MD |
| 7565CC15 | Green | LG |
| 7565CC14 | Red | XL |
| 7565CC13 | Orange | 2XL |
| 7565CC12 | Purple | 3XL |
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The subject device has the same intended use and similar technological characteristics to the predicate device. These technological characteristics have undergone testing to ensure the device is as safe and effective as the predicate.
{6}------------------------------------------------
ProPel SG3™ Surgical Gown has a similar design and dimensions and uses similar materials as the ComPel XTR cleared under K922753 as indicated in the table below and forms the basis for the determination of substantial equivalence.
| Device<br>Characteristic | ComPel XTR<br>(Predicate Device K922753) | ProPel SG3™ Surgical Gown<br>(Subject Device: K241214) | Comparison |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code/<br>Regulation | FYA<br>878.4040 | FYA<br>878.4040 | Same |
| Product<br>Classification | II | II | Same |
| Intended Use | ComPel XTR is intended to be<br>worn by operating room<br>personnel during surgical<br>procedures to protect both the<br>surgical patient and the<br>operating room personnel<br>from transfer of<br>microorganisms, body fluids,<br>and particulate material. | ProPel SG3™ Surgical Gown is<br>intended to be worn by operating room<br>personnel during surgical procedures to<br>protect both the surgical patient and the<br>operating room personnel from transfer<br>of microorganisms, body fluids, and<br>particulate material. | Same |
| Indication for<br>Use | ComPel XTR Surgical gowns<br>are intended to be worn by<br>operating room personnel<br>during surgical procedures to<br>protect them by inhibiting the<br>migration of liquids through<br>the surface of the critical areas<br>of the gown. When processed<br>according to instructions,<br>ComPel XTR Surgical gowns<br>will function and perform as a<br>surgical gown. The ComPel<br>XTR Surgical gowns are<br>reusable through 75 wash, dry,<br>and sterilization cycles. They<br>are manufactured and<br>distributed as non-sterile<br>surgical gowns that are<br>intended to be sterilized and<br>processed by health care<br>facilities and/or contract<br>sterilization/laundry<br>companies. | ProPel SG3™ Surgical Gown is<br>intended to protect both surgical<br>patients and operating room personnel<br>from the transfer of microorganisms,<br>body fluids, and particulate material.<br>They are classified as Level 3 surgical<br>gowns in accordance with ANSI/AAMI<br>PB70-2022.<br><br>The reusable Propel Surgical Gown is<br>provided non-sterile and must be<br>sterilized before use. Sterilization<br>parameters are as follows:<br><br>Prevacuum Steam Sterilization:<br>132°C/270°F (temperature)/ 4 minutes<br>(exposure time)/15 minute (dry time)<br><br>Item Number Gown Size<br>Tag Color Gown<br>Size<br>7565CC17 Yellow SM<br>7565CC16 Gold MD<br>7565CC15 Green LG<br>7565CC14 Red XL<br>7565CC13 Orange 2XL<br>7565CC12 Purple 3XL | Similar. Both devices are<br>surgical gowns that are<br>indicated to protect<br>operating room<br>personnel. |
| ANSI/AAMI<br>PB70 Level<br>Classification in<br>Critical Zones | Level 3* | Level 3 | Similar. Predicate<br>predated standards for<br>testing but was<br>subsequently confirmed |
| Device<br>Characteristic | ComPel XTR<br>(Predicate Device K922753) | ProPel SG3TM Surgical Gown<br>(Subject Device: K241214) | Comparison |
| | | | to meet ANSI/AAMI<br>PB70, which the subject<br>device also meets. |
| Design | ● Tape ties<br>● Metal snaps<br>● 100% polyester knit cuffs<br>● Overlapping back panels | ● Tape Ties<br>● Plastic interlocking snaps<br>● 100% polyester knit cuffs<br>● Overlapping back panels with vent<br>and mesh | Similar. Both devices<br>include tape ties and<br>snaps for fastening the<br>gown, 100% polyester<br>knit cuffs and<br>overlapping back panels.<br>The addition of a vent<br>and mesh in the subject<br>device does not impact<br>safety or effectiveness. |
| Materials | ● 100% dyed Dacron<br>polyester with silicone<br>coating (outer ply critical<br>zone)<br>● 100% polyester with<br>carbon yarns (inner ply<br>critical zone and back<br>panels) | ● 100% polyester with silicone<br>coating (outer ply front and<br>sleeves)<br>● Polyester with electrostatic<br>dissipative yarn (graphene) and<br>silicone coating (inner ply front,<br>sleeves, outer back panels) | Similar. Both gowns<br>contain 100% polyester<br>with silicone coating.<br>The subject device<br>includes additional<br>materials that passed<br>biocompatibility testing. |
| Color | Blue | Green | Different. Subject device<br>passed biocompatibility<br>testing. No impact on<br>safety or effectiveness. |
| Sizes | Small/medium<br>Large/extra large<br>2XL | Small<br>Medium<br>Large<br>Extra Large<br>2XL<br>3XL | Similar. Both gowns are<br>available in a variety of<br>sizes. |
| Reusable | Yes, up to 75 times | Yes, up to 60 times | Both gowns are reusable<br>with fewer reuses for the<br>subject device. |
| Sterility | Provided nonsterile to user<br>for steam sterilization | Provided nonsterile to user for<br>steam sterilization | Same |
| Tensile<br>Strength | Warp/Fill<br>150 lbf/123 lbf | $\ge$ 30 N ( $\ge$ 7 lbf) | Similar. Predicate<br>predated current<br>standards but has been<br>tested against the<br>requirements.<br>Performance of the<br>subject device<br>demonstrates the product<br>meets this requirement. |
| Tear Strength | N/A | $\ge$ 10 N ( $\ge$ 2.3 lbf) | Similar. Predicate<br>predated current<br>standards but has been<br>tested against the<br>requirements. |
| Device<br>Characteristic | ComPel XTR<br>(Predicate Device K922753) | ProPel SG3TM Surgical Gown<br>(Subject Device: K241214) | Comparison |
| Seam Strength | N/A | ≥ 30 N (≥ 7 lbf) | Performance of the<br>subject device<br>demonstrates the product<br>meets this requirement.<br>Different. Predicate<br>predated current<br>standards. Performance<br>of the subject device<br>demonstrates the product<br>meets this requirement. |
| Hydrostatic<br>Resistance | 25 lbs/14 lbs (min.)<br>Mullens Hydrostat (1X/75X) | ≥ 50 cm (Level 3) | Similar. Predicate<br>predated standards for<br>testing but was<br>subsequently confirmed<br>to meet ANSI/AAMI<br>PB70 Level 3<br>requirements, which the<br>subject device also<br>meets. |
| Impact<br>Penetration | N/A | ≤ 1.0 g (Level 3) | |
| Lint Generation | <100 (Linting coefficient of 3) | Linting coefficient of 3 | Same |
| Flammability | Class I | Class I | Same |
| Cytotoxicity | Under the conditions of the<br>study, the device is non-<br>cytotoxic | Under the conditions of the study, the<br>device is non-cytotoxic | Same |
| Sensitization | Under the conditions of<br>study, not a sensitizer | Under the conditions of study,<br>not a sensitizer | Same |
| Irritation | Under the conditions of<br>study, not an irritant | Under the conditions of study,<br>not an irritant | Same |
| Pyrogenicity | N/A | Non-pyrogenic | Different. Predicate<br>predated standards for<br>testing. Performance of<br>the subject device<br>demonstrates the product<br>is nonpyrogenic. |
| Acute Systemic<br>Toxicity | No acute systemic toxicity | No acute systemic toxicity | Same |
{7}------------------------------------------------
{8}------------------------------------------------
* At the time of 510(k) clearance, AAMI levels for surgical gowns had not yet been established. Testing subsequently performed established that the gown meets AAMI PB70 Level 3 requirements.
### SUMMARY OF NON-CLINICAL TESTING
The test results of non-clinical testing demonstrated that the proposed device met its acceptance criteria or testing endpoints with the use of following standards.
- . ANSI/AAMI PB70:2022: Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.
- . ASTM F2407-20: Standard Specification for Surgical Gowns Intended for
{9}------------------------------------------------
Use in Healthcare Facilities
- 16 CFR 1610: Standard for flammability of clothing textiles ●
- AATCC Method 42: Test Method for Water Resistance: Impact Penetration
- AATCC Method 127: Test Method for Water Resistance: Hydrostatic Pressure ●
- ASTM D1683: Standard Test Method for Failure in Sewn Seams of Woven . Apparel Fabrics
- . ASTM D5034: Standard Test Method for Breaking Strength and Elongation of Textile Fabrics
- ASTM D5587: Standard Test Method for Tearing Strength of Fabrics by . Trapezoid Procedure
- ASTM F1868-17: Standard Test Method for Thermal and Evaporative ● Resistance of Clothing Materials
- . ASTM D4846-96 (2011): Standard Test Method for Resistance to Unsnapping of Snap Fasteners
- . ISO 10993-5:2009: Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021: Biological evaluation of medical devices-Part 10: Tests for ● skin sensitization
- . ISO 10993-11:2017: Biological evaluation of medical devices-Part 11: Tests for systemic toxicity
- ISO 10993-23:2021 Biological evaluation of medical devices-Part 23: . Tests for irritation
- USP <151> Pyrogen Test
The table below provides a summary of non-clinical testing performed.
| Test<br>Methodology | Test Method/<br>Applicable Standard(s) | Acceptance Criteria | Result |
|---------------------------|----------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Tensile Strength | ASTM D5034/<br>ASTM F2407-20 | ≥ 30 N (≥ 7 lbf) | PASS |
| Tear Strength | ASTM D5587/<br>ASTM F2407-20 | ≥ 10 N (≥ 2.3 lbf) | PASS |
| Seam Strength | ASTM D1683/<br>ASTM F2407-20 | ≥ 30 N (≥ 7 lbf) | PASS |
| Hydrostatic<br>Resistance | AATCC Method 127/<br>ANSI/AAMI PB70 | ≥ 50 cm per AAMI PB70 for Level 3 | PASS |
| Impact<br>Penetration | ATCC Method 42/<br>ANSI/AAMI PB70 | ≤ 1.0 g per AAMI PB70 for Level 3 | PASS |
| Lint Generation | ISO 9073-10 | --- | Coefficient for<br>linting is 3 for<br>both predicate and<br>subject device |
| Flammability | 16 CFR 1610 | Pass Class I Flammability | PASS<br>Class I |
| Snap Strength | ASTM D4846-96 | Fasteners provide peel strength of ≤ 1.5 lbf<br>(easy doff) and shear force ≥ 4.5 lbf (secure) | PASS |
{10}------------------------------------------------
| Test<br>Methodology | Test Method/<br>Applicable Standard(s) | Acceptance Criteria | Result |
|-------------------------------------------|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Evaporative<br>Resistance | ASTM F1868-17 | Evaporative resistance of critical zone<br>(average of 3 specimens) ≤ 350 Pa m²/W;<br>evaporative resistance of the back mesh (2-<br>ply average of 3 specimens) ≤ 5 Pa m²/W | PASS |
| Critical Zone<br>Fabric Weight<br>Testing | ASTM D3776/D3776M-20 | --- | Gown critical<br>zone fabric<br>construction<br>weighs less than<br>7.83 OSY |
| Use Life<br>Tracking | --- | QCM label is legible and RFID Chip can be<br>read at the beginning and end of use life<br>(after 60 reuses) | PASS |
| Cytotoxicity | ISO 10993-5 | Less than/equal to grade 2 (mild reactivity) | PASS |
| Sensitization | ISO 10993-10 | No evidence of causing delayed dermal<br>contact sensitization | PASS |
| Irritation | ISO 10993-23 | No erythema, no edema | PASS |
| Pyrogenicity | USP <151>/<br>ISO 10993-11 | Temperature rise within acceptable limits | PASS |
| Acute Systemic<br>Toxicity | ISO 10993-11 | No mortality or evidence of systemic toxicity | PASS |
#### SUMMARY OF CLINICAL TESTING Not Applicable
# CONCLUSION
The conclusions drawn from the nonclinical tests demonstrate that ProPel SG3™ Surgical Gown is as safe, as effective, and performs as well as or better than the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.