Sterile Triplex Surgical Gown (S, M, L, XL, XXL, XXXL)
K230577 · Medcare Saglik Urunleri Sanayi VE Ticaret Anonim Sirketi · FYA · Aug 16, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230577
Device Name
Sterile Triplex Surgical Gown (S, M, L, XL, XXL, XXXL)
Applicant
Medcare Saglik Urunleri Sanayi VE Ticaret Anonim Sirketi
Product Code
FYA · General, Plastic Surgery
Decision Date
Aug 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Sterile Triplex Surgical Gown (S, M, L, XL, XXL, is a sterile single-use surgical gown intended to protect surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material. They have been tested and are classified as Level 4 in the critical zones per AAMI Standard PB70 Liquid barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities.
Device Story
Sterile single-use surgical gown; provides protective barrier for surgical patients and OR personnel. Functions by preventing transfer of microorganisms, body fluids, and particulate material. Classified as AAMI Level 4 for critical zones. Used in operating room environments; worn by healthcare staff. Benefits patient and staff by reducing risk of cross-contamination and exposure to infectious agents during surgical procedures.
Clinical Evidence
Bench testing only; device classified as AAMI Level 4 per AAMI PB70 standard for liquid barrier performance.
Technological Characteristics
Sterile, single-use surgical gown. Classified as AAMI Level 4 in critical zones per AAMI PB70 standard. Form factor: various sizes (S, M, L, XL, XXL, XXXL).
Indications for Use
Indicated for use by surgical patients and operating room personnel as a sterile, single-use protective barrier against microorganisms, body fluids, and particulate material in healthcare settings.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Related Devices
K192560 — Surgical Gown · Shandong Kangli Medical Equipment Technology Co., Ltd. · May 22, 2020
K222403 — Surgical Gown · Guangzhou Fuzelong Hygiene Material Co., Ltd. · Jan 26, 2023
K200824 — Cardinal Health SMARTGOWN Breathable Surgical Gown, Cardinal Health SMARTGOWN AIR Breathable Surgical Gown · Cardinal Health200, LLC · Jul 29, 2020
K211765 — Cardinal Health RoyalSilk Surgical Gown, Cardinal Health RoyalSilk Scrub Nurse Gown · Cardinal Health, LLC · Nov 23, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
August 16, 2023
Medcare Saglik Urunleri Sanayi Ve Ticaret Anonim Sirketi % Rhonda Alexander Sr. Consultant, Regulatory Strategy IUVO Consulting, LLC P.O. Box 56436 Virginia Beach, Virginia 23456
Re: K230577
Trade/Device Name: Sterile Triplex Surgical Gown (S, M, L, XL, XXL, XXXL) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: June 7, 2023 Received: June 7, 2023
Dear Rhonda Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230577
Device Name
Sterile Triplex Surgical Gown (S, M, L, XL, XXL, XXXL)
Indications for Use (Describe)
The Sterile Triplex Surgical Gown (S, M, L, XL, XXL, is a sterile single-use surgical gown intended to protect surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material.
They have been tested and are classified as Level 4 in the critical zones per AAMI Standard PB70 Liquid barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------|--|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | |
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.