← Product Code [FXN](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXN) · K070426

# SILHOUETTE, MODEL 1000.01 (K070426)

_Aranz Medical Limited · FXN · Jun 29, 2007 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FXN/K070426

## Device Facts

- **Applicant:** Aranz Medical Limited
- **Product Code:** [FXN](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXN.md)
- **Decision Date:** Jun 29, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4160
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The intended use of the Silhouette system is to measure and document the progression of external wounds over time. The Silhouette system is comprised of a camera (SilhouetteCamera) connected to a PDA running a software application (SilhouetteMobile Software) to capture and document images taken by the user. A software accessory (SilhouetteServer) on a PC may be used to transfer images and documentation to a third-party patient information database. The system is non-contact with respect to the patient.

## Device Story

Silhouette is a non-contact wound measurement and documentation system. It consists of a camera (SilhouetteCamera) connected to a PDA running mobile software (SilhouetteMobile) to capture wound images. A PC-based software accessory (SilhouetteServer) facilitates data transfer to third-party patient databases. The system is operated by healthcare providers to track wound progression over time. By capturing images and performing measurements, the device assists clinicians in documenting wound healing, providing more accurate measurements than predicate devices. The system does not contact the patient.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Non-contact wound measurement system comprising a camera, PDA, and PC software. Operates via image capture and digital measurement. Connectivity provided via PC-based server for database integration.

## Regulatory Identification

A surgical camera and accessories is a device intended to be used to record operative procedures.

## Predicate Devices

- Verge Videometer (Verg Incorportated)
- Visitrak (Smith & Nephew)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

|                                          | K070426                       |          |            |
|------------------------------------------|-------------------------------|----------|------------|
|                                          | 510(k) Premarket Notification |          | QF-00      |
|                                          | Silhouette                    |          | 2007-00001 |
| aranz<br>applied research associates ltd | 510(k) Summary of Safety and  | Rev: 1.0 | Approved   |

## 510(k) Summary of Safety and Effectiveness

: •

| Submitter:                | ARANZ Medical Limited                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | JUN 29 2007 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Address:                  | Ground Floor, St Elmo Courts<br>47 Hereford Street, PO Box 3894<br>Christchurch 8013                                                                                                                                                                                                                                                                                                                                                                                                                                                                |             |
| Phone:                    | +64 3 3746120 ext 217                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |             |
| Fax:                      | +64 3 3746130                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |             |
| Contact:                  | Shane Goodwin                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |             |
| Trade name:               | Silhouette                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |             |
| Common name:              | Wound measurement and documentation system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |             |
| Classification name:      | Surgical camera and accessories (21 CFR 878.4160)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |             |
| Class:                    | Class I (general controls)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |             |
| Predicate Devices:        | Verge Videometer (Verg Incorportated)<br>Visitrak (Smith & Nephew)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |             |
| Device Description:       | Silhouette consists of a camera connected to a PDA, for the<br>measurement and tracking of wounds                                                                                                                                                                                                                                                                                                                                                                                                                                                   |             |
| Indications for Use:      | Silhouette is indicated for wound measurement and documentation<br>and can be used on all external wound types.                                                                                                                                                                                                                                                                                                                                                                                                                                     |             |
| Intended Use:             | The intended use of the Silhouette system is to measure and<br>document the progression of external wounds over time. The<br>Silhouette system is comprised of a camera (SilhouetteCamera)<br>connected to a PDA running a software application<br>(SilhouetteMobile Software) to capture and document images taken<br>by the user. A software accessory (SilhouetteServer) on a PC may<br>be used to transfer images and documentation to a third-party<br>patient information database. The system is non-contact with<br>respect to the patient. |             |
| Comparison to Predicates: | Silhouette is substantially equivalent the predicate devices,<br>considered in light of the comparison of the indications for use, the<br>intended use, the workflow of the devices and the technological<br>principles being applied. Standard wound measurements made with<br>Silhouette are more accurate than those made with the predicate<br>devices. These differences do not raise questions safety and<br>effectiveness over the predicate devices.                                                                                        |             |
| Date of Summary:          | 17 January 2007                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |             |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 9 2007

Aranz Medical Limited % Shane Goodwin Regulatory Scientist St. Elmo Courts Ground Floor 47 Hereford Street P.O. Box 3894 Christchurch, New Zealand

Re: K070426

Trade/Device Name: Silhouette Wound Measurement and Documentation System Regulation Number: 21 CFR 878.4160 Regulation Name: Surgical camera and accessories Regulatory Class: I Product Code: FXN Dated: June 18, 2007 Received: June 20, 2007

Dear Shane Goodwin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{2}------------------------------------------------

Page 2 - Shane Goodwin

CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known): K070426

Device Name: Silhouette Wound Measurement and Documentation System

Indications For Use:

Silhouette is indicated for wound measurement and documentation and can be used on all external wound types.

| Prescription Use<br>(Part 21 CFR 801 Subpart D) | ✓ | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |  |
|-------------------------------------------------|---|--------|------------------------------------------------|--|
|-------------------------------------------------|---|--------|------------------------------------------------|--|

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of General, Restorative,

and Neurological Devices

| 510(k) Number | 467424 |
|---------------|--------|
| Page          | 1 of 1 |

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FXN/K070426](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FXN/K070426)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
