KARL STORZ DIRECT COUPLED INTERFACE (DCI 5) CAMERA SYSTEM

K990154 · Karl Storz Imaging, Inc. · FWF · Jun 1, 1999 · General, Plastic Surgery

Device Facts

Record IDK990154
Device NameKARL STORZ DIRECT COUPLED INTERFACE (DCI 5) CAMERA SYSTEM
ApplicantKarl Storz Imaging, Inc.
Product CodeFWF · General, Plastic Surgery
Decision DateJun 1, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4160
Device ClassClass 1

Indications for Use

The Karl Storz Imaging (KSI) Direct Coupled Interface (DCI 5) Camera System combines a color, television camera head with an integrated light source cable and camera control unit (CCU). These units are designed for attachment to any Karl Storz Tuttlingen (KST) DCI compatible rigid or flexible endoscope, including, but not limited to sinoscopes, colonoscopes, sigmoidoscopes, bronchoscopes, gastroscopes, laparoscopes, choledochoscopes, ureteroscopes, hysteroscopes and arthroscopes. The DCI 5 camera head is coupled to the endoscope by directly pressing the endoscope onto the front of the camera head until it locks into place. An optional adapter allows DCI 5 camera head attachment to non-DCI rigid and flexible endoscopes. The endoscopic image can be displayed on any standard operating room video monitor. All commonly available endoscopic light sources may be used with the DCI 5 System.

Device Story

DCI 5 Camera System consists of color television camera head, integrated light cable, and camera control unit (CCU). Device attaches to rigid/flexible endoscopes via direct coupling or optional adapter. User operates focus knob, zoom control, and programmable buttons on camera head; axial rotation allows image orientation adjustment. CCU processes video signals for display on standard OR monitors. System provides contrast enhancement, signal amplification, and electronic shutter control. Used by physicians in clinical/surgical settings for endoscopic visualization. Benefits include integrated light/camera interface and ergonomic control of image orientation and zoom during procedures.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Single CCD sensor; sealed aluminum camera housing; integrated light cable; programmable function buttons; focus and zoom controls; composite video output; electronic shutter control. Connectivity via standard video signal. Sterilization method not specified.

Indications for Use

Indicated for visualization during endoscopic procedures using rigid or flexible endoscopes (e.g., sinoscopes, colonoscopes, sigmoidoscopes, bronchoscopes, gastroscopes, laparoscopes, choledochoscopes, ureteroscopes, hysteroscopes, arthroscopes) in patients requiring endoscopic examination or surgery.

Regulatory Classification

Identification

A surgical camera and accessories is a device intended to be used to record operative procedures.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ R990154 ## 510(k) Summary of Safety and Effectiveness This summary of premarket notification safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. - Karl Storz Imaging, Incorporated Application: 175 Cremona Drive Goleta, California 93117 Contact: Mr. Terry Fernandez Registration: 2027009 Proprietary Name -- Karl Storz Imaging Direct Coupled Interface Camera Head Device Name: Common Name -- Color Television Camera Head Classification Name -- Camera, Television, Endoscopic Intended Use: The Karl Storz Imaging (KSI) Direct Coupled Interface (DCI 5) Camera System combines a color, television camera head with an integrated light source cable and camera control unit (CCU). These units are designed for attachment to any Karl Storz Tuttlingen (KST) DCI compatible rigid or flexible endoscope, including, but not limited to sinoscopes, colonoscopes, sigmoidoscopes, bronchoscopes, gastroscopes, laparoscopes, choledochoscopes, ureteroscopes, hysteroscopes and arthroscopes. The DCI 5 camera head is coupled to the endoscope by directly pressing the endoscope onto the front of the camera head until it locks into place. > An optional adapter allows DCI 5 camera head attachment to non-DCI rigid and flexible endoscopes. The endoscopic image can be displayed on any standard operating room video monitor. All commonly available endoscopic light sources may be used with the DCI 5 System. Device Description: The KSI DCI 5 incorporates two programmable function buttons (to select an accessory port(s) and to control the port output), a focus knob and a zoom in/zoom out control bar. The camera head also has an integrated light cable. The camera is designed to allow the user to axially rotate the camera head while maintaining proper image orientation on the monitor. 000014 {1}------------------------------------------------ Substantial Equivalence: The DCI 5 Camera System is a combination of features found in earlier KSI premarket notifications. The proposed camera head and its indications for use are based upon KSI's Direct Coupled Camera 510(k), K973252. Both devices feature direct connection to an endoscope, light guide camera connection, programmable buttons, camera head focus control, automatic image rotation, sealed aluminum camera housings, single CCD sensing and anti-fogging features. The proposed camera IPM CCU board is based upon KSI's Digivideo 510(k), K950974. Both units feature contrast enhancement, use of standard video signal, and digital processing. The proposed camera system's CCU is based upon KSI's Urocam/Hysterocam 510(k), K940283. Both units feature the same power supply, maximum signal amplification, electronic shutter control, composite video output and cable connection to a camera The proposed new KSI camera head is unique only in that it head. incorporates into one camera head variations of several features found in similar, earlier camera head models. Signed: Terry Fernandez Date: 18 January 1999 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 1 1999 Mr. Terry Fernandez Director, Regulatory and Standards Compliance Karl Storz Imaging, Inc. 175 Cremona Drive Goleta, California 93117 Re: K990154 > Trade Name: Karl Storz Imaging Direct Coupled Interface Camera Head Regulatory Class: II Product Code: FWF Dated: April 28, 1999 Received: April 29, 1999 Dear Mr. Fernandez: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Mr. Terry Fernandez This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page _ _1 K990/54 510(k) Number (if known): Device Name: KSI DIRECT COUPLED CAMERA SYSTEM Indication for Use: The Karl Storz Imaging (KSI) Direct Coupled Interface (DCI 5) Camera System is a color, television camera head with integrated light source cable, and separate video processor. The camera system is suitable for use with any KST DCI rigid or flexible endoscope, including, but not limited to sinoscopes, colonoscopes, sigmoidoscopes, gastroscopes, gastroscopes, laparoscopes, choledochoscopes, ureteroscopes, hysteroscopes and arthroscopes. The camera head is coupled to the endoscope by directly pressing the endoscope onto the front of the DCI 5 camera head until it locks into place. The endoscopic image can be displayed on any standard operating room video monitor. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) OR (Division Sign-Off) 510(k) Number K990154 Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use_ (Optional Format 1-2-96) 0000018
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...