← Product Code [FTC](/submissions/SU/subpart-e%E2%80%94surgical-devices/FTC) · K223882

# Narrowband UV Phototherapy Light Lamp (Model: HB-UPLL-01, HB-UPLL-02) (K223882)

_Shenzhen Honpal Optoelectronic Technology Co., Ltd. · FTC · Mar 27, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FTC/K223882

## Device Facts

- **Applicant:** Shenzhen Honpal Optoelectronic Technology Co., Ltd.
- **Product Code:** [FTC](/submissions/SU/subpart-e%E2%80%94surgical-devices/FTC.md)
- **Decision Date:** Mar 27, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4630
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Narrowband UV Phototherapy Light Lamp is intended for use, by or under the direction of a physician for the treatment of psoriasis, vitiligo, and atopic dermatitis (eczema). It is intended for use on all skin types (I-VI). It can be used in hospitals, clinics and households.

## Device Story

Portable medical device for localized phototherapy; treats psoriasis, vitiligo, and atopic dermatitis. Consists of LED lamp, irradiator, and control circuit. Operated by physician or patient (home use) under physician direction. User sets irradiation time (0-299s) via device buttons; light comb attachment facilitates treatment of scalp/hair-covered areas. Device emits UVB radiation (280-320nm) to target skin; excludes eyes. Provides therapeutic benefit through controlled UV exposure. Used in hospitals, clinics, and households.

## Clinical Evidence

No clinical data included. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and biocompatibility (ISO 10993-5, -10, -23).

## Technological Characteristics

Handheld phototherapy device; ABS shell; LED light source; UVB spectral output (280-320nm). Powered via DC inlet. IP22 rated. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, and IEC 62471. Biocompatibility per ISO 10993-5, 10993-10, 10993-23. Non-sterile.

## Regulatory Identification

An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.

## Predicate Devices

- UV Phototherapy Device ([K181805](/device/K181805.md))

## Reference Devices

- RESOLVE UVB PHOTOTHERAPY SYSTEM ([K072035](/device/K072035.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 27, 2023

Shenzhen Honpal Optoelectronic Technology Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161,East Lujiazui Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K223882

Trade/Device Name: Narrowband UV Phototherapy Light Lamp (Model: HB-UPLL-01, HB-UPLL-02) Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC Dated: December 6, 2022 Received: December 27, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223882

Device Name

Narrowband UV Phototherapy Light Lamp(Model:HB-UPLL-01, HB-UPLL-02)

Indications for Use (Describe)

The Narrowband UV Phototherapy Light Lamp is intended for use, by or under the direction of a physician for the treatment of psoriasis, vitiligo, and atopic dermatitis (eczema). It is intended for use on all skin types (I-VI). It can be used in hospitals, clinics and households.

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# K223882

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

# 1.0 Submitter's Information

| Name:    | Shenzhen Honpal Optoelectronic Technology Co., Ltd.                                                |  |  |  |  |
|----------|----------------------------------------------------------------------------------------------------|--|--|--|--|
| Address: | Room 505,Shenghui Building A,Xixiang Street,Baonan District,Shenzhen City,Guangdong Province,China |  |  |  |  |
| Tel:     | +86-15986705009                                                                                    |  |  |  |  |
| Contact: | Wang Tianming                                                                                      |  |  |  |  |

### Designated Submission Correspondent

| Contact: | Mr. Boyle Wang                                                         |
|----------|------------------------------------------------------------------------|
| Name:    | Shanghai Truthful Information Technology Co., Ltd.                     |
| Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai,<br>200120 China |
| Tel:     | +86-21-50313932                                                        |
| Email:   | Info@truthful.com.cn                                                   |

Date of Preparation: Mar.21,2023

### 2.0 Device Information

| Trade/Device name:   | Narrowband UV Phototherapy Light Lamp       |
|----------------------|---------------------------------------------|
| Common name:         | Ultraviolet Lamp for Dermatologic Disorders |
| Classification name: | Light, Ultraviolet, Dermatological          |
| Model(s):            | HB-UPLL-01, HB-UPLL-02                      |
| Production code:     | FTC                                         |
| Regulation number:   | 21 CFR 878.4630                             |
| Classification:      | Class II                                    |
| Panel:               | General & Plastic Surgery                   |

# 3.0 Predicate Device Information

Predicate Device#

Xuzhou Kernel Medical Equipment Co., LTD. Manufacturer:

{4}------------------------------------------------

UV Phototherapy Device: 510(k) number: K181805

Reference Device # Manufacturer: ALLUX MEDICAL Device: RESOLVE UVB PHOTOTHERAPY SYSTEM 510(k) number: K072035

## 4.0 Indication for Use Statement

The Narrowband UV Phototherapy Light Lamp is intended for use, by or under the direction of a physician for the treatment of psoriasis, vitiligo, and atopic dermatitis (eczema). It is intended for use on all skin types (I-VI). It can be used in hospitals, clinics and households.

### 5.0 Device Description

The proposed device, Narrowband UV Phototherapy Light Lamp is a portable medical device that consists of LED lamp, irradiator, control circuit. It is a therapeutic product under the direction of a physician for individuals who require ultraviolet radiation for diagnosed skin disorders. It is for the partial treatment excluding eyes. Irradiation time can be adjusted from 0~299s and the treatment status can be controlled by the button on the device. The light comb equipped on the device is intended for easier treatment of target skin covered by hair, such as the scalp. The device is available in two models: model HB-UPLL-01and model HB-UPLL-02.

The device can be used in hospitals, clinics and households.

### 6.0 _Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: IEC60601-1:2005+CORR.1:2006+CORR.2:2007+AM1:2012 electrical equipment Part 1: General requirements for basic safety and essential performance. IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.

IEC 60601-1-11:2015, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC 60601-2-57: 2011, Medical electrical equipment Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source 

{5}------------------------------------------------

equipment intended for therapeutic, diagnostic, monitoring and cosmetic/ aesthetic use.

IEC 62471:2006, Photobiological safety of lamps and lamp systems

#### Biocompatibility Test

ISO 10993-5:2009,Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity

ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization

ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation

### 7.0 _Clinical Test Conclusion

No clinical study is included in this submission.

#### 8.0 Technological Characteristic Comparison Table

#### Table1-General Comparison

{6}------------------------------------------------

| Item                                  | Subject Device<br>K223882                                                                                                                                                                                                                                                                                          | Predicate Device<br>K181805                                                                                                                                                                                                                                                               | Reference Device<br>K072035                                                                                                                                                                                                                                                                                                                                               | Remark    |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code                          | FTC                                                                                                                                                                                                                                                                                                                | FTC                                                                                                                                                                                                                                                                                       | FTC                                                                                                                                                                                                                                                                                                                                                                       | Same      |
| Regulation No.                        | 21 CFR 878.4630                                                                                                                                                                                                                                                                                                    | 21 CFR 878.4630                                                                                                                                                                                                                                                                           | 21 CFR 878.4630                                                                                                                                                                                                                                                                                                                                                           | Same      |
| Class                                 | II                                                                                                                                                                                                                                                                                                                 | II                                                                                                                                                                                                                                                                                        | II                                                                                                                                                                                                                                                                                                                                                                        | Same      |
| Intended<br>Use/Indication for<br>Use | The Narrowband UV<br>Phototherapy Light Lamp is<br>intended for use, by or under the<br>direction of a physician for the<br>treatment of psoriasis, vitiligo,<br>and atopic dermatitis (eczema). It<br>is intended for use on all skin<br>types (I-VI).<br>It can be used in hospitals, clinics<br>and households. | The UV Phototherapy is<br>intended for use, by or under the<br>direction of a physician for the<br>treatment of psoriasis, vitiligo,<br>and atopic dermatitis (eczema).<br>It is intended for use on all skin<br>types (I-VI). It can be<br>used in hospitals, clinics and<br>households. | The ResolveTM UVB<br>Phototherapy System is an<br>ultraviolet light emitting<br>medical device for localized<br>phototherapeutic treatment of<br>dermatologic conditions such<br>as psoriasis, vitiligo, atopic<br>dermatitis (eczema), and<br>seborrheic dermatitis. The<br>ResolveTM UVB Phototherapy<br>System is intended to be used<br>with all Skin Types (I - VI). | Same      |
| Prescriptive                          | Yes                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                       | Same      |
| Mode of operation                     | Handheld                                                                                                                                                                                                                                                                                                           | Handheld                                                                                                                                                                                                                                                                                  | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                    | Same      |
| Shell material                        | ABS                                                                                                                                                                                                                                                                                                                | ABS                                                                                                                                                                                                                                                                                       | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                    | Same      |
| Treatment Area                        | Partial treatment excluding eyes                                                                                                                                                                                                                                                                                   | Partial treatment excluding eyes                                                                                                                                                                                                                                                          | Partial treatment excluding<br>eyes                                                                                                                                                                                                                                                                                                                                       | Same      |
| Treatment Time                        | 0~299s                                                                                                                                                                                                                                                                                                             | 0~180s                                                                                                                                                                                                                                                                                    | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                    | Different |
| Test Frequency                        | 3~5 times per week                                                                                                                                                                                                                                                                                                 | 3~5 times per week                                                                                                                                                                                                                                                                        | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                    | Same      |
| Treatment Area                        | HB-UPLL-01: 30 cm²                                                                                                                                                                                                                                                                                                 | 60cm²/100cm²/104cm²                                                                                                                                                                                                                                                                       | 12.8 x 9.0 cm                                                                                                                                                                                                                                                                                                                                                             | Different |
|                                       | HB-UPLL-02: 7 cm²                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                           |           |
| Irradiation Intensity                 | HB-UPLL-01: UVB: 2.0 MW/cm²<br>HB-UPLL-02: UVB: 1.2 MW/cm²                                                                                                                                                                                                                                                         | UVA: 1~50 MW/cm²<br>UVB: 0.3~20 MW/cm²                                                                                                                                                                                                                                                    | 1 mW/cm²                                                                                                                                                                                                                                                                                                                                                                  | Similar   |
| Working distance                      | HB-UPLL-01:3cm;<br>HB-UPLL-02:2cm                                                                                                                                                                                                                                                                                  | 3 cm                                                                                                                                                                                                                                                                                      | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                    | Same      |
| UV spectral output                    | UVB                                                                                                                                                                                                                                                                                                                | UVA or UVB                                                                                                                                                                                                                                                                                | UVB                                                                                                                                                                                                                                                                                                                                                                       | Same      |
| Lamp configuration                    | 1 or 2 UV tube                                                                                                                                                                                                                                                                                                     | 1 or 2<br>9W UV tube                                                                                                                                                                                                                                                                      | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                    | Same      |
| Power Source                          | DC inlet                                                                                                                                                                                                                                                                                                           | AC outlet or DC jack                                                                                                                                                                                                                                                                      | AC outlet                                                                                                                                                                                                                                                                                                                                                                 | Same      |
| Wavelength range<br>(nm)              | UVB: 280~320                                                                                                                                                                                                                                                                                                       | UVA: 350~400<br>UVB: 310~315                                                                                                                                                                                                                                                              | 300-320nm                                                                                                                                                                                                                                                                                                                                                                 | Similar   |
| IPX - Rating / water<br>resistance    | IP22                                                                                                                                                                                                                                                                                                               | IP22                                                                                                                                                                                                                                                                                      | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                    | Same      |
| Application<br>Environment            | Hospitals, Clinics and<br>Households.                                                                                                                                                                                                                                                                              | Hospitals, Clinics and<br>Households.                                                                                                                                                                                                                                                     | Hospitals, Clinics and<br>Households.                                                                                                                                                                                                                                                                                                                                     | Same      |
| Electrical<br>Safety/Performance      | Comply with IEC60601-1 and IEC<br>60601-2-57                                                                                                                                                                                                                                                                       | Comply with IEC60601-1 and<br>IEC 60601-2-57                                                                                                                                                                                                                                              | Comply with IEC60601-1 and<br>IEC 60601-2-57                                                                                                                                                                                                                                                                                                                              | Same      |
| Home Use                              | Comply with IEC 60601-1-11                                                                                                                                                                                                                                                                                         | Comply with IEC 60601-1-11                                                                                                                                                                                                                                                                | Comply with IEC 60601-1-11                                                                                                                                                                                                                                                                                                                                                | Same      |
| Sterile                               | Non-sterile                                                                                                                                                                                                                                                                                                        | Non-sterile                                                                                                                                                                                                                                                                               | Non-sterile                                                                                                                                                                                                                                                                                                                                                               | Same      |
| Single Use                            | No                                                                                                                                                                                                                                                                                                                 | No                                                                                                                                                                                                                                                                                        | No                                                                                                                                                                                                                                                                                                                                                                        | Same      |
| EMC                                   | Comply with IEC 60601-1-2                                                                                                                                                                                                                                                                                          | Comply with IEC 60601-1-2                                                                                                                                                                                                                                                                 | Comply with IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                 | Same      |
| Biocompatibility                      | Cytotoxicity (ISO 10993-5:2009)<br>Sensitization (ISO<br>10993-10:2021)<br>Irritation (ISO 10993-23:2021)                                                                                                                                                                                                          | Under the conditions of the<br>study, Comply with the<br>requirements                                                                                                                                                                                                                     | Under the conditions of the<br>study, Comply with the<br>requirements                                                                                                                                                                                                                                                                                                     | Same      |

{7}------------------------------------------------

{8}------------------------------------------------

#### Analysis:

The Irradiation Intensity of the proposed devices, but the value range of UVB is within that of the predicate device K181805 and between the value of the reference device K072035. So the we can think the slight differences will not affect the substantive equivalence.

The output specifications as "Treatment time" of the proposed device are a little different from the predicate devices, but they are considered substantially equivaled the performance tests in accordance with same standards: IEC 6001-1, IEC 60601-1-2 and IEC 60601-2-57,So there are no safety and effectiveness aspects concerned.

In conclusion, the technological characteristics, features, mode of the device substantially equivalent to the predicate devices quoted above. The subject device and predicate devices do not raise new issues of safety of effectiveness.

## 9.0 Conclusion

The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device and raises no new questions of safety or effectiveness. The substantive equivalence to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FTC/K223882](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FTC/K223882)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
