K200971 · Xuzhou Kernel Medical Equipment Co., Ltd. · FTC · Dec 15, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200971
Device Name
308nm Excimer System
Applicant
Xuzhou Kernel Medical Equipment Co., Ltd.
Product Code
FTC · General, Plastic Surgery
Decision Date
Dec 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The 308nm Excimer System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
Device Story
308nm Excimer System is a handheld ultraviolet light therapy instrument; utilizes Xenon-Chlorine (XeCl) excimer lamp to emit monochromatic 308nm UVB light. Device features 8-inch touch LCD screen; multi-spec treatment handle filters; one-click light source output; user management; and fingerprint entry (KN-5000D). Operated by clinicians in clinical settings for targeted small-area treatment. Device performs manual Minimal Erythema Dose (MED) tests to determine patient-specific skin characteristics. Output is controlled via software to limit maximum dosage to 5 J/cm². Accurate calibration system (KN-5000D) ensures reliable dose output. Healthcare providers use device to deliver controlled UV radiation to skin lesions; intended to improve patient outcomes for psoriasis and vitiligo.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified with IEC 60601-1 (safety), IEC 60601-2-57 (non-laser light source), IEC 60601-1-2 (EMC), IEC 62471 (photobiological safety), and ISO 10993-5/10 (biocompatibility/cytotoxicity/irritation).
Technological Characteristics
Xenon-Chlorine (XeCl) excimer lamp; 308nm ± 2nm wavelength; handheld irradiation; air cooling; 8-inch touch LCD; AC100-240V power; software-controlled dose limits (max 5 J/cm²); automatic calibration (KN-5000D). Complies with IEC 60601-1, IEC 60601-2-57, IEC 60601-1-2, IEC 62471, ISO 10993-5, and ISO 10993-10.
Indications for Use
Indicated for treatment of psoriasis and vitiligo on intact skin.
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
K202827 — 308nm Excimer UV-light Skin Therapy System · Shenzhen Gsd Tech Co., Ltd. · Dec 15, 2020
K073066 — 308 DERMATOLOGICAL EXCIMER SYSTEM · Quantel Medical, Inc. · Dec 26, 2007
K024020 — THERALIGHT UV120-2 UVA/UVB PHOTOTHERAPY SYSTEM · Theralight, Inc. · Jan 17, 2003
K062963 — PHAROS EXCIMER LASER SYSTEM, MODEL EX-308 · Ra Medical Systems,Inc · Apr 3, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 15, 2020
Xuzhou Kernel Medical Equipment Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K200971
Trade/Device Name: 308nm Excimer System Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC Dated: October 15, 2020 Received: October 19, 2020
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200971
Device Name 308nm Excimer System
Indications for Use (Describe)
The 308nm Excimer System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
#### 1.0 Submitter's Information
Name: Xuzhou Kernel Medical Equipment Co., Ltd. Address: Kernel Mansion, Economic Development District, Xuzhou City, Jiangsu Province, 221004 China Tel: +86-516-87732209 Fax: +86-516-87732210 Contact: Jing Wang Date of Preparation: Dec. 10th, 2020
#### Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Address: Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
#### 2.0 Device Information
Trade name: 308nm Excimer System Ultraviolet Lamp for Dermatologic Disorders Common name: Classification name: Light, Ultraviolet, Dermatological Model(s): KN-5000C; KN-5000D
#### 3.0 Classification
Production code: FTC Regulation number: 21 CFR 878.4630 Classification: Class II Review Panel: General& Plastic Surgery
#### 4.0 Predicate Device Information
Predicate Device Information 1: Manufacturer: Chongqing Peninsula Medical Technology Co.,Ltd. 308nm Excimer System Device:
{4}------------------------------------------------
510(k) number: K172273
Predicate Device Information 2:
Manufacturer: Chongqing Peninsula Medical Technology Co.,Ltd. Device: 308nm Excimer System (Model: XECL-308E) 510(k) number: K192642
#### 5.0 Indication for Use Statement
The 308nm Excimer System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
#### 6.0 Device Description
The proposed device, 308nm Excimer System is a new type of ultraviolet light therapy instrument, which could emit high levels of UV light. It has the following features:
- High intensity UV light. -
- The instrument is suitable for small area treatment with good targeting. -
- -8 inches Super Large color touch LCD screen which make the operation display more convenient;
- Manual MED tests can also be performed according to the skin characteristics of the patient.
- One-click Control light source Output. -
- Powerful user management capabilities. -
- Can realize the export and view of user information; -
- -Multi-spec treatment handle light filter. The user can choose filter according to the size of the treatment area. Avoid excess radiation output.
- -Fingerprint entry makes it easy for doctors to quickly access patient information (for KN-5000D):
- Accurate calibration system. Enhanced reliability of instrument dose output -(KN-5000D applicable).
#### Non-Clinical Test Conclusion
Non clinical tests were conducted by lab bench testing to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2012, Medical electrical equipment-Part 1: General requirements for basic safety, and essential performance.
IEC 60601-2-57: 2011 Medical electrical equipment Part2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and
{5}------------------------------------------------
cosmetic/ aesthetic use.
IEC 60601-1-2:2014, Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
IEC 62471:2006, Photobiological safety of lamps and lamp systems
ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
### 7.0 Clinical Test Conclusion
No clinical study is included in this submission.
#### 8.0 Substantial Equivalence Comparison
| Table1-Technological Characteristic Comparison | | | | | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|---------------|
| Item | Proposed device<br>K200971 | Predicated device 1<br>K172273 | Predicated device 2<br>K192642 | Remark | |
| Product<br>Code | FTC | FTC | FTC | SE | |
| Regulation<br>No. | 21 CFR 878.4630 | 21 CFR 878.4630 | 21 CFR 878.4630 | SE | |
| Class | II | II | II | SE | |
| Intended<br>Use | The 308nm Excimer<br>System is intended to be<br>used for the treatment of<br>psoriasis and vitiligo. It is<br>to be used on intact skin<br>only. | The 308nm Excimer<br>System is intended to<br>be used for the<br>treatment of psoriasis<br>and vitiligo. It is to be<br>used on intact skin<br>only. | The 308nm Excimer<br>System(model:<br>XECL-308E) is an<br>Ultraviolet Light<br>Emitting<br>Medical<br>Device.It is intended<br>to be used for the<br>treatment of<br>psoriasis, vitiligo,<br>atopic dermatitis, and<br>leukoderma. It is to<br>be used on intact skin<br>only. | SE | |
| Prescription<br>Required | Yes | Yes | Yes | SE | |
| Weight | 5kg (Treatment<br>handle:1.4kg ) | < 2.6 kg | Not publicly available | -- | |
| Dimensions<br>(H x W x D) | 34.1cm×29.7cm×<br>22.1cm(For Host) | 26 cm x 24 cm x 27<br>cm(For lamp)<br>5 cm x 9.5 cm x 18 cm | 548.0mm*253.6<br>mm*181.4mm | -- | |
| | 28.0cm(After placing the<br>Treatment handle) | cm<br>(For Adapter) | | | |
| Mode of<br>Operation | Continuous operation | Continuous operation | Not publicly available | SE | |
| Wavelength | 308 nm ± 2nm | 308 nanometers (nm)<br>±3nm | 308±2 nm | SE | |
| Light Source | Xenon-Chlorine (XeCI)<br>excimer lamp produces<br>monochromatic UVB light | Xenon-Chlorine<br>(XeCI)<br>excimer lamp<br>produces<br>monochromatic UVB<br>light | Xenon-Chlorine<br>(XeCl) excimer<br>lamp produces<br>monochromatic<br>UVB light | SE | |
| Irradiation<br>mode | Handheld irradiation | Handheld irradiation | Handheld irradiation | SE | |
| Cooling of<br>light source | Air cooling | Air cooling | Cooling fan | SE | |
| Treatment<br>Area | $20cm^2±10%$ | 16 cm² (4 x 4 cm) | $6cm^2~28.8 cm^2$ | Analysis<br>1 | |
| Max.UV<br>Irradiation<br>Intensity | 50 mW/cm² | 50 mW/cm² | 50 mW/cm² | SE | |
| Max.<br>Irradiation<br>Power | 1000 mW | 800 mW | 1440 mW | Analysis<br>1 | |
| Treatment<br>Time | 0~120s | 1s to 40 s | 1-140s | Analysis<br>1 | |
| Maximum<br>dose | 5 J/cm² | 2 J/cm² | 7 J/cm² | Analysis<br>2 | |
| Power<br>Source | AC100-240V, 50/60Hz<br>Input Power:<br>KN-5000C: 160VA<br>KN-5000D: 200VA | Adapter Input:<br>100~240Vac, 50/60<br>Hz<br>Main unit input: 48<br>Vdc,2.94 A,135 VA<br>max | Not publicly available | Analysis<br>3 | |
| Power<br>Calibration<br>Method | KN-5000D:<br>Automatic calibration<br>KN-5000C:NA | Internal, automatic | Not publicly available | SE | |
| MED Dose<br>Determination | Manual MED test | Menu driven | Not publicly available | Analysis<br>4 | |
| Display | 8 inches touch LCD<br>screen | Touch Screen Control<br>Panel | Not publicly available | SE | |
| Security<br>Type | Classification by type of<br>anti-electric shock: Class | Classification by type<br>of anti-electric shock: | Not publicly available | SE | |
| | | I; Classification according<br>to the degree of<br>anti-electric shock: Type B. | Class I; Classification<br>according to the<br>degree of anti-electric<br>shock: Type B. | | |
| | Patient<br>Leakage<br>Current | Complied with IEC<br>60601-1 and IEC<br>60601-2-57 | Complied with IEC<br>60601-1 and IEC<br>60601-2-57 | Complied with IEC<br>60601-1 and IEC<br>60601-2-57 | SE |
| | Operating<br>Environment | Temperature: 5~35℃<br>Relative humidity: ≤85%<br>Atmospheric pressure::<br>700hPa~1060hPa | Temperature:<br>15~35°C,<br>Humidity: ≤ 80%RH,<br>Atmospheric<br>Pressure:<br>86 ~ 106 kPa | Temperature:<br>5℃~30℃<br>Humidity: ≤80%RH<br>Atmospheric<br>Pressure: 700 ~<br>1060 hPa | Analysis<br>4 |
| | Storage<br>Environment | Ambient temperature:<br>-40 ~ 55 °C<br>Relative humidity: ≤<br>90%<br>Atmospheric pressure:<br>500 ~ 1060hpa | Temperature:<br>-20~45°C,<br>Humidity: 10-85%RH<br>Atmospheric<br>Pressure:<br>50 ~ 106 kPa | Temperature:<br>0℃~40℃<br>Humidity:≤85%RH<br>Atmospheric<br>Pressure: 700 ~<br>1060 hPa | Analysis<br>4 |
| | Electrical<br>Safety/Performance | Comply with IEC60601-1<br>and IEC 60601-2-57 | Comply with<br>IEC60601-1 and IEC<br>60601-2-57 | Comply with<br>IEC60601-1 and IEC<br>60601-2-57 | SE |
| | Sterile | N/A | N/A | N/A | SE |
| | Single Use | No | No | No | SE |
| | EMC | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | SE |
| | Biocompatibility | Cytotoxicity (ISO<br>10993-5:2009)<br>Irritation (ISO<br>10993-10:2010<br>Sensitization (ISO<br>10993-10:2010) | Under the conditions<br>of the study, Comply<br>with the requirements | Under the conditions<br>of the study, Comply<br>with the requirements | SE |
| | Label and<br>Labeling | Conforms to FDA<br>Regulatory Requirements | Conforms to FDA<br>Regulatory<br>Requirements | Conforms to FDA<br>Regulatory<br>Requirements | SE |
# Table1 Tochnological Charactoristic Comparis
{6}------------------------------------------------
{7}------------------------------------------------
Comparison Discussion:
#### Analysis1:
Although some output specifications as "Max Power Output", "Treatment area", and "Treatment time" of the proposed device are a little different from the predicate devices, but they are considered substantially equivalent, and they all comply with the standards: IEC 60601-1, IEC 60601-1-2 and IEC 60601-2-57, and range between the predicate device K172273 and predicate
{8}------------------------------------------------
device K192642. So the we can think that the slight differences in specification will not raise any safety or effectiveness issue.
#### Analysis 2:
The software limits the maximum dosage of the proposed device to no more than 5 J/cm²,it is different from the predicate devices, but the value range between the predicate device K172273 and predicate device K192642. So the we can think that the slight differences in specification will not raise any safety or effectiveness issue.
#### Analvsis 3:
The " Power Source" is a little different from the predicate devices, but they all comply with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57 requirements. So the differences will not raise any safety or effectiveness issue.
#### Analvsis 4:
The " Power Source", "MED Dose Determination" and "Operating and Storage Environment" are a little different from the predicate devices, but they all comply with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57 requirements. So the differences of function specification will not raise any safety or effectiveness issue.
#### Final Conclusion:
Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device. The proposed device "308nm Excimer System" is Substantial Equivalent to the predicate devices.
#### 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the leqally marketed predicated device and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.