← Product Code [FSY](/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY) · K140930

# VIDA LED SURGICAL LIGHTING SYSTEM, V SERIES (K140930)

_Convida Healthcare & Systems Corporation · FSY · Jun 9, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY/K140930

## Device Facts

- **Applicant:** Convida Healthcare & Systems Corporation
- **Product Code:** [FSY](/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY.md)
- **Decision Date:** Jun 9, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4580
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.

## Device Story

VIDA LED Surgical Lighting System, V series is a ceiling-mounted surgical lamp providing illumination for operating rooms. Device utilizes high-efficacy LEDs as light source; features prism-based light delivery technology to focus light into 110cm focal cylinder, reducing shadow. System offers 5 color temperature settings (3500K-5000K) and adjustable intensity up to 160,000 lux. Operated by surgeons or OR staff via lighthead handle or wall-mounted control center. Includes ambient lighting modes (central luminance or green light) for endoscopy/diagnostic procedures. Supports optional HD 1080p camera handles and medical-grade LCD monitors via pre-wired suspension arms. Designed for durability, ease of cleaning, and continuous operation during power failure. Provides consistent, glare-free illumination to assist surgical teams in visualizing the operative site.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing and compliance with international standards (IEC 60601-1, IEC 60601-2-41, IEC 60601-1-2).

## Technological Characteristics

Ceiling-mounted surgical luminaire; LED light source; prism-based optical delivery; 5 color temperature settings (3500K-5000K); CRI 95; 160,000 lux max intensity. Modular design; pre-wired for HD video/camera integration. Materials: tough molded, sealed housing for cleaning. Standards: IEC 60601-1, IEC 60601-2-41, IEC 60601-1-2, ISO 14971, ISO 15223-1, ISO 17664, IEC 62304.

## Regulatory Identification

A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

## Predicate Devices

- VIDA LED Surgical Lighting System-X series ([K132307](/device/K132307.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains the logo for ConVida Healthcare & Systems. The logo features a geometric pattern resembling a flower or starburst on the left. To the right of the pattern is the text "ConVida" in a bold, sans-serif font. Below "ConVida" is the text "Healthcare & Systems" in a smaller font size.

ConVida Healthcare & Systems Corp. 510(K) Notification, K number: N/A

# JUN 0 9 2014

Integrate Visual Clarity at the Highest Precision

VIDA LED Surgical Lighting System, V series

## 510(K) Summary

| 1. | Type of Submission:  | Traditional                                                         |
|----|----------------------|---------------------------------------------------------------------|
| 2. | Preparation Date:    | April 3, 2014                                                       |
| 3. | Revised Date:        | N/A                                                                 |
| 4. | Submitter:           | ConVida Healthcare & Systems Corporation                            |
|    | Address:             | No.33, Dinghu Rd., Guishan Township, Taoyua<br>County 33378, Taiwan |
|    | Phone:               | +886-3-318-3777 #332                                                |
|    | Fax:                 | +886-3-318-7799                                                     |
|    | Contact:             | Sharon Lee / Regulatory Engineer                                    |
|    | Registration Number: | 3010402711                                                          |

#### 5. Identification of the Device:

| Proprietary/Trade name: | VIDA LED Surgical Lighting System, V series |
|-------------------------|---------------------------------------------|
| Common Name:            | Surgical lamp                               |
| Classification Name:    | Light, Surgical, Ceiling Mounted            |
| Device Classification:  | 2                                           |
| Regulation Number:      | 878.4580                                    |
| Panel:                  | General & Plastic Surgery                   |
| Product Code:           | FSY                                         |

#### б. Identification of the Predicate Device:

| Predicate Device Name: | VIDA LED Surgical Lighting System-X series |
|------------------------|--------------------------------------------|
| Manufacturer:          | ConVida Healthcare & Systems Corporation   |
| Product Code:          | FSY                                        |
| 510(K) Number:         | K132307                                    |

{1}------------------------------------------------

Integrate Visual Cinrity at the Highest Precision

Image /page/1/Picture/1 description: The image contains the logo for ConVida Healthcare & Systems. The logo consists of a grid-like pattern on the left and the word "ConVida" in bold, sans-serif font on the right. Below "ConVida" are the words "Healthcare & Systems" in a smaller font size.

ConVida Healthcare & Syst 510(K) Notification, K number: N/A VIDA LED Surgical Lighting System, V series

#### 7. Intended Use and Indications for Use of the Subject Device:

#### Intended Use:

The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.

### Indications for Use:

The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.

#### 8. Device Description:

The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.

VIDA LED Surgical Lighting System, V series uses high-efficacy light-emitting-diodes (LED) as light source which provides high brightness and no extra heat; the design of the optical mechanism gives a characteristic of less shadow which is significant to the surgeon during the operation. Given of its modular design, VIDA V provided many options for configurations and can be installed in an operating room with or without a laminar air flow system.

The features of VIDA LED Surgical Lighting System, V series are as followed:

- Prism-Based Light Delivery Technology: ConVida's patented optical system, that . combines an advanced refractive prism lens and innovative high-power LED technology, focuses light into an intense 110cm (43") focal cylinder, eliminating the need for refocusing when adjusting surgical table.
- . Consistent Illumination at Different Color Temperature Settings: VIDA V series lighthead's color temperature can be individually adjusted at any of the 5 settings (3500K, 4000K, 4250K, 4500K, 5000K) with a high CRI of 95, R9 94.
- . Optimal Brightness without Glare: Elegant, sophisticated optical system consists up to 2 types of high-power LEDs and multiple refractive prisms produces up to 160,000 lux of intensity without discomfort flare/glare.
- . VIDA V light gives the Surgeon Control: On the VIDA V, both lighthead focus and light intensity can easily be adjusted via turning the lighthead handle. They are also adjustable from the standard lighthead control or optional VIDA V Control Center on the wall.
- Long Lifetime: Under normal use, VIDA V LEDs enjoy an extremely long lifetime of . 50,000+ hours (70% lumen maintenance at 50,000 hours).
- . Two Types of Ambient Lights (Option): 5% central luminance (GUIDE lighting mode) c be configured through standard lighthead control. Guide lighting mode may ambient lighting for manipulating endoscopy instruments or lighting diagnos procedures. Green ambient light sheds low-level green light from the suspension hub

{2}------------------------------------------------

Integrate Visual Clarity at the Highest Precision

ConVida Healthcare & Systems Corp.

510(K) Notification, K number: N/A

VIDA LED Surgical Lighting System, V series

and is controllable on the optional VIDA V Control Center.

- Camera-Ready Lighthead is Built into the System: All VIDA V lightheads are pre-wired . with HD cabling for taking an optional HD 1080p Camera Handle. The HD 1080p Camera Handle is removable from one VIDA V lighthead to another providing superb flexibility for recording.
- Exceptional Maneuvering Flexibility: The exceptionally durable non-sterile handles allow easy positioning of all our lightheads. Each lighthead rotates a full 360°so that the system can be quickly and easily positioned for any surgical procedure. Individual lightheads can be lower to 810mm above the floor providing excellent low lateral illumination.
- . Sleek Non-Sterile Handle: The lighthead has is molded with a sleek all-around handle that facilitates easy repositioning and adds modern elegance to O.R. setting
- Easy-to-clean Compact Lighthead: Made of tough molded materials, the VIDA V series . lightheads are completely sealed which helps prevent biocontaminants from entering the operative site and lighthead itself, and facilitates cleaning.
- . HD Supported Suspensions: All VIDA V suspensions are prewired with video cables to deliver information users need within and without the surgical field. Images are in crisp and accurate HD format with no signal conversion necessary and no loss of video quality.
- Heavy Duty HD Flat Monitor Suspension: Flat Monitor Suspension Arms support single/dual monitors up to a total of 24 kg (53 lbs), making them ideal for surgical displays up to 32". Large display compatibility allows the surgical team to see more of the procedural area. Optional Medical Grade LCD Displays from SONY also offer uncompressed HD video at 1080p which shows true images from SD- and HD-endoscopes, ultrasound, C-arms and other equipment.
- Optimal HD Connectivity: The built-in fiber optic cable allows the suspension to support monitors featuring Full HD 1920 x 1080 video signals.
- . Low Profile Suspension: Low-profile cardinal lighthead suspension is offered for rooms with ceiling lower than 280mm.
- . Failsafe Design: The unit is protected against an electrical power failure, thus ensuring that all surgical procedures can continue without interruption.

Image /page/2/Picture/14 description: The image shows a collection of squares of varying sizes arranged in a pattern. The squares are outlined in black, and some of them have a gray fill. The squares are arranged in a way that creates a visually interesting composition. The text "page 5-4" is visible in the lower right corner of the image.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the ConVida Healthcare & Systems logo. The logo consists of a grid-like pattern on the left and the text "ConVida Healthcare & Systems" on the right. The text is in a bold, sans-serif font, with "ConVida" appearing larger than "Healthcare & Systems".

ConVida Healthcare & Syste 510(K) Notification, K number: N/A

Integrate Visual Clarity at the Highest Precision

VIDA LED Surgical Lighting System, V series

#### 9. Performance Summary:

This device conforms to the standards listed as below:

- IEC 60601-1:2005 Medical Electrical Equipment Part 1: General Requirements for Basic . Safety and Essential Performance
- . IEC 60601-2-41:2009 Medical Electrical Equipment - Part 2-41: Particular requirements for the basic and essential performance of surgical luminaires and luminaires for diagnosis
- . IEC 60601-1-2:2007 Medical Electrical Equipment, Part 1-2: General requirements for basic Safety and essential performance Collateral Standard: Electromagnetic Compatibility Requirements and tests
- IEC 62304:2006 Medical Device Software: Software Life Cycle Process •
- ISO 14971:2007 Medical Device: Application of Risk Management to Medical Devices .
- ISO 15223-1:2012 Medical Device: Symbols to be used with medical device labels, labeling and information to be supplied, part 1: General requirements
- . ISO 17664:2004 Sterilization of medical devices: information to be provided by the manufacturer for the processing of resterilizable medical devices

### 10. Safety and Effectiveness

The bench testing result shows that the proposed device is complied with IEC standards, and as safe as the predicate device.

### 11. Substantial Equivalence Determination:

The VIDA LED Surgical Lighting System, V series is substantially equivalent to VIDA LED Surgical System-X series. There are no negative effect on safety or effectiveness between VIDA LED Surgical Lighting System, V series and the predicate device. The following is the detail of the compared item.

Image /page/3/Picture/18 description: The image shows a pattern of squares with rounded corners. The squares are arranged in a grid-like fashion, with some squares being larger than others. The squares are all outlined in black, and the background is white. The text "page 5-5" is written in the bottom right corner of the image.

{4}------------------------------------------------

page 5-6

ﻨﺔ ﺍﻟ

Integrate Visual Clarity at the Highest Precision

Image /page/4/Picture/1 description: The image shows the logo for ConVida Healthcare & Systems. The logo consists of a geometric pattern on the left and the text "ConVida" on the right. Below "ConVida" is the text "Healthcare & Systems" in a smaller font. The geometric pattern is made up of small squares arranged in a grid-like pattern.

ConVida Healthcare & Systems Corp. 510(K) Notification, K number: N/A

VIDA LED Surgical Lighting System, V series

| Item                               | Proposed Device                                                                                                    |                                                               | Predicate Device                                                                                               |          |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|----------|
| Manufacturer                       | ConVida Healthcare & Systems Corp.                                                                                 |                                                               | ConVida Healthcare & Systems Corp.                                                                             |          |
| Classification                     | Class II                                                                                                           |                                                               | Class II                                                                                                       |          |
| Regulation Number                  | 878.4580                                                                                                           |                                                               | 878.4580                                                                                                       |          |
| Product Code                       | FSY                                                                                                                |                                                               | FSY                                                                                                            |          |
| Intended Use                       | The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room. |                                                               | VIDA LED Surgical Lighting System- X series provide illumination for surgery or examination at operating room. |          |
| Lighthead model                    | V70                                                                                                                | V50                                                           | X70                                                                                                            | X50      |
| Similarity                         |                                                                                                                    |                                                               |                                                                                                                |          |
| Over-the-counter Medical Device    | No                                                                                                                 |                                                               | No                                                                                                             |          |
| Light Source                       | LED                                                                                                                | LED                                                           | LED                                                                                                            | LED      |
| Color Rendering Index (CRI)        | 95                                                                                                                 | 95                                                            | 95                                                                                                             | 95       |
| LED Life time (hrs)                | 50,000                                                                                                             | 50,000                                                        | 50,000                                                                                                         | 50,000   |
| Difference                         |                                                                                                                    |                                                               |                                                                                                                |          |
| Diameter of Light Head (cm)        | 82                                                                                                                 | 72                                                            | 72                                                                                                             | 58       |
| Number of LEDs                     | 55                                                                                                                 | 44                                                            | 90                                                                                                             | 60       |
| Central illumination (at 1m) (lux) | 160,000                                                                                                            | 140,000                                                       | 160,000                                                                                                        | 130,000  |
| Color temperature, Kelvins         | 3500-5000K<br>(3500K,<br>4000K,<br>4250K,<br>4500K,<br>5000K)                                                      | 3500-5000K<br>(3500K,<br>4000K,<br>4250K,<br>4500K,<br>5000K) | 4200                                                                                                           | 4200     |
| Total irradiance (W/m²)            | 592                                                                                                                | 518                                                           | < 656                                                                                                          | 533      |
| Irradiance intensity (mW/m²lux)    | 3.7                                                                                                                | 3.7                                                           | 4.1                                                                                                            | 4.1      |
| d50/d10 (%)                        | 56                                                                                                                 | ਟੇਰੇ                                                          | > 50                                                                                                           | > 50     |
| Light field size, d10 (cm)         | 17~30                                                                                                              | 16~27                                                         | 20~32                                                                                                          | 18~28    |
| Depth of illumination              | 20%: 110                                                                                                           | 20%: 100                                                      | 20%: 140                                                                                                       | 20%: 145 |
| L1+L2 (cm)                         | 60%: 45                                                                                                            | 60%: 45                                                       | 60%: 71                                                                                                        | 60%: 82  |
| Power Consumption                  | 100W@24Vdc                                                                                                         | 100@24Vdc                                                     | 100W                                                                                                           | 80W      |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for ConVida Healthcare & Systems. The logo consists of a grid-like pattern on the left and the text "ConVida" in bold font on the right. Below "ConVida" is the text "Healthcare & Systems" in a smaller font.

Integrate Visual Clarity at the Highest Precision

510(K) Notification, K number: N/A

VIDA LED Surgical Lighting System, V series

### 12. Non-clinical Performance Testing:

Performance testing were conducted to verify that the VIDA LED Surgical Lighting System, V series meets the requirements for Medical Electrical Equipment as defined in IEC 60601-1, IEC60601-2-41, and IEC 60601-1-2

### 13. Clinical Performance Testing:

No Clinical testing is required for this device classification submission.

### 14. Conclusion:

The VIDA LED Surgical Lighting System, V series submitted in this 510(K) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared VIDA LED Surgical Lighting System-X series which is the subject of K132307. Differences between the devices cited in this section do not raise any new issue of safety or effectiveness. After analyzing bench test, safety testing data, it can be concluded that VIDA LED Surgical Lighting System, V series is substantially equivalent to the predicate device.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes resembling birds in flight, arranged in a diagonal line.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 9, 2014

ConVida Healthcare & Systems Corporation Sharon Lee, Regulatory Engineer No. 33, Dinghu Road . Guishan Township, Taoyuan County 33378, Taiwan

Re: K140930

Trade/Device Name: VIDA LED Surgical Lighting System, V Series Regulation Number: 21 CFR 878.4580 Regulatory Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: April 09, 2014 Received: April 11, 2014

Dear Ms. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a and regulations administered by other Federal agencies. You must of any I edital bandler and a courtements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{7}------------------------------------------------

Page 2 - Ms. Sharon Lee

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# David Krause -S

for

- Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
- cc: DMC 510(k) Staff Division D.O.

{8}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K140930

#### Device Name

VIDA LED Surgical Lighting System, V series

#### Indications for Use (Describe)

The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

NEW YOU WARE OF FOR FDA USE ONLY MASE ONLY CHANGE ON WARRER OF THE WORLD Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

Image /page/8/Picture/12 description: The image shows the name Neil R Ogden -S, followed by the date and time 2014.06.05 15:06:29 -04. There is also a logo to the right of the date and time. The logo is not clear, but it appears to be a seal or emblem.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY/K140930](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY/K140930)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
