MACH LED 2SC

K140460 · Dr. Mach GmbH & Co. KG · FSY · Apr 3, 2014 · General, Plastic Surgery

Device Facts

Record IDK140460
Device NameMACH LED 2SC
ApplicantDr. Mach GmbH & Co. KG
Product CodeFSY · General, Plastic Surgery
Decision DateApr 3, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4580
Device ClassClass 2

Indications for Use

The surgical light Mach LED 2SC is intended to illuminate the surgical field and the patient.

Device Story

Mach LED 2SC is a surgical lamp designed to illuminate surgical fields and patients in hospitals and clinics. System comprises lamp housing, LED modules, optical/electrical/mechanical components, and sterilizable handle sleeves. Mounted on ceiling suspension systems with horizontal and vertical rotation capabilities. Operated via keypad on lamp head or wall-mounted unit. Features light field merging and intensity control; no color temperature adjustment. Accessories include camera modules, remote controls, monitor yokes, instrument trays, battery backup, and laser pointers. Provides high-intensity illumination to assist healthcare providers during surgical or examination procedures.

Clinical Evidence

No clinical data required for this submission. Performance verified via bench testing against medical electrical equipment standards.

Technological Characteristics

LED-based surgical lamp. Components: lamp housing, LED modules, optical/electrical/mechanical parts, sterilizable handle sleeves. Connectivity: ceiling-mounted suspension with 360° horizontal rotation, 50° downward/45° upward vertical movement. Power: 24V DC battery backup available. Control: keypad-operated. Standards: EN 61000 group for medical electrical equipment. Accessories: camera modules, monitor yokes, laser pointers.

Indications for Use

Indicated for illumination of the surgical field and patient in hospital and doctor's practice settings. Prescription use only.

Regulatory Classification

Identification

A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.

Special Controls

*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows a logo for "Dr. Mach Medizinische Leuchten + Technik". The logo features the company name in a bold, sans-serif font, with the "Dr. Mach" portion being larger and more prominent. Below the name, the words "Medizinische Leuchten + Technik" are written in a smaller font. To the left of the logo is a partial illustration of a surgical lamp, suggesting the company's focus on medical lighting equipment. 03 ## 510(k) Summary for Mach LED 2SC ### as required by section 807.92(c) Submitter Dr. Mach GmbH & Co.KG Flossmannstrasse 28 85560 Ebersberg Germany . Contact Person Preparation Date Trade Name Common Name Classification Name Predicate Device Rainer Adams Technical Director Phone: +49 8092 209330 +49 8092 209350 Fax: February 14, 2014 Mach LED 2SC Surgical Lamp Surgical Lamp Regulation 21 CFR 878.4580, Class II Product Code: FSY Mach LED SC Surgical lamp 510(k) No. K093009, April 1, 2010 {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows a logo for "Dr. Mach Medizinische Leuchten + Technik". The logo features the text "Dr. Mach" in a stylized font, with the words "Medizinische Leuchten" and "+ Technik" appearing below in a smaller font. The image also includes a partial illustration of a medical lighting device. ### Device Description The surgical light Mach LED 2SC is intended to illuminate the surgical field and the patient. The Mach LED 2SC consists of lamp housing, LED modules, optical -/electrical and mechanical components, one sterilizable handle sleeve as well as the cables. One LED-module consists of one white LED. The light system can be added to the ceiling mounted suspension system supporting the horizontal arms and spring arms. The horizontal arms can be rotated horizontally with 360°, the spring arms can be rotated horizontally with 360° and moved vertically with 50° downwards and 45° upwards. The light system is operated by a kevpad on the lamp head or, by special request of the customer, by a keypad on the wall. The surgical light Mach LED 2SC will be marketed with merging of light fields and light intensity control. Color temperature adjustment will not be available. Available accessories for the Mach LED 2SC lighting systems are as follows: - Camera module . - Remote control of camera module . - Remote control with network interface for camera module . - Single monitor yoke for flat panel monitors . - Double monitor yoke for flat panel monitors . - Instrument trays . - Trays for CRT monitors - 24V DC battery backup support . - . Low profile wall control unit - Integrated laser pointer . - Sterilizable handle sleeves {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a logo for "Dr. Mach Medizinische Leuchten + Technik". The logo is in black and white and features a stylized font. The image also shows a partial view of a medical examination light. ### Intended Use The Mach LED 2SC lighting system is designed for illuminating an examination area and surgical field at the hospital and doctor's practice. The intended use is identical to the intended use of the predicate device Mach LED SC, K093009. ### Technological characteristics Comparison The Mach LED 2SC has the same technological characteristics (design, materials, chemical composition, energy source) as the predicate device. The Mach LED 2SC and the predicate device Mach LED SC are both based on the LED (light-emitting diode) technology. For the new device, the number of LEDs was reduced. Therefore, dimensions of the light head diameter, the illumination depth, and the light intensity have been reduced accordingly. | Technical Parameter | New Device | Predicate Devices | | |---------------------|--------------|-------------------|--------------| | | Mach LED 2SC | Mach LED 3SC | Mach LED 5SC | | Number of LEDs | 21 | 28 | 40 | ### Non-clinical data Performance testing was conducted to verify that Mach LED 2SC meet the requirements for Medical Electrical Equipment as defined in the relevant standards for Medical electrical equipment and the EN 61000 group. The accessories are identical to those used with the predicate device. The risk analysis method used to assess the impact of the modifications was a Failure Modes and Effects Analysis (FMEA). The design verification test methods used are the same as those submitted for the predicate device submission. ### Clinical data No clinical data is required for this device classification submission. ### Conclusions The modifications incorporated into the Mach LED 2SC design are minor (change in number of LEDs). Based on the information provided herein it is concluded from the performed testing that the device is as safe, as effective, and performs as well as the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a series of flowing lines, possibly representing the flow of health and human services. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 3, 2014 Dr. Mach GmbH & Co., KG % Gudrun Busch, Ph.D. Dr. Busch Regulatory Strategy GmbH Fehmarnstrasse 31 Norderstedt, Schleswig-Holstein 22846 Germany Re: K140460 Trade/Device Name: Mach LED 2SC Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: February 27, 2014 Received: March 05, 2014 Dear Dr. Busch: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Gudrun Busch, Ph.D. device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. for Sincerely yours, # Felipe Aguel Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K140460 ### Device Name Mach LED 2SC ### Indications for Use (Describe) The surgical light Mach LED 2SC is intended to illuminate the surgical field and the patient. Type of Use (Select one or both, as applicable) [区] Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) FORM FDA 3881 (1/14) Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...