← Product Code [FSY](/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY) · K102758

# TRULIGHT 5520 AND TRULIGHT 5320 (K102758)

_Trumpf Kreuzer Medizin Systeme GmbH + Co.Kg · FSY · Oct 26, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY/K102758

## Device Facts

- **Applicant:** Trumpf Kreuzer Medizin Systeme GmbH + Co.Kg
- **Product Code:** [FSY](/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY.md)
- **Decision Date:** Oct 26, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4580
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The TRUMPF Surgical lights are intended to locally illuminate an operating or examination area of the patient's body with high intensity light.

## Device Story

TruLight 5520 and 5320 are ceiling-mounted surgical lights consisting of LED modules; used in clinics and doctor's offices to illuminate surgical or examination sites. Features include adjustable illuminance (30-100%), endo mode (5%), and optional Adaptive Light Control (ALC) for working distance adjustment. Adaptive Light Control Plus (ALC+) uses motion detection to automatically measure distance between luminaire and wound area, selecting appropriate light settings. Operated via light head control panel or wall panel. Provides high-intensity illumination to assist healthcare providers during procedures. Safety verified through risk analysis and compliance with IEC 60601-1-2 and IEC 60825-1 (Class 2 laser for ALC+).

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through compliance with IEC 60601-1-2 (EMC) and IEC 60825-1 (laser safety) standards, alongside comprehensive risk analysis comparing the ALC+ laser functionality to predicate devices.

## Technological Characteristics

LED-based surgical lighting system; ceiling-mounted or mobile configurations. Features ALC+ motion detection using a Class 2 laser (IEC 60825-1:2007). Reusable hand grips sterilized via steam (121°C/20 min, 134°C/4 min, or 270°F/4 min). Connectivity via local control panels or optional wall-mounted control panels. Complies with IEC 60601-1-2:2007.

## Regulatory Identification

A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

## Predicate Devices

- Light Surgical Ceiling Mounted, iLED ([K061317](/device/K061317.md))
- Surgical Lamp, Mobile ([K091246](/device/K091246.md))
- TRUMPF Surgical Lights, Models 501 and 701 ([K011693](/device/K011693.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K102758

# E. 510 (k) Summarv

OCT 2 6 2010

| Submitter:            | TRUMPF Medizin Systemé GmbH + Co. KG<br>Benzstraße 26<br>82178 Puchheim<br>Germany                                |
|-----------------------|-------------------------------------------------------------------------------------------------------------------|
| Contact Person:       | Lindsey Ronneberg<br>Quality Manager, TRUMPF Medical Systems, Inc.                                                |
|                       | Phone: 843-416-1376<br>Fax: 843-534-0206                                                                          |
| Preparation Date:     | September 8, 2010                                                                                                 |
| Trade Name:           | Surgical Light, TruLight                                                                                          |
| Common Name:          | Surgical light, Ceiling Mounted                                                                                   |
| Classification Name:  | Surgical Lamp                                                                                                     |
| Product Code:         | FSY                                                                                                               |
| Device Relationships: | TRUMPF Medical Systems, Inc. distributes "Surgical light<br>Ceiling Mounted, ILED" systems cleared under K061317. |

with one to three arms per suspension. The devices are offered in a variety of combinations of the approved light heads. The light heads contain two to five modules each containing LEDs. The "ILED X" family of light heads will be transitioned to names involving "TruLight XXXX" over the next several years. The most complex of the ILED light heads is the ILED 5. The ILED 5 offers adjustable color temperature, dimming, focus, and shadow control.

TRUMPF Medical Systems, Inc. released the next versions of light heads including the TruLight 5500 and 5300 in 2009. These light heads in addition to the ILED 3 and ILED 5 are offered on the same suspension system as the "Surgical light, Ceiling Mounted, ILED". The TruLight 5500, 5300, and ILED 3 light heads are offered on a mobile stand as well. They are known as "Surgical Lamp Mobile". The mobile systems with the three light head options were cleared under K091246.

TRUMPF completed a letter to file stating that the TruLight 5500 and 5300 light heads offered on a ceiling mounted system (same mounting as the original ILED family light heads) did not require a 510K because there is no significant change from the cleared devices and there was no altering of the device's safety or effectiveness.

If Su

{1}------------------------------------------------

TRUMPF Medical Systems, Inc. listed a surgical light Helion 501 and 701 cleared under K011693. This light was available with a laser focus feature. The laser was a Class 2 laser and it was an optional feature on the Helion light systems. This item is primarily listed as a predicated device due to its use of a laser for assistance with the focusing of the device (similar to the ALC+ feature).

Details of the predicated devices in comparison to the new devices are provided in pages 5 and 6 of this section.

#### Device Description:

TruLight 5520 and 5320 light heads consist of 2 light modules containing LEDs which illuminate the area of a patient. The light heads are equipped with illuminated non-sterile hand grips. The light heads can be controlled on the control panel on the light head or on an optional wall control panel. Illuminance can be adjusted between 30 and 100%. Reducing the Illuminance does not change the color temperature of the light. The illuminance can also be adjusted to "endo" mode, which is 5%.

The optional Adaptive Light Control (ALC) feature allows the lighting settings to be selectively adjusted to the working distance. The three settings for a working distance of approx. 0.8, 1.0 and 1.2 meters are selected on the control panel or on the wall control panel. The light head then selects the light setting suitable for this working distance.

The Adaptive Light Control Plus (ALC+) offers automatic control of the light intensity. If the light head is moved during surgery, the motion detection function automatically measures the distance between the luminaire and the wound area. The light head then selects the light setting suitable for this working distance.

The surgical light family is for illuminating an examination and Intended Use. surgical site on the patient in the clinic and doctor's office.

The TRUMPF surgical lights are intended to locally illuminate Indication for use: an operating or examination area of the patient's body with high intensity light.

{2}------------------------------------------------

| Predicate Device:        | Light Surgical Ceiling Mounted, iLED<br>K# 061317                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                          | Surgical Lamp, Mobile<br>K# 019246                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                          | TRUMPF Surgical Lights, Models 501 and 701<br>K# 011693                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Substantial Equivalence: | Please see the Substantial Equivalence tables in page 5 and<br>6 of this section to review the technical details and<br>applications of the predicated and new devices. The main<br>difference is identified here.                                                                                                                                                                                                                                                                                                                 |
|                          | <b>The Main Difference:</b><br>The difference from the predicated TL5500 and 5300 is the<br>addition of the Adaptive Light Control Plus (ALC+) option.<br>Lighting systems prepared with the ALC+ option still have the<br>original ALC control system of the TruLight 5500 and 5300.<br>The ALC+ system can be activated and deactivated by the<br>user pressing the left and right ALC symbols on their control<br>panels. Therefore, the light can operate identically to the<br>predicated TruLight 5500 and 5300 light heads. |
|                          | Testing to the standards mentioned, risk analysis, and<br>comparison to devices with lasers similar to that of the ALC+<br>have verified the new device is not any less safe or effective<br>than the predicated devices.                                                                                                                                                                                                                                                                                                          |
| Sterilization:           | Reusable sterile handle grips are instructed to be sterilized by<br>traditional moist heat sterilization with the following<br>instructions:                                                                                                                                                                                                                                                                                                                                                                                       |
|                          | Steam sterilization: The hand grips can withstand at least 350<br>steam sterilization cycles without damage under the following<br>criteria:<br>- Steam sterilization at 121deg C, 1.3 bar 20 minutes, or<br>- Steam sterilization at 270deg F, 41.84 psi, 4 minutes or<br>- Steam sterilization at 134deg C, 2.3 bar, 4 minutes                                                                                                                                                                                                   |
|                          | - Stand the hand grips vertically with the open side facing<br>downwards.                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                          | - Do not exceed a sterilization temperature of 134deg C.                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                          | - The hand grips must be sterilized individually in packaging<br>suitable for steam sterilization.                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                          | Note: The sterilization instructions are the same for the<br>predicated devices, additional details can be found in each<br>Operator Manual.                                                                                                                                                                                                                                                                                                                                                                                       |

.

。

{3}------------------------------------------------

Electromagnetic Compatibility:

## Performance Data:

The TruLight 5520 and 5530, same as their predicated devices, comply with the requirements of the IEC 60601-1-2: 2007 "General requirements for safety-Collateral standard: electromagnetic compatibility-Requirements and tests"

The TruLight 5520 and 5320 lighting systems have been tested to the same standards as the predicated devices and are found to have similar or improved performance features to the predicated devices.

The TruLight 5520 and 5320 lighting systems were reviewed with a thorough risk analysis by TRUMPF and additional third party experts to ensure the safety of our customers and their patients. TRUMPF Medizin Systeme GmbH + Co. KG. risk analysis reviewed possible mechanical, electronic, radiological, and handling failures associated to the TruLight and TruLight ALC+. Additional details are provided in tab G.

The device has been tested to IEC60825-1:2007 (2"® Edition). The laser compares to similar devices such as the Helion 501 and 701 lighting systems with Class 2 lasers.

These comparisons and test results have verified the new device is not any less safe or effective than the predicated devices.

## Safe and Effective:

The TruLight 5520 and 5320 lighting systems have been tested to the same standards as the predicated devices. They have been subject to a thorough risk analysis. They have additionally been tested to IEC60825-1:2007 (2"d Edition). They compare to devices with Class 2 lasers similar to that of the ALC+, such as the Helion 501 and 701 lighting systems. These comparisons and test results have verified the new device is not any less safe or effective than the predicated devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wing. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

TRUMPF Medizin Systeme GmbH + Co. KG % Underwriters Laboratories, Inc. Mr. Jeff D. Rongero 12 Laboratory Drive Research Triangle, North Carolina 27709

OCT 2 6 2010

Re: K102758

Trade/Device Name: TruLight 5520 and TruLight 5320 Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FSY, FQP, FTD, FSQ Dated: October 13, 2010 Received: October 14, 2010.

Dear Mr. Rongero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{5}------------------------------------------------

Page 2 - Mr. Jeff D. Rongero

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Sincerely yours,

F-n Nulh Nn

N. Melkersson

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known):

Device Name: Surgical Light, Ceiling Mounted, TruLight

Indications for Use: The TRUMPF Surgical lights are intended to locally illuminate an operating or examination area of the patient's body with high intensity light.

Over-The-Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nishar Oslan for mxm

Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number_K102758

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY/K102758](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FSY/K102758)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
