K040015 · Alm S.A. · FSY · Jan 21, 2004 · General, Plastic Surgery
Device Facts
Record ID
K040015
Device Name
MODIFICATION TO PRISMALIX SURGICAL LIGHT
Applicant
Alm S.A.
Product Code
FSY · General, Plastic Surgery
Decision Date
Jan 21, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The PrismAlix® Surgical Light with the upgraded Energix™ Power Supply is intended to be used to provide visible illumination of the surgical area or the patient. The Energix™ Power Supply is intended for use with surgical light systems.
Device Story
Surgical light system; provides illumination for surgical area/patient. Upgraded Energix™ Power Supply incorporates software for lighthead control; features touch-pad units for on/off switching and intensity adjustment. System includes LED indicators for illumination level, bulb failure sensing, and power mode (mains/battery). Optional RS232 communication motherboard enables data transmission to remote LCD display or PC. Data includes maintenance schedules, operating parameters, hours of operation, bulb life, self-diagnosis, and voltage/current/temperature readings. Used in clinical/surgical settings; operated by surgical staff. Benefits include improved control of illumination, enhanced monitoring of system health, and simplified maintenance tracking.
Clinical Evidence
No clinical data required for this device classification submission; bench testing only.
Technological Characteristics
Surgical luminaire system. Energix™ Power Supply (WPS) includes power modules, filter PCB, and regulators. Connectivity via RS232 for remote monitoring. Software-controlled intensity regulation. Tested to UL 2601-1, CSA C22.2 No. 601.1, IEC 60601-2-41, and EN 60601-1-2. Software classified as Minor Level of Concern.
Indications for Use
Indicated for providing visible illumination of the surgical area or the patient during surgical procedures.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
ALM PrismAlix® (PRX) Series Surgical Light (K982063)
Related Devices
K100388 — 21ST CENTURY CENTURION EXCELED AND SYSTEM TWO LED LIGHTING SYSTEMS · Medical Illumination International, Inc. · Jun 22, 2010
K100395 — HARMONY VLED SURGICAL LIGHTNING SYSTEM · STERIS Corporation · Mar 17, 2010
K040735 — ALM X'TEN (X10) SURGICAL LIGHT SYSTEM · Alm S.A. · May 21, 2004
K072072 — HARMONY LED-1 SURGICAL LIGHTING SYSTEM · STERIS Corporation · Oct 5, 2007
K070442 — POWERLED SURGICAL LIGHT SYSTEM · Maquet S.A. · Mar 16, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
KO40015 (pg 1 of 3)
# 510(k) SUMMARY
## ALM PrismAlix® (PRX) Surgical Light with Energix™ Power Supply
| Submitted by: | Getinge USA, Inc. (as ALM S.A.'s US Agent)<br>1777 E Henrietta Road<br>Rochester, NY 14623-3133 |
|------------------------|-------------------------------------------------------------------------------------------------|
| Contact Person: | Frederick R. Catt<br>Senior Regulatory Engineer<br>Phone: (585) 272-5013<br>Fax: (585) 272-5299 |
| Date prepared: | December 31, 2003 |
| Proprietary Name: | ALM PrismAlix® (PRX) Surgical Light |
| Common Name: | Surgical Light |
| Device Classification: | Surgical Lamp (78 FSY)<br>Class II, as listed per 21 CFR 878.4580 |
| Predicate Device: | ALM PrismAlix® (PRX) Series Surgical Light [K982063] |
#### Description of Modified Device:
The PrismAlix® (PRX) Surgical Light System is an existing product line that will allow modification via the use of an updated Energix™ Power Supply as its means to control the surgical lighthead illumination levels. The updated Energix™ Power Supply contains software that enables additional features and capabilities not available on the earlier version of Energix™ Power Supplies. These include:
- a) The power supply and regulators,
- b) one or two touch-pad units that provide the user the ability to:
- i) turn on/off one or two lightheads,
- ii) adjust illumination intensity levels and
- iii) additional Light Emitting Diodes (LEDs) are used to indicate:
- (1) the illumination level,
- (2) bulb failure sensing and
- (3) power supply mode: Mains supply: green LED or the switching to a battery back-up system results in the LED indicator changing its color to red.
{1}------------------------------------------------
K040015 (pg 2 of 3)
ALM S.A. FDA 510(k) Summary Power Surgical Light with the upgraded Energix™ Power Supply
An optional RS232 Communication Motherboard is also a new feature that allows shifting An optional ROZOE Ochinanicaa specified distance (50m, 164 ft.) away from the surgical Controls to a romote bite are provided specific user and maintenance type information/data, displayed on a Liquid Crystal Display (LCD) window located on the front information and the Energix™ power supply. The user also has an option to have data communicated to a personal computer (PC).
The LCD can show information about preventive maintenance type items, including:
- 1. programming of maintenance inspections
- 2. storage of operating parameters
- 3. hours of operation
- 4. bulb life data
- 5. self-diagnosis information
- 6. direct reading of parameters voltage, current, temperature
- 7. traceable management data serial number, software version number, production and service dates)
#### Table 1
| Energix™ Power Supply – Designations/Configurations | |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Model | Description |
| WPS443 | Example: WPS Power Supply with two 360W Power Modules, two keypads and a RS232 Communication Motherboard |
| WPS | Base Model Designation [Transformer; plus filter PCB and regulator(s)] |
| | 2 150W Power Module and control keypad |
| | 4 360W Power Module and control keypad |
| | 0 Space holder – No #2 Power Module and 2nd control keypad (single lighthead) |
| | 2 150W Power Module #2 and 2nd control keypad |
| | 4 360W Power Module #2 and 2nd control keypad |
| | 0 Power Module(s) only |
| | 1 Power Module(s), RS232 Communication Motherboard and Battery Backup system/charger (Note: not currently planned for USA distribution) |
| | 3 Power Module(s), RS232 Communication Motherboard |
{2}------------------------------------------------
K040015 (pg 3 of 3)
ALM S.A. FDA 510(k) Summary Device: PrismAlix® Surgical Light with the upgraded Energix ™ Power Supply
Page 3 of 3 December 31, 2003
#### Intended Use:
The PrismAlix® Surgical Light with the upgraded Energix™ Power Supply is intended to be used to provide visible illumination of the surgical area or the patient. The Energix™ Power Supply is intended for use with surgical light systems.
#### Nonclinical Comparisons to Predicate Device
The PrismAlix® Surgical Light with the upgraded Energix™ Power Supply (subject device) is similar to the predicate device with the following modifications:
- Replacement of the older Energix™ design with an upgraded Energix™ Power . Supply (WPS) that includes use of software, and makes available a low energy remote touch-pad panel per lighthead.
- As the RS232 option, the Energix™ can provide the user with product condition . type information via a LCD screen or interfacing with a remote PC.
- Software is incorporated within the Energix™ design, which interfaces with input . signals sent from the dimmer control and ON/OFF lamp touch-pads. These switches provide output signals used by the intensity control regulator supplying voltage to the lamps (bulbs) within the corresponding lighthead.
- The Energix™ Power Supply can operate up to two lightheads, rated 360W each. ●
#### Test Data:
The test data supports conformance to:
- UL 2601-1 Standard for Medical Electrical Equipment, Part 1: General Requirements . for Safety
- CSA C22.2 No. 601.1 Medical Electrical Equipment, Part 1: General Requirements for . Safety
- IEC 60601-2-41 Medical electrical equipment Part 2-41: Particular requirements for . the safety of surgical luminaires and luminaires for diagnostics
- EN 60601-1-2 Medical electrical equipment Part 1-2: General requirements for safety . - Collateral standard: Electromagnetic compatibility - Requirements and tests
- Software used in the Energix™ Power Supply was tested according to the appropriate . FDA Software Guidance Documents, per its determination as a Minor Level of Concern.
#### Clinical Data:
No clinical data is required for this device classification submission.
#### Conclusion:
Based upon the information provided herein this 510(k) Premarket Notification, we conclude that the ALM PrismAlix® (PRX) Series Surgical Light System using the updated Energix™ Power Supply is substantially equivalent to the predicate device(s) and is safe and effective when used as intended.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 1 2004
ALM S.A. c/o Mr. Frederick R. Catt Sr. Regulatory Engineer Getinge USA, Inc. 1777 East Henrietta Road Rochester, New York 14623
Re: K040015
Trade/Devicc Name: PrismAlix®(PRX) Surgival Light with Energix™ Power Supply Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical Lamp Regulatory Class: II Product Codes: FSY Dated: December 31, 2003 Received: January 5, 2004
Dear: Mr. Catt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Frederick Catt
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
A Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): K040015
Device Name: PrismAlix® (PRX) Surgical Light with Energix™ Power Supply
Indications for Use:
ALM S.A. PrismAlix® (PRX) Surgical Lights with Energix™ Power Supply are intended to be used to provide visible illumination of the surgical area or the patient. The Energix™ Power Supply is intended for use with surgical light systems.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
Page 1 of _1
Sister Sign-Ofly inion of General, Restorative . Neurological Devices
K040015
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.