HARMONY VLED SURGICAL LIGHTNING SYSTEM

K100395 · STERIS Corporation · FST · Mar 17, 2010 · General, Plastic Surgery

Device Facts

Record IDK100395
Device NameHARMONY VLED SURGICAL LIGHTNING SYSTEM
ApplicantSTERIS Corporation
Product CodeFST · General, Plastic Surgery
Decision DateMar 17, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4580
Device ClassClass 2

Intended Use

The proposed Harmony vLED Surgical Lighting System is a variable pattern / intensity surgical light designed to provide visible illumination of the surgical field and the patient and to provide video-visual procedural support for the hospital staff during surgical procedures.

Device Story

Variable pattern/intensity surgical light system; provides visible illumination of surgical field/patient; includes video-visual procedural support. Components: STERIS ACT interface, camera module, single/dual flat panel monitors, 24VDC battery backup, remote power module, sterile disposable handle covers. Used in OR/surgical settings by hospital staff. Replaces existing surgical lights or installed in new/renovated facilities. Benefits include improved visualization of surgical site and integrated video support for clinical teams.

Clinical Evidence

Bench testing only. Performance verified against CEI/IEC 60601-1 and IEC 60601-2-41 standards for medical electrical equipment.

Technological Characteristics

Variable pattern/intensity LED surgical light. Includes camera module, flat panel monitors, and battery backup. Complies with CEI/IEC 60601-1 and IEC 60601-2-41 safety standards. Ceiling-mounted form factor.

Indications for Use

Indicated for hospital staff to provide visible illumination of the surgical field and patient, and to provide video-visual procedural support during surgical procedures.

Regulatory Classification

Identification

A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.

Special Controls

*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # STERIS® MAR ] 7 2010 # 510(k) Summary For Harmony vLED Surgical Lighting System K100395 STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (216) 354-2600 Fax No: (216) 639-4459 Contact: Robert F. Sullivan Senior Director, Regulatory Affairs 440 392 7695 Telephone: 440 357 9198 Fax No: Submission Date: ﻪ ﻣﻦ ﻣﺎﻳﻮ February 12, 2010 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {1}------------------------------------------------ # 1. Device Name | Trade Name: | Harmony vLED Surgical Lighting System | |------------------------|---------------------------------------| | Common Name: | Light, Surgical, Ceiling Mounted. | | Classification Name: | Light, Surgical, Ceiling Mounted. | | Classification Number: | 21 CFR 878.4580 | | Product Code: | FSY | # 2. Predicate Device Harmony LED-1 Surgical Lighting System (K072072) # 3. Device Description The Harmony vLED Surgical Lighting System is a variable pattern / intensity surgical light designed to provide visible illumination of the surgical field and the patient and to provide video-visual procedural support for the hospital staff during surgical procedures. The Harmony vLED Surgical Lighting System is designed to replace existing surgical lights, or to be installed as part of major renovations to existing facilities or in new facilities. The Harmony vLED Surgical Lighting System accessories are as follows: - STERIS ACT Interface o - o Camera Module - Dual or Single Flat Panel Monitors o - o 24VDC Battery Backup support - Remote Power Module Unit 0 - Sterile Disposable Handle Covers o # 4. Intended Use The proposed Harmony vLED Surgical Lighting System is a variable pattern / intensity surgical light designed to provide visible illumination of the surgical field {2}------------------------------------------------ and the patient and to provide video-visual procedural support for the hospital staff during surgical procedures. The performance of the Harmony vLED Surgical Lighting System meets the general requirements for safety as defined in CEI/IEC 60601-1 and IEC 60601-2-41 for Medical Electrical Equipment. # Description of Safety and Substantial Equivalence 5. The proposed Harmony vLED Surgical Lighting System, like its predicate device, is a variable pattern, variable intensity surgical light designed to provide visible illumination of the surgical field and the patient and to provide optional videovisual procedural support for the hospital staff during surgical procedures. The proposed device has the same function, intended use, components, technology, and performance as the predicate device, Harmony LED-1 Surgical Lighting System (K072072) manufactured and owned by STERIS Corporation. The differences between the proposed and predicate devices are limited to differences in design and operation. These differences do not raise any new issues of safety and efficacy. #### Performance Testing 6. Performance testing was conducted to verify that the Harmony vLED Surgical Lighting System meets the requirements for Medical Electrical Equipment as defined in CEI/IEC 60601-1 and IEC 60601-2-41. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, surrounded by a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is inscribed along the border of the seal. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 MAR 1 7 2010 STERIS Corporation % Mr. Robert F. Sullivan Senior Director. Regulatory Affairs 5960 Heisley Road Mentor, Ohio 44060 Re: K100395 Trade/Device Name: Harmony vLED Surgical Lighting System Regulation Number: 21 CFR 878,4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FST Dated: February 12, 2010 Received: February 16, 2010 Dear Mr. Sullivan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 - Mr. Robert F. Sullivan CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, incerely yours, Mark N. Melkerson Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use K100395 510(k) Number (if known): Device Name: Harmony vLED Surgical Lighting System Indications For Use: The Harmony vLED Surgical Lighting System is a variable pattern / intensity surgical light designed to provide visible illumination of the surgical field and the patient and to mgnide video-visual procedural support for the hospital staff during surgical procedures. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR X Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number. K100398 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...