Maestro System (REF100)

K240598 · Moon Surgical · FQO · Jun 3, 2024 · General, Plastic Surgery

Device Facts

Record IDK240598
Device NameMaestro System (REF100)
ApplicantMoon Surgical
Product CodeFQO · General, Plastic Surgery
Decision DateJun 3, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4960
Device ClassClass 1

Indications for Use

The Maestro System is intended to hold and position laparoscopes and laparoscopic instruments during laparoscopic surgical procedures.

Device Story

Maestro System is a 2-arm, electrically actuated surgical support device used in laparoscopic procedures. It utilizes DC motors to compensate for gravitational forces on instruments, allowing surgeons to maneuver tools freely; when hand force is removed, the system maintains the instrument's position and orientation. The system includes a surgical base, arms, control unit, and instrument coupling. New features include positioning guidance sensors, bedside joint control via a forearm collar, arm proximity indicators with audible/visual alerts, and tactile vibration feedback for hold status. The system is operated by surgeons in the OR. It benefits patients by providing stable instrument positioning, reducing surgeon fatigue. Output is physical support of the instrument; status is communicated via LEDs, touch screen, and tactile vibration.

Clinical Evidence

No clinical data from human trials was required. Evidence consists of bench-level design verification (payload capacity, force accuracy, collision detection, system latency, electrical safety, EMC) and design validation, including cadaver testing and human factors testing to confirm system usability and safety.

Technological Characteristics

Electrically actuated 2-arm system; DC motors for gravity compensation; sensors for positioning guidance and proximity detection; LED status indicators; tactile vibration feedback; touch screen interface. Sterilization for single-use coupling devices via Ethylene Oxide (EtO). Connectivity includes system console and bedside controls. Software-controlled motion and fault monitoring.

Indications for Use

Indicated for holding and positioning laparoscopes and laparoscopic instruments during laparoscopic surgical procedures.

Regulatory Classification

Identification

Operating tables and accessories and operating chairs and accessories are AC-powered or air-powered devices, usually with movable components, intended for use during diagnostic examinations or surgical procedures to support and position a patient.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. June 3, 2024 Moon Surgical % Michael Daniel Consultant Daniel & Daniel Consulting 340 Jones Lane Gardnerville, Nevada 89460 Re: K240598 Trade/Device Name: Maestro System (REF100) Regulation Number: 21 CFR 878.4960 Regulation Name: Operating tables and accessories and operating chairs and accessories Regulatory Class: Class I Product Code: QZB Dated: March 2, 2024 Received: March 4, 2024 Dear Michael Daniel: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long H. Chen =S Digitally signed by Long H. Chen-S Digitally signed Date: 2024.06.03 14:50:58 -04'00' Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use Submission Number (if known) K240589 Device Name Maestro System (REF100) Indications for Use (Describe) The Maestro System is intended to hold and position laparoscopes and laparoscopic instruments during laparoscopic surgical procedures. Type of Use (Select one or both, as applicable) (Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 5. Premarket Notification 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. #### 510(k) Number: K240589 #### Applicant Information: | Date Prepared: | July 6, 2023 | |-----------------|-------------------------------------------------------------| | Name: | Moon Surgical | | Address: | 9 rue d'Enghien<br>75010 Paris France | | Contact Person: | Michael A Daniel, Consultant<br>madaniel@clinregconsult.com | | Mobile Number: | (415) 407-0223 | #### Device Information: | Device Trade Name: | Maestro System | |---------------------------|-----------------------------------| | Common Name: | Maestro System | | Classification Name(s): | Table, Operating-Room, AC-Powered | | Product Code/ Regulation: | QZB 21 CFR 870.4960 | | Classification: | Class I | ### Predicate Device: Maestro System (K221410) # Subject Device Description The Moon Maestro System is a 2-arm system which utilizes software and hardware to provide support to surgeons for manipulating and maintaining instrument position. Motors compensate for gravitational force applied to laparoscopic instruments, while surgeon control is not affected. Conventional laparoscopic tools are exclusively controlled and maneuvered by the surgeon, who grasps the handle of the surgical laparoscopic instrument and moves it freely until the instrument is brought to the desired position. Once surgeon hand force is removed, the Maestro system reverts to maintenance of the specified tool position and instrument tip location. This 510(k) is being submitted to implement design changes to the previously cleared Maestro System. The following modifications have been implemented to the Maestro System: - · System Positioning Guidance - · System Hold Status Indication - Instrument Coupling - · System Setup - Bedside Setup Joint Control {4}------------------------------------------------ # Subject Device Intended Use / Indications for Use The Maestro System is intended to hold and position laparoscopes and laparoscopic instruments during laparoscopic surgical procedures. # Predicate Device and Subject Device Comparison The table below compares the Maestro System to the predicate device. | ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON | |---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------| | Device Name | Maestro | Maestro | N/A | | 510(k) Number | K221410 | Not yet assigned | N/A | | Product Code /<br>Regulation | FQO / 878.4960<br>Table Operating-room, AC-<br>Powered. | FQO / 878.4960<br>Table Operating-room, AC-<br>Powered. | Identical | | Brief<br>Description | The Moon Surgical Maestro<br>device is an electrically<br>actuated device with<br>movable components<br>intended for laparoscopic<br>surgical procedures to<br>support and position<br>laparoscopic instruments.<br>Its major components consist<br>of the Surgical Base,<br>Surgical Arms, Control Unit,<br>Instrument Coupling, and<br>Control Station. | The Moon Surgical Maestro<br>device is an electrically<br>actuated device with<br>movable components<br>intended for laparoscopic<br>surgical procedures to<br>support and position<br>laparoscopic instruments.<br>Its major components<br>consist of the Surgical<br>Base, Surgical Arms,<br>Control Unit, Instrument<br>Coupling, and Control<br>Station. | Identical | | Intended Use /<br>Indications for<br>Use | The Maestro System is<br>intended to hold and position<br>laparoscopic Instruments<br>during laparoscopic surgical<br>procedures. | Same | Identical | | Mechanism of<br>Action (System<br>Actuation /<br>Positioning) | The Moon Maestro System<br>utilizes DC motors to<br>compensate for gravitational<br>forces applied to<br>laparoscopic instruments. | Same | Identical | | ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON | | | Software and hardware are<br>designed to maintain<br>instrument position.<br>Surgeon control is achieved<br>by grasping the handle of<br>conventional laparoscopic<br>tools and moving the tool to<br>the desired position. Once<br>surgeon hand force is<br>removed, the Maestro<br>system reverts to<br>maintenance of the specified<br>location including<br>orientation. | | | | Endoscope<br>Movement | Control algorithm applying<br>motor current for instrument<br>motion triggered by<br>physician direct movement. | Same | Identical | | Platform /<br>Mechanism of<br>stable<br>attachment of<br>arms | Jointed arms are attached to<br>stages located on a portable<br>base that can be locked in<br>position. The bases are<br>locked in position during<br>system use. | Same | Identical | | Cart Setup prior<br>to procedure<br>initiation | No positioning guidance | Sensors have been added to<br>improve system positioning<br>and setup prior to procedure<br>initiation. These sensors<br>monitor OR table position<br>and distance as the user<br>advances the system toward<br>the OR table. The user<br>always has the freedom to<br>position the system where<br>they want, but the guidance<br>provides the recommended<br>window of placement | Substantially<br>equivalent with no<br>new or different<br>questions of safety or<br>effectiveness. | | System Setup | User-controlled powered x &<br>z translation of the arms, and<br>manual translation of the | User-selectable preset<br>position for each specific<br>clinical procedure to be | Substantially<br>equivalent with no<br>new or different | | ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON | | | shoulder roll via a toggle on<br>the system console. | performed, chosen off a<br>menu on the system<br>console. | questions of safety or<br>effectiveness. | | | | Discrete manual control of<br>these degrees of freedom<br>always remains available to<br>the user on the system<br>console for fine adjustments<br>via user-controlled powered<br>x & z translation of the<br>arms, and user-controlled<br>powered translation of the<br>shoulder roll | | | Bedside Setup<br>Control | Lack of complete control<br>over the setup joints from the<br>surgeon's bedside position.<br>Adjusting x & z setup joints<br>required interfacing with the<br>non-sterile console on the<br>system. | A control collar was added<br>to each Maestro arm's<br>forearm. Adjusting x, z and<br>shoulder roll setup joints no<br>longer requires interfacing<br>with the non-sterile<br>console.<br><br>In all cases, if an instrument<br>is attached during<br>positioning adjustments, the<br>system maintains the<br>instrument's tip position<br>during any setup joint<br>motion. | Substantially<br>equivalent with no<br>new or different<br>questions of safety or<br>effectiveness. | | Arm Proximity<br>Indicator | No Arm Proximity Indicator | Sensors have been added to<br>improve arm positioning by<br>monitoring for objects<br>within an established range.<br>When an object enters that<br>range, the system emits an<br>audible chime, and provides<br>a notification on the<br>Console. | | | Method of<br>instrument<br>attachment | Mechanical clamp around<br>laparoscopic instrument and<br>magnetic attachment to arm | Mechanical clamp around<br>laparoscopic instrument and | Substantially<br>equivalent with no<br>new or different | | ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON | | | | mechanical attachment to<br>arm | questions of safety or<br>effectiveness | | Instrument<br>Coupling | Self-orienting sterile<br>couplers attached to the<br>instrument's shaft. Two<br>coupler sizes were used, 10<br>mm and 5 mm, to<br>accommodate the most<br>typical surgical instrument<br>shaft calibers | A modified Coupler design<br>maintains the same<br>functionality and sizes (5<br>and 10 mm).<br>Alignment and constraints<br>are achieved through new<br>mechanical methods on the<br>modified coupler. | Substantially<br>equivalent with no<br>new or different<br>questions of safety or<br>effectiveness | | Mechanisms<br>involved in<br>physical<br>positioning | Electro-mechanical jointed<br>arms driven by internal<br>motors that follow the<br>physician movements. | Same | Identical | | Method of<br>control | Direct manipulation of<br>conventional laparoscopic<br>instrument handles.<br>Endoscope and Instrument<br>positions are unlocked via<br>surgeon applied force to<br>laparoscope or instrument<br>handles. This unlocking<br>force is on the order of<br>magnitude of what would be<br>expected in holding or<br>repositioning the endoscope<br>or instruments themselves.<br>The endoscope or<br>instruments are repositioned<br>as desired by the surgeon.<br>Once the surgeon releases<br>the instruments or scope the<br>system maintains the new<br>position with motor driven<br>forces sufficient to<br>compensate for gravity. | Same | Identical | | System Hold<br>Status Indicator | Two LED indicators on the<br>system forearms are used to<br>indicate transitions between | The three LED locations<br>have been expanded and<br>pushed further proximal on | Substantially<br>equivalent with no<br>new or different | | ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON | | | co-manipulation and hold<br>states as they change<br>between purple and blue,<br>respectively | the device. LED indicators<br>are present at the elbow and<br>shoulder of the support arm.<br>In addition to the visual<br>indicators, the system now<br>has light tactile feedback<br>upon transition from co-<br>manipulation to hold for<br>information purposes to the<br>user. The system provides<br>short vibrations along the<br>axis of the instrument,<br>which the surgeon feels<br>while holding the<br>instrument. | questions of safety or<br>effectiveness. | | Alarm and<br>Warning<br>Condition<br>Indicators | The Maestro has two<br>primary LEDs for<br>communication to the<br>surgeon. These LEDs change<br>colors and pulse to<br>communicate system status.<br>In the case of a critical<br>system fault, the LEDs will<br>change to red. | Same | Identical | | | The Maestro is also<br>equipped with a touch screen<br>graphical interface. This<br>screen provides messages to<br>the operating room staff<br>regarding system state. | | | | Single Fault<br>Tolerance | Automatic System<br>Performance Monitoring in<br>real-time with multiple fault<br>tiers + visual supervision by<br>surgeon via LED indicating<br>system status. | Same | Identical | | | Examples include:<br>- Redundant encoders on<br>motorized axes | | | | ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON | | | - Velocity, acceleration,<br>current and torque limits<br>- Brakes engage if power is<br>removed | | | | Back-up fault<br>response | Brakes engage on motorized<br>axis in the event of a fault<br>state to prohibit any arm<br>motion | Same | Identical | | Maximum<br>Applied Load | 4.4 lbs tested. | Same | Identical | | Sterilization<br>Method | Steam - for coupling devices<br>used to attach laparoscopic<br>instruments | Ethylene Oxide (EtO) – for<br>single use coupling devices<br>used to attach laparoscopic<br>instruments | Substantially<br>equivalent with no<br>new or different<br>questions of safety or<br>effectiveness | | Sterility barrier | Drape | Same | Identical | {5}------------------------------------------------ {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ # Testing Performed Design verification testing included the following: - . Payload capacity - Single fault condition ● - Force accuracy ● - Drape integrity ● - Hold position accuracy. - Positioning guidance and collision detection ● - System positioning accuracy ● - Bedside joint control ● - System end-to-end workflow ● - Design inspection ● - System setup ● - System latency ● - System cleaning ● - LED status ● - Emergency stop ● - Coupler performance - IFU inspection ● - Software verification . {10}------------------------------------------------ - Electrical safety - EMC Design validation testing included the following: - Cadaver testing ● - Human factors testing ● Testing described in this 510(k) consisted of verification of all system input requirements and product specifications. All clinical input requirements were validated. #### Summary Based upon the Intended Use, Indications for Use, product technical information, performance evaluation, and standards compliance provided in this premarket notification, the Maestro System has been shown to be substantially equivalent to the cited predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...